DysphagiaNCT07780981
Impact of Body Positioning on EndoFLIP Measurements: A Comparative Study of Supine and Left Lateral Decubitus Positions
A study testing EndoFlip in people with Dysphagia and Esophageal Stricture, run by Johns Hopkins University at 1 site in United States.
Trial summary
This is a prospective, single-center, crossover study comparing EndoFLIP measurements obtained in two patient positions, supine and left lateral decubitus, in pediatric and young adult patients. Read more about this study
- Who is it for?
- Ages 5 to 21 years
- Conditions the study is for
- Dysphagia, Esophageal Stricture, Epidermolysis Bullosa (EB)
- What is being tested?
- EndoFlip
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 25 participants
- Where
- 1 siteUnited States
- Study dates
- 24 Jul 2026to 1 Jul 2028
- Phase
- Interventional
- Are there results?
- Not yet: the study is under way
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 5 to 21 years
- All sexes
- No healthy volunteers
- Child, Adult
Who may be able to join
- must be between 5 years and 21 years of age at the time of the procedure, which is within the FDA-approved age range for EndoFLIP use.
- participants must be scheduled to undergo a clinically indicated EndoFLIP procedure at Johns Hopkins Children's Center.
- participants must have a parent or legal guardian who is able to provide written informed consent in English, or with the assistance of translation services for non-English speakers.
- participants who are of appropriate age and developmental capacity must be able to provide assent for study participation.
- participants must be candidates for general anesthesia as determined by the anesthesia team.
Who cannot join
- patients younger than 5 years of age or 22 years of age and older will be excluded.
- patients who have contraindications to EndoFLIP will be excluded, including situations where upper endoscopy is generally contraindicated such as known or suspected gastrointestinal perforation, hemodynamic instability, severe coagulopathy, or inability to safely undergo general anesthesia.
- patients who cannot be safely positioned in both supine and left lateral decubitus positions due to medical or anatomical considerations will be excluded.
- pregnant patients will be excluded. Pregnancy status will be determined by patient or guardian report and, when clinically indicated, by urine or serum pregnancy testing per institutional protocols for patients of childbearing potential.
- patients whose parent or legal guardian declines to provide informed consent or patients who decline to provide assent when age-appropriate will be excluded.
- patients for whom the proceduralist or anesthesia team determines that participation in the research protocol would pose additional risk beyond routine clinical care will be excluded.
Trial locations and contacts
1 site in 1 country. Contact the study team or a site near you to ask about taking part.
United States1 site, 1 recruiting
- Recruiting
Charlotte R. Bloomberg Children's Center
Baltimore, Maryland, 21287
About this study
This is a prospective, single-center, crossover study comparing EndoFLIP measurements obtained in two patient positions, supine and left lateral decubitus, in pediatric and young adult patients. The Endoluminal Functional Lumen Imaging Probe (EndoFLIP) is an FDA-approved device for patients aged 5 and older that utilizes high-resolution impedance planimetry and volume-controlled distension to assess esophagogastric junction function. Currently, no standardized positioning protocol exists, and proceduralists perform EndoFLIP based on individual preference, either supine or left lateral decubitus. This variability may introduce measurement inconsistencies, particularly when comparing preoperative and postoperative assessments performed in different positions.
Read the detailed description
The Endoluminal Functional Lumen Imaging Probe (EndoFLIP) is an FDA-approved device for patients aged 5 years and older and it utilizes high-resolution impedance planimetry and volume-controlled distension to evaluate the GI lumen including the esophagogastric junction. EndoFLIP has become an invaluable tool in pediatric gastroenterology helping to characterize various luminal conditions including but not limited to esophageal stenosis and compression, screening for dysmotility, guiding esophageal dilation through the selection of the optimal dilator size, among others. Despite its increasing clinical utility, there is currently no standardized protocol for its use and much of the technique is derived from adult practice.
Proceduralists perform measurements based on individual preference, with some favoring the supine position and others preferring the left lateral decubitus position. This lack of consensus has been an active topic of discussion among pediatric advanced endoscopists who perform EndoFLIP, yet no published studies exist comparing measurements obtained in different positions. This variability presents a significant clinical problem because positioning may influence EndoFLIP parameters through several physiological mechanisms. In the supine position, abdominal viscera shift posteriorly, potentially altering esophagogastric junction geometry and increasing intra-abdominal pressure. In the left lateral decubitus position, the gastric fundus falls away from the esophagogastric junction due to gravity, possibly reducing external compression and changing the angle of His. Additionally, diaphragmatic crural tension, thoracic mechanics, and vascular engorgement patterns may differ between positions, all of which could affect luminal measurements. These positional differences have important clinical implications, particularly when comparing preoperative and postoperative measurements, assessing dilation durability within the same procedure, or monitoring disease progression longitudinally over multiple visits. If measurements are obtained in different positions across these clinical scenarios, apparent changes in esophagogastric junction function may be artifactual rather than reflective of true physiological change.
The investigators hypothesize that patient positioning, supine versus left lateral decubitus, impacts EndoFLIP measurements. This is a non-directional hypothesis, as there is currently insufficient evidence to predict whether measurements will be higher or lower in one position compared to the other.
- Conditions
- Dysphagia, Esophageal Stricture, Epidermolysis Bullosa (EB), Achalasia, Esophageal Motility Disorders, Eosinophilic Esophagitis (EoE)
- Being tested
- EndoFlip
- Keywords
- Pediatric, EndoFlip, Position, Lateral, supine, measurement
- Medical subject headings
- Deglutition Disorders, Esophageal Stenosis, Epidermolysis Bullosa, Esophageal Achalasia, Esophageal Motility Disorders, Eosinophilic Esophagitis, Deception
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
EndoFlipDiagnostic test
The Endoluminal Functional Lumen Imaging Probe (EndoFLIP) is an FDA-approved device for patients aged 5 years and older and it utilizes high-resolution impedance planimetry and volume-controlled distension to evaluate the GI lumen.
| Study group | Role | What they receive |
|---|---|---|
| Left lateral position first, supine position second | Other | Not described Receives: Diagnostic Test: EndoFlip |
| supine position first, Left lateral position second | Other | Not described Receives: Diagnostic Test: EndoFlip |
What is being measured
The primary outcome is the main question the study is designed to answer.
Diameter
Diameter of esophagus (mm)
Measured: During procedure only (less than 10 minutes)
Cross sectional area
Esophageal cross sectional area measured in square millimeters (mm²)
Measured: During procedure only (less than 10 minutes)
Compliance
esophageal compliance (mm3/mmHg)
Measured: During procedure only (less than 10 minutes)
distensibility index
esophageal distensibility index(mm2/mmHg)
Measured: During procedure only (less than 10 minutes)
Pressure
intraluminal esophageal pressure (mmHg)
Measured: During procedure only (less than 10 minutes)
Secondary outcomes (5)
Differences in diameter
calculated difference in diameter between supine vs left lateral position (mm), percentage difference from left lateral position to supine position and also supine to left lateral position
Measured: During procedure only (less than 10 minutes)
Differences in Cross sectional area
calculated difference in Cross sectional area between supine vs left lateral position (mm2), percentage difference from left lateral position to supine position and also supine to left lateral position
Measured: During procedure only (less than 10 minutes)
Differences in esophageal compliance (mm3/mmHg)
calculated difference in esophageal compliance (mm3/mmHg) between supine vs left lateral position, percentage difference from left lateral position to supine position and also supine to left lateral position
Measured: During procedure only (less than 10 minutes)
Differences in esophageal distensibility index(mm2/mmHg)
calculated difference in esophageal distensibility index(mm2/mmHg) between supine vs left lateral position, percentage difference from left lateral position to supine position and also supine to left lateral position
Measured: During procedure only (less than 10 minutes)
Differences in intraluminal esophageal pressure (mmHg)
calculated difference in intraluminal esophageal pressure (mmHg) between supine vs left lateral position, percentage difference from left lateral position to supine position and also supine to left lateral position
Measured: During procedure only (less than 10 minutes)
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted24 Aug 2026
- Study start24 Jul 2026
- Primary completion1 Jul 2027 (estimated)
- Study completion1 Jul 2028 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for DysphagiaStudies of the same condition, open and under way.
- More trials run by Johns Hopkins UniversityThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Diagnosis
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- Double blind
- Who is masked
- Participant, Outcomes assessor
- Enrolment
- 25 participants (estimated)
Sponsor and investigators
- Lead sponsor
- Johns Hopkins University
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor
- Investigators
- Kenneth Ng, DO, Principal investigator, Johns Hopkins University
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT07780981
- Sponsor's protocol ID
- IRB00560539
- Status last verified
- Aug 2026
- First submitted
- 19 Aug 2026
- Last updated
- 24 Aug 2026
- Results posted
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
Study at a glance
From the registry record, last updated by the sponsor on 24 Aug 2026. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Impact of Body Positioning on EndoFLIP Measurements: A Comparative Study of Supine and Left Lateral Decubitus Positions
- ClinicalTrials.gov ID
- NCT07780981
- Acronym / sponsor ID
- IRB00560539
- Conditions
- Dysphagia, Esophageal Stricture, Epidermolysis Bullosa (EB), Achalasia, Esophageal Motility Disorders, Eosinophilic Esophagitis (EoE)
- Keywords
- Pediatric, EndoFlip, Position, Lateral, supine, measurement
- Study type
- Interventional
- Recruitment status
- Recruiting
- Enrolment
- 25 participants (estimated)
- Lead sponsor
- Johns Hopkins UniversityUniversity, hospital or other
- Study start
- 24 Jul 2026 (actual)
- Primary completion
- 1 Jul 2027 (estimated)
- Study completion
- 1 Jul 2028 (estimated)
- First posted
- 24 Aug 2026 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
This is a prospective, single-center, crossover study comparing EndoFLIP measurements obtained in two patient positions, supine and left lateral decubitus, in pediatric and young adult patients. The Endoluminal Functional Lumen Imaging Probe (EndoFLIP) is an FDA-approved device for patients aged 5 and older that utilizes high-resolution impedance planimetry and volume-controlled distension to assess esophagogastric junction function. Currently, no standardized positioning protocol exists, and proceduralists perform EndoFLIP based on individual preference, either supine or left lateral decubitus. This variability may introduce measurement inconsistencies, particularly when comparing preoperative and postoperative assessments performed in different positions.
Detailed description
The Endoluminal Functional Lumen Imaging Probe (EndoFLIP) is an FDA-approved device for patients aged 5 years and older and it utilizes high-resolution impedance planimetry and volume-controlled distension to evaluate the GI lumen including the esophagogastric junction. EndoFLIP has become an invaluable tool in pediatric gastroenterology helping to characterize various luminal conditions including but not limited to esophageal stenosis and compression, screening for dysmotility, guiding esophageal dilation through the selection of the optimal dilator size, among others. Despite its increasing clinical utility, there is currently no standardized protocol for its use and much of the technique is derived from adult practice.
Proceduralists perform measurements based on individual preference, with some favoring the supine position and others preferring the left lateral decubitus position. This lack of consensus has been an active topic of discussion among pediatric advanced endoscopists who perform EndoFLIP, yet no published studies exist comparing measurements obtained in different positions. This variability presents a significant clinical problem because positioning may influence EndoFLIP parameters through several physiological mechanisms. In the supine position, abdominal viscera shift posteriorly, potentially altering esophagogastric junction geometry and increasing intra-abdominal pressure. In the left lateral decubitus position, the gastric fundus falls away from the esophagogastric junction due to gravity, possibly reducing external compression and changing the angle of His. Additionally, diaphragmatic crural tension, thoracic mechanics, and vascular engorgement patterns may differ between positions, all of which could affect luminal measurements. These positional differences have important clinical implications, particularly when comparing preoperative and postoperative measurements, assessing dilation durability within the same procedure, or monitoring disease progression longitudinally over multiple visits. If measurements are obtained in different positions across these clinical scenarios, apparent changes in esophagogastric junction function may be artifactual rather than reflective of true physiological change.
The investigators hypothesize that patient positioning, supine versus left lateral decubitus, impacts EndoFLIP measurements. This is a non-directional hypothesis, as there is currently insufficient evidence to predict whether measurements will be higher or lower in one position compared to the other.
Study design
- Primary purpose
- Diagnosis
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- Double blind (participant, outcomes assessor)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Left lateral position first, supine position second | Other | Diagnostic Test: EndoFlip |
| supine position first, Left lateral position second | Other | Diagnostic Test: EndoFlip |
Who may qualify
- Ages
- 5 to 21 years
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Child, Adult
Key inclusion criteria (5 listed)
- must be between 5 years and 21 years of age at the time of the procedure, which is within the FDA-approved age range for EndoFLIP use.
- participants must be scheduled to undergo a clinically indicated EndoFLIP procedure at Johns Hopkins Children's Center.
- participants must have a parent or legal guardian who is able to provide written informed consent in English, or with the assistance of translation services for non-English speakers.
- participants who are of appropriate age and developmental capacity must be able to provide assent for study participation.
- participants must be candidates for general anesthesia as determined by the anesthesia team.
Key exclusion criteria (6 listed)
- patients younger than 5 years of age or 22 years of age and older will be excluded.
- patients who have contraindications to EndoFLIP will be excluded, including situations where upper endoscopy is generally contraindicated such as known or suspected gastrointestinal perforation, hemodynamic instability, severe coagulopathy, or inability to safely undergo general anesthesia.
- patients who cannot be safely positioned in both supine and left lateral decubitus positions due to medical or anatomical considerations will be excluded.
- pregnant patients will be excluded. Pregnancy status will be determined by patient or guardian report and, when clinically indicated, by urine or serum pregnancy testing per institutional protocols for patients of childbearing potential.
- patients whose parent or legal guardian declines to provide informed consent or patients who decline to provide assent when age-appropriate will be excluded.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| EndoFlipThe Endoluminal Functional Lumen Imaging Probe (EndoFLIP) is an FDA-approved device for patients aged 5 years and older and it utilizes high-resolution impedance planimetry and volume-controlled distension to evaluate the GI lumen. | Diagnostic test | Left lateral position first, supine position second, supine position first, Left lateral position second |
Primary outcome
Diameter
Time frame: During procedure only (less than 10 minutes)
Diameter of esophagus (mm)
Cross sectional area
Time frame: During procedure only (less than 10 minutes)
Esophageal cross sectional area measured in square millimeters (mm²)
Compliance
Time frame: During procedure only (less than 10 minutes)
esophageal compliance (mm3/mmHg)
distensibility index
Time frame: During procedure only (less than 10 minutes)
esophageal distensibility index(mm2/mmHg)
Pressure
Time frame: During procedure only (less than 10 minutes)
intraluminal esophageal pressure (mmHg)
5 secondary outcomes
Differences in diameter
Time frame: During procedure only (less than 10 minutes)
Differences in Cross sectional area
Time frame: During procedure only (less than 10 minutes)
Differences in esophageal compliance (mm3/mmHg)
Time frame: During procedure only (less than 10 minutes)
Differences in esophageal distensibility index(mm2/mmHg)
Time frame: During procedure only (less than 10 minutes)
Differences in intraluminal esophageal pressure (mmHg)
Time frame: During procedure only (less than 10 minutes)
Results and safety
This study is still under way. Results are not available in the source record yet.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 24 Aug 2026.
Related trials for Dysphagia
Not yet recruitingPhase 4
Effects of Tirzepatide on Esophagogastric Junction Mechanics, Reflux Burden, and Gastric Physiology in Adults With Obesity
Obese Adults
Not yet recruiting
EEG Feedback-Based Vagus Nerve Stimulation for Post-Stroke Dysphagia
Ischemic Stroke, Post-stroke Dysphagia
Recruiting
Mesh Vs Pledgets for Repair of Paraesophageal Hernia
Paraesophageal Hernia, GERD
Active, not recruitingObservational study
Dexmedetomidine Nasal Spray and Postoperative Sore Throat After Bronchoscopy
Postoperative Sore Throat, Bronchoscopy
Recruiting
The Esophageal String Test as a Diagnostic Screening Tool for Eosinophilic Esophagitis Among Africans With Dysphagia in Mali and the United States
Dysphagia, Eosinophilic Esophagitis
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part.
Kenneth Ng, DO principal investigator
Johns Hopkins University
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing EndoFlip in people with Dysphagia and Esophageal Stricture, run by Johns Hopkins University at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 5 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: looking for participants now, as of the sponsor's last update on 24 Aug 2026. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
This study is still under way. Results are not available in the source record yet.
How can a doctor refer a patient or contact the research team?
Contact Kenneth Ng, DO (410-955-8769, kng13@jhmi.edu), or a site near the patient. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Diameter (time frame: During procedure only (less than 10 minutes)). The study lists 5 primary outcomes in all. It is what the study was designed to measure, not a finding.