Clinical Trial of Blood Pressure Monitor in Conformance With the ANSI/AAMI ISO 81060-2 Standard

A study testing Blood Pressure Measurement in people with Blood Pressure, run by BTS International at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The clinical protocol of the trial: 1. Objective of the trial: To assess the accuracy of device. 2. Test methods and procedures: Perform comparation test on seated position by two devices at the same time. 3. Read more about this study

Who is it for?
Ages 3 years and olderHealthy volunteers can take part
Conditions the study is for
Blood Pressure
What is being tested?
Blood Pressure Measurement
Could I receive a placebo?
No, everyone receives an active treatment
How many people
88 participants
Where
1 siteUnited States
Study dates
30 May 2017to 23 Apr 2018
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 3 years and older
  • All sexes
  • Healthy volunteers accepted
  • Child, Adult, Older adult

Inclusion Criteria: male,female. Exclusion Criteria: below 3 years old.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Clinimark, LLC

    Louisville, Colorado, 80027

Other trials in United States

About this study

Official title: Assess the Accuracy of Transtek TMB-1591-A-002 Blood Pressure Monitor in Conformance With the ANSI/AAMI ISO 81060-2 Standard

The clinical protocol of the trial:

  1. Objective of the trial: To assess the accuracy of device.
  2. Test methods and procedures: Perform comparation test on seated position by two devices at the same time.
  3. DUT: Transtek Blood Pressure Monitor, Model TMB-1591-A-002, with Cuff (sizes, cm) 15-21, 20-26, 25-34, 25-34(L), 32-43, 32-43(L), 40-55.
  4. Reference device: Baumanometer Desk Mercury Sphygmomanometer.
  5. Study endpoints: Comply with ANSI/AAMI ISO 81060-2:2013 standard.
  6. Statistical methodology used: Standard deviation, Mean error.
Read the detailed description
  1. Introduction: Automated blood pressure cuff measurement is the standard of care in numerous medical settings today. This study was designed to assess the accuracy of a previously cleared automated blood pressure cuff system using the same arm sequential dual observer auscultation method. The study was conducted in accordance to ANSI/AAMI/ISO 81060-2:2013 Non-invasive sphygmomanometers - Part 2: Clinical investigation of automated measurement type, the Code of Federal Regulations requirements for Non-Significant Risk Device Investigations and as appropriate ISO 14155: Clinical investigation of medical devices for human subjects - Good clinical practice.
  2. Purpose of Clinical Investigation: The purpose of this study is to provide supporting documentation for non-invasive blood pressure measurement Accuracy claims for the TMB-1591-A-002 Blood Pressure System with Reusable Cuffs on the intended adult and pediatric population.
  3. Clinical Investigation Population: Following Salus Independent Review Board (IRB) approval, a total of 90 subjects were screened and enrolled into the study and data was collected. Two subjects were withdrawn from the study and the data was not used in the data analysis. These subjects were removed due to difficulty in hearing the reference blood pressure range (systolic and/or diastolic) and the other was removed due an unstable systolic range. A total of 88 healthy subjects, (53 adult / adolescent and 35 pediatric) ages 3-75yr, arm circumference 15-49.5 cm, systolic blood pressures ranges 79-176.5 and diastolic blood pressure ranges of 49-116 were included in the study for analysis. The study was conducted May 30 - June 16, 2017 to perform an accuracy validation on TMB-1591-A-002 Blood Pressure System with FlexiPortTM Reusable Cuffs.
  4. Method:

The requirements for the study were to include a minimum of 85 subjects with a minimum of 35 pediatric subjects 3-12 years of age and the remainder over 12 years of age, with at least 30% being male and 30% being female, and an arm circumference range of 15-55cm where at least six subjects were evaluated in each of the seven cuffs. For the data collection, at least 90% of the subjects needed to contribute 3 paired observations allowing the remaining 10% to be 2 paired observations. A minimum of 255 valid paired observations were needed for the analysis.

This study utilized the same arm sequential method (Dual Observer) in which the observers alternated the measurements for the reference sphygmomanometer and the sphygmomanometer-under test. Each paired reading incorporated the average of the reference reading prior to the device under test and after the device under test. (ANSI/AAMI/ISO 81060-2:2013, Section 5.2.4.2).

The TMB-1591-A-002 Blood Pressure System with FlexiPortTM Reusable Cuffs, (sphygmomanometer-under-test) was designed for use on the upper arm of patients with arm circumferences that range from 15-55cm. Seven cuff sizes were included in the investigation requiring that each cuff size have data collection on at least 6 subjects. Two trained observers (auscultators) simultaneously observed (listened to) the Korotkoff sounds at the brachial artery during the automated blood pressure cuff measurement. The 5th Korotkoff (K5) sound was used to identify diastole for the adult/adolescent subjects and the 4th Korotkoff (K4) sound was used for the pediatric subjects.

The reference sphygmomanometer used in this study [Baumanometer Desk Mercury Sphygmomanometer, SN: CT1573] complied with the requirements of ISO 81060-1 (the maximum error was 1 mmHg per NIST traceable calibration verification.

Conditions
Blood Pressure
Being tested
Blood Pressure Measurement
Keywords
Monitors, Blood Pressure

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Blood Pressure MeasurementDiagnostic test

    Also called Transtek TMB-1591-A-002 and reference device, blood pressure

    Measurement blood pressure by Transtek BPM TMB-1591-A-002 and reference device.

Study groups in NCT03588559
Study groupRoleWhat they receive
BPM, TMB-1591-A-002 and Reference DeviceExperimentalDUT: Transtek Blood Pressure Monitor TMB-1591-A-002 Reference Device: Baumanometer Desk Mercury Sphygmomanometer. Blood Pressure Measurement with the Transtek BPM TMB-1591-A-002 and with reference device.
Receives: Diagnostic Test: Blood Pressure Measurement

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Blood pressure measurement data

    Systolic Pressure and Diastolic Pressure

    Measured: 30 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted17 Jul 2018
  2. Study start30 May 2017
  3. Primary completion16 Jun 2017
  4. Study completion23 Apr 2018

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
88 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03588559
Sponsor's protocol ID
TM-BPM-1801
Other IDs
BTS-TM-BPM-1801
Acronym
CT_1591-A
Status last verified
Jul 2018
First submitted
24 Jun 2018
Last updated
17 Jul 2018
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Blood Pressure Measurement in people with Blood Pressure, run by BTS International at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 17 Jul 2018. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Blood pressure measurement data (time frame: 30 minutes). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 Jul 2018.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in