Accuracy of Two CGM Systems Tested Simultaneously in Ambulatory Patients With Type 1 Diabetes

A study testing Guardian Enlite sensor and Dexcom G4 platinum sensor in people with Diabetes Mellitus, Type 1 and Blood Glucose, run by Vastra Gotaland Region at 2 sites in Sweden.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Continuous glucose monitoring (CGM) is a tool used in the treatment of patients with type 1 diabetes. A continuous glucose monitor (CGM) is a subcutaneous tissue sensor, which provides a tissue fluid glucose measurement every 1 to 5 minutes. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Diabetes Mellitus, Type 1, Blood Glucose
What is being tested?
Guardian Enlite sensor and Dexcom G4 platinum sensor
Could I receive a placebo?
No, everyone receives an active treatment
How many people
38 participants
Where
2 sitesSweden
Study dates
May 2013to 1 Sep 2014
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Type 1 diabetes
  • Adult patients, age 18 or older and <75 years.
  • Written Informed Consent

Who cannot join

  • Pregnancy
  • Patients with severe cognitive dysfunction or other disease which makes CGM use difficult.
  • Patients requiring continuous use of paracetamol. Paracetamol must not have been used
  • the week before the study and shall not be used during the duration because it disturbs the
  • interpretation of blood glucose levels estimated by the DexCom4G. However, other pain
  • killers can be used throughout the study period.
Show all 14Show fewer
  • Current CGM use
  • History of allergic reaction to any of the CGMS materials
  • or adhesives in contact with the skin.
  • History of allergic reaction to chlorhexidine or alcohol
  • anti-septic solution.
  • Abnormal skin at the anticipated glucose sensor
  • attachment sites (excessive hair, burn, inflammation,
  • infection, rash, and/or tattoo).

Trial locations and contacts

2 sites in 1 country.

Sweden2 sites
  • NU-hospital Organisation

    Trollhättan, 46173

  • Uddevalla hospital

    Uddevalla, 45180

Other trials in Sweden

About this study

Continuous glucose monitoring (CGM) is a tool used in the treatment of patients with type 1 diabetes. A continuous glucose monitor (CGM) is a subcutaneous tissue sensor, which provides a tissue fluid glucose measurement every 1 to 5 minutes. Since CGM measure the glucose level by a sensor in subcutaneous tissue, there is uncertainty in the estimation of blood glucose levels. The accuracy of a certain CGM system can be assessed by comparing the glucose levels estimated by CGM with measured glucose levels in plasma. In the current study, we will compare the accuracy of the 2 CGM systems available on the market for clinical use in patients with type 1 diabetes. The aim of the current study is to evaluate effectiveness, safety and treatment satisfaction with 2 different CGM systems among adult type 1 diabetic patients.

Read the detailed description

Each study patient will have both subcutaneous tissue CGM sensors inserted. Each sensor will produce a maximum of 1,440 tissue fluid glucose measurements per 24 hours and 8,460 measurements during the 6 days in study. The plan is to study 36 to 50 ambulatory patients during the hole study period. The two CGM data sets (DexCom4G and Medtronic Enlite) will be compared to each other and the time-matched reference blood glucose measurements.

The HemoCue Analyser- venous blood and finger-stick blood in cuvette, will be used to measure the concentration of glucose.

Each ambulatory patient will sample capillary blood and measure the concentration of glucose 6 to 10 times per day for max 6 days. The concentration of finger-stick capillary blood glucose will be measured using the HemoCue meter in their daily living and additionally a HemoCue Analyser at research visits using a lancet. Each patient will be admitted to the clinical research unit (CRU) for 7 measurements of venous samples at two occasions: on day 1-3 and 4-6 of the study with an interval of at least 15 minutes. An intravenous catheter will be inserted to facilitate blood sample acquisition each of these days of study. Three capillary finger-stick blood will be sampled at these 2 occasions, 1 at the first venous measurement (one day of days 1-3) and one at the last venous sample (one day of days 4-6).

Blood sample acquisition and handling will be standardized to minimize pre-analytical error.

The sensor insertion sites will be observed to detect bleeding, inflammation and infection of the skin or subcutaneous tissue. Insertion sites will be photographed if any abnormal findings exist at the end of the study for the individual patient.

The subjects will record self monitoring of blood glucose (SMBG), meal, and activity data in a written diary.

Being tested
Guardian Enlite sensor, Dexcom G4 platinum sensor
Keywords
Diabetes Mellitus, Diabetes Mellitus, Type 1, Diabetes Mellitus, Insulin-Dependent, Endocrine System Diseases, Blood glucose, Blood Glucose, Self-Monitoring
Medical subject headings
Diabetes Mellitus, Type 1, Diabetes Mellitus, Endocrine System Diseases

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Guardian Enlite sensorDevice

    Sensor insertion

  • Dexcom G4 platinum sensorDevice

    Sensor insertion

Study groups in NCT02159638
Study groupRoleWhat they receive
comparisson CGM accuracyOtherEach patient will have both subcutaneous tissue CGM sensors (Guardian Enlite sensor and Dexcom G4 Platinum sensor) inserted at the same time. The HemoCue Analyser- venous blood and finger-stick blood in cuvette, will be used to measure the concentration of glucose
Receives: Device: Guardian Enlite sensor; Device: Dexcom G4 platinum sensor

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Difference between the two CGM systems compared to capillary glucose value

    Primary efficacy analysis will be investigating whether there is a statistically significant difference between the two CGM systems with respect to mean absolute relative difference (MARD) of estimated capillary glucose value over the whole study period.

    Measured: 6 days

Secondary outcomes (5)
  1. the accuracy of the 2 CGM systems during each studied time interval (day 1-3 and day 4-6)

    Accuracy of all estimated capillary glucose values (registered both during days 1-3 and days 4-6) of DexCom 4G system (DexCom sensor) vs Guardian Real-Time system (Enlite sensor) system compared to the capillary reference blood glucose measured with HemoCue.

    Measured: day 1-3 and day 4-6

  2. the accuracy of the 2 CGM systems for hypoglycaemia

    Accuracy in estimating capillary glucose levels <4 registered both during days 1-3 and days 4-6 of DexCom 4G system (DexCom sensor) vs Guardian Real-Time system (Enlite sensor) system compared to the capillary reference blood glucose measured with HemoCue(hypo, normoglycaemia and hyperglycaemia)

    Measured: 6 days

  3. the accuracy of the 2 CGM systems for normoglycaemia

    Accuracy in estimating capillary glucose levels of 4.0-10.0 mmol/l registered both during days 1-3 and days 4-6 of DexCom 4G system (DexCom sensor) vs Guardian Real-Time system (Enlite sensor) system compared to the capillary reference blood glucose measured with HemoCue

    Measured: 6 days

  4. the accuracy of the 2 CGM systems for hyperglycaemia

    Accuracy in estimating capillary glucose levels >14.0 mmol/l registered both during days 1-3 and days 4-6 of DexCom 4G system (DexCom sensor) vs Guardian Real-Time system (Enlite sensor) system compared to the capillary reference blood glucose measured with HemoCue

    Measured: 6 days

  5. Evaluation of two CGM systems from questionnaire

    Evaluation of DexCom 4G system (DexCom sensor) vs Guardian Real-Time system (Enlite sensor) from questionnaire

    Measured: day 6

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted10 Jun 2014 (estimated)
  2. Study startMay 2013
  3. Primary completion1 Jul 2014
  4. Study completion1 Sep 2014

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
38 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02159638
Sponsor's protocol ID
VGFOUREG-387191
Status last verified
Jun 2014
First submitted
5 Jun 2014
Last updated
19 Dec 2017
Results posted
No
Data monitoring committee
No

Publications

  • Matuleviciene V, Joseph JI, Andelin M, Hirsch IB, Attvall S, Pivodic A, Dahlqvist S, Klonoff D, Haraldsson B, Lind M. A clinical trial of the accuracy and treatment experience of the Dexcom G4 sensor (Dexcom G4 system) and Enlite sensor (guardian REAL-time system) tested simultaneously in ambulatory patients with type 1 diabetes. Diabetes Technol Ther. 2014 Nov;16(11):759-67. doi: 10.1089/dia.2014.0238. Epub 2014 Sep 18.

    PubMed 25233297

Questions about this trial

What is this clinical trial studying?

A study testing Guardian Enlite sensor and Dexcom G4 platinum sensor in people with Diabetes Mellitus, Type 1 and Blood Glucose, run by Vastra Gotaland Region at 2 sites in Sweden.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 14 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 19 Dec 2017. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in Sweden. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Difference between the two CGM systems compared to capillary glucose value (time frame: 6 days). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT02159638, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.