Vital signs are critically important to modern emergency medical care. These metrics (heart rate (HR), respiratory rate (RR), blood pressure, temperature, and pulse oximetry) have been shown to predict a patient's morbidity, mortality, need for ICU-level admission, and suitability for discharge. Read more about this study

Who is it for?
Ages 18 to 75 yearsHealthy volunteers can take part
Conditions the study is for
Vital Sign Evaluation
What is being tested?
SmartSpectra Vital Signs Monitor measurement
Type of treatment
Medical device
How many people
111 participants
Where
1 siteUnited States
Study dates
28 Mar 2025to 21 Apr 2025
Phase
Observational
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 75 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Study population: Located in Ben Taub Emergency Department of Harris County Hospital District

Who may be able to join

  • Subject decision making capacity;
  • Subject not undergoing medical treatment;
  • Subject oriented to person, place, time;
  • Subject is a patient, visitor or employee to Ben Taub Hospital;
  • Subject greater than 18 years of age and less than 75 years old.

Who cannot join

  • Subject is employee of Presage Technologies;
  • Subject pregnant;
  • Subject does not pass screening questionnaire.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Ben Taub Hospital

    Houston, Texas, 77030

Other trials in United States

About this study

Official title: Clinical Validation of Non-Contact Vital Signs in an Emergency Department Setting

Vital signs are critically important to modern emergency medical care. These metrics (heart rate (HR), respiratory rate (RR), blood pressure, temperature, and pulse oximetry) have been shown to predict a patient's morbidity, mortality, need for ICU-level admission, and suitability for discharge. As such, vital signs are used at multiple points of emergency department (ED) care, from triage through treatment through disposition.

Unfortunately, the acquisition of vital signs can be a time-intensive process. Time-and-motion studies have shown the measurement and recording of vital signs may take over five minutes per patient and longer with interruptions.5 Increased crowding, lack of available staff, and length-of-stay have all been shown to delay or decrease the frequency of vital sign measurements. Some studies have shown a majority of both children and adults may be missing vital signs during their ED stays. Other studies have shown repeat vital signs to be taken less frequently every two hours. With increasing boarding and crowding in ED settings, the frequency of vital sign acquisition may be significantly worse than this reported data.

Read more

Many emergency departments have sought to improve vital sign acquisition through various quality improvement initiatives. Some researchers have also evaluated alternative means of vital sign acquisition. Telemedical providers, for instance, have evaluated the accuracy of patients' self-reported vital signs. Device studies have also looked at non-contact vital signs via radar and thermal imaging. More recently, studies have evaluated the feasibility of acquiring vital signs using cameras.

These studies of camera-acquired vital signs have shown promise, though they are limited by small sample sizes. Here, we present the clinical validation of camera-acquired vital signs obtained from commercially available products over a larger sample size.

Read the detailed description

Vitals signs are a key component of modern emergency medical assessment. They are acquired at the initiation of most emergency department ("ED") visits and inform triage, diagnostic workup, and therapeutic care. Unfortunately, vital signs are often resource-intensive. Acquisition of vital signs may take a provider several minutes. Accuracy of vital signs may be affected by the noise and other environmental factors within the ED.

Decreased frequency of vital sign acquisition affects ED care in two separate ways. First, patient care is directly affected. Increased frequency of vital signs would allow for better re-evaluation of a patient's condition and a given therapy's effect. Conversely, failure to acknowledge abnormal discharge vital signs has been correlated to worse outcomes in both adult and pediatric patient populations. Secondly, ED operations are also affected. Vital signs are often a key component of disposition. Patients with abnormal vital signs are often held in the ED until their vital signs improve. Patients with more abnormal vital signs may similarly be held for ICU evaluation instead of floor-level admission. Delayed vital signs thus impede patient disposition and increase ED crowding. Of note, ED crowding has also been shown to delay vital sign acquisition, itself.

Presage Technologies has developed software which can acquire heart rate and respiratory rate non-invasively using commonly available cameras. This technology can theoretically allow for decreased time to vital sign acquisition on patient arrival, increased frequency of vital sign reassessment during patient stays, increased patient safety as the device is non-contact, and decreased time to vital sign acquisition at time of discharge. It may also allow for increased vital sign acquisition for agitated and other potentially dangerous patient encounters. These benefits could theoretically allow for better single-patient care and ED operations, ultimately improving patient outcomes.

Presage's SmartSpectra Vital Signs Monitor is a software only medical device ("SaMD") which uses optical camera-based measurements for estimation of pulse rate and breathing rate. The SmartSpectra device includes a preprocessing client that is a software development kit ("SDK") installed into a host software application and includes algorithms to analyze video signals and output pulse rate and breathing rate measurement results.

The SmartSpectra Vital Signs Monitor is intended for non-invasive spot measurement of pulse rate and breathing rate for humans aged 18 and older of all skin types, who do not require critical care or continuous vital signs monitoring. The device assesses imaging of an individual's face and chest wall movements after installation on an approved, statically mounted mobile platform - either the Apple iPhone 16 Pro or a Samsung Galaxy S24e.

The SmartSpectra device is intended for prescription only use ("POU") in hospitals, general care, and secured environments, from a statically mounted position installed by approved Presage Technologies personnel, where a single human face is visible in the frame of the camera and where a framework exists for periodic checks by a trained professional to ensure subject safety. It is not intended for use in an at-home environment. The SmartSpectra device is intended for use by appropriately trained individuals and should not be used by untrained users.

The SmartSpectra device is not intended for use in transportation or domestic (at-home) environments. The device is not intended to be the sole method of checking the physical health of a subject. Additionally, it is not meant for continuous monitoring or for over-the-counter use.

CLINICAL STUDY The objective of the study was to evaluate the safety and effectiveness of the SmartSpectra Vital Signs Monitor version 1.0.0 vis-a-vis standard-of-care, FDA-cleared Philips Intellivue MX450 patient monitor device (FDA Clearance K130849)("MX450"). The MX450 measures end-tidal CO2 ("EtCO2") and three-lead electrocardiogram ("ECG") waveforms.

This study compared the SmartSpectra Vital Signs Monitor pulse rate and breath rate acquired from two mobile devices (iOS and Android systems), as well as from one laptop + USB camera, against the MX450 pulse rate and breath rate data.

The FDA approved MX450 measures heart rate and respiratory rate (clinician scored electrocardiogram and capnography, respectively) within a 30 second spot check period. A concurrent 30 second spot check period was used for each of the iOS mobile device, Android mobile device and MX450 device. A third party validation of rate measurements will provide data for FDA to use in evaluation of SmartSpectra for clearance as a class II SaMD.

The secondary objective of this study was to ensure that the safety, effectiveness, and performance of the SmartSpectra device was consistent across demographic subgroups of skin color, age, sex, body mass index ("BMI"), and device types.

Study Design A study design was chosen to directly compare algorithm-estimated vital signs from SmartSpectra with concurrently collected clinician-scored reference data under standardized conditions, reducing inter-subject variability. By using each subject as their own control, the design maximizes statistical power and minimizes confounding variables without requiring external control groups.

The ECG and etCO2 systems serve as clinical gold standards for heart rate and respiration rate, respectively. No placebo or inactive comparator is included, as the goal is performance validation against these accepted references rather than clinical outcome measurement. The stratified subgroup analysis evaluated any performance degradation or demographic bias in a real-world, intended use environment.

Selection of Study Subjects Baylor College of Medicine ("BCM") conducted a non-randomized enrollment to ensure capture of statistically significant sample sizes of target subgroup population. All study related activities took place in a clinical room at Harris County Hospital District ("HCHD") Ben Taub Emergency Department. Initial enrollment criteria was based on convenience sampling in order to target the observable demographic subgroups. Therefore, relatively healthy patients, visitors, or employees located in the Ben Taub Emergency Department of HCHD were requested to participate voluntarily. Potential subjects were screened based on exclusion/inclusion criteria.

Subject decision making capacity was determined by the research staff approaching the subject. No one was approached for participation while undergoing medical treatment, and a subject must be oriented to person, place, time to participate. Other than decision making capacity, inclusion criteria was patients, visitors or employees greater than 18 years of age and less than 75 years old.

Exclusion criteria, other than lack of decision making capacity, were patients requiring immediate medical care, persons with facial scarring, age less than 18 years old, age greater than 75 years old, employees of Presage Technologies, self-reported pregnancy and anyone who does not pass the screening questionnaire.

Removal of Subjects from Study It was intended that subjects requiring immediate medical attention or those withdrawing consent would be removed from the study. Zero persons who were assigned a unique identifier number were removed from the study due to requiring immediate medical attention or upon withdrawing consent. No subjects required medical treatment in the course of the study.

Data from subject 56 was excluded after review. Data appears to have been recorded for this patient twice, with differing baseline vital signs between the measurements. The data recorded in this row is a mixture of the two measurements and appears to be a clerical error. The cause of this is unknown.

Study Methodology This study is a prospective, non-randomized, within-subject observational evaluation of a contactless vital signs monitoring software (Presage Technologies SmartSpectra Vital Sign Monitor). The software under investigation was used in an iPhone, an Android as well as a laptop with a USB webcam (described interchangeably as C++) to estimate pulse rate and respiratory rate. These estimates are compared to reference standards: electrocardiogram (ECG) for pulse and capnography (etCO2) for respiration; the reference standards measured by a MX450 device.

Each subject underwent a protocol-defined measurement session lasting approximately 10 minutes in a clinical setting. The study assessed accuracy, precision, and failure modes across a demographically diverse sample, stratified by age, sex, skin tone phenotype (Fitzpatrick scale), and Body Mass Index ("BMI"). Data collection took place in a controlled clinical setting and involves no investigational drug or invasive procedure. Subjects serve as their own control in all comparisons.

Being tested
SmartSpectra Vital Signs Monitor measurement
Keywords
rPPG, remote photoplethysmography, PPG from video signal, No contact measurement respiration rate and pulse rate, Vital signs measurement, Respiration rate, Pulse rate, heart rate

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • SmartSpectra Vital Signs Monitor measurementDevice

    The objective of the study was to evaluate the safety and effectiveness of the SmartSpectra Vital Signs Monitor version 1.0.0 vis-a-vis standard-of-care, FDA-cleared Philips Intellivue MX450 patient monitor device (FDA Clearance K130849)("MX450"). The MX450 measures end-tidal CO2 ("EtCO2") and three-lead electrocardiogram ("ECG") waveforms.

Study groups in NCT07362641
Study groupRoleWhat they receive
Relatively healthy patients, visitors, or employees located in a hospital ERPotential subjects were screened based on exclusion/inclusion criteria. Subject decision making capacity was determined by the research staff approaching the subject. No one was approached for participation while undergoing medical treatment, and a subject must be oriented to person, place, time to participate. Other than decision making capacity, inclusion criteria was patients, visitors or employees greater than 18 years of age and less than 75 years old. Exclusion criteria, other than lack of decision making capacity, were patients requiring immediate medical care, persons with facial scarring, age less than 18 years old, age greater than 75 years old, employees of Presage Technologies, self-reported pregnancy and anyone who does not pass the screening questionnaire.
Receives: Device: SmartSpectra Vital Signs Monitor measurement

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Difference in Pulse Rate Measurements (Beats Per Minute) Between SmartSpectra Vital Sign Monitor App and the Standard-of-Care Philips Patient Monitor

    One co-primary endpoint was evaluation of the accuracy and robustness of pulse rate measurements from the SmartSpectra Vital Sign Monitor app versus a standard-of-care, FDA cleared Philips Intellivue MX450 patient monitor. That is, the Outcome Measure is the difference between the SmartSpectra Vital Sign Monitor App and the standard-of-care Philips patient monitor. The pulse rate from the Philips patient monitor is the 'actual' pulse rate and the pulse rate from the SmartSpectra App is the 'measured' pulse rate and the Outcome Measure is, therefore, the difference between the actual and measured pulse rate. Pulse rate is reported as beats per minute.

    Measured: From enrollment to completion of pulse rate measurement was between 15 and 30 minutes.

  2. Difference in Breathing Rate Measurements (Breaths Per Minute) Between SmartSpectra Vital Sign Monitor App and the Standard-of-Care Philips Patient Monitor

    One co-primary endpoint was evaluation of the accuracy and robustness of breathing rate measurements from the SmartSpectra Vital Sign Monitor app versus a standard-of-care, FDA cleared Philips Intellivue MX450 patient monitor. That is, the Outcome Measure is the difference between the SmartSpectra Vital Sign Monitor App and the standard-of-care Philips patient monitor. The breathing rate from the Philips patient monitor is the 'actual' breathing rate and the breathing rate from the SmartSpectra App is the 'measured' breathing rate and the Outcome Measure is, therefore, the difference between the actual and measured breathing rate. This measurement may also be referred to as respiration rate. Breathing rate is reported in units of breaths per minute.

    Measured: From enrollment to completion of breathing rate measurement was between 15 and 30 minutes.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted23 Jan 2026
  2. Study start28 Mar 2025
  3. Primary completion21 Apr 2025
  4. Study completion21 Apr 2025

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Other
Time perspective
Cross-sectional
Enrolment
111 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07362641
Sponsor's protocol ID
H-54364
Other IDs
HQ08452390071
Status last verified
Mar 2026
First submitted
6 Jan 2026
Last updated
24 Aug 2026
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No: Will not be shared due to privacy concerns.

Publications

  • Padaki AS, Zarzour AL, Keene KR, Canepa CA, Levin DR, Antonsen EL. Clinical validation of non-contact vital signs in an emergency department setting. Front Med Technol. 2026 Jan 20;7:1728913. doi: 10.3389/fmedt.2025.1728913. eCollection 2025.

    PubMed 41640807

Questions about this trial

What is this clinical trial studying?

An observational study of Vital Sign Evaluation, run by Presage Technologies, Inc. at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 24 Aug 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 24 Aug 2026). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Difference in Pulse Rate Measurements (Beats Per Minute) Between SmartSpectra Vital Sign Monitor App and the Standard-of-Care Philips Patient Monitor (time frame: From enrollment to completion of pulse rate measurement was between 15 and 30 minutes.). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT07362641, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 24 Aug 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in