SpO2 Accuracy Comparison of Smart Sock V. 2 SpO2 to Arterial Blood CO-Oximetry

A study testing Pulse oximetry with Owlet BabySat v1.0 sensor in people with Hypoxia, run by Owlet Baby Care, Inc. at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. There are two devices used in this investigational study: BabySat v.1.0 and Owlet Baby Care, Inc. Read more about this study

Who is it for?
Ages 18 to 50 yearsHealthy volunteers can take part
Conditions the study is for
Hypoxia
What is being tested?
Pulse oximetry with Owlet BabySat v1.0 sensor
Could I receive a placebo?
No, everyone receives an active treatment
How many people
11 participants
Where
1 siteUnited States
Study dates
13 Dec 2017to 17 Dec 2017
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 50 years
  • All sexes
  • Healthy volunteers accepted
  • Adult

Who may be able to join

  • 10-15 Adults with a minimum of 3 males and a minimum of 3 females, with the balance, made up of either.
  • Subject must have the ability to understand and provide written informed consent
  • Subject is 18 to 50 years of age
  • At least 2 darkly pigmented subjects or 15% of the subject pool, whichever is larger
  • Subject must be willing and able to comply with study procedures and duration
  • Subject is a non-smoker or who have not smoked within 2 days prior to the study.

Who cannot join

  • Subject is considered as being morbidly obese (defined as BMI >39.5)
  • Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study
  • Females who are pregnant, who are trying to get pregnant
  • Smoker Subjects who have refrained will be screened for COHb levels
  • Subjects with known heart or cardiovascular conditions
  • Subjects with known clotting disorders

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Owlet Baby Care, Inc.

    Lehi, Utah, 84043

Other trials in United States

About this study

Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. There are two devices used in this investigational study: BabySat v.1.0 and Owlet Baby Care, Inc. Smart SockTM 2 (currently available over-the-counter), with specific emphasis on its pulse oximetry accuracy. The BabySat and Smart SockTM 2 are non-invasive home care devices for use with infants. The purpose of this study was to validate the SpO2 accuracy of BabySat v1.0 and Owlet Smart SockTM 2, OSS v1.1 sensors pulse oximetry, during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry.

Read the detailed description

Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. A pulse oximeter is a device that measures the oxygen saturation of arterial blood non-invasively.

The purpose of this study was to evaluate the SpO2 accuracy performance of the Owlet BabySat v1.0 and Smart SockTM 2 pulse oximetry OSS v1.1 sensors during non-motion conditions over the range of 70-100% SaO2 to arterial blood samples assessed by CO-Oximetry for a SpO2 validation.

The goal, in its entirety, was to show the SpO2 accuracy performance for the investigational device Owlet Baby Care, Inc. Smart SockTM V. 2.

It was expected that the Accuracy Root Mean Square (ARMS) performance would meet the required specification of ARMS 3% or lower in non-motion conditions for the range of 70 - 100% SaO2 thereby demonstrating an acceptable SpO2 accuracy performance specification.

No risks or adverse device effects were expected. There were no contraindications for use in the proposed study population.

The study was conducted in accordance to the code of federal regulations for non-significant risk medical device studies and applicable ISO 14155 (2nd edition 2011-02-01), applicable sections of ISO80601-2-61 (1st edition 2011-04- 01), and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013)

Conditions
Hypoxia
Being tested
Pulse oximetry with Owlet BabySat v1.0 sensor
Medical subject headings
Hypoxia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Pulse oximetry with Owlet BabySat v1.0 sensorDiagnostic test

    A comparative, single-center, non-randomized study to evaluate the SpO2 accuracy guidelines for pulse oximetry over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison with Owlet BabySat v1.0 and Owlet Smart Sock 2 sensors

Study groups in NCT03630016
Study groupRoleWhat they receive
Reference CO-OximetryActive comparatorReference Co-Oximetry
Receives: Diagnostic Test: Pulse oximetry with Owlet BabySat v1.0 sensor
Owlet BabySat v1.0ExperimentalOwlet BabySat v1.0
Receives: Diagnostic Test: Pulse oximetry with Owlet BabySat v1.0 sensor
Owlet Smart Sock V2 v1.1ExperimentalOwlet Smart SockTM 2 , OSS v1.1 Sensor and Custom Adult Thumb Sock
Receives: Diagnostic Test: Pulse oximetry with Owlet BabySat v1.0 sensor

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. It Was Expected That the Accuracy Root Mean Square (ARMS) Between Measured SpO2 of Owlet Smart Sock V2 v1.1 and Reference SpO2 Would Meet the Required Specification of ARMS 3% or Lower in Non-motion Conditions for the Range of 70 - 100%

    The purpose of the study was to evaluate the accuracy of the blood oxygen saturation (SpO2) at rest over the range of 70-100% compared to arterial blood samples assessed by standard methods [Co-Oximetry]. A standard pulse oximeter accuracy is evaluated by the Accuracy Root Mean Square (ARMS). The results of accurate performance of the oximetry system require ARMS of 3% or less in non-motion conditions for the range of 70-100% SpO2 per standard.

    Measured: Acute immediate assessment of the sensor accuracy compared to CO-oximetry

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted14 Aug 2018
  2. Study start13 Dec 2017
  3. Primary completion17 Dec 2017
  4. Study completion17 Dec 2017

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Non-randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
11 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03630016
Sponsor's protocol ID
PR2017-263
Status last verified
Jul 2019
First submitted
3 Aug 2018
Last updated
23 Jul 2019
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Pulse oximetry with Owlet BabySat v1.0 sensor in people with Hypoxia, run by Owlet Baby Care, Inc. at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 23 Jul 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 23 Jul 2019). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

It Was Expected That the Accuracy Root Mean Square (ARMS) Between Measured SpO2 of Owlet Smart Sock V2 v1.1 and Reference SpO2 Would Meet the Required Specification of ARMS 3% or Lower in Non-motion Conditions for the Range of 70 - 100% (time frame: Acute immediate assessment of the sensor accuracy compared to CO-oximetry). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 23 Jul 2019.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in