Heart FailureDEFEAT-PENCT00916929
Detect Fluid Early From Intra-thoracic Impedance Monitoring
A study testing Impedance Monitoring Feature in people with Heart Failure, run by Abbott Medical Devices at 34 sites in United States.
Trial summary
To demonstrate the safety and effectiveness of the Impedance Monitoring Feature in St Jude Medical cardiac devices. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Heart Failure
- What is being tested?
- Impedance Monitoring Feature
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 162 participants
- Where
- 34 sitesUnited States
- Study dates
- May 2009to Mar 2012
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Must have a St. Jude Medical (SJM) ICD or CRT-D for at least 31 days
- Must have had an episode of acute decompensated heart failure (ADHF) within the past 6 months
Who cannot join
- History of kidney disease requiring hemodialysis
- Refractory end stage heart failure
Trial locations and contacts
34 sites in 1 country.
34 sites in 1 country
United States34 sites
The Hope Heart Institute
Bellevue, Washington, 98004
University of Alabama at Birmingham
Birmingham, Alabama, 35249
Massachusetts General Hospital
Boston, Massachusetts, 02114
Deborah Heart and Lung Center
Browns Mills, New Jersey, 08015
Medical University of South Carolina
Charleston, South Carolina, 29425
Mid Carolina Cardiology
Charlotte, North Carolina, 28204
Associated Cardiovascular Consultants
Cherry Hill, New Jersey, 08034
The Cleveland Clinic Foundation
Cleveland, Ohio, 44195
Colorado Cardiac Alliance, LLC
Colorado Springs, Colorado, 80907
Ohio State University
Columbus, Ohio, 43210
Saint Vincent Consultants in Cardiovascular Diseases
Erie, Pennsylvania, 16502
Mercy Hospital Fairfield
Fairfield, Ohio, 45014
Cardiology Consultants of East Michigan
Flint, Michigan, 48532
Glendale Memorial Hospital and MC
Glendale, California, 91204
Hattiesburg Clinic
Hattiesburg, Mississippi, 39401
Shands Jacksonville
Jacksonville, Florida, 32209
Watson Clinic Center
Lakeland, Florida, 33805
Thoracic Cardiovascular HC Found. (Sparrow Research)
Lansing, Michigan, 48910
Arkansas Cardiology, PA
Little Rock, Arkansas, 72005
Little Rock Cardiology Clinic, PA
Little Rock, Arkansas, 72211
Methodist University Hospital
Memphis, Tennessee, 38104
St Thomas Hospital
Nashville, Tennessee, 37205
New York Presbyterian Hospital / Cornell University
New York, New York, 10021
Sentara Norfolk General Hospital
Norfolk, Virginia, 23507
Cardiac Rhythm Specialists, Inc
Northridge, California, 91325
Orlando Heart Center
Orlando, Florida, 32806
Hospital of the University of PA
Philadelphia, Pennsylvania, 19104
Thomas Jefferson University Hospital
Philadelphia, Pennsylvania, 19107
Arizona Arrhythmia Research Center
Phoenix, Arizona, 85016
South Texas Cardiovascular Consultants
San Antonio, Texas, 78201
Tallahassee Research Institute
Tallahassee, Florida, 32308
North Mississippi Medical Center
Tupelo, Mississippi, 38801
Iowa Heart Center
West Des Moines, Iowa, 50266
York Hospital
York, Pennsylvania, 17403
About this study
To demonstrate the safety and effectiveness of the Impedance Monitoring Feature in St Jude Medical cardiac devices.
- Conditions
- Heart Failure
- Being tested
- Impedance Monitoring Feature
- Medical subject headings
- Heart Failure
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Impedance Monitoring FeatureDevice
Algorithm in cardiac device that measures intra-thoracic impedance from the implanted leads
| Study group | Role | What they receive |
|---|---|---|
| Implantable Cardioverter Defibrillator (ICD) | Other | Impedance Monitoring Feature in an Implantable Cardioverter Defibrillator (ICD). Receives: Device: Impedance Monitoring Feature |
| Cardiac Resynchronization Therapy (CRT-D) | Other | Impedance Monitoring Feature in a Cardiac Resynchronization Therapy (CRT-D) device. Receives: Device: Impedance Monitoring Feature |
What is being measured
The primary outcome is the main question the study is designed to answer.
False Positive Rate
False Positive Rate is the the number of departures from the device's programmed threshold that are unrelated to a heart failure event. The False Positive Rate per patient year of follow up should be less than 1.5.
Measured: 6-months
Secondary outcomes (1)
Sensitivity
Sensitivity is defined as the ability of the algorithm to detect heart failure events. Sensitivity is calculated as the number of heart failure events detected by the algorithm (true positives) divided by the total number of heart failure events (true positives + false positives).
Measured: 6-months
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted10 Jun 2009 (estimated)
- Study startMay 2009
- Primary completionFeb 2011
- Study completionMar 2012
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Heart FailureStudies of the same condition, open and under way.
- More trials run by Abbott Medical DevicesThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Diagnosis
- Allocation
- Non-randomised
- Intervention model
- Parallel groups
- Masking
- Triple blind
- Who is masked
- Participant, Care provider, Investigator
- Enrolment
- 162 participants (actual)
Sponsor and investigators
- Lead sponsor
- Abbott Medical Devices
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
- Investigators
- Edwin K Heist, MD, PhD, Study chair, Massachusetts General Hospital
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT00916929
- Sponsor's protocol ID
- 40006062/D
- Acronym
- DEFEAT-PE
- Status last verified
- Feb 2019
- First submitted
- 8 Jun 2009
- Last updated
- 19 Feb 2019
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
Study at a glance
From the registry record, last updated by the sponsor on 19 Feb 2019. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Detect Fluid Early From Intra-thoracic Impedance Monitoring
- ClinicalTrials.gov ID
- NCT00916929
- Acronym / sponsor ID
- DEFEAT-PE / 40006062/D
- Conditions
- Heart Failure
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 162 participants (actual)
- Lead sponsor
- Abbott Medical DevicesIndustry (pharma, biotech, devices)
- Study start
- May 2009
- Primary completion
- Feb 2011 (actual)
- Study completion
- Mar 2012 (actual)
- First posted
- 10 Jun 2009 (estimated)
- Results
- Posted 24 Jun 2013
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
To demonstrate the safety and effectiveness of the Impedance Monitoring Feature in St Jude Medical cardiac devices.
Study design
- Primary purpose
- Diagnosis
- Allocation
- Non-randomised
- Intervention model
- Parallel groups
- Masking
- Triple blind (participant, care provider, investigator)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Implantable Cardioverter Defibrillator (ICD) | Other | Device: Impedance Monitoring FeatureImpedance Monitoring Feature in an Implantable Cardioverter Defibrillator (ICD). |
| Cardiac Resynchronization Therapy (CRT-D) | Other | Device: Impedance Monitoring FeatureImpedance Monitoring Feature in a Cardiac Resynchronization Therapy (CRT-D) device. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (2 listed)
- Must have a St. Jude Medical (SJM) ICD or CRT-D for at least 31 days
- Must have had an episode of acute decompensated heart failure (ADHF) within the past 6 months
Key exclusion criteria (2 listed)
- History of kidney disease requiring hemodialysis
- Refractory end stage heart failure
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Impedance Monitoring FeatureAlgorithm in cardiac device that measures intra-thoracic impedance from the implanted leads | Device | Cardiac Resynchronization Therapy (CRT-D), Implantable Cardioverter Defibrillator (ICD) |
Primary outcome
False Positive Rate
Time frame: 6-months
False Positive Rate is the the number of departures from the device's programmed threshold that are unrelated to a heart failure event. The False Positive Rate per patient year of follow up should be less than 1.5.
1 secondary outcome
Sensitivity
Time frame: 6-months
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 19 Feb 2019.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
Related trials for Heart Failure
RecruitingPhase 4
The Effects of Ferric Derisomaltose in Patients With Acute Heart Failure and Iron Deficiency on Exercise Capacity and Quality of Life(COREVIVE-HFrEF)
Heart Failure With Reduced Ejection Fraction, Iron Deficiency
Not yet recruitingPhase 2
A Study of Bimagrumab in Participants With HFpEF
Heart Failure, Diastolic, Diastolic Heart Failure
Recruiting
Evaluation of a French Hospital Discharge Therapeutic Education Program by Nurses for Heart Failure Patient (EduSor)
Heart Failure
Active, not recruiting
Assessment of Combined CCM and ICD Device in HFrEF
Heart Failure, Heart Failure With Reduced Ejection Fraction
Not yet recruiting
RELIEVE-HFrEF TRIAL: REducing Lung congestIon Symptoms Using the v-wavE Shunt in adVancEd Heart Failure With Reduced EF
Heart Failure and Reduced Ejection Fraction, Heart Failure Chronic
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Edwin K Heist, MD, PhD study chair
Massachusetts General Hospital
34 sites in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Impedance Monitoring Feature in people with Heart Failure, run by Abbott Medical Devices at 34 sites in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 2 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 19 Feb 2019. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 34 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 24 Jun 2013). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
False Positive Rate (time frame: 6-months). It is what the study was designed to measure, not a finding.