AmetropiaNCT03835221

Performance of Toric Hydrogel Lenses Following A Refit With Toric Silicone Hydrogel Lenses for 1 Month

A study testing methafilcon A toric contact lenses and fanfilcon A toric contact lenses in people with Ametropia, run by CooperVision, Inc. at 1 site in Mexico.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The aim of this prospective study is to evaluate the clinical performance of habitual wearers of methafilcon A toric lenses after a refit with fanfilcon A toric lenses for 1-month of daily wear. Read more about this study

Who is it for?
Ages 18 to 40 years
Conditions the study is for
Ametropia
What is being tested?
methafilcon A toric contact lenses and fanfilcon A toric contact lenses
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
43 participants
Where
1 siteMexico
Study dates
1 Jan 2019to 6 Jul 2019
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 40 years
  • All sexes
  • No healthy volunteers
  • Adult

Who may be able to join

  • Is between 18 and 40 years of age (inclusive)
  • Has had a self-reported visual exam in the last two years
  • Is an adapted soft contact lens wearer
  • Has a contact lens spherical prescription between +6.00 to - 8.00 (inclusive)-Have no less than -0.75D of astigmatism and no more than -2.25 D in both eyes
  • Can achieve best corrected spectacle distance visual acuity of 20/25 (0.10 logMAR) or better in each eye.
  • Can achieve a distance visual acuity of 20/30 (0.18 logMAR) or better in each eye with the study contact lenses.
Show all 11Show fewer
  • Has clear corneas and no active ocular disease
  • Has read, understood and signed the information consent letter
  • Patient contact lens refraction should fit within the available parameters of the study lenses
  • Is willing to comply with the wear schedule (at least 5 days per week, >8 hours/day assuming there are no contraindications for doing so)
  • Is willing to comply with the visit schedule

Who cannot join

  • Has a Contact Lens (CL) prescription outside the range of the available parameters of the study lenses
  • Has a spectacle cylinder less than -0.75D or more than -2.50 D of cylinder in either eye
  • Has a history of not achieving comfortable Contact Lens (CL) wear (5 days per week; > 8 hours/day)
  • Has contact lens best corrected distance vision worse than 20/25 (0.10 logMAR) in either eye
  • Presence of clinically significant (grade 2-4) anterior segment abnormalities
  • Presence of ocular or systemic disease or need of medications which might interfere with contact lens wear
Show all 20Show fewer
  • Slit lamp findings that would contraindicate contact lens wear such as:
  • Pathological dry eye or associated findings
  • Pterygium, pinguecula, or corneal scars within the visual axis
  • Neovascularization > 0.75 mm in from of the limbus
  • Giant papillary conjunctivitis (GCP) worse than grade 1
  • Anterior uveitis or iritis (past or present)
  • Seborrheic eczema, Seborrheic conjunctivitis
  • History of corneal ulcers or fungal infections
  • Poor personal hygiene
  • Has a known history of corneal hypoesthesia (reduced corneal sensitivity)
  • Has aphakia, keratoconus or a highly irregular cornea
  • Has Presbyopia or has dependence on spectacles for near work over the contact lenses
  • Has undergone corneal refractive surgery
  • Is participating in any other type of eye related clinical or research study.

Trial locations and contacts

1 site in 1 country.

Mexico1 site
  • Optometry Clinic, National Autonomous University

    Mexico City

Other trials in Mexico

About this study

The aim of this prospective study is to evaluate the clinical performance of habitual wearers of methafilcon A toric lenses after a refit with fanfilcon A toric lenses for 1-month of daily wear.

Conditions
Ametropia
Being tested
methafilcon A toric contact lenses, fanfilcon A toric contact lenses
Medical subject headings
Refractive Errors

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • methafilcon A toric contact lensesDevice

    Bilateral daily wear of methafilcon A toric contact lenses

  • fanfilcon A toric contact lensesDevice

    Bilateral daily wear of fanfilcon A toric contact lenses

Study groups in NCT03835221
Study groupRoleWhat they receive
methafilcon A toric / fanfilcon A toric contact lensesExperimentalAll subjects will first wear methafilcon A toric contact lenses for four (4) weeks of daily wear, then refitted with fanfilcon A toric contact lenses for four (4) weeks of daily wear.
Receives: Device: methafilcon A toric contact lenses; Device: fanfilcon A toric contact lenses

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses

    Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).

    Measured: Baseline

  2. Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses

    Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).

    Measured: 4 weeks

  3. Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).

    Measured: Baseline

  4. Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).

    Measured: 2 weeks

  5. Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable).

    Measured: 4 weeks

  6. Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses

    Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea

    Measured: Baseline

  7. Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses

    Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea

    Measured: 4 weeks

  8. Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea

    Measured: Baseline

  9. Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea

    Measured: 2 weeks

  10. Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea

    Measured: 4 weeks

  11. Post-blink Movement Assessment by the Investigator - Methafilcon A Toric Contact Lenses

    Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).

    Measured: Baseline

  12. Post-blink Movement Assessment by the Investigator - Methafilcon A Toric Contact Lenses

    Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).

    Measured: 4 weeks

  13. Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).

    Measured: Baseline

  14. Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).

    Measured: 2 weeks

  15. Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement).

    Measured: 4 weeks

  16. Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Methalfilcon A Toric Contact Lenses

    Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position.

    Measured: Baseline

  17. Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Methalfilcon A Toric Contact Lenses

    Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position.

    Measured: 4 weeks

  18. Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position.

    Measured: Baseline

  19. Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position.

    Measured: 2 weeks

  20. Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position.

    Measured: 4 weeks

  21. Rotational Recovery Assessment by the Investigator - Methafilcon A Toric Contact Lenses

    Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position

    Measured: Baseline

  22. Rotational Recovery Assessment by the Investigator - Methafilcon A Toric Contact Lenses

    Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position

    Measured: 4 weeks

  23. Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position

    Measured: Baseline

  24. Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position

    Measured: 2 weeks

  25. Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position

    Measured: 4 weeks

  26. Overall Stability Assessment by the Investigator - Methafilcon A Toric Contact Lenses

    Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability).

    Measured: Baseline

  27. Overall Stability Assessment by the Investigator - Methafilcon A Toric Contact Lenses

    Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability).

    Measured: 4 weeks

  28. Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability).

    Measured: Baseline

  29. Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability).

    Measured: 2 weeks

  30. Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses

    Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability).

    Measured: 4 weeks

  31. Overall Lens Fit Acceptance by the Investigator - Methafilcon A Toric Contact Lenses

    Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect.

    Measured: Baseline

  32. Overall Lens Fit Acceptance by the Investigator - Methafilcon A Toric Contact Lenses

    Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect.

    Measured: 4 weeks

  33. Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses

    Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect.

    Measured: Baseline

  34. Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses

    Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect.

    Measured: 2 weeks

  35. Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses

    Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect.

    Measured: 4 weeks

Secondary outcomes (6)
  1. Average Daily Wearing Time - Methafilcon A Toric Contact Lenses

    Average daily wearing time response by the subject

    Measured: 4 weeks

  2. Average Daily Wearing Time - Fanfilcon A Toric Contact Lenses

    Average daily wearing time response by the subject

    Measured: 2 weeks

  3. Average Daily Wearing Time - Fanfilcon A Toric Contact Lenses

    Average daily wearing time response by the subject

    Measured: 4 weeks

  4. Average Comfortable Wearing Time - Methafilcon A Toric Contact Lenses

    Average comfortable wearing time - response by the subject

    Measured: 4 weeks

  5. Average Comfortable Wearing Time - Fanfilcon A Toric Contact Lenses

    Average comfortable wearing time - response by the subject

    Measured: 2 weeks

  6. Average Comfortable Wearing Time - Fanfilcon A Toric Contact Lenses

    Average comfortable wearing time - response by the subject

    Measured: 4 weeks

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted8 Feb 2019
  2. Study start1 Jan 2019
  3. Primary completion27 Mar 2019
  4. Study completion6 Jul 2019

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
43 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03835221
Sponsor's protocol ID
EX-MKTG-96
Status last verified
May 2020
First submitted
31 Jan 2019
Last updated
11 Jun 2020
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes

Questions about this trial

What is this clinical trial studying?

A study testing methafilcon A toric contact lenses and fanfilcon A toric contact lenses in people with Ametropia, run by CooperVision, Inc. at 1 site in Mexico.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 11 inclusion and 20 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 11 Jun 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Mexico. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 11 Jun 2020). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses (time frame: Baseline). The study lists 35 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 11 Jun 2020.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in