Surface Roughness and Thickness Clear Aligner SheatNCT07449598

Evaluation of Physical Changes in Single-Layer and Multi-Layer Clear Aligners

A study testing clear aligner in people with Surface Roughness and Thickness Clear Aligner Sheat, run by Mansoura University at 1 site in Egypt.

Active, not recruiting

Under way, not looking for participants. Everyone needed has enrolled. The study is still running.

Official record
Search all trials

Trial summary

aim of the study: to evaluate and compare the surface roughness (Ra, Rq, Rz), Vickers microhardness (VHN), and thickness of single-layer and multi-layer clear aligner sheets. material and methods: the study will be conducted on clear aligners fabricated from two commercial thermoplastic materials: single-layer (0.76… Read more about this study

Who is it for?
Ages 16 to 24 years
What is being tested?
clear aligner
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
20 participants
Where
1 siteEgypt
Study dates
1 Jan 2026to 1 Feb 2027
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 16 to 24 years
  • All sexes
  • No healthy volunteers
  • Child, Adult

Who may be able to join

  • Class I malocclusion with mild-moderate crowding (≤5 mm per arch).
  • Full permanent dentition (excluding third molars).
  • No restorations, caries, or attachments planned on the labial surface of upper/lower central incisors.

Who cannot join

  • Systemic disease affecting saliva.
  • Parafunctional habits (bruxism, nail-biting).
  • Smoking.
  • Pregnancy.
  • Previous orthodontic treatment within 1 year.

Trial locations and contacts

1 site in 1 country.

Egypt1 site
  • Mansoura university

    Al Mansurah

Other trials in Egypt

About this study

Official title: Evaluation of Physical Changes in Single-Layer and Multi-Layer Clear Aligners After Thermoforming and Intraoral Wear

aim of the study: to evaluate and compare the surface roughness (Ra, Rq, Rz), Vickers microhardness (VHN), and thickness of single-layer and multi-layer clear aligner sheets.

material and methods: the study will be conducted on clear aligners fabricated from two commercial thermoplastic materials: single-layer (0.76 mm, ) and multi-layer (0.76 mm, ). Each material will be evaluated at three stages: before thermoforming, after thermoforming on 3D-printed casts, and after 2 weeks of intraoral wear.

Read more

Specimen Preparation For each group, specimens will be prepared at three stages: before thermoforming (T0), after thermoforming (T1), and after intraoral wear (T2). At T0, square-shaped specimens (approximately 5 × 5 mm) will be cut directly from the unused raw sheets. At T1 and T2, specimens of the same dimensions will be obtained from the mid-labial surface of the central incisor region of the thermoformed aligners, both immediately after thermoforming and after two weeks of intraoral wear. The full thickness of the material will be preserved in all specimens to ensure consistency across groups.

Outcomes:

Primary Outcome:

Surface Roughness (Ra):

Measured at three stages:

T0: Raw, unused sheet specimens (baseline) T1: After thermoforming on dental casts T2: After 2 weeks of intraoral use

Secondary Outcomes:

Hardness:

Measured at T0, T1, and T2. Tested using Vickers hardness method (ISO 6507 compliant).

Read the detailed description

aim of the study: To evaluate and compare the surface roughness (Ra, Rq, Rz), Vickers microhardness (VHN), and thickness of single-layer and multi-layer clear aligner sheets at three stages: before thermoforming, after thermoforming, and following intraoral wear.

material and methods: This in vivo study will be conducted on clear aligners fabricated from two commercial thermoplastic materials: single-layer (0.76 mm, ) and multi-layer (0.76 mm, ). Each material will be evaluated at three stages: before thermoforming (raw sheet), after thermoforming on 3D-printed casts, and after 2 weeks of intraoral wear. At each stage, specimens will be retrieved and tested for changes in mechanical properties, specifically surface roughness and hardness.

Study Design Type: A prospective, randomized, controlled, two-arm parallel clinical study

Design Details: participants will be randomly assigned to one of two groups:

Group A:

Upper arch: Multi-layer clear aligners Lower arch: Multi-layer clear aligners

Group B:

Upper arch: Single-layer clear aligners Lower arch: Single-layer clear aligners Baseline control: Unused aligner sheets of both types to serve as reference for measurements.

Sample Size Calculation: Sample size calculation was calculated using G*Power v3.1.9.7, a paired t-test with an expected effect size dz = 0.65 (based on Bucci et al.), α error = 0.05, and power = 80% indicated a minimum of 19 subjects. To simplify group allocation and ensure adequate power, the sample size was rounded up to 20 patients (10 patients in each group: single-layer vs. multi-layer).

Ethical Considerations:

Informed Consent: All participants (or their legal guardians) will sign an informed.

Appliance Fabrication and Treatment Protocol

  • Digital impressions will be taken for all patients.
  • Treatment setup will be performed with standardized staging: maximum 0.25-0.3 mm linear movement or 2-3° rotation per stage.
  • Aligners will be fabricated by thermoforming using single and multilayer sheets (0.76 mm), cut and finished identically.

Specimen Preparation For each group, specimens will be prepared at three stages: before thermoforming (T0), after thermoforming (T1), and after intraoral wear (T2). At T0, square-shaped specimens (approximately 5 × 5 mm) will be cut directly from the unused raw sheets. At T1 and T2, specimens of the same dimensions will be obtained from the mid-labial surface of the central incisor region of the thermoformed aligners, both immediately after thermoforming and after two weeks of intraoral wear. The full thickness of the material will be preserved in all specimens to ensure consistency across groups.

Outcome Measures:

Primary Outcome:

Surface Roughness (Ra):

Measured at three stages:

T0: Raw, unused sheet specimens (baseline) T1: After thermoforming on dental casts T2: After 2 weeks of intraoral use Assessed using contact profilometry. Roughness defined as the arithmetic average surface deviation (Ra, in µm).

Secondary Outcomes:

Hardness:

Measured at T0, T1, and T2. Tested using Vickers hardness method (ISO 6507 compliant). Higher VHN values indicate greater resistance to indentation and deformation.

Thickness Change:

Measured at T0, T1, and T2. Evaluated using a digital micrometer or 3D scanning system. Calculated as the difference in thickness between the original sheet, thermoformed aligners, and post-use aligner

Being tested
clear aligner

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • clear alignerDevice

    Appliance Fabrication and Treatment Protocol

    • Digital impressions will be taken for all patients.
    • Treatment setup will be performed with standardized staging: maximum 0.25-0.3 mm linear movement or 2-3° rotation per stage.
    • Aligners will be fabricated by thermoforming using single and multilayer sheets (0.76 mm), cut and finished identically.
Study groups in NCT07449598
Study groupRoleWhat they receive
Multi-layer clear alignersExperimentalclear aligners fabricated from thermoplastic materials: multi-layer (0.76 mm). the material will be evaluated at three stages: before thermoforming (raw sheet), after thermoforming on 3D-printed casts, and after 2 weeks of intraoral wear. At each stage, specimens will be retrieved and tested for changes in mechanical properties, specifically surface roughness and hardness. Baseline control: Unused aligner sheets of both types to serve as reference for measurements.
Receives: Device: clear aligner
Single-layer clear alignersExperimentalclear aligners fabricated from thermoplastic materials: single-layer (0.76 mm) . the material will be evaluated at three stages: before thermoforming (raw sheet), after thermoforming on 3D-printed casts, and after 2 weeks of intraoral wear. At each stage, specimens will be retrieved and tested for changes in mechanical properties, specifically surface roughness and hardness. Baseline control: Unused aligner sheets of both types to serve as reference for measurements.
Receives: Device: clear aligner

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. surface roughness of clear aligner

    evaluate and compare the surface roughness of single-layer and multi-layer clear aligner sheets at three stages: before thermoforming, after thermoforming, and following intraoral wear.

    Quantitative evaluation of the arithmetic mean surface roughness (Ra) using profilometer

    Measured: 2 weeks

Secondary outcomes (2)
  1. microhardness of clear aligner

    evaluate and compare the microhardness of single-layer and multi-layer clear aligner sheets at three stages: before thermoforming, after thermoforming, and following intraoral wear.

    Quantitative evaluation of surface microhardness using the Vickers hardness test.

    Measured: 2 weeks

  2. thickness change of clear aligner

    evaluate and compare the thickness of single-layer and multi-layer clear aligner sheets at three stages: before thermoforming, after thermoforming, and following intraoral wear.

    Thickness will be measured using a calibrated digital caliper with an accuracy of ±0.01 mm

    Measured: 2 weeks

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted4 Mar 2026
  2. Study start1 Jan 2026
  3. Primary completion1 Jan 2027 (estimated)
  4. Study completion1 Feb 2027 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
20 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07449598
Sponsor's protocol ID
MS.25.11.85
Status last verified
Dec 2025
First submitted
15 Feb 2026
Last updated
4 Mar 2026
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

A study testing clear aligner in people with Surface Roughness and Thickness Clear Aligner Sheat, run by Mansoura University at 1 site in Egypt.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: under way, not looking for participants, as of the sponsor's last update on 4 Mar 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Egypt. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

surface roughness of clear aligner (time frame: 2 weeks). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 4 Mar 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in