The purpose of this study is the verification of Cloud-DX Vitaliti™ Continuous Automated Non-Invasive Blood Pressure Monitor for Clinical Use and Self Measurement according to standards laid out by IEEE Std 1708 and AAMI-ISO 81060-2. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Blood Pressure
What is being tested?
Vitaliti
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
78 participants
Where
1 siteCanada
Study dates
9 Aug 2018to 6 Nov 2019
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patients ≥18 years of age;
  • Patient has undergone major cardiac or vascular surgery and has been admitted post-op to the cardiac surgical intensive care unit (CICU) or vascular surgery step down unit with an arterial line in place

Who cannot join

  • Patients < 18 years of age;
  • Pregnant;
  • Current arrhythmia
  • Blood pressure outside of test range for phase (refer to table 1)
  • Allergy to device contact adhesive (e.g. ECG leads)

Trial locations and contacts

1 site in 1 country.

Canada1 site
  • Hamilton General Hospital

    Hamilton, Ontario, L8L 2X2

Other trials in Canada

About this study

Official title: Verification Protocol for The Cloud DX Vitaliti™ Continuous Vital Sign Monitor (CVSM): Non-invasive Blood Pressure Component for Clinical Use and Self Measurement According to the American Safety in Medical Technology and the IEEE

The purpose of this study is the verification of Cloud-DX Vitaliti™ Continuous Automated Non-Invasive Blood Pressure Monitor for Clinical Use and Self Measurement according to standards laid out by IEEE Std 1708 and AAMI-ISO 81060-2.

Read the detailed description

Several automated vital sign measurement systems have been developed recently; however, non-invasive continuous blood pressure monitoring continues to be a major limitation. To circumvent the limitations of current systems of interval hemodynamic measurement and increasing demands on health care providers, a non-invasive automated continuous vital sign monitor that can integrate into hospital early warning systems and warn health care providers of deteriorating vital sign parameters would be of significant value and a major advance for current hospital monitoring systems.

Cloud-DX Vitaliti™ is a new low profile automated vital sign monitor, which includes a continuous non-invasive blood pressure measurement system. Additional parameters are also measured. Paired with companion software from cloud based computing, Vitaliti™ features a user friendly interface, onboard radios for communication with smartphones/tablets/personal computers, and several sensors.

Cloud-DX Vitaliti™ has undergone preliminary testing and is now ready for verification to achieve compliance with AAMI/ISO and IEEE standards of medical grade device. At the Hamilton General Hospital (HGH), Cardiac Surgical Intensive Care Unit (CICU) and Vascular Surgical Step Down Units, arterial catheters are standard of care and performed on most patients for variable periods of post-operative recovery. Within this setting, the IEEE Std 1708 and AAMI-ISO 81060-2 would be fully met. Furthermore, the cardiovascular surgical caseloads at HGH would provide sufficient test subjects to validate VitalitiTM in a timely fashion.

Conditions
Blood Pressure
Being tested
Vitaliti
Keywords
Non-invasive blood pressure monitoring, Wearable device, Medical device verification study

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • VitalitiDevice

    Measurements recorded by the Vitaliti device will be compared to the recordings from the arterial catheter.

Study groups in NCT03493867
Study groupRoleWhat they receive
VitalitiOtherThe subject's blood pressure will be simultaneously determined and recorded using the invasive arterial line blood pressure reading and the test Vitaliti device readings.
Receives: Device: Vitaliti

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Blood pressure

    Determining the accuracy of non-invasive blood pressure measurement compared to invasive blood pressure. A detailed analysis will be conducted on the invasive blood pressure to verify accuracy as defined in the ISO 81060-2 Standard.

    Measured: 30 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted11 Apr 2018
  2. Study start9 Aug 2018
  3. Primary completion6 Nov 2019
  4. Study completion6 Nov 2019

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Model description
Vitaliti Continuous Vital Sign Monitoring Device - Version 2.0
Masking
None (open label)
Enrolment
78 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03493867
Sponsor's protocol ID
STP-VIT-002
Status last verified
Mar 2020
First submitted
16 Mar 2018
Last updated
3 Mar 2020
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided: The data which will be collected is only valid for the verification of this particular device.

Publications

  • McGillion MH, Dvirnik N, Yang S, Belley-Cote E, Lamy A, Whitlock R, Marcucci M, Borges FK, Duceppe E, Ouellette C, Bird M, Carroll SL, Conen D, Tarride JE, Harsha P, Scott T, Good A, Gregus K, Sanchez K, Benoit P, Owen J, Harvey V, Peter E, Petch J, Vincent J, Graham M, Devereaux PJ. Continuous Noninvasive Remote Automated Blood Pressure Monitoring With Novel Wearable Technology: A Preliminary Validation Study. JMIR Mhealth Uhealth. 2022 Feb 28;10(2):e24916. doi: 10.2196/24916.

    PubMed 34876396

Questions about this trial

What is this clinical trial studying?

A study testing Vitaliti in people with Blood Pressure, run by Population Health Research Institute at 1 site in Canada.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 3 Mar 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Canada. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Blood pressure (time frame: 30 minutes). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT03493867, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 3 Mar 2020.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in