Fluid Leakage Past Tracheal Tube Cuff : Effect of Suctioning Manoeuvre
A study testing Hi-Lo Evac tube (Mallinckrodt Inc, USA) (Instilled blue dye) in people with Deep Sedation and Respiration, Artificial, run by Centre Hospitalier de Roanne at 1 site in France.
Trial summary
The leakage of oropharyngeal secretions around high-volume low-pressure tracheal tube cuffs is usually considered as a major risk factor for bacterial tracheal colonization and subsequent development of ventilator-associated pneumonia. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Deep Sedation, Respiration, Artificial
- What is being tested?
- Hi-Lo Evac tube (Mallinckrodt Inc, USA) (Instilled blue dye)
- Type of treatment
- Medical device
- How many people
- 25 participants
- Where
- 1 siteFrance
- Study dates
- Sep 2010to Jan 2011
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- intubated with a Hi-Lo Evac tube since less than 48 hours
- under continuous sedation
- written consent signed
Who cannot join
- hemodynamic instability
- allergy of blue dye
Trial locations and contacts
1 site in 1 country.
About this study
The leakage of oropharyngeal secretions around high-volume low-pressure tracheal tube cuffs is usually considered as a major risk factor for bacterial tracheal colonization and subsequent development of ventilator-associated pneumonia. The rate of leakage around the cuff is related to the pressure differential across the cuff, namely the difference between the pressure of the subglottic fluid above the cuff and the tracheal pressure under the cuff. Consequently, positive end-expiratory pressure (PEEP) improves the sealing around the cuff towards fluid leakage. However, this preventive effect of PEEP is compromised during prolonged mechanical ventilation by tracheal suctioning manoeuvre, which may enhance fluid leakage, by decreasing tracheal pressure. Indeed, in a benchtop model, a suctioning manoeuvre, without disconnection of the ventilator, induced a constant fluid leakage past a high-volume low-pressure tracheal tube cuff when performed with a high level of suction pressure (- 400 mbar) and a large size of suction catheter size (16 French).
Read the detailed description
This clinical study aims to confirm these experimental data in patients under mechanical ventilation.
- Conditions
- Deep Sedation, Respiration, Artificial
- Being tested
- Hi-Lo Evac tube (Mallinckrodt Inc, USA) (Instilled blue dye)
- Keywords
- deep sedation, respiratory resuscitation, Hi-Lo Evac tube, suctioning manoeuvre
- Medical subject headings
- Respiratory Aspiration
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Hi-Lo Evac tube (Mallinckrodt Inc, USA) (Instilled blue dye)Device
Also called Hi-Lo Evac tube (Mallinckrodt Inc, USA), Blue dye 2.5% (Guerbet, France)
The cuff pressure will be first checked and reset at 30 cm H2O if needed. Next, blue dye will be instilled just above the cuff through the lumen ending in the subglottic area of the Hi-Lo Evac tube. Then a suctioning manoeuvre will be performed with a suction pressure of - 400 mbar through a 16 French suction catheter. Thereafter, a fiberoptic bronchoscopy will be performed, looking for the presence of blue dye in the trachea and/or the bronchi.
What is being measured
The primary outcome is the main question the study is designed to answer.
Presence of blue dye in the trachea and/or the bronchi after the suctioning manoeuvre.
The cuff pressure will be first checked and reset at 30 cm H2O if needed. Next, blue dye will be instilled just above the cuff through the lumen ending in the subglottic area of the Hi-Lo Evac tube. Then a suctioning manoeuvre will be performed with a suction pressure of - 400 mbar through a 16 French suction catheter. Thereafter, a fiberoptic bronchoscopy will be performed, looking for the presence of blue dye in the trachea and/or the bronchi.
Measured: 48 hours after the beginning of the mechanical ventilation
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted27 Jul 2010 (estimated)
- Study startSep 2010
- Primary completionJan 2011
- Study completionJan 2011
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (1)Papers the sponsor has linked to this study.
- Other trials for Deep SedationStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Prevention
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 25 participants (actual)
Sponsor and investigators
- Lead sponsor
- Centre Hospitalier de Roanne
- Sponsor type
- University, hospital or other
- Investigators
- Pascal BEURET, MD, Principal investigator, CH de Roanne
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT01170156
- Sponsor's protocol ID
- 2010REACHR01
- Other IDs
- 2010-A00539-30
- Status last verified
- Mar 2011
- First submitted
- 22 Jul 2010
- Last updated
- 7 Mar 2011
- Results posted
- No
- Data monitoring committee
- No
Publications
Beuret P, Philippon B, Fabre X, Kaaki M. Effect of tracheal suctioning on aspiration past the tracheal tube cuff in mechanically ventilated patients. Ann Intensive Care. 2012 Nov 7;2(1):45. doi: 10.1186/2110-5820-2-45.
PubMed 23134813
Study at a glance
From the registry record, last updated by the sponsor on 7 Mar 2011. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Fluid Leakage Past Tracheal Tube Cuff : Effect of Suctioning Manoeuvre
- ClinicalTrials.gov ID
- NCT01170156
- Acronym / sponsor ID
- 2010REACHR01
- Conditions
- Deep Sedation, Respiration, Artificial
- Keywords
- deep sedation, respiratory resuscitation, Hi-Lo Evac tube, suctioning manoeuvre
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 25 participants (actual)
- Lead sponsor
- Centre Hospitalier de RoanneUniversity, hospital or other
- Study start
- Sep 2010
- Primary completion
- Jan 2011 (actual)
- Study completion
- Jan 2011 (actual)
- First posted
- 27 Jul 2010 (estimated)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The leakage of oropharyngeal secretions around high-volume low-pressure tracheal tube cuffs is usually considered as a major risk factor for bacterial tracheal colonization and subsequent development of ventilator-associated pneumonia. The rate of leakage around the cuff is related to the pressure differential across the cuff, namely the difference between the pressure of the subglottic fluid above the cuff and the tracheal pressure under the cuff. Consequently, positive end-expiratory pressure (PEEP) improves the sealing around the cuff towards fluid leakage. However, this preventive effect of PEEP is compromised during prolonged mechanical ventilation by tracheal suctioning manoeuvre, which may enhance fluid leakage, by decreasing tracheal pressure. Indeed, in a benchtop model, a suctioning manoeuvre, without disconnection of the ventilator, induced a constant fluid leakage past a high-volume low-pressure tracheal tube cuff when performed with a high level of suction pressure (- 400 mbar) and a large size of suction catheter size (16 French).
Detailed description
This clinical study aims to confirm these experimental data in patients under mechanical ventilation.
Study design
- Primary purpose
- Prevention
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (3 listed)
- intubated with a Hi-Lo Evac tube since less than 48 hours
- under continuous sedation
- written consent signed
Key exclusion criteria (2 listed)
- hemodynamic instability
- allergy of blue dye
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Hi-Lo Evac tube (Mallinckrodt Inc, USA) (Instilled blue dye)Also: Hi-Lo Evac tube (Mallinckrodt Inc, USA), Blue dye 2.5% (Guerbet, France)The cuff pressure will be first checked and reset at 30 cm H2O if needed. Next, blue dye will be instilled just above the cuff through the lumen ending in the subglottic area of the Hi-Lo Evac tube. Then a suctioning manoeuvre will be performed with a suction pressure of - 400 mbar through a 16 French suction catheter. Thereafter, a fiberoptic bronchoscopy will be performed, looking for the presence of blue dye in the trachea and/or the bronchi. | Device | – |
Primary outcome
Presence of blue dye in the trachea and/or the bronchi after the suctioning manoeuvre.
Time frame: 48 hours after the beginning of the mechanical ventilation
The cuff pressure will be first checked and reset at 30 cm H2O if needed. Next, blue dye will be instilled just above the cuff through the lumen ending in the subglottic area of the Hi-Lo Evac tube. Then a suctioning manoeuvre will be performed with a suction pressure of - 400 mbar through a 16 French suction catheter. Thereafter, a fiberoptic bronchoscopy will be performed, looking for the presence of blue dye in the trachea and/or the bronchi.
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 7 Mar 2011.
- Derived from this study: PubMed 23134813Beuret P, Philippon B, Fabre X, Kaaki M. Effect of tracheal suctioning on aspiration past the tracheal tube cuff in mechanically ventilated patients. Ann Intensive Care. 2012 Nov 7;2(1):45. doi: 10.1186/2110-5820-2-45.
Related trials for Deep Sedation
Not yet recruiting
Bispectral Index-Guided Propofol Sedation During Endoscopic Retrograde Cholangiopancreatography
Deep Sedation, Endoscopic Retrograde Cholangiopancreatography
Active, not recruiting
Weighted Blankets for Agitation in Adults in the Intensive Care Unit
Psychomotor Agitation, Critical Illness
Recruiting
A Comparison of Pulsed Field Ablation (PFA) vs Cryoablation for Atrial Fibrillation Under Deep Sedation
Atrial Fibrillation (AF)
Not yet recruitingObservational study
Sugammadex Timing and Recovery Profile in Pediatric Adenotonsillectomy
Neuromuscular Blockade, Delayed Emergence From Anesthesia
Recruiting
Restoration of Consciousness With Ultrasonic Deep Brain Stimulation During Anesthetic Sedation
Mental Function
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Pascal BEURET, MD principal investigator
CH de Roanne
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Hi-Lo Evac tube (Mallinckrodt Inc, USA) (Instilled blue dye) in people with Deep Sedation and Respiration, Artificial, run by Centre Hospitalier de Roanne at 1 site in France.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 7 Mar 2011. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Presence of blue dye in the trachea and/or the bronchi after the suctioning manoeuvre. (time frame: 48 hours after the beginning of the mechanical ventilation). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT01170156, which also lists 1 linked publication.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 7 Mar 2011.