Electric Impedance TomographyNCT01168557
Stress Echo and Electric Impedance Tomography (EIT) Pilot Study
A study testing EIT (Electrical Impedance Tomography) in people with Electric Impedance Tomography and Stress Echocardiography, run by RWTH Aachen University at 1 site in Germany.
Trial summary
Electric Impedance Tomography (EIT) is a technique based on the injection of small currents and voltage measurements using electrodes on the skin surface generating cross-sectional images representing impedance change in a slice of the thorax.It is a radiation free, non- invasive and portable lung imaging technique. Read more about this study
- Who is it for?
- Ages 18 years and older, men only
- Conditions the study is for
- Electric Impedance Tomography, Stress Echocardiography, Cardiac Output
- What is being tested?
- EIT (Electrical Impedance Tomography)
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 8 participants
- Where
- 1 siteGermany
- Study dates
- Jul 2010to Sep 2012
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- Men only
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- male patients >18 years who routinely have to undergo stress echocardiography
- ejection fraction >50%
- able to give informed consent
Who cannot join
- restrictive lung diseases
- active implants and heart pace makers and implanted converter defibrillator
- foreign metallic objects in the thorax
- Not adequate quality of imaging of stress echocardiography
Trial locations and contacts
1 site in 1 country.
Germany1 site
Medical Clinic I, University Hospital Aachen
Aachen, North Rhine-Westphalia, 52074
About this study
Official title: Stress Echo and Electric Impedance Tomography (EIT) Pilot Study to Investigate the Influence of Cardiac Output on Parameters of the EIT
Electric Impedance Tomography (EIT) is a technique based on the injection of small currents and voltage measurements using electrodes on the skin surface generating cross-sectional images representing impedance change in a slice of the thorax.It is a radiation free, non- invasive and portable lung imaging technique. Impedance changes in lung ventilation as well as in lung perfusion can be shown. To investigate lung perfusion by EIT measurements, cardiac output measured by routinely undergone stress echocardiography at rest and stress are compared to parameters measured by EIT.
Read the detailed description
In 15 patients who routinely have to undergo stress echocardiography Electric Impedance Tomography (EIT) measurements are performed parallel to measurements of cardiac output at rest, low stress generated by a dosage of 10µg dobutamine, at peak generated by a dosage between 20-40µg dobutamine (individually)and at relaxation(pulse <100). Therefore prior to the examination a rubber belt with 16 integrated electrocardiographic electrodes is placed around the thorax at the fifth or sixth intercostal space, connected with an EIT device. EIT data are generated by application of a small alternating electrical current of 5mA and 50kHz and are stored and analyzed offline on a personal computer. Changes due to ventilation as well as changes due to perfusion will be measured by EIT. Cardiac output will be retrospectively compared to parameters of the EIT. EIT data are stored in a database and pseudonymized evaluated in relationship to routinely measured stress echocardiography data.
- Conditions
- Electric Impedance Tomography, Stress Echocardiography, Cardiac Output, Ventilation, Perfusion
- Being tested
- EIT (Electrical Impedance Tomography)
- Keywords
- electric impedance tomography, stress-echocardiography, ventilation, perfusion
- Medical subject headings
- Respiratory Aspiration
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
EIT (Electrical Impedance Tomography)Device
EIT measurement; concomitant to stress echocardiography (about 45 minutes) EIT measurements (1 minute) are performed at rest, at low dobutamine stress, at peak and at relaxation
| Study group | Role | What they receive |
|---|---|---|
| Stress-Echo and EIT | Experimental | Patients who routinely undergo stress-echocardiography will additionally be measured by EIT using a rubber belt, which will be placed around their chest Receives: Device: EIT (Electrical Impedance Tomography) |
What is being measured
The primary outcome is the main question the study is designed to answer.
influence of cardiac output on EIT measurements (feasibility)
It will be investigated if a change in cardiac output, which is measured by stress-echocardiography by dobutamine, has an influence on the parameters determined by EIT measurements.
Measured: single measurement of 45 minutes
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted23 Jul 2010 (estimated)
- Study startJul 2010
- Primary completionSep 2012
- Study completionSep 2012
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Electric Impedance TomographyStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Basic science
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 8 participants (actual)
Sponsor and investigators
- Lead sponsor
- RWTH Aachen University
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor
- Investigators
- Stefan Krueger, MD, Assoc. Prof., Principal investigator, RWTH Aachen University of Cardiology, Pneumology and Angiology, Aachen, NRW, Germany
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT01168557
- Sponsor's protocol ID
- Stress Echo - EIT
- Status last verified
- Nov 2013
- First submitted
- 22 Jul 2010
- Last updated
- 27 Nov 2013
- Results posted
- No
- Data monitoring committee
- No
Study at a glance
From the registry record, last updated by the sponsor on 27 Nov 2013. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Stress Echo and Electric Impedance Tomography (EIT) Pilot Study to Investigate the Influence of Cardiac Output on Parameters of the EIT
- ClinicalTrials.gov ID
- NCT01168557
- Acronym / sponsor ID
- Stress Echo - EIT
- Conditions
- Electric Impedance Tomography, Stress Echocardiography, Cardiac Output, Ventilation, Perfusion
- Keywords
- electric impedance tomography, stress-echocardiography, ventilation, perfusion
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 8 participants (actual)
- Lead sponsor
- RWTH Aachen UniversityUniversity, hospital or other
- Study start
- Jul 2010
- Primary completion
- Sep 2012 (actual)
- Study completion
- Sep 2012 (actual)
- First posted
- 23 Jul 2010 (estimated)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Electric Impedance Tomography (EIT) is a technique based on the injection of small currents and voltage measurements using electrodes on the skin surface generating cross-sectional images representing impedance change in a slice of the thorax.It is a radiation free, non- invasive and portable lung imaging technique. Impedance changes in lung ventilation as well as in lung perfusion can be shown. To investigate lung perfusion by EIT measurements, cardiac output measured by routinely undergone stress echocardiography at rest and stress are compared to parameters measured by EIT.
Detailed description
In 15 patients who routinely have to undergo stress echocardiography Electric Impedance Tomography (EIT) measurements are performed parallel to measurements of cardiac output at rest, low stress generated by a dosage of 10µg dobutamine, at peak generated by a dosage between 20-40µg dobutamine (individually)and at relaxation(pulse <100). Therefore prior to the examination a rubber belt with 16 integrated electrocardiographic electrodes is placed around the thorax at the fifth or sixth intercostal space, connected with an EIT device. EIT data are generated by application of a small alternating electrical current of 5mA and 50kHz and are stored and analyzed offline on a personal computer. Changes due to ventilation as well as changes due to perfusion will be measured by EIT. Cardiac output will be retrospectively compared to parameters of the EIT. EIT data are stored in a database and pseudonymized evaluated in relationship to routinely measured stress echocardiography data.
Study design
- Primary purpose
- Basic science
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Stress-Echo and EIT | Experimental | Device: EIT (Electrical Impedance Tomography)Patients who routinely undergo stress-echocardiography will additionally be measured by EIT using a rubber belt, which will be placed around their chest |
Who may qualify
- Ages
- 18 years and older
- Sex
- Male only
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (3 listed)
- male patients >18 years who routinely have to undergo stress echocardiography
- ejection fraction >50%
- able to give informed consent
Key exclusion criteria (4 listed)
- restrictive lung diseases
- active implants and heart pace makers and implanted converter defibrillator
- foreign metallic objects in the thorax
- Not adequate quality of imaging of stress echocardiography
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| EIT (Electrical Impedance Tomography)EIT measurement; concomitant to stress echocardiography (about 45 minutes) EIT measurements (1 minute) are performed at rest, at low dobutamine stress, at peak and at relaxation | Device | Stress-Echo and EIT |
Primary outcome
influence of cardiac output on EIT measurements (feasibility)
Time frame: single measurement of 45 minutes
It will be investigated if a change in cardiac output, which is measured by stress-echocardiography by dobutamine, has an influence on the parameters determined by EIT measurements.
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 27 Nov 2013.
Related trials for Electric Impedance Tomography
Not yet recruitingObservational study
Development and Prospective Validation of the AutoEIT Foundation Model for ARDS
ARDS (Acute Respiratory Distress Syndrome), Respiratory Failure
Not yet recruiting
Perioperative High-flow Nasal Oxygen in Patients Undergoing Robotic Surgery
Robotic Surgery
RecruitingObservational study
Observational Study of Electrical Impedance Tomography for Non-Invasive Assessment of Cerebral Autoregulation in Neurocritical Patients
Homeostasis
Not yet recruitingPhase 1/2
Influence of Lung Volume Optimization Maneuver on Cardiac Output and Lung Compliance in Ventilated Children With Congenital Heart Disease Undergoing Surgical Repair
Congenital Heart Disease, Cardiopulmonary Bypass
RecruitingObservational study
A Smart Wearable Device for Monitoring Urination Function in Benign Prostatic Hyperplasia
Benign Prostatic Hyperplasia
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Stefan Krueger, MD, Assoc. Prof. principal investigator
RWTH Aachen University of Cardiology, Pneumology and Angiology, Aachen, NRW, Germany
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing EIT (Electrical Impedance Tomography) in people with Electric Impedance Tomography and Stress Echocardiography, run by RWTH Aachen University at 1 site in Germany.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 27 Nov 2013. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Germany. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
influence of cardiac output on EIT measurements (feasibility) (time frame: single measurement of 45 minutes). It is what the study was designed to measure, not a finding.