Comparing 3D-Printed and Milled Titanium Plates in Jaw Surgery (Le Fort I): Accuracy and Operative Time

A study testing 3D-Printed Patient-Specific Titanium Plates and Milled Patient-Specific Titanium Plates in people with Malocclusion, Angle Class II and Malocclusion, Angle Class III, run by Future University in Egypt at 1 site in Egypt.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

The purpose of this study is to find out if 3D printed patient-specific titanium plates improve surgical accuracy and decrease operative time compared to milled patient-specific titanium plates in patients having Le Fort I osteotomy for jaw correction. Read more about this study

Who is it for?
Ages 18 years and older
What is being tested?
3D-Printed Patient-Specific Titanium Plates and Milled Patient-Specific Titanium Plates
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
12 participants
Where
1 siteEgypt
Study dates
1 Oct 2026to Apr 2027
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Skeletally mature patients (typically ≥ 18 years old) with fused growth plates.
  • Non-syndromic skeletal Class II or Class III malocclusion requiring a Le Fort I osteotomy (with or without mandibular involvement).
  • Patients with no signs or symptoms of active temporomandibular disorders (TMD).
  • Patients with good oral hygiene and no active periodontal disease that could affect tooth-borne guides.
  • Highly motivated patients.

Who cannot join

  • Patients who refused to be included in the research and patients unable to comply with follow up Protocol.
  • Patients with systemic diseases may hinder the normal healing process or render the patient not fit for general anesthesia.
  • Craniofacial anomalies or syndromes (cleft lip/palate, syndromic deformities).
  • Previous orthognathic surgery or revision cases.
  • Patients with intra-bony lesions or infections that may retard the osteotomy healing.
  • History of maxillofacial trauma altering anatomy.

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

Egypt1 site, 1 recruiting
  • Oral and Maxillofacial Surgery Department, Faculty of Oral and Dental Medicine, Future University in Egypt

    Cairo, New Cairo, Fifth Settlement, 11835

    Mohamed A Khedre, BDS, MSc candidate, +201091102408, 20253897@fue.edu.eg

    Mohamed A Khedre, BDS, MSc candidate

    Recruiting

Other trials in Egypt

About this study

Official title: Assessment of the Three-Dimensional Surgical Accuracy and Operative Time of 3D-Printed Versus Milled Patient-Specific Titanium Plates in Le Fort I Osteotomy: A Randomized Clinical Trial

The purpose of this study is to find out if 3D printed patient-specific titanium plates improve surgical accuracy and decrease operative time compared to milled patient-specific titanium plates in patients having Le Fort I osteotomy for jaw correction. The main questions it aims to answer are:

  • Do 3D-printed patient specific titanium plates improve surgical accuracy (as measured by ability to transfer the computer plan to the patient's jaw)?
  • Do 3D-printed patient specific titanium plates decrease operative time when compared to milled patient specific titanium plates? This study will compare two groups of patients. One group will receive 3D printed plates (study group) and the other will receive milled plates (control group) to see if the different methods of making the plates changes the surgical accuracy or operative time.
Read more

All participants in the study will:

  • Have Le Fort I osteotomy surgery using patient specific titanium plates (either 3D printed or milled).
  • Have CT scans before and after surgery to evaluate accuracy. Have their operative time recorded.
  • Have a post-surgical follow-up to evaluate healing, stability and orthodontic treatment.
Read the detailed description
  • Background and Clinical Rationale Orthognathic surgeries, specifically Le Fort I osteotomy, requires precise transfer of the virtual surgical plan to the operative field to achieve predictable functional and facial outcomes. Conventional splint-based fixation techniques may introduce cumulative inaccuracies arising from laboratory fabrication errors, intraoperative handling, and manual plate adaptation. To overcome these limitations, computer-aided design and computer-aided manufacturing (CAD/CAM) workflows allow for the fabrication of patient-specific titanium plates via 3D printing (additive manufacturing) or milling (subtractive manufacturing). While both fabrication methods aim to enhance surgical accuracy and workflow efficiency, direct randomized controlled trial evidence comparing the three-dimensional accuracy and operative performance of 3D-printed versus milled maxillary patient-specific plates remains limited.
  • Preoperative Diagnostic Protocol \& 3D Virtual Surgical Planning
  1. Presurgical Orthodontic \& Diagnostic Setup: Participants undergo presurgical orthodontic treatment for dental decompensation to establish a stable occlusion in at least three points. Preoperative diagnostic records include standardized facial and dental photographs along with dental casts obtained from alginate impressions.
  2. Image Acquisition \& 3D Skull Reconstruction: Computed Tomography (CT) scans are acquired in the natural head position utilizing a Planmeca ProFace imaging machine. Diagnostic dental models are digitized using an optical scanner (Shera operating system 7Series; Dental Wing Inc., Montreal, Quebec, Canada). The digitized dental models will be manipulated using point-based matching iterative closest point (ICP) registration based on anatomical landmarks followed by manual surface-based registration to generate an artifact-free composite 3D skull model.
  3. Surgical Simulation \& Guide Design: 3D cephalometric analysis of the bony and soft tissue landmarks are conducted using surgical simulation software (Mimics 19.0; Materialise NV, Leuven, Belgium). Virtual Le Fort I osteotomy and mandibular sagittal split osteotomy (SSO) are simulated, adjusting the maxillary segment relative to facial symmetry planes and clinical anthropometric analysis. The cutting guides will be designed on the maxilla to orient the osteotomy and mark reference holes (16 holes with 8 holes on each side) to be used later for the repositioning/ fixation plate, using CAD software (3-matic 11.0; Materialise NV, Leuven, Belgium).
  4. CAD/CAM Manufacturing:

Surgical Guides: Exported in stereolithography (STL) file format and printed in white polyamide (PA2200; EOS e-manufacturing solutions, Munich, Germany) via fused deposition modeling (FDM) on an additive CAM machine (FORMIGA P 110 printer; EOS e-manufacturing solutions, Munich, Germany). The guides are cold-sterilized by overnight immersion in 2% glutaraldehyde.

Patient-Specific Osteosynthesis Plates: Designed to fix the maxilla in the desired position making use of the previously established reference holes. The designed plates will be exported in STL file format to be manufactured in grade 5 titanium alloy using either 3D printing or milling according to randomized allocation .

Intraoperative Protocol \& Interventions Surgical procedures are performed under general anesthesia then intraoral local anesthetic infiltration along incision line with (lidocaine 2%, 1/100,000 adrenaline) for hemostasis. Access is established via intraoral vestibular maxillary and sagittal split mandibular incisions.

Osteotomy \& Reference Hole Drilling: The maxillary cutting guide is seated on the exposed bone, verified for passive fit, and secured using four 2.0-mm screws. Sixteen reference holes (8 per side) are established through the guide. A reciprocating saw is used to perform the Le Fort I osteotomy, followed by maxillary mobilization and removal of bony interferences. Bilateral sagittal split osteotomies of the mandible are subsequently performed using a mandibular guide.

Fixation Protocol (Randomized Allocation):

Study Arm (Group 1): Maxillary repositioning and fixation are executed using 3D-printed patient-specific titanium plates, aligned with the pre-established reference holes and secured with 2.0-mm screws.

Control Arm (Group 2): Maxillary repositioning and fixation are executed using milled patient-specific titanium plates, aligned with the pre-established reference holes and secured with 2.0-mm screws.

Mandibular Fixation \& Closure: In both arms, the mandible is repositioned and fixated relative to the maxilla using traditional tri-poding bi-cortical screws. Incisions are sutured using 4-0 resorbable sutures in a continuous running fashion.

Postoperative Care \& Evaluation Schedule Postoperative Medications: Systemic antibiotic coverage (Amoxicillin / Clavulanic acid 625 mg every 8 hours) is initiated 4 hours postoperatively and continued for 5 days. Analgesics (NSAIDs every 6 hours) are provided for 3 days, and 0.12% Chlorhexidine mouthwash is prescribed for 2 weeks.

Follow-Up Schedule: Clinical evaluations occur weekly for the first month, followed by monthly visits for 5 additional months.

Radiographic Evaluation \& Orthodontics: A postoperative CT scan is acquired within the first postoperative week using parameters identical to the baseline scan to analyze 3D plan transfer accuracy. Post-surgical orthodontic treatment resumes 4 to 6 weeks postoperatively.

Quality Assurance, Monitoring, and Auditing Supervision \& Governance: All CAD/CAM virtual planning and surgical procedures are conducted under the direct supervision of senior academic staff (Prof. Lobna Abdelaziz Aly and Dr. Aya Magdy AbdelRady) at the Oral and Maxillofacial Surgery Department, Future University in Egypt.

Harms: Any temporary or permanent adverse effect will be recorded and documented.

Auditing: Auditing of the study design will be done by the Research Ethics Council, Future University in Egypt.

Protocol Amendments: Any modifications impacting study conduct, patient safety, or benefits require formal approval by the Council of the Oral and Maxillofacial Surgery Department, Future University in Egypt.

Discontinuation Criteria: Interventions are discontinued if technical difficulty with cutting guides/plates arises or if inaccurate translation of the surgical plan occurs intraoperatively.

Confidentiality: All study-related information will be stored securely. All participant information will be stored in locked file cabinets in areas with limited access. All patients' records will be digitized and stored securely on personal computer.

Ancillary and Post Trial Care: All patients will be followed up until complete healing and satisfactory results occur.

Declaration of interests: The study is self-funded. No conflict of interest Consent: The entire procedure will be explained to the patients and a written consent will be obtained.

Standard Operating Procedures \& Protocol Workflows

  1. Patient Recruitment \& Consent: Inclusion criteria target skeletally mature patients requiring Le Fort I osteotomy for Class II/III malocclusion with good oral hygiene and no active TMDs. Fully informed written consent is obtained prior to enrollment.
  2. CAD/CAM \& Manufacturing: Preoperative CT scans and digitized dental models. Cutting guides and Patient-specific titanium plates (Grade 5 titanium alloy) are manufactured.
  3. Intraoperative Fixation Procedures
  4. Postoperative Care \& Follow-Up
  5. Adverse Event (Harms) Reporting:

Any temporary or permanent adverse effects occurring during the trial are documented and recorded in patient files.

Plan for Missing Data \& Participant Retention Participant Retention Strategies: Patient phone numbers are recorded during written consent, and structured telephone reminders are made prior to every follow-up visit.

Adherence Verification: Patients are recalled at 1-week intervals and instructed to bring their medication packaging to verify postoperative drug compliance.

Statistical Analysis Plan Statistical analysis will be carried out. Data will be analyzed using IBM SPSS advanced statistics (Statistical Package for Social Sciences), version 21 (SPSS Inc., Chicago, IL). Numerical data will be described as mean and standard deviation or median and range, as appropriate. Because of the small sample size, the comparisons between pre and post treatment will be done by Wilcoxon signed rank test. A p-value less than or equal to 0.05 will be considered statistically significant. All tests will be two tailed.

Being tested
3D-Printed Patient-Specific Titanium Plates, Milled Patient-Specific Titanium Plates
Keywords
Orthognathic Surgery, Le Fort I Osteotomy, Dentofacial Deformities, Malocclusion, Angle Class II, Malocclusion, Angle Class III, Milled Titanium Plates, 3D Printed Titanium Plates
Medical subject headings
Malocclusion, Angle Class II, Malocclusion, Angle Class III, Dentofacial Deformities

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • 3D-Printed Patient-Specific Titanium PlatesDevice

    Following a Le Fort I osteotomy performed with a reciprocating saw, the maxilla is repositioned and rigidly fixated using the 3D-printed patient-specific plates. Plate alignment is guided by 16 pre-drilled reference holes (8 per side) established using a customized cutting guide, and plates are secured using 2.0-mm screws

  • Milled Patient-Specific Titanium PlatesDevice

    Following a Le Fort I osteotomy performed with a reciprocating saw, the maxilla is repositioned and rigidly fixated using the milled patient-specific plates. Plate alignment is guided by the same 16 pre-drilled reference holes (8 per side) established via a customized cutting guide, and plates are secured using 2.0-mm screws

Study groups in NCT07841067
Study groupRoleWhat they receive
3D-Printed Patient-Specific Titanium PlatesExperimentalParticipants in this group undergo Le Fort I osteotomy with maxillary repositioning and fixation using 3D-printed (additive manufactured) patient-specific grade 5 titanium alloy plates. Plate placement is guided by pre-drilled reference holes established via CAD/CAM cutting guides and secured with 2.0-mm screws.
Receives: Device: 3D-Printed Patient-Specific Titanium Plates
Milled Patient-Specific Titanium PlatesActive comparatorParticipants in this group undergo Le Fort I osteotomy with maxillary repositioning and fixation using milled (subtractive manufactured) patient-specific grade 5 titanium alloy plates. Plate placement is guided by pre-drilled reference holes established via CAD/CAM cutting guides and secured with 2.0-mm screws.
Receives: Device: Milled Patient-Specific Titanium Plates

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Accuracy of Surgical Transfer

    The primary outcome measures the precision with which the virtual surgical plan (P0/T0) is transferred to the actual surgical outcome (P1/T1).

    • Image Registration: The virtual surgical plan is superimposed onto the postoperative Computed Tomography (CT) scan using point-based matching Iterative Closest Point (ICP) registration, followed by manual surface-based best-fit registration targeting the stable skull base in Mimics 19.0 software.
    • Linear Measurement: Discrepancies between the planned and actual positions are measured from hard tissue reference points specifically the upper central incisor (U1) and upper first molars (16, 26):

    Vertical Deviation: Distance measured relative to the Frankfort Horizontal Plane (FHP).

    Anteroposterior Deviation: Distance measured relative to the Coronal Plane (CP).

    Mediolateral Deviation: Distance measured relative to the Mid-Facial Plane (MFP).

    Measured: Immediate postoperative period (within the first week postoperatively)

Secondary outcomes (2)
  1. Operative Time

    Operative time is defined as the total intraoperative duration required for all surgical procedures, including the positioning and fixation of the patient-specific titanium plate on the maxilla during Le Fort I osteotomy. The time is measured and recorded in minutes for each participant using a digital stopwatch, starting when the surgeon performs the first surgical incision and ending when the surgeon completes final closure suturing

    Measured: Intraoperatively (from initial surgical incision to final suture closure)

  2. Postoperative Stability

    Evaluates the extent to which skeletal structures maintain their corrected position relative to the virtual surgical plan following Le Fort I osteotomy. Postoperative CT scans obtained within 1 week are imported into Mimics software and superimposed onto the virtual plan using stable skull base reference points. Linear deviations are evaluated at key hard tissue landmarks (central incisor U1, first molars 16 and 26) across three spatial dimensions

    Measured: Immediate postoperative period (within 1 week postoperatively)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted25 Sep 2026
  2. Study start1 Oct 2026 (estimated)
  3. Primary completionDec 2026 (estimated)
  4. Study completionApr 2027 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Model description
This study uses a parallel assignment model. Eligible participants will be randomized into two groups: one group will undergo Le Fort I osteotomy with fixation using 3D-printed patient-specific titanium plates, and the other group will undergo the same procedure with milled patient-specific titanium plates. Each participant will receive only one intervention, and outcomes will be compared between the two groups.
Masking
Single blind
Who is masked
Outcomes assessor
Enrolment
12 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07841067
Sponsor's protocol ID
FUE.REC(66)/5-2026
Acronym
TiPlateRCT
Status last verified
Sep 2026
First submitted
19 Sep 2026
Last updated
25 Sep 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Publications

  • Abdelhamid AM, Hassan AM, El-Mohandes WA. Accuracy Assessment of Customized Titanium Plates Compared to 3D-Printed Splints in Le Fort I Osteotomy: A Randomized Clinical Trial Evaluating Clinical and Radiographic Outcomes. Cureus. 2025 May 15;17(5):e84151. doi: 10.7759/cureus.84151. eCollection 2025 May.

    PubMed 40519468
  • Cunningham SJ, Garratt AM, Hunt NP. Development of a condition-specific quality of life measure for patients with dentofacial deformity: II. Validity and responsiveness testing. Community Dent Oral Epidemiol. 2002 Apr;30(2):81-90. doi: 10.1034/j.1600-0528.2002.300201.x.

    PubMed 12000348
  • Cunningham SJ, Garratt AM, Hunt NP. Development of a condition-specific quality of life measure for patients with dentofacial deformity: I. Reliability of the instrument. Community Dent Oral Epidemiol. 2000 Jun;28(3):195-201. doi: 10.1034/j.1600-0528.2000.280305.x.

    PubMed 10830646
  • Zinser MJ, Sailer HF, Ritter L, Braumann B, Maegele M, Zoller JE. A paradigm shift in orthognathic surgery? A comparison of navigation, computer-aided designed/computer-aided manufactured splints, and "classic" intermaxillary splints to surgical transfer of virtual orthognathic planning. J Oral Maxillofac Surg. 2013 Dec;71(12):2151.e1-21. doi: 10.1016/j.joms.2013.07.007.

    PubMed 24237776
  • Buchanan EP, Hyman CH. LeFort I Osteotomy. Semin Plast Surg. 2013 Aug;27(3):149-54. doi: 10.1055/s-0033-1357112.

    PubMed 24872761
  • Turvey T, Hall DJ, Fish LC, Epker BN. Surgical-orthodontic treatment planning for simultaneous mobilization of the maxilla and mandible in the correction of dentofacial deformities. Oral Surg Oral Med Oral Pathol. 1982 Nov;54(5):491-8. doi: 10.1016/0030-4220(82)90185-2.

    PubMed 6960302
  • Fish LC, Epker BN. Dentofacial deformities related to midface deficiencies. Integrated orthodontic-surgical correction. J Clin Orthod. 1987 Sep;21(9):654-64. No abstract available.

    PubMed 3482095
  • Hanafy M, Akoush Y, Abou-ElFetouh A, Mounir RM. Precision of orthognathic digital plan transfer using patient-specific cutting guides and osteosynthesis versus mixed analogue-digitally planned surgery: a randomized controlled clinical trial. Int J Oral Maxillofac Surg. 2020 Jan;49(1):62-68. doi: 10.1016/j.ijom.2019.06.023. Epub 2019 Jun 29.

    PubMed 31262680
  • Heufelder M, Wilde F, Pietzka S, Mascha F, Winter K, Schramm A, Rana M. Clinical accuracy of waferless maxillary positioning using customized surgical guides and patient specific osteosynthesis in bimaxillary orthognathic surgery. J Craniomaxillofac Surg. 2017 Sep;45(9):1578-1585. doi: 10.1016/j.jcms.2017.06.027. Epub 2017 Jul 8.

    PubMed 28793965
  • Bai S, Bo B, Bi Y, Wang B, Zhao J, Liu Y, Feng Z, Shang H, Zhao Y. CAD/CAM surface templates as an alternative to the intermediate wafer in orthognathic surgery. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010 Nov;110(5):e1-7. doi: 10.1016/j.tripleo.2010.05.052. Epub 2010 Sep 1.

    PubMed 20813559
  • Levine JP, Patel A, Saadeh PB, Hirsch DL. Computer-aided design and manufacturing in craniomaxillofacial surgery: the new state of the art. J Craniofac Surg. 2012 Jan;23(1):288-93. doi: 10.1097/SCS.0b013e318241ba92.

    PubMed 22337427
  • Kusnoto B. Two-dimensional cephalometry and computerized orthognathic surgical treatment planning. Clin Plast Surg. 2007 Jul;34(3):417-26. doi: 10.1016/j.cps.2007.04.005.

    PubMed 17692701
  • Ellis E 3rd. Accuracy of model surgery: evaluation of an old technique and introduction of a new one. J Oral Maxillofac Surg. 1990 Nov;48(11):1161-7. doi: 10.1016/0278-2391(90)90532-7.

    PubMed 1698956
  • Ellis E 3rd. Bimaxillary surgery using an intermediate splint to position the maxilla. J Oral Maxillofac Surg. 1999 Jan;57(1):53-6. doi: 10.1016/s0278-2391(99)90633-x. No abstract available.

    PubMed 9915396
  • Bai S, Shang H, Liu Y, Zhao J, Zhao Y. Computer-aided design and computer-aided manufacturing locating guides accompanied with prebent titanium plates in orthognathic surgery. J Oral Maxillofac Surg. 2012 Oct;70(10):2419-26. doi: 10.1016/j.joms.2011.12.017. Epub 2012 Apr 18. No abstract available.

    PubMed 22516840

Questions about this trial

What is this clinical trial studying?

A study testing 3D-Printed Patient-Specific Titanium Plates and Milled Patient-Specific Titanium Plates in people with Malocclusion, Angle Class II and Malocclusion, Angle Class III, run by Future University in Egypt at 1 site in Egypt.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 25 Sep 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Egypt. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Each site lists its own contact: choose the nearest one under Trial locations. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Accuracy of Surgical Transfer (time frame: Immediate postoperative period (within the first week postoperatively)). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT07841067, which also lists 18 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 25 Sep 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in