Safety and Effectiveness Study of a Closed Loop System Maintaining Patients' Glucose Levels During an Overnight Period

A study testing Closed Loop Procedure (Medtronic) in people with Type 1 Diabetes, run by Medtronic MiniMed, Inc. at 1 site in United Kingdom.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Currently, patients diagnosed with type 1 diabetes rely on either finger stick or sensor glucose readings when making their insulin dosage decisions. Designing a computerized system that mimics the way insulin is produced naturally in a person who does not have type 1 diabetes holds many challenges; all of which… Read more about this study

Who is it for?
Ages 18 to 70 years
Conditions the study is for
Type 1 Diabetes
What is being tested?
Closed Loop Procedure (Medtronic)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
9 participants
Where
1 siteUnited Kingdom
Study dates
May 2013to Dec 2014
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 70 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • A clinical diagnosis of Type 1 diabetes as determined by the Investigator for a minimum of 36 months prior to enrollment
  • Weight is between 50 and 120 kg
  • Subject has used a Medtronic insulin pump for at least the previous 6 months
  • Subject has an glycated hemoglobin (A1C) value ≤ 9.0% demonstrated at the time of enrollment
  • Subject uses a rapid-acting analogue insulin in his/her pump

Who cannot join

  • Female subject who has a positive serum pregnancy screening test, or who plans to become pregnant during the course of the study
  • Subject has a history of hypoglycemic seizure or hypoglycemic coma within the past 12 months
  • Subject has an exclusionary cardiac disorder

Trial locations and contacts

1 site in 1 country.

United Kingdom1 site
  • King's College London

    London, SE5 9PJ

Other trials in United Kingdom

About this study

Official title: Feasibility and Safety of a Closed Loop Insulin Delivery System (Aka AAGC) With an Artificially Induced Calibration Error During the Overnight Period

Currently, patients diagnosed with type 1 diabetes rely on either finger stick or sensor glucose readings when making their insulin dosage decisions. Designing a computerized system that mimics the way insulin is produced naturally in a person who does not have type 1 diabetes holds many challenges; all of which cannot be addressed in just one study. The purpose of the Medtronic Overnight Closed Loop study is to assess the performance of a system designed to automatically infuse the correct insulin dose during the hours that the patient is sleeping.

The system consists of an insulin pump that provides insulin to the patient through an infusion set. A sensor inserted just under the patient's skin measures glucose levels and a transmitter sends this information to the pump. To enable the sensor to register the glucose information correctly, it must be set (calibrated) by a finger stick blood glucose 3-4 times a day. In the commercially released system, the physician would recommend the continuous background (basal) insulin rates and the patient would be required to make decisions regarding extra insulin (bolus) for meals or as a response to high glucose levels.

Read more

In the Closed Loop System, a mobile control system is added to these devices. This consists of an Android phone, a closed loop algorithm and a translator. This system is designed to translate the sensor information and direct the pump to provide the required dosage of insulin automatically without requiring input from the patient.

Reliable calibration has proved challenging, and so it is important that the system function safely, even when calibration is inaccurate. In this study, a calibration error will be introduced under very controlled circumstances. This testing will identify if the system can maintain acceptable overnight glucose levels, regardless of whether or not calibration is ideal.

Conditions
Type 1 Diabetes
Being tested
Closed Loop Procedure (Medtronic)
Keywords
United Kingdom Overnight closed loop
Medical subject headings
Diabetes Mellitus, Type 1

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Closed Loop Procedure (Medtronic)Device

    Also called AAGC

    The procedure A is the device running under normal conditions, procedure B will incorporate an induced calibration error.

Study groups in NCT01712594
Study groupRoleWhat they receive
Closed Loop Procedure ABActive comparatorClosed loop procedure with normal calibration first followed by closed loop procedure B with calibration error induced.
Receives: Device: Closed Loop Procedure (Medtronic)
Closed loop Procedure BAActive comparatorClosed loop procedure with an induced calibration error first followed by closed loop procedure with normal calibration.
Receives: Device: Closed Loop Procedure (Medtronic)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Amount of time spent in the target range defined as 3.8-8.8 mmol/l (70 to 160 mg/dL) YSI glucose during in-clinic overnight visits

    Subjects will arrive at the research facility between 17:00 and 18:00. Plasma glucose will be measured and they will have their evening meal of no more than 80 g of carbohydrate. They will bolus for this meal using their usual Bolus Wizard (Medtronic) settings. After the meal, an IV catheter will be inserted for the collection of venous blood samples. The Mobile Control System will be connected and closed-loop control will be started approximately 4 hours after dinner. The subject will remain under closed-loop control overnight (approximately 11 p.m.-6 a.m.). During one of the Closed Loop nights the system will be challenged by a calibration error of 30%. At approximately 7 am (before breakfast), the Mobile Control System will be turned off and the patient will be provided with a breakfast (approximately 40-70 gm of carbohydrate) and will take a bolus according to the Bolus Wizard. The patient will return to their usual pump treatment with the MiniMed Paradigm Veo System.

    Measured: Two in-patient nights

Secondary outcomes (2)
  1. Glucose Area under the curve (AUC) below 3.8 mmol/l (70 mg/dL ) YSI glucose obtained during in-clinic overnight visits

    Blood glucose samples will be obtained from the line approximately every 30 minutes, or every 5-15 minutes at treating physician's discretion, if either the sensor or a previous glucose reading is < 81mg/dL or > 360mg/dL (< 3.5 mmol/l or > 20 mmol/l). Plasma samples will be analyzed using a YSI 2300 Glucose and Lactate Analyzer between 18:00 and 6:00.

    Measured: Two overnight closed loop sessions

  2. Glucose AUC above 8.8mmol/l (160 mg/dL) YSI glucose obtained during in-clinic overnight visits

    Blood glucose samples will be obtained from the line approximately every 30 minutes, or every 5-15 minutes at treating physician's discretion, if either the sensor or a previous glucose reading is < 81mg/dL or > 360mg/dL (< 3.5 mmol/l or > 20 mmol/l). Plasma samples will be analyzed using a YSI 2300 Glucose and Lactate Analyzer (YSI, Inc., OH, USA) between 18:00 and 6:00.

    Measured: Two overnight closed loop sessions

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted23 Oct 2012 (estimated)
  2. Study startMay 2013
  3. Primary completionDec 2013
  4. Study completionDec 2014

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Non-randomised
Intervention model
Crossover
Masking
None (open label)
Enrolment
9 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01712594
Sponsor's protocol ID
CEP223
Status last verified
May 2017
First submitted
19 Oct 2012
Last updated
12 May 2017
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Closed Loop Procedure (Medtronic) in people with Type 1 Diabetes, run by Medtronic MiniMed, Inc. at 1 site in United Kingdom.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 12 May 2017. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Amount of time spent in the target range defined as 3.8-8.8 mmol/l (70 to 160 mg/dL) YSI glucose during in-clinic overnight visits (time frame: Two in-patient nights). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.