Hydrophobic Acrylic Toric IOL Designed to Reduce the Effects of Preoperative Corneal Astigmatism

A study testing enVista® One-Piece Hydrophobic Acrylic Toric IOL in people with Astigmatism, run by Bausch & Lomb Incorporated at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The investigational enVista IOL, Model MX60Tis equipped with toric optics and axis marks on the posterior surface to minimize the possibility of surgically induced astigmatism. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Astigmatism
What is being tested?
enVista® One-Piece Hydrophobic Acrylic Toric IOL
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
191 participants
Where
1 siteUnited States
Study dates
6 Mar 2013to Dec 2016
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Subjects must have a clinically significant cataract in the study eye that is considered amenable to treatment with standard phacoemulsification cataract extraction.
  • Subjects must require a lens power from 16.0 D to 27.0 D in the study eye.
  • Subjects must be willing to discontinue contact lens wear for the duration of the study and, demonstrate stability prior to biometry and surgery in the study eye.
  • Subjects must have a best corrected distance visual acuity (BCDVA) projected to be better than 20/30 after toric IOL implantation in the study eye.
  • Subjects must have predicted postoperative corneal astigmatism between 0.90D and 2.40D in the study eye, as determined by the Toric Calculator.
  • Subjects must have a minimum preoperative corneal astigmatism of 0.90 D in the study eye.
  • Subjects must have a dilated preoperative pupil size greater than or equal to 5.0 mm in the study eye.

Who cannot join

  • Subjects who have had previous intraocular or corneal surgery in the study eye.
  • Subjects with any clinically significant ocular pathology which would be expected to reduce potential BCDVA to 20/30 or worse in the study eye during the course of the study.
  • Subjects with any anterior segment pathology in the study eye likely to increase the risk of complications from phacoemulsification cataract surgery.
  • Subjects with associated ocular conditions which could affect the stability of the IOL in the study eye.
  • Subjects with a traumatic cataract in the study eye.
  • Subjects with uncontrolled glaucoma in the study eye.
Show all 12Show fewer
  • Subjects with clinically significant irregular corneal astigmatism in the study eye, as determined by the medical judgment of the Investigator.
  • Subjects with clinically significant corneal pathology potentially affecting topography of either eye.
  • Subjects with any ocular condition that could prevent the possibility of a visual outcome better than 20/30 in the study eye.
  • Subjects with a difference of greater than 0.50D of corneal astigmatism as measured with the IOL Master and the topographer in the study eye using vector analysis.
  • Subjects with an anterior chamber depth <2mm in the study eye.
  • Subjects with potential BCDVA of 20/200 or worse in the fellow eye.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Bausch & Lomb

    Irvine, California, 92618

Other trials in United States

About this study

Official title: To Evaluate the Safety and Effectiveness of the EnVista® One-Piece Hydrophobic Acrylic Toric IOL Designed to Reduce the Effects of Preoperative Corneal Astigmatism on Postoperative Refraction Following Cataract Surgery

The investigational enVista IOL, Model MX60Tis equipped with toric optics and axis marks on the posterior surface to minimize the possibility of surgically induced astigmatism.

Read the detailed description

The investigational enVista Toric IOL, Model MX60T, and the currently approved Model MX60, are identical with the exception of the addition of toric optics and axis marks on the posterior optic surface for the correction of astigmatism on the Model MX60T. Both lenses have an aspheric biconvex optic with a modified C haptic monoblock, single-piece design. The design and material of both the MX60 and MX60T allow them to be folded and inserted through small incisions to minimize the possibility of surgically induced astigmatism (SIA).

Conditions
Astigmatism
Being tested
enVista® One-Piece Hydrophobic Acrylic Toric IOL, enVista control lens
Medical subject headings
Astigmatism

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • enVista® One-Piece Hydrophobic Acrylic Toric IOLDevice

    Test lens having a toric cylinder power of either 1.25 diopters (D), 2.00 D, or 2.75 D

  • enVista control lensDevice

    Spherical Control lens

Study groups in NCT01852084
Study groupRoleWhat they receive
enVista® One-Piece Hydrophobic Acrylic Toric IOLExperimentalToric cylinder power of either 1.25 diopters (D), 2.00 D, or 2.75 D
Receives: Device: enVista® One-Piece Hydrophobic Acrylic Toric IOL
enVista control lensExperimentalSpherical control lens
Receives: Device: enVista control lens

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mean Dioptric Change in Cylinder

    Dioptric change in cylinder from the preoperative keratometric cylinder magnitude to postoperative manifest refraction cylinder magnitude (referenced to the corneal plane).

    Measured: Baseline, Final Visit (scheduled within window: days 120-180)

  2. Percentage of Participants With Intraocular Lens (IOL) Rotation of Less Than or Equal to Five Degrees

    Intraocular Lens (IOL) rotation of less than or equal to five degrees between Forms 3 (scheduled within window: days 30-60) and final visit (scheduled within window: days 120-180).

    Measured: Form 3 Visit (window: days 30-60), Final Visit (window: days 120-180)

  3. Mean Best-corrected Distance Visual Acuity (BCDVA) at Final Visit

    Best-corrected distance visual acuity (BCDVA) at Final Visit, expressed as logarithm of the minimum angle of resolution (logMAR). Corrected means spectacle corrected. Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity. The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters. The total number of correct letters were recorded. A decrease in the number of correct letters indicates a decrease in visual acuity.

    Measured: Final Visit (scheduled within window: days 120-180)

  4. Percentage of Participants With Reduction in Cylinder Within 0.50 Diopter (D) and Within 1.00 D of Intended.

    Measured: Final Visit (scheduled within window: days 120-180)

Secondary outcomes (1)
  1. Mean Uncorrected Distance Visual Acuity (UCDVA) at Final Visit

    Uncorrected distance visual acuity (UCDVA), expressed as logarithm of the minimum angle of resolution (logMAR). Uncorrected means not spectacle-corrected. Visual acuity (VA) was reported as letters read correctly, and the letters read were converted to logarithm of the minimum angle of resolution (logMAR) acuity. The participant read rows of letters on a standard LogMAR chart until the participant misses 3 letters on a row with 5 letters. The total number of correct letters were recorded. A decrease in the number of correct letters indicates a decrease in visual acuity.

    Measured: Final Visit (scheduled within window: days 120-180)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted13 May 2013 (estimated)
  2. Study start6 Mar 2013
  3. Primary completion18 Jul 2016
  4. Study completionDec 2016

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Model description
A Prospective, Multicenter Clinical Trial to Evaluate the Safety and Effectiveness
Masking
Double blind
Who is masked
Participant, Investigator
Enrolment
191 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01852084
Sponsor's protocol ID
761
Status last verified
Aug 2020
First submitted
9 May 2013
Last updated
16 Sep 2020
Data monitoring committee
No
Individual data sharing
Yes

Questions about this trial

What is this clinical trial studying?

A study testing enVista® One-Piece Hydrophobic Acrylic Toric IOL in people with Astigmatism, run by Bausch & Lomb Incorporated at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 12 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 16 Sep 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 16 Sep 2020). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mean Dioptric Change in Cylinder (time frame: Baseline, Final Visit (scheduled within window: days 120-180)). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 16 Sep 2020.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in