Dental Implant MacrodesignNCT07744776
Effect of Implant Macro-Design and Surface Treatment on Insertion Torque and Early Implant Stability Quotient (ISQ)
A study testing Chairside implant surface treatment with Plasma X Motion device, SLA implant surface and Straumann BLT implant macro design in people with Dental Implant Macrodesign and Dental Implant Surface Treatment, run by Dantu implantologijos centras at 1 site in Lithuania.
Trial summary
This prospective, randomized pilot clinical trial will evaluate the effect of dental implant macro-design and surface treatment on insertion torque values (ITV) and early implant stability quotient (ISQ) trajectories in healed posterior mandibular bone. Read more about this study
- Who is it for?
- Ages 18 years and olderHealthy volunteers can take part
- Conditions the study is for
- Dental Implant Macrodesign, Dental Implant Surface Treatment, Dental Implant Primary Stability
- What is being tested?
- Chairside implant surface treatment with Plasma X Motion device, SLA implant surface and Straumann BLT implant macro design
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 50 participants
- Where
- 1 siteLithuania
- Study dates
- 10 Aug 2026to 30 Nov 2026
- Phase
- Interventional
- Are there results?
- Not yet: the study is under way
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- Healthy volunteers accepted
- Adult, Older adult
Who may be able to join
- Age ≥ 18 years
- Requiring a dental implant in a healed mandibular molar site (≥ 3 months post-extraction)
- Site anatomy suitable for placement of a 4.1 × 10 mm implant
- Bleeding on probing < 20%
- Plaque index < 25%
Who cannot join
- Smoking ≥ 10 cigarettes/day
- History of uncontrolled periodontitis
- Uncontrolled diabetes
- Alcoholism
- Use of medication known to impair bone healing
Trial locations and contacts
1 site in 1 country. Contact the study team or a site near you to ask about taking part.
Lithuania1 site, 1 recruiting
- Recruiting
Dantu implantologijos centras
Vilnius, 08216
Marius Svedas, 069829987, svedasm@gmail.com
About this study
Official title: Effect of Implant Macro-Design and Surface Treatment on Insertion Torque and Early Implant Stability Quotient (ISQ): A Pilot Randomized Clinical Trial
This prospective, randomized pilot clinical trial will evaluate the effect of dental implant macro-design and surface treatment on insertion torque values (ITV) and early implant stability quotient (ISQ) trajectories in healed posterior mandibular bone. The study has been approved by the Vilnius University Regional Bioethics Committee (approval no. 2025/11-1722-1165) and will be conducted at a private dental clinic (Dantų implantologijos centras, Vilnius, Lithuania).
Fifty implants will be randomly allocated to five groups of ten implants each, comparing two implant macro-designs - Straumann Bone Level Tapered (BLT) and Megagen BlueDiamond (BD) - combined with different surface treatments: (1) BLT with a sandblasted, large-grit, acid-etched (SLA®) surface plus chairside vacuum plasma activation (XActive®); (2) BLT with a manufacturer-hydrophilic SLActive® surface; (3) BD with a nanostructured calcium-incorporated Xpeed® surface plus chairside plasma activation; (4) BD with the standard Xpeed® surface (no plasma activation); and (5) BLT with a standard SLA surface (control, no plasma activation). Randomization will be performed by drawing a sealed envelope indicating group assignment immediately after osteotomy preparation and before implant placement.
Read more
Eligible participants will be adults aged 18 years or older requiring a single dental implant in a healed mandibular molar site (at least 3 months post-extraction) suitable for a 4.1 × 10 mm implant.
Insertion torque will be recorded in Newton centimeters (Ncm) using a prosthetic torque wrench at the time of implant placement. Implant stability will be measured non-invasively by resonance frequency analysis (RFA) using a SmartPeg, with ISQ recorded at four timepoints: at implant placement (baseline) and at 2, 4, and 6 weeks after placement. The primary comparisons will be (1) insertion torque according to implant macro-design (BLT vs. BD) and (2) the trajectory of ISQ change from baseline through 6 weeks according to surface treatment/plasma activation status. Correlation between insertion torque and ISQ will also be assessed. Also marginal bone maintenance (loss) will be documented from periapical radiographs obtained during prosthetic and 1 year follow-up visits.
As a pilot study, this trial is designed to assess feasibility, estimate effect sizes, and inform sample-size calculations for a future, adequately powered randomized controlled trial comparing implant macro-design and surface activation strategies for early implant stability.
Read the detailed description
Background and Rationale Dental implants play a central role in the replacement of missing teeth, and defining the earliest point at which an implant can be safely loaded functionally remains a key unresolved issue in implant dentistry. Minimizing the time to functional and aesthetic rehabilitation may improve patient satisfaction and reduce the likelihood of migration of neighboring teeth into an edentulous site. Consequently, strategies that accelerate osseointegration remain of substantial clinical interest.
Implant stability can be divided into primary stability - mechanical fixation of the implant immediately after placement - and secondary stability, which reflects the transition to biological integration through new bone apposition on the implant surface. Two non-invasive parameters are commonly used to guide decisions about loading protocols: insertion torque value (ITV), measured in Newton centimeters (Ncm), which reflects the mechanical friction between the implant and surrounding bone at placement, and implant stability quotient (ISQ), obtained through resonance frequency analysis (RFA), which indirectly reflects the density and extent of bone-to-implant contact. ITV and ISQ may correlate at the time of implant placement, and both are considered useful indicators for decisions regarding loading timing.
Primary implant stability is influenced by bone density at the implant site, osteotomy preparation technique, and implant macrogeometry. Tapered implant designs, such as the Straumann Bone Level Tapered (BLT) and Megagen BlueDiamond (BD) systems, generally achieve higher insertion torque than cylindrical designs, which is particularly relevant in low-density bone. Although both BLT and BD implants are tapered, they differ in thread depth: the BD design features deeper threads that increase implant surface area and mechanical engagement with bone, which may increase insertion torque.
Secondary stability is influenced by the same local factors as primary stability, as well as by oral hygiene, infection control, mechanical loading protocol, and patient-related biological factors affecting bone metabolism (e.g., overall health, age, smoking status). A further determinant of secondary stability is implant surface treatment, which affects the rate of protein adsorption and osteoblast adhesion and therefore the onset of osseointegration. Additionally, titanium implant surfaces are subject to biological aging through hydrocarbon deposition, which reduces surface hydrophilicity and biological activity and may slow osseointegration. Implant macro-design and surface treatment are among the few variables in implant therapy that remain fully under clinician control.
With early machined implant surfaces, healing times of 3 months (mandible) and 6 months (maxilla) prior to prosthetic loading were historically recommended. The introduction of surface treatments such as sandblasted, large-grit, acid-etched (SLA) surfaces shortened proposed healing times to 6-12 weeks. A limitation of conventional SLA surfaces is their hydrophobicity. SLActive is a chemically modified, hydrophilic variant of the SLA surface: following sandblasting and acid-etching, implants are stored in an isotonic saline solution under nitrogen to prevent hydrocarbon contamination and to preserve a highly hydrophilic titanium oxide surface, which is intended to enhance early biological response and accelerate osseointegration. Clinical and histological studies have reported that SLActive implants outperform conventional SLA surfaces in early bone apposition, bone-to-implant contact, and early ISQ values.
XActive is a low-temperature vacuum plasma surface activation treatment (Plasma X Motion system, MegaGen Implant Co., Ltd.) applied chairside immediately prior to implant placement. Unlike manufacturer-applied hydrophilic treatments such as SLActive, this activation is performed at the point of care rather than during manufacturing, and is intended to reduce implant surface hydrocarbons and increase surface energy, thereby improving surface hydrophilicity and early implant stability. Because plasma activation is applied chairside rather than integrated into the manufacturing process, it can in principle be used with implant systems from any manufacturer, and has been shown in vitro to enhance protein adsorption and osteoblast response.
Separately, the nanostructured calcium-incorporated Xpeed® surface (used on Megagen BlueDiamond implants) has been reported to accelerate early bone formation compared with SLA surfaces in preclinical and early clinical studies, independent of plasma activation.
To date, no human randomized trial has directly compared XActive chairside plasma activation with manufacturer-applied SLActive treatment, nor has any trial investigated the combined effect of implant macro-design (Straumann BLT vs. Megagen BD) and surface treatment/activation status on ITV and early ISQ trajectories. This pilot randomized controlled trial is designed to address this gap by evaluating five implant macro-design and surface treatment combinations placed in healed posterior mandibular bone.
Study Objectives
The primary objectives of this pilot trial are:
To compare insertion torque values (ITV) between two implant macro-designs - Straumann Bone Level Tapered (BLT) and Megagen BlueDiamond (BD) - placed in healed posterior mandibular bone.
To compare early ISQ trajectories (from implant placement through 6 weeks) among five implant macro-design and surface treatment/activation combinations.
To explore the correlation between insertion torque and ISQ values at the time of implant placement.
As a pilot study, to estimate effect sizes and variability to inform the sample size calculation for a future, adequately powered randomized controlled trial.
Null Hypothesis There will be no statistically significant differences in early ISQ trajectories or insertion torque values among the five implant macro-design and surface treatment combinations placed in healed posterior mandibular bone.
Study Design This will be a prospective, randomized, pilot clinical trial with five parallel arms, conducted at a single private dental clinic (Dantų implantologijos centras, Kernavės g. 88, Vilnius, LT-08216, Lithuania). The protocol has been approved by the Vilnius University Regional Bioethics Committee (approval no. 2025/11-1722-1165).
A total of 50 implants will be enrolled, with 10 implants allocated to each of five treatment groups. This sample size is consistent with methodological recommendations for pilot studies, which suggest that small sample sizes are appropriate for feasibility assessment and estimation of variability ahead of a definitively powered trial.
Randomization and Allocation Implants will be randomly allocated to one of five groups (n=10 per group) using a sealed-envelope method. The envelope containing the group assignment will be drawn immediately after osteotomy preparation and immediately prior to implant placement, in order to standardize the timing of randomization relative to the surgical sequence.
Study Groups
Group 1 - BLT SLA XActive: Straumann Bone Level Tapered implant with a conventional SLA surface, activated chairside with the XActive plasma treatment immediately before placement.
Group 2 - BLT SLActive: Straumann Bone Level Tapered implant with the manufacturer-applied hydrophilic SLActive surface (no chairside plasma activation).
Group 3 - BD XpeedActive: Megagen BlueDiamond implant with the nanostructured calcium-incorporated Xpeed® surface, activated chairside with the XActive plasma treatment immediately before placement.
Group 4 - BD Xpeed: Megagen BlueDiamond implant with the standard Xpeed® surface (no plasma activation).
Group 5 - BLT SLA (control): Straumann Bone Level Tapered implant with the conventional SLA surface (no plasma activation).
This design allows comparison of implant macro-design independent of surface treatment (BLT vs. BD, pooling across surface/activation status) and comparison of surface treatment/activation status within each macro-design (activated vs. non-activated in both the BLT and BD systems), as well as a direct comparison of chairside plasma activation (XActive) against manufacturer-integrated hydrophilic treatment (SLActive).
Plasma Activation Procedure For implants assigned to plasma activation, the implant will be retrieved from its sterile container using the handpiece implant driver. The handpiece with the implant attached will be placed in the magnetic holder of the Plasma X Motion device, and the activation cycle will be initiated according to the manufacturer's instructions, completing automatically within approximately 50 seconds. Immediately following completion of the cycle, the magnetic holder will be removed from the device and the implant will be inserted into the prepared osteotomy according to the implant manufacturer's surgical protocol, without delay.
Surgical Procedure Two grams of amoxicillin (Ospamox, Sandoz GmbH, Austria) will be administered as a single preoperative dose; no antibiotics will be prescribed postoperatively. All surgeries will be performed under local anesthesia by the same surgeon to minimize operator variability. A mid-crestal incision will be made and a buccal mucoperiosteal flap elevated. Vertical soft tissue thickness will be measured with a periodontal probe applied to the unelevated lingual flap, after which the lingual flap will be elevated. Osteotomy preparation will follow the standard protocol recommended by the respective implant manufacturer, followed by implant insertion positioned 4 mm apical to the gingival margin. A healing abutment will be placed and sutures applied. Postoperative 0.12% chlorhexidine mouthwash will be prescribed; no postoperative antibiotics will be given.
Statistical Analysis Plan Statistical analysis will be performed using R (v. 4.0.4). Between-group comparisons of quantitative variables for two groups will use Student's t-test or the non-parametric Mann-Whitney U test, as appropriate. Comparisons among more than two groups will use one-way analysis of variance (ANOVA) or the non-parametric Kruskal-Wallis test, as appropriate. Normality of distributions will be assessed using the Shapiro-Wilk test. Paired comparisons (baseline vs. follow-up timepoints within the same implant) will use the paired t-test or Wilcoxon signed-rank test, as appropriate. Associations between insertion torque and ISQ will be assessed using Spearman correlation. Descriptive statistics will include mean, standard deviation, quartiles (Q1, Q3), median, and available number of observations for continuous variables, and absolute counts and percentages for categorical variables. A p-value <0.05 will be considered statistically significant.
Sample Size Justification As a pilot study, the planned enrollment of 10 implants per group (50 implants total) is intended to assess feasibility and estimate effect sizes and variability rather than to achieve definitive statistical power for between-group comparisons, consistent with established pilot study methodology. Data generated by this pilot will be used to perform a formal power analysis to determine the sample size required for a future, adequately powered confirmatory trial.
Significance This trial is intended to provide preliminary data on whether implant macro-design (thread geometry) and surface activation strategy (chairside plasma activation vs. manufacturer-integrated hydrophilic treatment) meaningfully influence early implant stability dynamics in healed, high-quality mandibular bone. Because chairside plasma activation systems such as XActive can, in principle, be applied to implants from any manufacturer, a favorable comparison with proprietary hydrophilic surfaces such as SLActive could have practical implications for clinical protocol selection and implant loading timelines. Findings from this pilot are also intended to inform the design, eligibility criteria, and sample size of a future confirmatory trial, potentially including sites with lower primary stability (e.g., softer bone or immediate implant placement), where surface-treatment effects on ISQ dynamics may be more pronounced.
- Conditions
- Dental Implant Macrodesign, Dental Implant Surface Treatment, Dental Implant Primary Stability, Dental Implant Secondary Stability, Dental Implant Stability Quotient (ISQ), Effects of Dental Implant Macrodesign and Surface Treatment on Primary and Secondary Stability
- Being tested
- Chairside implant surface treatment with Plasma X Motion device, SLA implant surface, Straumann BLT implant macro design, SLActive implant surface treatment, Megagen BD implant macro design
- Keywords
- dental implant macro design, dental implant surface treatment, early implant stability, osseointegration, primary stability, secondary stability, SLActive, Xpeed, Xactive, XpeedActive, SLA, Plasma X motion, ISQ, ITV, insertion torque value, Implant Stability Quotient, Vacuum plasma, Plasma treatment, hydrophilic surface, hydrocarbon
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Chairside implant surface treatment with Plasma X Motion deviceDevice
XActive® chairside plasma surface activation (Plasma X Motion system) applied
SLA implant surfaceDevice
Manufacturer SLA implant surface
Straumann BLT implant macro designDevice
Straumann Bone Level Tapered implant macro design
SLActive implant surface treatmentDevice
Manufacturer SLActive surface treatment
Megagen BD implant macro designDevice
Megagen BlueDiamond implant macro design
Megagen Xpeed implant surface treatmentDevice
Megagen Xpeed manufacturer implant surface treatment
| Study group | Role | What they receive |
|---|---|---|
| Straumann Bone Level Tapered implants SLA with Plasma X Motion treatment (BLT SLA XActive) | Experimental | Participants will receive a Straumann Bone Level Tapered (BLT) implant with a conventional sandblasted, large-grit, acid-etched (SLA) surface. Immediately prior to placement, the implant will undergo chairside low-temperature vacuum plasma surface activation using the Plasma X Motion system (XActive®), performed according to the manufacturer's protocol (approximately 50-second activation cycle) immediately before insertion into the prepared osteotomy. Receives: Device: Chairside implant surface treatment with Plasma X Motion device; Device: SLA implant surface; Device: Straumann BLT implant macro design |
| Straumann Bone Level Tapeded implants with SLActive surface (BLT SLActive) | Experimental | Participants will receive a Straumann Bone Level Tapered (BLT) implant with the manufacturer hydrophilic SLActive surface. No chairside plasma activation will be performed; the hydrophilic, hydrocarbon-free surface state is achieved during manufacturing and preserved through storage in isotonic saline under nitrogen until use Receives: Device: Straumann BLT implant macro design; Device: SLActive implant surface treatment |
| MegaGen BlueDiamond implants Xpeed surface treated using Plasma X Motion (BD XpeedActive) | Experimental | Participants will receive a Megagen BlueDiamond (BD) implant with the nanostructured calcium-incorporated Xpeed® surface. Immediately prior to placement, the implant will undergo chairside low-temperature vacuum plasma surface activation using the Plasma X Motion system (XActive®), performed according to the manufacturer's protocol immediately before insertion into the prepared osteotomy. Receives: Device: Chairside implant surface treatment with Plasma X Motion device; Device: Megagen BD implant macro design; Device: Megagen Xpeed implant surface treatment |
| MegaGen BlueDiamond implants Xpeed surface (BD Xpeed) | Experimental | Participants will receive a Megagen BlueDiamond (BD) implant with the manufacturer nanostructured calcium-incorporated Xpeed® surface, without any chairside plasma activation. Receives: Device: Megagen BD implant macro design; Device: Megagen Xpeed implant surface treatment |
| Straumann Bone Level Tapeded implants with SLActive surface (BLT SLA) | Active comparator | Participants will receive a Straumann Bone Level Tapered (BLT) implant with a conventional sandblasted, large-grit, acid-etched (SLA) surface, without any chairside plasma activation and without a manufacturer-integrated hydrophilic treatment. This arm serves as the reference/control group for surface-treatment comparisons. Receives: Device: SLA implant surface; Device: Straumann BLT implant macro design |
What is being measured
The primary outcome is the main question the study is designed to answer.
Implant Insertion Torque Value (ITV)
Insertion torque will be recorded in Newton-centimeters (Ncm) using a prosthetic torque wrench at the moment of implant seating, prior to healing abutment connection. ITV will be compared between the two implant macro-designs (Straumann Bone Level Tapered vs. Megagen BlueDiamond), pooling across surface treatment/activation status, to evaluate the effect of implant macrogeometry on primary mechanical stability.
Measured: Day 1
Implant Stability Quotient (ISQ)
Implant stability will be measured non-invasively by resonance frequency analysis (RFA) using a SmartPeg transducer attached to the implant's internal connection. Two perpendicular readings (bucco-lingual and mesio-distal) will be obtained at each timepoint using a manufacturer-recommended 45-degree probe angle, and their average will be recorded as the ISQ value for that visit. The change in ISQ (ΔISQ) from baseline will be calculated at each follow-up timepoint (W2, W4, W6) for each of the five study arms, to evaluate the trajectory of early implant stability according to implant macro-design and surface treatment/activation.
Measured: Baseline (day 1) and at 2, 4, and 6 weeks post-placement
Secondary outcomes (4)
Vertical gingival thickness
Measurement will be taken with a milimeter periodontal probe after midcrestal incision and reflection of the buccal full thickness flap
Measured: Day 1
Implant placement depth
Implant will be inserted 4mm deeper than the gingival margin and implant placement depth below the bony crest will be measured with a milimeter periodontal probe right after the implant insertion
Measured: Day 1
Buccal vertical gingival thickness
At the time of impression making, vertical gingival thickness from the neck of the implant to the buccal gingival margin will be measured with a periodontal probe
Measured: 8 weeks
Marginal bone level
Radiographic evaluation of marginal bone level from the periapical radiographs obtained routinely during prosthetic and 1 year follow-up visits. Radiographic evaluation would be performed anyway at the time of these visits for other reasons than the clinical trial.
Measured: At the time of prosthetic rehabilitation of the implant and at the time of the 1 year follow-up visit.
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted4 Aug 2026
- Study start10 Aug 2026
- Primary completion30 Oct 2026 (estimated)
- Study completion30 Nov 2026 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (23)Papers the sponsor has linked to this study.
- Other trials for Dental Implant MacrodesignStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- None (open label)
- Enrolment
- 50 participants (estimated)
Sponsor and investigators
- Lead sponsor
- Dantu implantologijos centras
- Sponsor type
- University, hospital or other
- Collaborators
- MegaGen
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT07744776
- Sponsor's protocol ID
- SLActive vs SLA Xactive
- Status last verified
- Sep 2026
- First submitted
- 13 Jul 2026
- Last updated
- 9 Sep 2026
- Results posted
- No
- Data monitoring committee
- Yes
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- No: Small sample size raising re-identification risk given the specific implant/surface combinations.
Publications
Bashutski JD, D'Silva NJ, Wang HL. Implant compression necrosis: current understanding and case report. J Periodontol. 2009 Apr;80(4):700-4. doi: 10.1902/jop.2009.080581.
PubMed 19335092Makary C, Menhall A, Zammarie C, Lombardi T, Lee SY, Stacchi C, Park KB. Primary Stability Optimization by Using Fixtures with Different Thread Depth According To Bone Density: A Clinical Prospective Study on Early Loaded Implants. Materials (Basel). 2019 Jul 27;12(15):2398. doi: 10.3390/ma12152398.
PubMed 31357620Julious, S.A. (2005) Sample size of 12 per group rule of thumb for a pilot study. Pharmaceutical Statistics, 4 (4). pp. 287-291. ISSN: 1539-1604
Chun HJ, Cheong SY, Han JH, Heo SJ, Chung JP, Rhyu IC, Choi YC, Baik HK, Ku Y, Kim MH. Evaluation of design parameters of osseointegrated dental implants using finite element analysis. J Oral Rehabil. 2002 Jun;29(6):565-74. doi: 10.1046/j.1365-2842.2002.00891.x.
PubMed 12071926Stacchi C, Rapani A, Montanari M, Martini R, Lombardi T. Effect of Vacuum Plasma Activation on Early Implant Stability: a Single-Blind Split-Mouth Randomized Clinical Trial. J Oral Maxillofac Res. 2025 Jun 30;16(2):e4. doi: 10.5037/jomr.2025.16205. eCollection 2025 Apr-Jun.
PubMed 40693123Oates TW, Valderrama P, Bischof M, Nedir R, Jones A, Simpson J, Toutenburg H, Cochran DL. Enhanced implant stability with a chemically modified SLA surface: a randomized pilot study. Int J Oral Maxillofac Implants. 2007 Sep-Oct;22(5):755-60.
PubMed 17974109Menhall A, Lahoud P, Yang KR, Park KB, Razukevicius D, Traini T, Makary C. The Mineral Apposition Rate on Implants with Either a Sandblasted Acid-Etched Implant Surface (SLA) or a Nanostructured Calcium-Incorporated Surface (XPEED(R)): A Histological Split-Mouth, Randomized Case/Control Human Study. Materials (Basel). 2024 Jul 5;17(13):3341. doi: 10.3390/ma17133341.
PubMed 38998421Lozano-Carrascal N, Salomo-Coll O, Gilabert-Cerda M, Farre-Pages N, Gargallo-Albiol J, Hernandez-Alfaro F. Effect of implant macro-design on primary stability: A prospective clinical study. Med Oral Patol Oral Cir Bucal. 2016 Mar 1;21(2):e214-21. doi: 10.4317/medoral.21024.
PubMed 26827067Lee SY, Kim SJ, An HW, Kim HS, Ha DG, Ryo KH, Park KB. The effect of the thread depth on the mechanical properties of the dental implant. J Adv Prosthodont. 2015 Apr;7(2):115-21. doi: 10.4047/jap.2015.7.2.115. Epub 2015 Apr 23.
PubMed 25932309Tissue-integrated prostheses : osseointegration in clinical dentistry by Brånemark, Per-Ingvar; Zarb, George A. (George Albert), 1938-; Albrektsson, Tomas Publication date 1985
Makary C, Menhall A, Lahoud P, Yang KR, Park KB, Razukevicius D, Traini T. Bone-to-Implant Contact in Implants with Plasma-Treated Nanostructured Calcium-Incorporated Surface (XPEEDActive) Compared to Non-Plasma-Treated Implants (XPEED): A Human Histologic Study at 4 Weeks. Materials (Basel). 2024 May 14;17(10):2331. doi: 10.3390/ma17102331.
PubMed 38793397Chhabra K, Rajasekar A. Comparison of Roughness, Wettability, and SEM Features between Sandblasted Acid-Etched and Oxidized Titanium Dental Implants. J Long Term Eff Med Implants. 2024;34(4):57-63. doi: 10.1615/JLongTermEffMedImplants.2023049632.
PubMed 38842233Canullo L, Penarrocha D, Clementini M, Iannello G, Micarelli C. Impact of plasma of argon cleaning treatment on implant abutments in patients with a history of periodontal disease and thin biotype: radiographic results at 24-month follow-up of a RCT. Clin Oral Implants Res. 2015;26(1):8-14. doi: 10.1111/clr.12290. Epub 2013 Nov 6.
PubMed 24191873H H, G W, E H. The clinical significance of implant stability quotient (ISQ) measurements: A literature review. J Oral Biol Craniofac Res. 2020 Oct-Dec;10(4):629-638. doi: 10.1016/j.jobcr.2020.07.004. Epub 2020 Aug 14.
PubMed 32983857Patel R, Patel S, Girgis W, Ahmed W, Barrak F. A systematic assessment of the stability of SLA(R) vs. SLActive(R) implant surfaces over 12 weeks. Evid Based Dent. 2025 Mar;26(1):67-68. doi: 10.1038/s41432-024-01097-1. Epub 2025 Jan 7.
PubMed 39775154
Study at a glance
From the registry record, last updated by the sponsor on 9 Sep 2026. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Effect of Implant Macro-Design and Surface Treatment on Insertion Torque and Early Implant Stability Quotient (ISQ): A Pilot Randomized Clinical Trial
- ClinicalTrials.gov ID
- NCT07744776
- Acronym / sponsor ID
- SLActive vs SLA Xactive
- Conditions
- Dental Implant Macrodesign, Dental Implant Surface Treatment, Dental Implant Primary Stability, Dental Implant Secondary Stability, Dental Implant Stability Quotient (ISQ), Effects of Dental Implant Macrodesign and Surface Treatment on Primary and Secondary Stability
- Keywords
- dental implant macro design, dental implant surface treatment, early implant stability, osseointegration, primary stability, secondary stability, SLActive, Xpeed
- Study type
- Interventional
- Recruitment status
- Recruiting
- Enrolment
- 50 participants (estimated)
- Lead sponsor
- Dantu implantologijos centrasUniversity, hospital or other
- Collaborators
- MegaGen
- Study start
- 10 Aug 2026 (actual)
- Primary completion
- 30 Oct 2026 (estimated)
- Study completion
- 30 Nov 2026 (estimated)
- First posted
- 4 Aug 2026 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
This prospective, randomized pilot clinical trial will evaluate the effect of dental implant macro-design and surface treatment on insertion torque values (ITV) and early implant stability quotient (ISQ) trajectories in healed posterior mandibular bone. The study has been approved by the Vilnius University Regional Bioethics Committee (approval no. 2025/11-1722-1165) and will be conducted at a private dental clinic (Dantų implantologijos centras, Vilnius, Lithuania).
Fifty implants will be randomly allocated to five groups of ten implants each, comparing two implant macro-designs - Straumann Bone Level Tapered (BLT) and Megagen BlueDiamond (BD) - combined with different surface treatments: (1) BLT with a sandblasted, large-grit, acid-etched (SLA®) surface plus chairside vacuum plasma activation (XActive®); (2) BLT with a manufacturer-hydrophilic SLActive® surface; (3) BD with a nanostructured calcium-incorporated Xpeed® surface plus chairside plasma activation; (4) BD with the standard Xpeed® surface (no plasma activation); and (5) BLT with a standard SLA surface (control, no plasma activation). Randomization will be performed by drawing a sealed envelope indicating group assignment immediately after osteotomy preparation and before implant placement.
Eligible participants will be adults aged 18 years or older requiring a single dental implant in a healed mandibular molar site (at least 3 months post-extraction) suitable for a 4.1 × 10 mm implant.
Insertion torque will be recorded in Newton centimeters (Ncm) using a prosthetic torque wrench at the time of implant placement. Implant stability will be measured non-invasively by resonance frequency analysis (RFA) using a SmartPeg, with ISQ recorded at four timepoints: at implant placement (baseline) and at 2, 4, and 6 weeks after placement. The primary comparisons will be (1) insertion torque according to implant macro-design (BLT vs. BD) and (2) the trajectory of ISQ change from baseline through 6 weeks according to surface treatment/plasma activation status. Correlation between insertion torque and ISQ will also be assessed. Also marginal bone maintenance (loss) will be documented from periapical radiographs obtained during prosthetic and 1 year follow-up visits.
As a pilot study, this trial is designed to assess feasibility, estimate effect sizes, and inform sample-size calculations for a future, adequately powered randomized controlled trial comparing implant macro-design and surface activation strategies for early implant stability.
Detailed description
Background and Rationale Dental implants play a central role in the replacement of missing teeth, and defining the earliest point at which an implant can be safely loaded functionally remains a key unresolved issue in implant dentistry. Minimizing the time to functional and aesthetic rehabilitation may improve patient satisfaction and reduce the likelihood of migration of neighboring teeth into an edentulous site. Consequently, strategies that accelerate osseointegration remain of substantial clinical interest.
Implant stability can be divided into primary stability - mechanical fixation of the implant immediately after placement - and secondary stability, which reflects the transition to biological integration through new bone apposition on the implant surface. Two non-invasive parameters are commonly used to guide decisions about loading protocols: insertion torque value (ITV), measured in Newton centimeters (Ncm), which reflects the mechanical friction between the implant and surrounding bone at placement, and implant stability quotient (ISQ), obtained through resonance frequency analysis (RFA), which indirectly reflects the density and extent of bone-to-implant contact. ITV and ISQ may correlate at the time of implant placement, and both are considered useful indicators for decisions regarding loading timing.
Primary implant stability is influenced by bone density at the implant site, osteotomy preparation technique, and implant macrogeometry. Tapered implant designs, such as the Straumann Bone Level Tapered (BLT) and Megagen BlueDiamond (BD) systems, generally achieve higher insertion torque than cylindrical designs, which is particularly relevant in low-density bone. Although both BLT and BD implants are tapered, they differ in thread depth: the BD design features deeper threads that increase implant surface area and mechanical engagement with bone, which may increase insertion torque.
Secondary stability is influenced by the same local factors as primary stability, as well as by oral hygiene, infection control, mechanical loading protocol, and patient-related biological factors affecting bone metabolism (e.g., overall health, age, smoking status). A further determinant of secondary stability is implant surface treatment, which affects the rate of protein adsorption and osteoblast adhesion and therefore the onset of osseointegration. Additionally, titanium implant surfaces are subject to biological aging through hydrocarbon deposition, which reduces surface hydrophilicity and biological activity and may slow osseointegration. Implant macro-design and surface treatment are among the few variables in implant therapy that remain fully under clinician control.
With early machined implant surfaces, healing times of 3 months (mandible) and 6 months (maxilla) prior to prosthetic loading were historically recommended. The introduction of surface treatments such as sandblasted, large-grit, acid-etched (SLA) surfaces shortened proposed healing times to 6-12 weeks. A limitation of conventional SLA surfaces is their hydrophobicity. SLActive is a chemically modified, hydrophilic variant of the SLA surface: following sandblasting and acid-etching, implants are stored in an isotonic saline solution under nitrogen to prevent hydrocarbon contamination and to preserve a highly hydrophilic titanium oxide surface, which is intended to enhance early biological response and accelerate osseointegration. Clinical and histological studies have reported that SLActive implants outperform conventional SLA surfaces in early bone apposition, bone-to-implant contact, and early ISQ values.
XActive is a low-temperature vacuum plasma surface activation treatment (Plasma X Motion system, MegaGen Implant Co., Ltd.) applied chairside immediately prior to implant placement. Unlike manufacturer-applied hydrophilic treatments such as SLActive, this activation is performed at the point of care rather than during manufacturing, and is intended to reduce implant surface hydrocarbons and increase surface energy, thereby improving surface hydrophilicity and early implant stability. Because plasma activation is applied chairside rather than integrated into the manufacturing process, it can in principle be used with implant systems from any manufacturer, and has been shown in vitro to enhance protein adsorption and osteoblast response.
Separately, the nanostructured calcium-incorporated Xpeed® surface (used on Megagen BlueDiamond implants) has been reported to accelerate early bone formation compared with SLA surfaces in preclinical and early clinical studies, independent of plasma activation.
To date, no human randomized trial has directly compared XActive chairside plasma activation with manufacturer-applied SLActive treatment, nor has any trial investigated the combined effect of implant macro-design (Straumann BLT vs. Megagen BD) and surface treatment/activation status on ITV and early ISQ trajectories. This pilot randomized controlled trial is designed to address this gap by evaluating five implant macro-design and surface treatment combinations placed in healed posterior mandibular bone.
Study Objectives
The primary objectives of this pilot trial are:
To compare insertion torque values (ITV) between two implant macro-designs - Straumann Bone Level Tapered (BLT) and Megagen BlueDiamond (BD) - placed in healed posterior mandibular bone.
To compare early ISQ trajectories (from implant placement through 6 weeks) among five implant macro-design and surface treatment/activation combinations.
To explore the correlation between insertion torque and ISQ values at the time of implant placement.
As a pilot study, to estimate effect sizes and variability to inform the sample size calculation for a future, adequately powered randomized controlled trial.
Null Hypothesis There will be no statistically significant differences in early ISQ trajectories or insertion torque values among the five implant macro-design and surface treatment combinations placed in healed posterior mandibular bone.
Study Design This will be a prospective, randomized, pilot clinical trial with five parallel arms, conducted at a single private dental clinic (Dantų implantologijos centras, Kernavės g. 88, Vilnius, LT-08216, Lithuania). The protocol has been approved by the Vilnius University Regional Bioethics Committee (approval no. 2025/11-1722-1165).
A total of 50 implants will be enrolled, with 10 implants allocated to each of five treatment groups. This sample size is consistent with methodological recommendations for pilot studies, which suggest that small sample sizes are appropriate for feasibility assessment and estimation of variability ahead of a definitively powered trial.
Randomization and Allocation Implants will be randomly allocated to one of five groups (n=10 per group) using a sealed-envelope method. The envelope containing the group assignment will be drawn immediately after osteotomy preparation and immediately prior to implant placement, in order to standardize the timing of randomization relative to the surgical sequence.
Study Groups
Group 1 - BLT SLA XActive: Straumann Bone Level Tapered implant with a conventional SLA surface, activated chairside with the XActive plasma treatment immediately before placement.
Group 2 - BLT SLActive: Straumann Bone Level Tapered implant with the manufacturer-applied hydrophilic SLActive surface (no chairside plasma activation).
Group 3 - BD XpeedActive: Megagen BlueDiamond implant with the nanostructured calcium-incorporated Xpeed® surface, activated chairside with the XActive plasma treatment immediately before placement.
Group 4 - BD Xpeed: Megagen BlueDiamond implant with the standard Xpeed® surface (no plasma activation).
Group 5 - BLT SLA (control): Straumann Bone Level Tapered implant with the conventional SLA surface (no plasma activation).
This design allows comparison of implant macro-design independent of surface treatment (BLT vs. BD, pooling across surface/activation status) and comparison of surface treatment/activation status within each macro-design (activated vs. non-activated in both the BLT and BD systems), as well as a direct comparison of chairside plasma activation (XActive) against manufacturer-integrated hydrophilic treatment (SLActive).
Plasma Activation Procedure For implants assigned to plasma activation, the implant will be retrieved from its sterile container using the handpiece implant driver. The handpiece with the implant attached will be placed in the magnetic holder of the Plasma X Motion device, and the activation cycle will be initiated according to the manufacturer's instructions, completing automatically within approximately 50 seconds. Immediately following completion of the cycle, the magnetic holder will be removed from the device and the implant will be inserted into the prepared osteotomy according to the implant manufacturer's surgical protocol, without delay.
Surgical Procedure Two grams of amoxicillin (Ospamox, Sandoz GmbH, Austria) will be administered as a single preoperative dose; no antibiotics will be prescribed postoperatively. All surgeries will be performed under local anesthesia by the same surgeon to minimize operator variability. A mid-crestal incision will be made and a buccal mucoperiosteal flap elevated. Vertical soft tissue thickness will be measured with a periodontal probe applied to the unelevated lingual flap, after which the lingual flap will be elevated. Osteotomy preparation will follow the standard protocol recommended by the respective implant manufacturer, followed by implant insertion positioned 4 mm apical to the gingival margin. A healing abutment will be placed and sutures applied. Postoperative 0.12% chlorhexidine mouthwash will be prescribed; no postoperative antibiotics will be given.
Statistical Analysis Plan Statistical analysis will be performed using R (v. 4.0.4). Between-group comparisons of quantitative variables for two groups will use Student's t-test or the non-parametric Mann-Whitney U test, as appropriate. Comparisons among more than two groups will use one-way analysis of variance (ANOVA) or the non-parametric Kruskal-Wallis test, as appropriate. Normality of distributions will be assessed using the Shapiro-Wilk test. Paired comparisons (baseline vs. follow-up timepoints within the same implant) will use the paired t-test or Wilcoxon signed-rank test, as appropriate. Associations between insertion torque and ISQ will be assessed using Spearman correlation. Descriptive statistics will include mean, standard deviation, quartiles (Q1, Q3), median, and available number of observations for continuous variables, and absolute counts and percentages for categorical variables. A p-value <0.05 will be considered statistically significant.
Sample Size Justification As a pilot study, the planned enrollment of 10 implants per group (50 implants total) is intended to assess feasibility and estimate effect sizes and variability rather than to achieve definitive statistical power for between-group comparisons, consistent with established pilot study methodology. Data generated by this pilot will be used to perform a formal power analysis to determine the sample size required for a future, adequately powered confirmatory trial.
Significance This trial is intended to provide preliminary data on whether implant macro-design (thread geometry) and surface activation strategy (chairside plasma activation vs. manufacturer-integrated hydrophilic treatment) meaningfully influence early implant stability dynamics in healed, high-quality mandibular bone. Because chairside plasma activation systems such as XActive can, in principle, be applied to implants from any manufacturer, a favorable comparison with proprietary hydrophilic surfaces such as SLActive could have practical implications for clinical protocol selection and implant loading timelines. Findings from this pilot are also intended to inform the design, eligibility criteria, and sample size of a future confirmatory trial, potentially including sites with lower primary stability (e.g., softer bone or immediate implant placement), where surface-treatment effects on ISQ dynamics may be more pronounced.
Study design
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- None (open label)
- Number of arms
- 5
| Arm | Type | Receives |
|---|---|---|
| Straumann Bone Level Tapered implants SLA with Plasma X Motion treatment (BLT SLA XActive) | Experimental | Device: Chairside implant surface treatment with Plasma X Motion device, Device: SLA implant surface, Device: Straumann BLT implant macro designParticipants will receive a Straumann Bone Level Tapered (BLT) implant with a conventional sandblasted, large-grit, acid-etched (SLA) surface. Immediately prior to placement, the implant will undergo chairside low-temperature vacuum plasma surface activation using the Plasma X Motion system (XActive®), performed according to the manufacturer's protocol (approximately 50-second activation cycle) immediately before insertion into the prepared osteotomy. |
| Straumann Bone Level Tapeded implants with SLActive surface (BLT SLActive) | Experimental | Device: Straumann BLT implant macro design, Device: SLActive implant surface treatmentParticipants will receive a Straumann Bone Level Tapered (BLT) implant with the manufacturer hydrophilic SLActive surface. No chairside plasma activation will be performed; the hydrophilic, hydrocarbon-free surface state is achieved during manufacturing and preserved through storage in isotonic saline under nitrogen until use |
| MegaGen BlueDiamond implants Xpeed surface treated using Plasma X Motion (BD XpeedActive) | Experimental | Device: Chairside implant surface treatment with Plasma X Motion device, Device: Megagen BD implant macro design, Device: Megagen Xpeed implant surface treatmentParticipants will receive a Megagen BlueDiamond (BD) implant with the nanostructured calcium-incorporated Xpeed® surface. Immediately prior to placement, the implant will undergo chairside low-temperature vacuum plasma surface activation using the Plasma X Motion system (XActive®), performed according to the manufacturer's protocol immediately before insertion into the prepared osteotomy. |
| MegaGen BlueDiamond implants Xpeed surface (BD Xpeed) | Experimental | Device: Megagen BD implant macro design, Device: Megagen Xpeed implant surface treatmentParticipants will receive a Megagen BlueDiamond (BD) implant with the manufacturer nanostructured calcium-incorporated Xpeed® surface, without any chairside plasma activation. |
| Straumann Bone Level Tapeded implants with SLActive surface (BLT SLA) | Active comparator | Device: SLA implant surface, Device: Straumann BLT implant macro designParticipants will receive a Straumann Bone Level Tapered (BLT) implant with a conventional sandblasted, large-grit, acid-etched (SLA) surface, without any chairside plasma activation and without a manufacturer-integrated hydrophilic treatment. This arm serves as the reference/control group for surface-treatment comparisons. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (5 listed)
- Age ≥ 18 years
- Requiring a dental implant in a healed mandibular molar site (≥ 3 months post-extraction)
- Site anatomy suitable for placement of a 4.1 × 10 mm implant
- Bleeding on probing < 20%
- Plaque index < 25%
Key exclusion criteria (5 listed)
- Smoking ≥ 10 cigarettes/day
- History of uncontrolled periodontitis
- Uncontrolled diabetes
- Alcoholism
- Use of medication known to impair bone healing
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Chairside implant surface treatment with Plasma X Motion deviceXActive® chairside plasma surface activation (Plasma X Motion system) applied | Device | MegaGen BlueDiamond implants Xpeed surface treated using Plasma X Motion (BD XpeedActive), Straumann Bone Level Tapered implants SLA with Plasma X Motion treatment (BLT SLA XActive) |
| SLA implant surfaceManufacturer SLA implant surface | Device | Straumann Bone Level Tapeded implants with SLActive surface (BLT SLA), Straumann Bone Level Tapered implants SLA with Plasma X Motion treatment (BLT SLA XActive) |
| Straumann BLT implant macro designStraumann Bone Level Tapered implant macro design | Device | Straumann Bone Level Tapeded implants with SLActive surface (BLT SLA), Straumann Bone Level Tapeded implants with SLActive surface (BLT SLActive), Straumann Bone Level Tapered implants SLA with Plasma X Motion treatment (BLT SLA XActive) |
| SLActive implant surface treatmentManufacturer SLActive surface treatment | Device | Straumann Bone Level Tapeded implants with SLActive surface (BLT SLActive) |
| Megagen BD implant macro designMegagen BlueDiamond implant macro design | Device | MegaGen BlueDiamond implants Xpeed surface (BD Xpeed), MegaGen BlueDiamond implants Xpeed surface treated using Plasma X Motion (BD XpeedActive) |
| Megagen Xpeed implant surface treatmentMegagen Xpeed manufacturer implant surface treatment | Device | MegaGen BlueDiamond implants Xpeed surface (BD Xpeed), MegaGen BlueDiamond implants Xpeed surface treated using Plasma X Motion (BD XpeedActive) |
Primary outcome
Implant Insertion Torque Value (ITV)
Time frame: Day 1
Insertion torque will be recorded in Newton-centimeters (Ncm) using a prosthetic torque wrench at the moment of implant seating, prior to healing abutment connection. ITV will be compared between the two implant macro-designs (Straumann Bone Level Tapered vs. Megagen BlueDiamond), pooling across surface treatment/activation status, to evaluate the effect of implant macrogeometry on primary mechanical stability.
Implant Stability Quotient (ISQ)
Time frame: Baseline (day 1) and at 2, 4, and 6 weeks post-placement
Implant stability will be measured non-invasively by resonance frequency analysis (RFA) using a SmartPeg transducer attached to the implant's internal connection. Two perpendicular readings (bucco-lingual and mesio-distal) will be obtained at each timepoint using a manufacturer-recommended 45-degree probe angle, and their average will be recorded as the ISQ value for that visit. The change in ISQ (ΔISQ) from baseline will be calculated at each follow-up timepoint (W2, W4, W6) for each of the five study arms, to evaluate the trajectory of early implant stability according to implant macro-design and surface treatment/activation.
4 secondary outcomes
Vertical gingival thickness
Time frame: Day 1
Implant placement depth
Time frame: Day 1
Buccal vertical gingival thickness
Time frame: 8 weeks
Marginal bone level
Time frame: At the time of prosthetic rehabilitation of the implant and at the time of the 1 year follow-up visit.
Results and safety
This study is still under way. Results are not available in the source record yet.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 9 Sep 2026.
- Background: PubMed 19335092Bashutski JD, D'Silva NJ, Wang HL. Implant compression necrosis: current understanding and case report. J Periodontol. 2009 Apr;80(4):700-4. doi: 10.1902/jop.2009.080581.
- Background: PubMed 31357620Makary C, Menhall A, Zammarie C, Lombardi T, Lee SY, Stacchi C, Park KB. Primary Stability Optimization by Using Fixtures with Different Thread Depth According To Bone Density: A Clinical Prospective Study on Early Loaded Implants. Materials (Basel). 2019 Jul 27;12(15):2398. doi: 10.3390/ma12152398.
- BackgroundJulious, S.A. (2005) Sample size of 12 per group rule of thumb for a pilot study. Pharmaceutical Statistics, 4 (4). pp. 287-291. ISSN: 1539-1604
- Background: PubMed 12071926Chun HJ, Cheong SY, Han JH, Heo SJ, Chung JP, Rhyu IC, Choi YC, Baik HK, Ku Y, Kim MH. Evaluation of design parameters of osseointegrated dental implants using finite element analysis. J Oral Rehabil. 2002 Jun;29(6):565-74. doi: 10.1046/j.1365-2842.2002.00891.x.
- Background: PubMed 40693123Stacchi C, Rapani A, Montanari M, Martini R, Lombardi T. Effect of Vacuum Plasma Activation on Early Implant Stability: a Single-Blind Split-Mouth Randomized Clinical Trial. J Oral Maxillofac Res. 2025 Jun 30;16(2):e4. doi: 10.5037/jomr.2025.16205. eCollection 2025 Apr-Jun.
- Background: PubMed 17974109Oates TW, Valderrama P, Bischof M, Nedir R, Jones A, Simpson J, Toutenburg H, Cochran DL. Enhanced implant stability with a chemically modified SLA surface: a randomized pilot study. Int J Oral Maxillofac Implants. 2007 Sep-Oct;22(5):755-60.
- Background: PubMed 38998421Menhall A, Lahoud P, Yang KR, Park KB, Razukevicius D, Traini T, Makary C. The Mineral Apposition Rate on Implants with Either a Sandblasted Acid-Etched Implant Surface (SLA) or a Nanostructured Calcium-Incorporated Surface (XPEED(R)): A Histological Split-Mouth, Randomized Case/Control Human Study. Materials (Basel). 2024 Jul 5;17(13):3341. doi: 10.3390/ma17133341.
- Background: PubMed 26827067Lozano-Carrascal N, Salomo-Coll O, Gilabert-Cerda M, Farre-Pages N, Gargallo-Albiol J, Hernandez-Alfaro F. Effect of implant macro-design on primary stability: A prospective clinical study. Med Oral Patol Oral Cir Bucal. 2016 Mar 1;21(2):e214-21. doi: 10.4317/medoral.21024.
- Background: PubMed 25932309Lee SY, Kim SJ, An HW, Kim HS, Ha DG, Ryo KH, Park KB. The effect of the thread depth on the mechanical properties of the dental implant. J Adv Prosthodont. 2015 Apr;7(2):115-21. doi: 10.4047/jap.2015.7.2.115. Epub 2015 Apr 23.
- BackgroundTissue-integrated prostheses : osseointegration in clinical dentistry by Brånemark, Per-Ingvar; Zarb, George A. (George Albert), 1938-; Albrektsson, Tomas Publication date 1985
- Background: PubMed 38793397Makary C, Menhall A, Lahoud P, Yang KR, Park KB, Razukevicius D, Traini T. Bone-to-Implant Contact in Implants with Plasma-Treated Nanostructured Calcium-Incorporated Surface (XPEEDActive) Compared to Non-Plasma-Treated Implants (XPEED): A Human Histologic Study at 4 Weeks. Materials (Basel). 2024 May 14;17(10):2331. doi: 10.3390/ma17102331.
- Background: PubMed 38842233Chhabra K, Rajasekar A. Comparison of Roughness, Wettability, and SEM Features between Sandblasted Acid-Etched and Oxidized Titanium Dental Implants. J Long Term Eff Med Implants. 2024;34(4):57-63. doi: 10.1615/JLongTermEffMedImplants.2023049632.
- Individual participant dataSharing planned: No. Small sample size raising re-identification risk given the specific implant/surface combinations.
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Chairside implant surface treatment with Plasma X Motion device, SLA implant surface and Straumann BLT implant macro design in people with Dental Implant Macrodesign and Dental Implant Surface Treatment, run by Dantu implantologijos centras at 1 site in Lithuania.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 5 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: looking for participants now, as of the sponsor's last update on 9 Sep 2026. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Lithuania. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
This study is still under way. Results are not available in the source record yet.
How can a doctor refer a patient or contact the research team?
Contact Marius Svedas (+37069829987, svedasm@gmail.com), or a site near the patient. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Implant Insertion Torque Value (ITV) (time frame: Day 1). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT07744776, which also lists 23 linked publications.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 9 Sep 2026.