Clinical Trial of Blood Pressure Monitor in Conformance With the ANSI/AAMI ISO 81060-2 Standard
A study testing Blood Pressure Measurement in people with Blood Pressure, run by BTS International at 1 site in United States.
Trial summary
The clinical protocol of the trial: 1. Objective of the trial: To assess the accuracy of device. 2. Test methods and procedures: Perform comparation test on seated position by two devices at the same time. 3. Read more about this study
- Who is it for?
- Ages 3 years and olderHealthy volunteers can take part
- Conditions the study is for
- Blood Pressure
- What is being tested?
- Blood Pressure Measurement
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 88 participants
- Where
- 1 siteUnited States
- Study dates
- 30 May 2017to 23 Apr 2018
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 3 years and older
- All sexes
- Healthy volunteers accepted
- Child, Adult, Older adult
Inclusion Criteria: male,female. Exclusion Criteria: below 3 years old.
Trial locations and contacts
1 site in 1 country.
About this study
Official title: Assess the Accuracy of Transtek TMB-1591-A-002 Blood Pressure Monitor in Conformance With the ANSI/AAMI ISO 81060-2 Standard
The clinical protocol of the trial:
- Objective of the trial: To assess the accuracy of device.
- Test methods and procedures: Perform comparation test on seated position by two devices at the same time.
- DUT: Transtek Blood Pressure Monitor, Model TMB-1591-A-002, with Cuff (sizes, cm) 15-21, 20-26, 25-34, 25-34(L), 32-43, 32-43(L), 40-55.
- Reference device: Baumanometer Desk Mercury Sphygmomanometer.
- Study endpoints: Comply with ANSI/AAMI ISO 81060-2:2013 standard.
- Statistical methodology used: Standard deviation, Mean error.
Read the detailed description
- Introduction: Automated blood pressure cuff measurement is the standard of care in numerous medical settings today. This study was designed to assess the accuracy of a previously cleared automated blood pressure cuff system using the same arm sequential dual observer auscultation method. The study was conducted in accordance to ANSI/AAMI/ISO 81060-2:2013 Non-invasive sphygmomanometers - Part 2: Clinical investigation of automated measurement type, the Code of Federal Regulations requirements for Non-Significant Risk Device Investigations and as appropriate ISO 14155: Clinical investigation of medical devices for human subjects - Good clinical practice.
- Purpose of Clinical Investigation: The purpose of this study is to provide supporting documentation for non-invasive blood pressure measurement Accuracy claims for the TMB-1591-A-002 Blood Pressure System with Reusable Cuffs on the intended adult and pediatric population.
- Clinical Investigation Population: Following Salus Independent Review Board (IRB) approval, a total of 90 subjects were screened and enrolled into the study and data was collected. Two subjects were withdrawn from the study and the data was not used in the data analysis. These subjects were removed due to difficulty in hearing the reference blood pressure range (systolic and/or diastolic) and the other was removed due an unstable systolic range. A total of 88 healthy subjects, (53 adult / adolescent and 35 pediatric) ages 3-75yr, arm circumference 15-49.5 cm, systolic blood pressures ranges 79-176.5 and diastolic blood pressure ranges of 49-116 were included in the study for analysis. The study was conducted May 30 - June 16, 2017 to perform an accuracy validation on TMB-1591-A-002 Blood Pressure System with FlexiPortTM Reusable Cuffs.
- Method:
The requirements for the study were to include a minimum of 85 subjects with a minimum of 35 pediatric subjects 3-12 years of age and the remainder over 12 years of age, with at least 30% being male and 30% being female, and an arm circumference range of 15-55cm where at least six subjects were evaluated in each of the seven cuffs. For the data collection, at least 90% of the subjects needed to contribute 3 paired observations allowing the remaining 10% to be 2 paired observations. A minimum of 255 valid paired observations were needed for the analysis.
This study utilized the same arm sequential method (Dual Observer) in which the observers alternated the measurements for the reference sphygmomanometer and the sphygmomanometer-under test. Each paired reading incorporated the average of the reference reading prior to the device under test and after the device under test. (ANSI/AAMI/ISO 81060-2:2013, Section 5.2.4.2).
The TMB-1591-A-002 Blood Pressure System with FlexiPortTM Reusable Cuffs, (sphygmomanometer-under-test) was designed for use on the upper arm of patients with arm circumferences that range from 15-55cm. Seven cuff sizes were included in the investigation requiring that each cuff size have data collection on at least 6 subjects. Two trained observers (auscultators) simultaneously observed (listened to) the Korotkoff sounds at the brachial artery during the automated blood pressure cuff measurement. The 5th Korotkoff (K5) sound was used to identify diastole for the adult/adolescent subjects and the 4th Korotkoff (K4) sound was used for the pediatric subjects.
The reference sphygmomanometer used in this study [Baumanometer Desk Mercury Sphygmomanometer, SN: CT1573] complied with the requirements of ISO 81060-1 (the maximum error was 1 mmHg per NIST traceable calibration verification.
- Conditions
- Blood Pressure
- Being tested
- Blood Pressure Measurement
- Keywords
- Monitors, Blood Pressure
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Blood Pressure MeasurementDiagnostic test
Also called Transtek TMB-1591-A-002 and reference device, blood pressure
Measurement blood pressure by Transtek BPM TMB-1591-A-002 and reference device.
| Study group | Role | What they receive |
|---|---|---|
| BPM, TMB-1591-A-002 and Reference Device | Experimental | DUT: Transtek Blood Pressure Monitor TMB-1591-A-002 Reference Device: Baumanometer Desk Mercury Sphygmomanometer. Blood Pressure Measurement with the Transtek BPM TMB-1591-A-002 and with reference device. Receives: Diagnostic Test: Blood Pressure Measurement |
What is being measured
The primary outcome is the main question the study is designed to answer.
Blood pressure measurement data
Systolic Pressure and Diastolic Pressure
Measured: 30 minutes
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted17 Jul 2018
- Study start30 May 2017
- Primary completion16 Jun 2017
- Study completion23 Apr 2018
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Blood PressureStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 88 participants (actual)
Sponsor and investigators
- Lead sponsor
- BTS International
- Sponsor type
- University, hospital or other
- Collaborators
- Clinimark, LLC, Guangdong Transtek Medical Electronics Co., Ltd.
- Responsible party
- Sponsor
- Investigators
- Nagulinie Shukla, Study chair, Clinimark, LLC
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT03588559
- Sponsor's protocol ID
- TM-BPM-1801
- Other IDs
- BTS-TM-BPM-1801
- Acronym
- CT_1591-A
- Status last verified
- Jul 2018
- First submitted
- 24 Jun 2018
- Last updated
- 17 Jul 2018
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- No
Study at a glance
From the registry record, last updated by the sponsor on 17 Jul 2018. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Assess the Accuracy of Transtek TMB-1591-A-002 Blood Pressure Monitor in Conformance With the ANSI/AAMI ISO 81060-2 Standard
- ClinicalTrials.gov ID
- NCT03588559
- Acronym / sponsor ID
- CT_1591-A / TM-BPM-1801
- Conditions
- Blood Pressure
- Keywords
- Monitors, Blood Pressure
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 88 participants (actual)
- Lead sponsor
- BTS InternationalUniversity, hospital or other
- Collaborators
- Clinimark, LLC, Guangdong Transtek Medical Electronics Co., Ltd.
- Study start
- 30 May 2017 (actual)
- Primary completion
- 16 Jun 2017 (actual)
- Study completion
- 23 Apr 2018 (actual)
- First posted
- 17 Jul 2018 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The clinical protocol of the trial:
- Objective of the trial: To assess the accuracy of device.
- Test methods and procedures: Perform comparation test on seated position by two devices at the same time.
- DUT: Transtek Blood Pressure Monitor, Model TMB-1591-A-002, with Cuff (sizes, cm) 15-21, 20-26, 25-34, 25-34(L), 32-43, 32-43(L), 40-55.
- Reference device: Baumanometer Desk Mercury Sphygmomanometer.
- Study endpoints: Comply with ANSI/AAMI ISO 81060-2:2013 standard.
- Statistical methodology used: Standard deviation, Mean error.
Detailed description
- Introduction: Automated blood pressure cuff measurement is the standard of care in numerous medical settings today. This study was designed to assess the accuracy of a previously cleared automated blood pressure cuff system using the same arm sequential dual observer auscultation method. The study was conducted in accordance to ANSI/AAMI/ISO 81060-2:2013 Non-invasive sphygmomanometers - Part 2: Clinical investigation of automated measurement type, the Code of Federal Regulations requirements for Non-Significant Risk Device Investigations and as appropriate ISO 14155: Clinical investigation of medical devices for human subjects - Good clinical practice.
- Purpose of Clinical Investigation: The purpose of this study is to provide supporting documentation for non-invasive blood pressure measurement Accuracy claims for the TMB-1591-A-002 Blood Pressure System with Reusable Cuffs on the intended adult and pediatric population.
- Clinical Investigation Population: Following Salus Independent Review Board (IRB) approval, a total of 90 subjects were screened and enrolled into the study and data was collected. Two subjects were withdrawn from the study and the data was not used in the data analysis. These subjects were removed due to difficulty in hearing the reference blood pressure range (systolic and/or diastolic) and the other was removed due an unstable systolic range. A total of 88 healthy subjects, (53 adult / adolescent and 35 pediatric) ages 3-75yr, arm circumference 15-49.5 cm, systolic blood pressures ranges 79-176.5 and diastolic blood pressure ranges of 49-116 were included in the study for analysis. The study was conducted May 30 - June 16, 2017 to perform an accuracy validation on TMB-1591-A-002 Blood Pressure System with FlexiPortTM Reusable Cuffs.
- Method:
The requirements for the study were to include a minimum of 85 subjects with a minimum of 35 pediatric subjects 3-12 years of age and the remainder over 12 years of age, with at least 30% being male and 30% being female, and an arm circumference range of 15-55cm where at least six subjects were evaluated in each of the seven cuffs. For the data collection, at least 90% of the subjects needed to contribute 3 paired observations allowing the remaining 10% to be 2 paired observations. A minimum of 255 valid paired observations were needed for the analysis.
This study utilized the same arm sequential method (Dual Observer) in which the observers alternated the measurements for the reference sphygmomanometer and the sphygmomanometer-under test. Each paired reading incorporated the average of the reference reading prior to the device under test and after the device under test. (ANSI/AAMI/ISO 81060-2:2013, Section 5.2.4.2).
The TMB-1591-A-002 Blood Pressure System with FlexiPortTM Reusable Cuffs, (sphygmomanometer-under-test) was designed for use on the upper arm of patients with arm circumferences that range from 15-55cm. Seven cuff sizes were included in the investigation requiring that each cuff size have data collection on at least 6 subjects. Two trained observers (auscultators) simultaneously observed (listened to) the Korotkoff sounds at the brachial artery during the automated blood pressure cuff measurement. The 5th Korotkoff (K5) sound was used to identify diastole for the adult/adolescent subjects and the 4th Korotkoff (K4) sound was used for the pediatric subjects.
The reference sphygmomanometer used in this study [Baumanometer Desk Mercury Sphygmomanometer, SN: CT1573] complied with the requirements of ISO 81060-1 (the maximum error was 1 mmHg per NIST traceable calibration verification.
Study design
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| BPM, TMB-1591-A-002 and Reference Device | Experimental | Diagnostic Test: Blood Pressure MeasurementDUT: Transtek Blood Pressure Monitor TMB-1591-A-002 Reference Device: Baumanometer Desk Mercury Sphygmomanometer. Blood Pressure Measurement with the Transtek BPM TMB-1591-A-002 and with reference device. |
Who may qualify
- Ages
- 3 years and older
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Child, Adult, Older adult
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Blood Pressure MeasurementAlso: Transtek TMB-1591-A-002 and reference device, blood pressureMeasurement blood pressure by Transtek BPM TMB-1591-A-002 and reference device. | Diagnostic test | BPM, TMB-1591-A-002 and Reference Device |
Primary outcome
Blood pressure measurement data
Time frame: 30 minutes
Systolic Pressure and Diastolic Pressure
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 17 Jul 2018.
- Individual participant dataSharing planned: No
Related trials for Blood Pressure
Active, not recruiting
Digital Reinforcement of Motivational Interviewing Micro-Skills Among Primary Care Healthcare Professionals
Hypertension (Without Type 2 Diabetes Mellitus)
RecruitingPhase 3
A Phase III Renal Outcomes and Cardiovascular Mortality Study to Investigate the Efficacy and Safety of Baxdrostat in Combination With Dapagliflozin in Participants With Chronic Kidney Disease and High Blood Pressure
Chronic Kidney Disease and Hypertension
Recruiting
Guiding Optimal Blood Pressure Management in Peritoneal Dialysis Pilot Trial
End Stage Kidney Disease (ESKD), Hypertension (HTN)
RecruitingPhase 4
Hematological Safety of Dipyrone Versus Paracetamol After Dental Extractions
Hematologic Test, Agranulocytosis
Not yet recruiting
Acute Effects of Mat Pilates on Blood Pressure Response to Mental Stress in Higher Education Teachers
Behavior
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Nagulinie Shukla study chair
Clinimark, LLC
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Blood Pressure Measurement in people with Blood Pressure, run by BTS International at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 17 Jul 2018. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Blood pressure measurement data (time frame: 30 minutes). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 Jul 2018.