Intraocular PressureNCT07770178

Safety, Tolerability, and Accuracy of a Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor in Healthy Adults

A study testing Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor in people with Intraocular Pressure, run by Shanghai General Hospital, Shanghai Jiao Tong University School of Medicine at 1 site in China.

Not yet recruiting

Opening soon. The study is approved but enrolment has not started yet. Sites may already take enquiries.

Sites and contacts
Official record
Search all trials

Trial summary

The occurrence and progression of eye diseases like glaucoma are closely related to intraocular pressure fluctuations. Traditional single-point intraocular pressure measurement methods cannot reflect daily dynamic changes and are prone to missing hidden intraocular pressure peaks, limiting precise diagnosis and… Read more about this study

Who is it for?
Ages 20 to 40 yearsHealthy volunteers can take part
Conditions the study is for
Intraocular Pressure
What is being tested?
Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
20 participants
Where
1 siteChina
Study dates
30 Aug 2026to 18 Feb 2027
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 20 to 40 years
  • All sexes
  • Healthy volunteers accepted
  • Adult

Who may be able to join

  • Voluntarily agrees to participate in the study and provides written informed consent.
  • Has previously worn contact lenses.
  • Has no ocular disease other than refractive error.
  • Central corneal thickness between 540 and 560 μm, inclusive.
  • Corneal curvature between 42 and 44 D, inclusive.

Who cannot join

  • Unable to tolerate the hydrogel-based wearable dynamic intraocular pressure sensor.
  • Pregnant or breastfeeding.
  • History of allergy to relevant medical materials or known hypersensitivity to materials used in the investigational device.

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

China1 site
  • Shanghai General Hospital

    Shanghai, Shanghai Municipality, 200080

Other trials in China

About this study

Official title: Safety, Tolerability, and Accuracy of a Novel Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor in Healthy Adults: A Single-Center, Single-Arm Exploratory Clinical Trial

The occurrence and progression of eye diseases like glaucoma are closely related to intraocular pressure fluctuations. Traditional single-point intraocular pressure measurement methods cannot reflect daily dynamic changes and are prone to missing hidden intraocular pressure peaks, limiting precise diagnosis and individualized treatment. This project aims to evaluate the wearing safety and measurement accuracy of a new low-modulus, high-strength hydrogel wearable dynamic intraocular pressure sensor. This product adopts the form of smart contact lenses, fully embedding a micro sensor into a hydrogel lens, combined with an external antenna mount and a portable recorder to continuously detect and wirelessly record even slight changes in corneal curvature, thereby obtaining real-time intraocular pressure fluctuation trend data. Based on this, the objective of this study is to evaluate the safety, measurement accuracy and tolerabilityof a wearable dynamic intraocular pressure sensor based on a novel hydrogel in healthy volunteers.

Being tested
Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Hydrogel-Based Wearable Dynamic Intraocular Pressure SensorDevice

    The investigational device is a wearable hydrogel-based contact lens sensor designed to record electrical signals associated with changes in intraocular pressure. The CLS is worn on the study eye and allowed to stabilize for 10 minutes before signal acquisition. Sensor signals are obtained at three predefined measurement points and compared with Goldmann applanation tonometry measurements to assess measurement accuracy.

Study groups in NCT07770178
Study groupRoleWhat they receive
Hydrogel-Based Wearable Dynamic Intraocular Pressure SensorExperimentalParticipants will undergo a single-arm evaluation of a hydrogel-based wearable dynamic intraocular pressure sensor (CLS). After 2 hours of fasting from food and water, baseline ocular safety examinations will be performed. Intraocular pressure will first be measured using non-contact tonometry (NCT) and Goldmann applanation tonometry (GAT). Participants will then wear the CLS and undergo a 10-minute adaptation period before the first sensor signal is recorded. Participants will drink 1 L of water within 5 minutes. At 10-15 minutes after completion of water intake, a second sensor signal will be recorded, followed by CLS removal and repeat GAT and NCT measurements. After an interval of at least 45 minutes, NCT and GAT will be repeated, the CLS will be worn again for a 10-minute adaptation period, and a third sensor signal will be recorded. After completion of the measurement procedure, the CLS will be removed and ocular safety examinations will be repeated to evaluate changes associate
Receives: Device: Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Absolute Error of CLS-Predicted Intraocular Pressure Compared With Goldmann Applanation Tonometry

    The first two paired measurements of the contact lens sensor (CLS) signal and Goldmann applanation tonometry (GAT) intraocular pressure will be used to establish an individualized linear calibration equation for each participant. The third CLS signal will be converted to a predicted intraocular pressure using the individualized calibration equation. Absolute error is defined as the absolute difference between the CLS-predicted intraocular pressure and the corresponding GAT measurement, expressed in mmHg.

    Measured: Day 1, at the third validation measurement

Secondary outcomes (6)
  1. Agreement Between CLS-Predicted Intraocular Pressure and Non-Contact Tonometry

    At the third validation assessment, intraocular pressure predicted from the contact lens sensor (CLS) using the individualized calibration equation will be compared with the corresponding non-contact tonometry (NCT) measurement.

    Measured: Day 1, at the third validation measurement

  2. Increased incidence of corneal fluorescent staining grade

    Corneal fluorescence staining grading (National Eye Institute (NEI) standards) The cornea is divided into five regions (central, top, and above).Square, inferior, nasal side, temporal side), each area scores 0-3 points (0 = unstained, 1 = sparse dots, 2 = dense dots, 3 = fused or patchy)

    Measured: Baseline and at the post-study ocular examination on Day 1

  3. Conjunctival congestion score

    The Ora Calibra® scale was used to measure conjunctival redness in each eye. Each eye will be rated on a scale from 0 to 4, with 0 being best and 4 being the worst.

    Measured: Baseline and at the post-study ocular examination on Day 1

  4. Change From Baseline in Tear Break-Up Time After CLS Wear

    Tear break-up time (TBUT) will be measured before and after CLS wear. After blinking several times, the participant will keep the eye open while looking straight ahead, and the time until the first dry spot appears in the tear film will be recorded in seconds.

    Measured: Baseline and at the post-study ocular examination on Day 1

  5. Change From Baseline in Best-Corrected Visual Acuity After CLS Wear

    Best-corrected visual acuity (BCVA) will be assessed before and after CLS wear. Change from baseline will be calculated by comparing the post-study BCVA with the baseline BCVA.

    Measured: Baseline and at the post-study ocular examination on Day 1

  6. CLDEQ-8-Based Immediate Ocular Tolerability Questionnaire

    Immediate ocular tolerability during wear of the investigational contact lens sensor (CLS) will be assessed using a study-specific questionnaire adapted from the Contact Lens Dry Eye Questionnaire-8 (CLDEQ-8).

    The questionnaire contains eight items assessing: frequency and intensity of ocular discomfort, frequency and intensity of ocular dryness, frequency and intensity of fluctuating or blurred vision, frequency of wanting to close the eye because of discomfort, and the desire to remove the CLS because of discomfort.

    Frequency items are scored from 0 to 4, intensity items from 0 to 5, and the item assessing the desire to remove the CLS from 1 to 6. Higher scores indicate worse ocular tolerability.

    Measured: Immediately after completion of CLS wear on Day 1

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted18 Aug 2026
  2. Study start30 Aug 2026 (estimated)
  3. Primary completion30 Nov 2026 (estimated)
  4. Study completion18 Feb 2027 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
20 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07770178
Sponsor's protocol ID
20260424074539870
Status last verified
Aug 2026
First submitted
12 Aug 2026
Last updated
18 Aug 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Yes

Questions about this trial

What is this clinical trial studying?

A study testing Hydrogel-Based Wearable Dynamic Intraocular Pressure Sensor in people with Intraocular Pressure, run by Shanghai General Hospital, Shanghai Jiao Tong University School of Medicine at 1 site in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: opening soon, as of the sponsor's last update on 18 Aug 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Peiyao Jin (+8613564534911, peiyaojin@hotmail.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Absolute Error of CLS-Predicted Intraocular Pressure Compared With Goldmann Applanation Tonometry (time frame: Day 1, at the third validation measurement). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 18 Aug 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in