A Pivotal Study to Evaluate Enlite™ Sensor Performance With iPro™2 in China

A study testing Enlite 1 connected to iPro2 recorder in people with Diabetes Mellitus, run by Medtronic MiniMed, Inc. at 3 sites in China.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to demonstrate the performance and safety of the Enlite Sensor over 144 hours (6 days) when inserted in the abdomen and used with the iPro2 in subjects age 14 - 75 years. Read more about this study

Who is it for?
Ages 14 to 75 years
Conditions the study is for
Diabetes Mellitus
What is being tested?
Enlite 1 connected to iPro2 recorder
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
72 participants
Where
3 sitesChina
Study dates
28 Jun 2018to 20 Sep 2018
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 14 to 75 years
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Who may be able to join

  • Subject is 14 - 75 years of age at time of screening
  • Subject has a clinical diagnosis of type 1 or 2 diabetes as determined via medical record or source documentation by an individual qualified to make a medical diagnosis
  • Subject has adequate venous access as assessed by investigator or appropriate staff
  • Subject is willing to follow the study procedures and willing to come to study visits.
  • Subject is willing to perform at least 4 self-monitoring of blood glucose (SMBG) per day for 6 days

Who cannot join

  • Subject will not tolerate tape adhesive in the area of Enlite Sensor placement as assessed by qualified individual.
  • Subject has any unresolved adverse skin condition in the area of study device or device placement (e.g., psoriasis, rash, Staphylococcus infection)
  • Subject is actively participating in an investigational study (drug or device) wherein they have received treatment from an investigational study (drug or device) in the last 2 weeks
  • Subject is female and has a positive pregnancy screening test
  • Females of child bearing age and who are sexually active should be excluded if they are not using a form of contraception deemed reliable by investigator
  • Subject is female and plans to become pregnant during the course of the study
  • Subject has a hematocrit (Hct) lower than the normal reference range
  • Subject may not be on the research staff of those performing this study

Trial locations and contacts

3 sites in 1 country.

China3 sites
  • PLA Army General Hospital

    Beijing, 100700

  • Sir Run Run Shaw Hospital School of Medicine Zhejiang University

    Hangzhou, 310016

  • Shanghai Sixth People's Hospital

    Shanghai, 200233

Other trials in China

About this study

The purpose of this study is to demonstrate the performance and safety of the Enlite Sensor over 144 hours (6 days) when inserted in the abdomen and used with the iPro2 in subjects age 14 - 75 years.

Read the detailed description

This study is a multi-center, randomized, prospective single-sample correlational design without controls. Up to 72 subjects will be enrolled in order to have approximately 60 subjects complete the study. Three investigational centers in China will be used during this study.

Each subject will wear the following devices:

  • Two Enlite Sensors each connected to an iPro2 for approximately 6 days

Sensor Location:

  • The 2 Enlite Sensors will be worn in the abdomen area. Investigational center staff will insert sensors and connect to the iPro2s.

During the study, each subject will be randomized and undergo one Yellow Springs Instrument (YSI™) frequent sample testing (FST) (Day 1, 3-4, or 6).

During the study, subjects will continue with their current diabetes regimen independent of the study devices. Subjects will be instructed by the investigational center that they are not to use the study devices (except for the study meter) for the management of their diabetes. The Study Meter may be used for treatment decisions and calibration of Enlite Sensor.

Being tested
Enlite 1 connected to iPro2 recorder
Keywords
Diabetes Type I Type II
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Enlite 1 connected to iPro2 recorderDevice

    Use Enlite Sensor over 144 hours (6 days) when inserted in the abdomen and used with the iPro2 in subjects age 14 - 75 years.

Study groups in NCT03417466
Study groupRoleWhat they receive
Study armExperimentalUse Enlite Sensor over 144 hours (6 days) when inserted in the abdomen and used with the iPro2 and undergo one Yellow Springs Instrument (YSI™) frequent sample testing (FST) (Day 1, 3-4, or 6).
Receives: Device: Enlite 1 connected to iPro2 recorder

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mean Percentage of Enlite Sensor Values That Are Within 20% Agreement of Gold Standard (Yellow Springs Instrument (YSI) YSI Plasma Glucose Values)

    Primary endpoint is the mean percentage of Enlite Sensor Values from primary sensor that are within 20% agreement of YSI plasma glucose values (within 20 mg/dL if YSI <= 80 mg/dL) during YSI frequent sampling testing (FST) days.

    Measured: YSI FST days (Day 1, 3-4, or 6 of sensor wear)

Secondary outcomes (3)
  1. Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD)

    Enlite sensor values from primary sensor were compared to YSI plasma glucose values during YSI FST days (Day 1, 3-4, or 6). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Therefore, the unit of MARD is percentage (%).

    Measured: YSI FST days (Day 1, 3-4, or 6 of sensor wear)

  2. Clarke Error Grid Analysis of Paired Sensor and YSI Plasma Glucose Values

    Clarke Error Grid compared the paired primary sensor and YSI reference glucose values. Zone A represents a region where sensor values that are within 20% of YSI reference values, or both sensor and YSI reference values are less than 70 mg/dL. Values in Zone A are "clinically accurate in that they would lead to clinically correct treatment decisions". Zone B represents sensor values that deviate YSI reference values by more than 20%, but "would lead to benign or no treatment". Ideal situation is 100% in Zone A + B.

    Measured: YSI FST days (Day 1, 3-4, or 6 of sensor wear)

  3. Consensus Error Grid Analysis of Paired Sensor and YSI Plasma Glucose Values

    Consensus Error Grid (or Parkes error grid) compared the paired primary sensor and YSI reference glucose values. Zone A is defined in the Parkes error grid as the zone of "clinical accurate measurements with no effect on clinical action", Zone B as "altered clinical action with little or no effect on clinical outcome". Ideal situation is 100% in Zone A + B.

    Measured: YSI FST days (Day 1, 3-4, or 6 of sensor wear)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted31 Jan 2018
  2. Study start28 Jun 2018
  3. Primary completion20 Sep 2018
  4. Study completion20 Sep 2018

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
72 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03417466
Sponsor's protocol ID
CIP313
Status last verified
Jul 2019
First submitted
25 Jan 2018
Last updated
20 Aug 2019
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Enlite 1 connected to iPro2 recorder in people with Diabetes Mellitus, run by Medtronic MiniMed, Inc. at 3 sites in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 20 Aug 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 3 sites in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 20 Aug 2019). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mean Percentage of Enlite Sensor Values That Are Within 20% Agreement of Gold Standard (Yellow Springs Instrument (YSI) YSI Plasma Glucose Values) (time frame: YSI FST days (Day 1, 3-4, or 6 of sensor wear)). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 20 Aug 2019.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in