Model 4396 Left Ventricular (LV) Lead Study

A study testing Pacing Lead in people with Heart Failure, run by Medtronic Cardiac Rhythm and Heart Failure at 31 sites in 8 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This study is designed to show that a new lead, which paces the left bottom chamber (left ventricle) of the heart, is safe and effective. Using a lead on the left side of the heart has been shown to potentially improve heart failure symptoms. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Heart Failure
What is being tested?
Pacing Lead
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
197 participants
Where
31 sitesin 8 countries
Study dates
Mar 2009to Aug 2010
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • CRT/Cardiac Resynchronization Therapy-Defibrillator (CRT-D) indications, New York Heart Association (NYHA) III and IV, indicated for implantable cardiac defibrillator (ICD) if CRT-D is implanted, patient consent, geographically stable

Who cannot join

  • LV lead implant attempt in last 30 days, unstable angina or acute myocardial infarction (MI) in past 30 days, coronary artery bypass graft (CABG) or percutaneous transluminal coronary angioplasty (PTCA) in past 3 months, contraindicated for transvenous pacing, heart transplant

Trial locations and contacts

31 sites in 8 countries.

31 sites in 8 countries

United States19 sites
  • Study site

    Atlanta, Georgia

  • Study site

    Atlantis, Florida

  • Study site

    Austin, Texas

  • Study site

    Baton Rouge, Louisiana

  • Study site

    Davenport, Iowa

  • Study site

    Durham, North Carolina

  • Study site

    Hinsdale, Illinois

  • Study site

    Kansas City, Kansas

  • Study site

    Lexington, Kentucky

  • Study site

    Liverpool, New York

  • Study site

    Morgantown, West Virginia

  • Study site

    Nashville, Tennessee

  • Study site

    New York, New York

  • Study site

    Pasadena, California

  • Study site

    Portland, Oregon

  • Study site

    Sacramento, California

  • Study site

    Saint Louis Park, Minnesota

  • Study site

    Saint Paul, Minnesota

  • Study site

    York, Pennsylvania

Other trials in United States

Canada5 sites
  • Study site

    Calgary, Alberta

  • Study site

    Montreal, Quebec

  • Study site

    Newmarket, Ontario

  • Study site

    Québec, Quebec

  • Study site

    Vancouver, British Columbia

Other trials in Canada

France2 sites
  • Study site

    Nantes

  • Study site

    Toulouse

Other trials in France

Australia1 site
  • Study site

    Bendigo

Other trials in Australia

Austria1 site
  • Study site

    Linz

Other trials in Austria

Denmark1 site
  • Study site

    Aarhus

Other trials in Denmark

Italy1 site
  • Study site

    Treviso

Other trials in Italy

Saudi Arabia1 site
  • Study site

    Riyadh

Other trials in Saudi Arabia

About this study

Official title: Model 4396 Left Ventricular Lead Study

This study is designed to show that a new lead, which paces the left bottom chamber (left ventricle) of the heart, is safe and effective. Using a lead on the left side of the heart has been shown to potentially improve heart failure symptoms. The shape and size of this new lead may make it a good choice for patients or physicians.

Read the detailed description

This study is a prospective, multi-center, non-randomized design. All consented patients who meet all inclusion and no exclusion criteria may receive the investigational lead. The study design has statistical power to show safety at one-month after implant. It also has statistical power to show effectiveness at one-month and three-months after implant.

Conditions
Heart Failure
Being tested
Pacing Lead
Keywords
heart failure, cardiac pacing, cardiac resynchronization therapy, left ventricular lead
Medical subject headings
Heart Failure

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Pacing LeadDevice

    implant and follow-up of study device

Study groups in NCT00853593
Study groupRoleWhat they receive
Model 4396 LV LeadExperimentalNon-randomized study.
Receives: Device: Pacing Lead

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Safety (Subjects Without a Model 4396 Lead Related Complication)

    A subject who was free of a Model 4396 lead related complication by the one month visit. All adverse events (AE) in the time frame were recorded at the subject's center and assessed by a centralized Adverse Event Advisory Committee (AEAC). The AEAC determined whether an AE was a complication and whether the event was related to the Model 4396 lead. A complication is an AE that results in death, termination of significant device function or invasive intervention (any therapy that penetrates the skin including administration of intramuscular (IM) and parenteral (IV) fluids).

    Measured: One month

  2. Efficacy: Distal Tip Electrode Voltage Threshold

    Subjects' distal tip electrode voltage threshold was collected at the one month visit. The Model 4396 was considered effective if the mean voltage threshold was less than 3.0 Volts. Voltage threshold was collected using LV tip to Right Ventricular (RV) coil configuration at 0.5 milliseconds [ms]. Voltage threshold is the minimum energy required from the device to consistently pace the ventricle.

    Measured: One month

  3. Efficacy: Proximal Ring Voltage Threshold

    Subject's proximal ring electrode voltage threshold was collected at the three months visit. The Model 4396 was considered effective if the mean voltage threshold was less than 3.0 Volts.

    Measured: Three months

Secondary outcomes (19)
  1. Subjects Successfully Implanted With Model 4396 Lead

    A successful implant occurs when the Model 4396 lead is implanted in a left ventricular vein and functions appropriately. A Model 4396 implant attempt was defined as any time when a Model 4396 lead was introduced into the body.

    Measured: During implant procedure.

  2. Subjects Successfully Implanted With Any Transvenous LV Lead After Cannulation

    A successful implant after cannulation occurs when the coronary sinus (CS) is successfully cannulated and a left ventricular lead (any transvenous LV lead) is implanted in a left ventricular vein and functions appropriately. An implant attempt of any transvenous LV lead was defined as any time when a transvenous LV lead was introduced into the body.

    Measured: During implant procedure.

  3. Subjects Successfully Implanted With Any Transvenous LV Lead

    A successful implant occurs when any transvenous LV lead is implanted in a left ventricular vein functions appropriately.

    Measured: During implant procedure.

  4. Subjects Successfully Implanted With Any Medtronic Attain Family LV Lead

    A successful implant occurs when any Medtronic Attain Family LV Lead is implanted in a left ventricular vein and functions appropriately. The Attain Family leads include the following models: 4193, 4194, 4195, 4196, and 4396.

    Measured: During implant procedure.

  5. Cannulation Time

    Cannulation time was defined as the time from insertion of the first coronary sinus (CS) cannulation catheter to the first successful CS cannulation.

    Measured: During implant procedure.

  6. Fluoroscopy Time

    The total time the fluoroscope was imaging (not including biplane fluoroscopy time).

    Measured: During implant procedure.

  7. Model 4396 Lead Placement Time

    Model 4396 lead placement time was defined as the time from insertion of the successfully implanted lead to the time when it was placed in the first acceptable pacing location.

    Measured: During implant procedure.

  8. Total Operation Time

    Total operation time was defined as time from initial incision to final closure.

    Measured: During implant procedure.

  9. Assessment of Lead Handling Characteristics Reported as Acceptable

    Implant lead handling characteristics were qualitatively assessed through physician feedback on the Implant Case Report Form (CRF). Physicians were asked for their overall assessment of the lead and results were categorized as acceptable or unacceptable. The number of acceptable responses are summarized.

    Measured: During implant procedure.

  10. Efficacy: Bipolar Voltage Threshold

    Subjects' voltage threshold in the bipolar configuration was collected at the one month visit. The Model 4396 was considered effective if the mean voltage threshold (at 0.5 milliseconds [ms]) is less than or equal to 4.0 Volts.

    Measured: 1 month

  11. Characterize Model 4396 Electrical Performance- Tip Electrode: Voltage Threshold

    Tip electrode voltage threshold at 0.5 ms was collected at implant, pre-hospital discharge and all scheduled follow-up visits. Voltage threshold values at the 6 month visit are summarized.

    Measured: 6 month

  12. Characterize Model 4396 Electrical Performance- Tip Electrode: Pacing Impedance

    Subjects' tip electrode pacing impedance (a measure of electrical resistance) was measured at implant, pre-hospital discharge and all scheduled follow-up visits. Pacing impedance at the 6 month visit is presented here.

    Measured: 6 month

  13. Characterize Model 4396 Electrical Performance- Tip Electrode: Sensing

    Tip electrode sensing, measured by R-wave amplitude, for the Model 4396 was collected at implant, pre-hospital discharge and all scheduled follow-up visits. Measurements at the 6 month visit are presented here. Sensing is the minimum energy produced by the left ventricle of the heart that the device can sense.

    Measured: 6 month

  14. Characterize Model 4396 Electrical Performance- Ring Electrode: Voltage Threshold

    Ring electrode voltage threshold at 0.5 ms was collected at implant, pre-hospital discharge and all scheduled follow-up visits. Voltage threshold at the 6 month visit is presented here.

    Measured: 6 month

  15. Characterize Model 4396 Electrical Performance- Ring Electrode: Pacing Impedance

    Subjects' ring electrode pacing impedance was measured at implant, pre-hospital discharge and all scheduled follow-up visits. Pacing impedance at the 6 month visit is presented here.

    Measured: 6 month

  16. Characterize Model 4396 Electrical Performance- Ring Electrode: Sensing

    Ring electrode sensing, measured by R-wave amplitude, for the Model 4396 was collected only at the implant procedure because the devices allowed in this study are not programmable to collect sensing measurements using the ring electrode. The analyzer was used to collect measurements.

    Measured: During implant procedure.

  17. Characterize Model 4396 Electrical Performance- Bipolar Configuration: Voltage Threshold

    Bipolar voltage threshold at 0.5 ms was collected at implant, pre-hospital discharge and all scheduled follow-up visits. Voltage threshold at the 6 month visit is reported here.

    Measured: 6 month

  18. Characterize Model 4396 Electrical Performance- Bipolar Configuration: Pacing Impedance

    Subjects' bipolar pacing impedance was measured at implant, pre-hospital discharge and all scheduled follow-up visits. Pacing impedance at the 6 month visit is presented here.

    Measured: 6 month

  19. Characterize Model 4396 Electrical Performance- Bipolar Configuration: Sensing

    Bipolar sensing, measured by R-wave amplitude, for the Model 4396 was collected at implant, pre-hospital discharge and all scheduled follow-up visits. Measurements at the 6 month visit are presented here.

    Measured: 6 month

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted2 Mar 2009 (estimated)
  2. Study startMar 2009
  3. Primary completionJul 2009
  4. Study completionAug 2010

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
197 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT00853593
Sponsor's protocol ID
116
Acronym
4396
Status last verified
Jan 2019
First submitted
26 Feb 2009
Last updated
18 Feb 2019
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

A study testing Pacing Lead in people with Heart Failure, run by Medtronic Cardiac Rhythm and Heart Failure at 31 sites in 8 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 18 Feb 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 31 sites in United States, Canada, France, Australia, Austria and 3 more countries. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 5 Dec 2011). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Safety (Subjects Without a Model 4396 Lead Related Complication) (time frame: One month). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 18 Feb 2019.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in