Cardiac Conduction Disorders Requiring Permanent Pacing in Patients With Prior or Planned Tricuspid Valve InterventionMCV3830NCT07805525

Early Feasibility Study of 3830 Lead Implantation Via the Middle Cardiac Vein in Patients With Prior or Planned Tricuspid Valve Intervention

A study testing Pacemaker in people with Cardiac Conduction Disorders Requiring Permanent Pacing in Patients With Prior or Planned Tricuspid Valve Intervention, run by Minneapolis Heart Institute Foundation at 1 site in United States.

Not yet recruiting

Opening soon. The study is approved but enrolment has not started yet. Sites may already take enquiries.

Sites and contacts
Official record
Search all trials

Trial summary

Prospective, single center, early feasibility study designed to evaluate the initial safety and device performance of the Medtronic 3830 Lead when implanted via the Middle Cardiac Vein in patients with prior or planned tricuspid valve interventions. Read more about this study

Who is it for?
Ages 18 years and older
What is being tested?
Pacemaker
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
10 participants
Where
1 siteUnited States
Study dates
Nov 2026to Jan 2028
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Subjects who meet current clinical practice guidelines for implantation of a PPM or CRT-P system and will undergo one of the following procedures are eligible for enrollment, provided all inclusion and no exclusion criteria are met:
  • Medtronic 3830 DDD pacing mode (dual-chamber pacemaker) OR
  • Medtronic 3830 VVI pacing mode (single-chamber ventricular pacemaker) OR
  • Medtronic 3830 CRT-P system implantation
  • Anatomical suitability for middle cardiac vein lead placement based on pre-procedural CT imaging.
  • Patient must have prior or planned tricuspid valve intervention, including any of the following:
Show all 14Show fewer
  • Surgical tricuspid valve repair (e.g., annuloplasty ring, leaflet repair) OR
  • Surgical tricuspid valve replacement with a bioprosthetic or mechanical valve OR
  • Transcatheter tricuspid valve repair (e.g., edge-to-edge repair devices) OR
  • Transcatheter tricuspid valve replacement (orthotopic or heterotopic) OR
  • Patients with pre-existing transvalvular RV lead undergoing lead extraction for the management of tricuspid regurgitation who require reimplantation of a new ventricular pacing lead
  • A clinical need to avoid crossing the tricuspid valve, as determined by the treating heart team (e.g., risk of prosthesis-lead interaction, leaflet trauma, or the need to maintain a lead-free tricuspid valve)
  • Age 18 years or older.
  • Signed Patient Informed Consent Form (ICF).

Who cannot join

  • Life expectancy < 6 months due to non-cardiac comorbidities or advanced frailty where the procedure is not aligned with goals of care
  • Contraindication to transvenous lead implantation, including:
  • Severe or occlusive disease of the axillary/subclavian venous system
  • Coronary sinus or middle cardiac vein occlusion, severe stenosis, or anatomy prohibiting safe lead delivery
  • Known hypercoagulable state or bleeding disorder not manageable with standard therapy.
  • Subject is contraindicated for a single dose of 40.0 µg of beclomethasone dipropionate
Show all 14Show fewer
  • Patient with evidence of active infection or undergoing treatment for an infection.
  • Inability to provide informed consent, including:
  • Cognitive impairment without a legally authorized representative
  • Language or communication barriers that prevent adequate understanding of the study
  • Unable to follow through study protocol for any reasons (social support, distance from research center, insurance denial, mental illness etc.) in the investigator's judgement
  • Pregnant women or breastfeeding women, or women of childbearing potential and who are not on a reliable form of birth regulation method or abstinence
  • End-Stage Renal Disease (ESRD) on dialysis
  • NYHA IV

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

Efrah M Hussein, BS contact

612-863-0666

Efrah.Hussein@allina.com

Dawn R Witt, PhD contact

612-385-8299

Dawn.Witt@allina.com

United States1 site
  • Minneapolis Heart Institute Foundation

    Minneapolis, Minnesota, 55407

    Efrah M Hussein, BS, 612-863-0666, Efrah.Hussein@allina.com

    Dawn R Witt, PhD, 612-385-8299, Dawn.Witt@allina.com

Other trials in United States

About this study

Official title: Early Feasibility Study of Medtronic 3830 Lead Implantation Via the Middle Cardiac Vein in Patients With Prior or Planned Tricuspid Valve Intervention: MCV 3830 EFS

Prospective, single center, early feasibility study designed to evaluate the initial safety and device performance of the Medtronic 3830 Lead when implanted via the Middle Cardiac Vein in patients with prior or planned tricuspid valve interventions.

Being tested
Pacemaker

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • PacemakerDevice

    Required electrical testing for pacing, sensing, impedance will be completed

Study groups in NCT07805525
Study groupRoleWhat they receive
InterventionalExperimentalPatients will undergo implant of the Medtronic Model 3830 pacemaker lead via the Middle Cardiac Vein and undergo required electrical testing
Receives: Device: Pacemaker

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Acute Procedural Success

    Acute Procedural Success: defined as successful implantation of the SelectSecure™ Model 3830 lead in the targeted proximal septal segment of the middle cardiac vein, absent of any serious adverse events related to the MCV implantation site, and achievement of the following:

    1. Successful lead delivery and fixation at the intended MCV location using ≤ 4 deployment attempts
    2. Stable lead position at completion of the implant procedure
    3. Acceptable acute electrical performance, including:
    • Capture threshold ≤ 2.0 V at 0.4 ms
    • Sensing amplitude ≥ 4.0 mV
    • Lead impedance between 200-1500 Ω
    1. No device- or procedure-related Major Adverse Cardiovascular Events (MACE) through hospital discharge
    2. Absence of oversensing, including:
    • T-wave oversensing (TWOS)
    • P-wave oversensing (PWOS)
    • Myopotential oversensing
    • Diaphragm stimulation

    Measured: Index Procedure

Secondary outcomes (3)
  1. Incidence of Device- or Procedure-Related Serious Adverse Events and Complications

    Number of participants experiencing device- or procedure-related complications through 90 days following implantation including but not limited to:

    • Pericardial effusion
    • Clinically significant perforation
    • Infection
    • Coronary sinus or venous injury
    • Clinically significant phrenic nerve/diaphragmatic stimulation
    • Lead revision or system extraction due to complications

    Measured: Up to 90 days following implant

  2. Lead Electrical Performance and Stability Through 12 Months

    1. Evaluation of pacing and sensing performance from the MCV lead at implant and during routine follow-up visits (pre-discharge, 3, 6, 9 and 12 months), including:
    • Stability of capture thresholds
    • Stability of sensing amplitudes
    • Lead impedance trends
    • Percentage pacing
    • Lead stability (absence of dislodgement or migration)
    • Need for lead revision or system modification due to performance issues
    • Presence of conduction system capture at implant
    • Persistence of conduction system capture at follow up

    Measured: From Implant through 12 Months

  3. Freedom From Device-or Procedure-Related Serious Adverse Events

    Short Term Safety will be assessed through 90 days following implantation, including:

    1. Freedom from device- or procedure-related serious adverse events (SAEs) through 90 days

    Measured: Up to 90 days following implant

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted4 Sep 2026
  2. Study startNov 2026 (estimated)
  3. Primary completionJan 2028 (estimated)
  4. Study completionJan 2028 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Device feasibility
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
10 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07805525
Sponsor's protocol ID
MCV3830
Acronym
MCV3830
Status last verified
Sep 2026
First submitted
27 Aug 2026
Last updated
4 Sep 2026
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Pacemaker in people with Cardiac Conduction Disorders Requiring Permanent Pacing in Patients With Prior or Planned Tricuspid Valve Intervention, run by Minneapolis Heart Institute Foundation at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 14 inclusion and 14 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: opening soon, as of the sponsor's last update on 4 Sep 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Efrah M Hussein, BS (612-863-0666, Efrah.Hussein@allina.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Acute Procedural Success (time frame: Index Procedure). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.