Valid-B.Well BP25 Study

A study testing Blood pressure measurement in people with Volunteers and Study Focus: Blood Pressure Measures, run by B.Well Swiss AG.

Not yet recruiting

Opening soon. The study is approved but enrolment has not started yet. Sites may already take enquiries.

Sites and contacts
Official record
Search all trials

Trial summary

Prospective, single-center, single-arm, open, comparative, validation of a commercialized device, repeated measures accuracy study designed in accordance with the requirements of the ISO 81060-2:2018 AAMI/ESH/ISO Universal Standard for the clinical validation of automated non-invasive sphygmomanometers (Non-invasive… Read more about this study

Who is it for?
Ages 12 years and olderHealthy volunteers can take part
What is being tested?
Blood pressure measurement
Could I receive a placebo?
No, everyone receives an active treatment
How many people
100 participants
Where
No sites listed yet
Study dates
Jan 2026to Mar 2026
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 12 years and older
  • All sexes
  • Healthy volunteers accepted
  • Child, Adult, Older adult

Who may be able to join

  • Age > 12 years
  • Signed Informed Consent

Who cannot join

  • Any medical condition preventing light to moderate arm compression
  • Absence of the upper arm
  • Subjects with upper arm circumference <22 cm or >42 cm
  • Subjects with severe shock;
  • Subjects with a history of mental illness or current mental disorders;
  • Patients with arrhythmias (atrial premature beats, ventricular premature beats, atrial fibrillation, etc.) or upper limb arteriosclerosis (at the direct measurement site);
Show all 15Show fewer
  • Subjects using extracorporeal circulation devices;
  • Subjects with excessive limb asymmetry;
  • Subjects who have undergone mastectomy or lymph node dissection;
  • Subjects with arm injuries or exposed wounds, or those with circulatory disorders;
  • Subjects with vascular accesses in the arms or arteriovenous shunt tubes;
  • Subjects who have participated in other clinical trials within the past 1 month;
  • Subjects for whom the procedures during the clinical trial may pose excessive medical risks;
  • Subjects determined by the investigator to be unsuitable for participating in the clinical study.
  • Subject under judicial protection, guardianship or curatorship or participant deprived of their liberty by judicial or administrative decision

Trial locations and contacts

No sites are listed yet. Contact the study team or a site near you to ask about taking part.

About this study

Official title: Validation of the B. Well Digital Upper Arm Blood Pressure Monitor Pro-25 According to ISO 81060- 2

Prospective, single-center, single-arm, open, comparative, validation of a commercialized device, repeated measures accuracy study designed in accordance with the requirements of the ISO 81060-2:2018 AAMI/ESH/ISO Universal Standard for the clinical validation of automated non-invasive sphygmomanometers (Non-invasive sphygmomano-meters - Part 2: Clinical investigation of intermittent automated measurement, including Amendments 1 and 2)).

The validation will be conducted with volunteers rather than necessarily with patients, and all measurements will be performed outside of the subjects' normal diagnosis or treatment.

Being tested
Blood pressure measurement
Keywords
blood pressure, iso 81060-2, upper arm blood pressure monitor, commercialized device validation

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Blood pressure measurementDiagnostic test

    All measurements will be performed by trained observers in line with the requirements of the ISO 81060-2:2018 AAMI/ESH/ISO Universal Standard (1), ensuring accuracy, consistency, and reproducibility.

    The investigation shall be performed in an isolated room under comfortable ambient temperature.

    • The cuff shall be applied on the bare arm with no compression proximal to the cuff.
    • The subject shall rest quietly for at least 10 minutes prior to the start of measurements.
    • The subject should avoid talking during the entire measurement procedure.
    • BP shall be measured on the subject's left arm at heart level.
    • A minimum of 5 minutes shall elapse prior to obtaining the first reference blood pressure measurement.
Study groups in NCT07340957
Study groupRoleWhat they receive
Sigle armExperimentalSingle-arm. All subjects will undergo blood pressure measurement with the reference manual method and with upper arm automated sphygmomanometer (3 times each).
Receives: Diagnostic Test: Blood pressure measurement

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. The study will assess compliance with the following validation criteria defined by the standard ISO 81060-2:2018: Criterion 1 - Group-level accuracy (paired-difference analysis). Criterion 2 - Subject-level accuracy

    The study will assess compliance with the following validation criteria defined by the standard ISO 81060-2:2018 for SBP and DBP (in mmHg):

    Criterion 1 - Group-level accuracy (paired-difference analysis). For systolic blood pressure (SBP) and diastolic blood pressure (DBP) separately, the mean difference between the test device and the reference method (observer auscultation with a mercury sphygmomanometer) across all valid paired determinations shall be within ±5 mmHg, and the standard deviation (SD) of these differences shall be ≤ 8 mmHg.

    Criterion 2 - Subject-level accuracy For SBP and DBP, each subject's mean difference (test device - reference) shall be calculated from that subject's repeated paired determinations. The standard deviation of these subject-level means across all subjects shall comply with the requirements specified in Table 1 Averaged subject data acceptance in mmHg of ISO 81060-2:2018 (1) for SBP and DBP.

    Measured: Day 1

Secondary outcomes (6)
  1. Agreement by Bland-Altman analysis

    Calculation of the bias (average difference between test device and reference method) and the 95% limits of agreement (LoA) for systolic blood pressure (SBP) and diastolic blood pressure (DBP), overall and across the blood pressure range

    Measured: Day 1

  2. Population characteristics- Age

    Age (descriptive summary)

    Measured: Day 1

  3. Population Characteristics - Sex

    Sex (descriptive summaries)

    Measured: day 1

  4. Population Characteristics- arm circumferences

    arm circumferences (descriptive summaries)

    Measured: Day 1

  5. Population Characteristics- baseline blood pressure categories

    Normotensive/hypertensive ranges for both systolic and diastolic pressures

    Measured: Day 1

  6. Observser agreement

    percentage of paired measurements (within +/- 4 mmHg)

    Measured: Day 1

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted14 Jan 2026
  2. Study startJan 2026 (estimated)
  3. Primary completionMar 2026 (estimated)
  4. Study completionMar 2026 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Model description
Prospective, single-center, single-arm, open, comparative, validation of a commercialized device, with repeated measures accuracy study designed in accordance with the requirements of the ISO 81060-2:2018 AAMI/ESH/ISO Universal Standard for the clinical validation of automated non-invasive sphygmomanometers (Non-invasive sphygmomano-meters - Part 2: Clinical investigation of intermittent automated measurement, including Amendments 1 and 2)). The study aims to clinically evaluate the accuracy of the B-Well PRO-25 upper arm automated sphygmomanometer for systolic blood pressure (SBP) and diastolic blood pressure (DBP) in adults and adolescents in accordance with ISO 81060-2. The comparative measurements are obtained through auscultation by an observer using a mercury sphygmomanometer (reference method).
Masking
None (open label)
Enrolment
100 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07340957
Sponsor's protocol ID
BWE-01
Status last verified
Jan 2026
First submitted
26 Nov 2025
Last updated
14 Jan 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Questions about this trial

What is this clinical trial studying?

A study testing Blood pressure measurement in people with Volunteers and Study Focus: Blood Pressure Measures, run by B.Well Swiss AG.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 15 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: opening soon, as of the sponsor's last update on 14 Jan 2026. Status can change between updates, so confirm with the study contact before making plans.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Thomas Fassler Fassler (+41 79 309 93 79, thomas.faessler@bwell-swiss.ch), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

The study will assess compliance with the following validation criteria defined by the standard ISO 81060-2:2018: Criterion 1 - Group-level accuracy (paired-difference analysis). Criterion 2 - Subject-level accuracy (time frame: Day 1). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 14 Jan 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in