Type 1 Diabetes MellitisDODGENCT07583238

Diathermy On Diabetes Glucose Monitoring Effectiveness

An observational study of Type 1 Diabetes Mellitis, run by University of Nottingham at 1 site in United Kingdom.

Not yet recruiting

Opening soon. The study is approved but enrolment has not started yet. Sites may already take enquiries.

Sites and contacts
Official record
Search all trials

Trial summary

The goal of this observational study is to investigate if diathermy (a surgical tool that uses electrical energy to control bleeding) has any affect on the accuracy and functioning of continuous glucose monitoring systems in adults and young people with Type 1 Diabetes. Read more about this study

Who is it for?
Ages 4 years and older
Conditions the study is for
Type 1 Diabetes Mellitis
How many people
126 participants
Where
1 siteUnited Kingdom
Study dates
May 2026to Feb 2027
How long will I take part?
1 Day
Phase
Observational
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 4 years and older
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Study population: Patients aged 4 or older with type 1 diabetes mellitus scheduled to undergo elective surgery at Nottingham Universities Hospitals NHS Trust.

Who may be able to join

  • Diagnosis of type 1 diabetes mellitus.
  • Planned admission for elective surgery at Nottingham University Hospitals NHS Trust.
  • Currently routinely using a prescribed Abbott Libre-2, Abbott Libre-3, Dexcom G6 or Dexcom G7 glucose monitoring system to monitor glucose concentrations.
  • Aged 4 years or older on the day of surgery.
  • Ability to give informed consent / ability of parent/carer to give informed consent (as appropriate).

Who cannot join

  • Use of paracetamol above maximum dose within 7 days before the scheduled date of surgery (in adults with body weight under 51kg - maximum of 60mg/kg per day, in adults with body weight 51kg and above - maximum of 4g per day, in children - maximum dose as per their age as stated in the British National Formulary for Children)
  • Any use of hydroxyurea within 7 days before the scheduled date of surgery
  • Use of more than 500mg per day of ascorbic acid / vitamin C within 7 days before the scheduled date of surgery
  • Currently taking medications as part of a clinical trial.

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

United Kingdom1 site
  • Queen's Medical Centre

    Nottingham

    James Armstrong, BMBS, 0115 924 9924, james.armstrong8@nhs.net

    James M Law, PhD, BMBS, james.law@nottingham.ac.uk

Other trials in United Kingdom

About this study

Official title: Assessing the Impact of Intraoperative Diathermy on Accuracy and Functioning of Glucose Monitoring Systems

The goal of this observational study is to investigate if diathermy (a surgical tool that uses electrical energy to control bleeding) has any affect on the accuracy and functioning of continuous glucose monitoring systems in adults and young people with Type 1 Diabetes. The main questions this study aims to answer is -

  • Does the accuracy of continuous glucose monitoring systems change after use of diathermy?
Read more

Participants will:

  • Have their height and weight checked.
  • Provide information about their medical history including type of diabetes, other medical conditions and any current medications they take.
  • Have paired glucose meter and sensor glucose measurements taken every 15-75 minutes from up to 4 hours before surgery until up to 4 hours after the end of surgery.
  • Have two blood samples taken to measure glucose levels, The first one will be before the use of diathermy and the second will be after the use of diathermy.
Read the detailed description

Background and rationale:

The incidence of type 1 diabetes is rising and technology is an increasingly important part of monitoring and responding to glucose levels. Contemporary monitoring of glucose levels using continuous glucose monitoring (CGM) systems provide users with information about their glucose levels in real time without finger-prick capillary glucose testing. Crucially, these new systems provide information about the trend in blood glucose levels and is now standard in type 1 diabetes. Use of CGM systems is also needed for patients who use "closed loop" systems which adjust insulin at, typically, 5-minute intervals in response to glucose levels. Use of these strategies in glucose level monitoring and management improves long-term outcomes for persons living with diabetes and is now standard of care in type 1 diabetes. Their importance during inpatient admissions and surgery is increasingly recognised but data is limited.

During surgery, diathermy is often used to control bleeding by using high-frequency electrical energy to seal small blood vessels. Available literature supports the hypothesis that CGM systems may not be affected by diathermy, but there is limited information on the actual impact of diathermy on the accuracy of CGM systems. At present, CGM systems warranties are invalidated by diathermy use. Consequently, patients undergoing surgery require a period of blood glucose testing to monitor and respond to glucose levels which are often rendered unstable in the period around surgery. This exposes patients to potentially unnecessary interventions and, for those patients who are needle-phobic or have learning difficulties, the need to renew repeated finger-prick capillary blood testing around surgery may be a significant barrier to effective and optimal glucose monitoring and control. In addition, there may be a delay in responding to hypoglycaemia and hyperglycaemia, due to the time between measurements.

This study aims to assess the impact of intraoperative diathermy use on the accuracy and functioning of CGM systems. Demonstrating that glucose monitoring systems could be effectively used during surgery would:

  • Prevent/reduce the need for the anaesthetic team to access the patients' fingers/toes and venous/arterial lines for blood glucose testing during surgery.
  • Provide easy, convenient and accurate glucose monitoring to the anaesthetic team, with data transmitted from monitoring systems which can remain underneath surgical coverings.
  • Provide prompt, real-time information on glucose trends during surgery, leading to better control of glucose levels including earlier recognition of low and high glucose levels (hypoglycaemia and hyperglycaemia, respectively).

Study design:

This is a single-centre, prospective, single-arm, observational, non-inferiority study of adults and young people with diabetes using CGM systems. 126 participants will be recruited. Sample size was calculated using simulation.

Participants will undergo paired glucometer and sensor glucose measurements every 15-75 minutes from up to 4 hours before the planned time of surgery until up to 4 hours after the completion of surgery, for not more than 24 hours. They will also have two venous blood samples taken for glucose concentration analysis.

Patient recruitment:

Participants will be recruited from patients due to have an elective surgical procedure at Nottingham University Hospital NHS Trust. The initial approach may be from a member of the patient's usual care team with consent to share contact details with the study team. In addition, the Trust's Theatre Management system (Nervecentre) will be reviewed to identify patients with type 1 diabetes who are due to attend for surgery.

Data management and analysis:

Data will be collected by members of the research team using study specific case report forms (CRFs). An ANOVA mixed effects model will be used with participant as a random effect and pre/post diathermy as a fixed effect. An equivalence F-test with an ARD equivalence value of 0.05 will be performed to determine effect of diathermy on ARD.

Quality assurance plan:

Monitoring of study data shall include confirmation of informed consent; source data verification; data storage and data transfer procedures; local quality control checks and procedures, back-up and disaster recovery of any local databases and validation of data manipulation. Entries on CRFs will be verified by inspection against the source data. A 10% sample of CRFs will be checked on a regular basis for verification of all entries made. In addition, the subsequent capture of the data on the study database will be checked. Where corrections are required, these will carry a full audit trail and justification. Study data and evidence of monitoring and systems audits will be made available for inspection by the REC as required.

Reporting adverse effects:

Intraoperative glucose monitoring is currently undertaken using blood glucose meters. The same monitoring will be continued through this study and this study involves no deviation from standard care. Sensor glucose measurements are passively recorded from the participant's receiving device and eligible participants would already be routinely using sensors as part of their diabetes care.

Equipment failure issues, where a sensor or blood glucose measurement is not able to be obtained, will be recorded on the CRF. Adverse events as a result of participation within this study are not expected but will be recorded on an Adverse Event Form if any occurs. Clinical deterioration and surgical complications, not related to the study, will be considered expected in terms of classification.

Keywords
Continuous Glucose Monitoring Systems, Diathermy, Type 1 Diabetes Mellitus, Capillary blood glucose, Sensor blood glucose, Peri-operative care, Intra-operative T1DM management, pre-operative, post-operative, intraoperative
Medical subject headings
Fever, Diabetes Mellitus, Type 1

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

Study groups in NCT07583238
Study groupRoleWhat they receive
Patients with T1DM aged 4 or older routinely using CGM systems undergoing elective surgeryPatients (aged 4 or above) with a diagnosis of type 1 diabetes mellitus who routinely use a prescribed Abbott Libre-2, Abbott Libre-3, Dexcom G6 or Dexcom G7 glucose monitoring system to monitor glucose concentrations. Patients will scheduled to have a planned admission for elective surgery.

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. To assess if the Absolute Relative Differences (ARDs) of continuous glucose monitoring systems change after use of diathermy.

    The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. The null hypothesis is that there is no difference (non-inferiority) following diathermy (where equivalence is defined as less than a 0.05 difference in ARD).

    Measured: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

Secondary outcomes (9)
  1. To assess whether ARD changes following diathermy depending on type of diathermy used (bipolar/monopolar).

    The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.

    Measured: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

  2. To assess whether ARD changes following diathermy when the model is adjusted for glucose trend (rate of change).

    The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.

    Measured: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

  3. To assess whether ARD changes following diathermy when the model is adjusted for duration of surgery (minutes).

    The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.

    Measured: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

  4. To assess whether ARD changes following diathermy when the model is adjusted for site of surgery.

    The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.

    Measured: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

  5. To assess whether ARD changes following diathermy according to participant characteristic of age (years).

    The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.

    Measured: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

  6. To assess whether ARD changes following diathermy according to participant characteristic of weight (kg).

    The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.

    Measured: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

  7. To assess whether ARD changes following diathermy according to participant characteristic of BMI (or BMI SDS in under 18s).

    The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (glucometer) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.

    Measured: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

  8. To assess sensor accuracy before and after diathermy using error grid analysis.

    Error grid analysis will be used to assess the effect of CGM systems accuracy after diathermy use on clinical decision making.

    Measured: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

  9. To assess whether the ARD of each system changes after use of diathermy using laboratory glucose concentration results as the reference readings.

    The ARD is the absolute difference between the sensor reading (mmol/L) and the reference (laboratory glucose concentration) glucose reading (mmol/L) divided by the reference reading. Equivalence is defined as less than a 0.05 difference in ARD.

    Measured: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted13 May 2026
  2. Study startMay 2026 (estimated)
  3. Primary completionFeb 2027 (estimated)
  4. Study completionFeb 2027 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
126 participants (estimated)
Target duration
1 Day
IDs, dates and publications
ClinicalTrials.gov ID
NCT07583238
Sponsor's protocol ID
25018
Acronym
DODGE
Status last verified
Mar 2026
First submitted
13 Apr 2026
Last updated
13 May 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • Using technology to support diabetes care in hospital (2024). Joint British Diabetes Societies for Inpatient Care (JBDS-IP) Group and Diabetes Technology Network (DTN). Available at: https://abcd.care/sites/default/files/resources/JBDS_20_Using_Technology_to_Support_Diabetes_Care_in_Hospital_1.pdf (Accessed: 16 May 2024).

  • Type 1 diabetes in adults: QS208 (2023). National Institute for Health and Care Excellence. Available at: https://www.nice.org.uk/guidance/qs208.

  • Parkes JL, Slatin SL, Pardo S, Ginsberg BH. A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care. 2000 Aug;23(8):1143-8. doi: 10.2337/diacare.23.8.1143.

    PubMed 10937512
  • Kovatchev BP, Gonder-Frederick LA, Cox DJ, Clarke WL. Evaluating the accuracy of continuous glucose-monitoring sensors: continuous glucose-error grid analysis illustrated by TheraSense Freestyle Navigator data. Diabetes Care. 2004 Aug;27(8):1922-8. doi: 10.2337/diacare.27.8.1922.

    PubMed 15277418
  • Hybrid closed loop systems for managing blood glucose levels in type 1 diabetes. London: National Institute for Health and Care Excellence (NICE); 2023 Dec 19. No abstract available. Available from http://www.ncbi.nlm.nih.gov/books/NBK613531/

    PubMed 40245209
  • Gregory GA, Robinson TIG, Linklater SE, Wang F, Colagiuri S, de Beaufort C, Donaghue KC; International Diabetes Federation Diabetes Atlas Type 1 Diabetes in Adults Special Interest Group; Magliano DJ, Maniam J, Orchard TJ, Rai P, Ogle GD. Global incidence, prevalence, and mortality of type 1 diabetes in 2021 with projection to 2040: a modelling study. Lancet Diabetes Endocrinol. 2022 Oct;10(10):741-760. doi: 10.1016/S2213-8587(22)00218-2. Epub 2022 Sep 13.

    PubMed 36113507
  • Campbell H, Lakens D. Can we disregard the whole model? Omnibus non-inferiority testing for R2 in multi-variable linear regression and etâ2 in ANOVA. Br J Math Stat Psychol. 2021 Feb;74(1):64-89. doi: 10.1111/bmsp.12201. Epub 2020 Feb 13.

    PubMed 32056209
  • Krutkyte G, Rolfes ED, Herzig D, Guensch DP, Schweizer T, Wuethrich PY, Beldi G, Vogt AP, Bally L. Performance of the DEXCOM G7 CGM system during and after major surgery. Diabetes Obes Metab. 2025 Aug;27(8):4586-4589. doi: 10.1111/dom.16462. Epub 2025 May 21. No abstract available.

    PubMed 40400093
  • Avari P, Lumb A, Flanagan D, Rayman G, Misra S, Dhatariya K, Choudhary P. Continuous Glucose Monitoring Within Hospital: A Scoping Review and Summary of Guidelines From the Joint British Diabetes Societies for Inpatient Care. J Diabetes Sci Technol. 2023 May;17(3):611-624. doi: 10.1177/19322968221137338. Epub 2022 Nov 28.

    PubMed 36444418

Questions about this trial

What is this clinical trial studying?

An observational study of Type 1 Diabetes Mellitis, run by University of Nottingham at 1 site in United Kingdom.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: opening soon, as of the sponsor's last update on 13 May 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Each site lists its own contact: choose the nearest one under Trial locations. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

To assess if the Absolute Relative Differences (ARDs) of continuous glucose monitoring systems change after use of diathermy. (time frame: From up to 4 hours prior to the start of surgery to 4 hours after the end of surgery, for no more than a total of 24 hours.). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT07583238, which also lists 9 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 13 May 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in