Type 2 DiabetesAP-RenalNCT04025775
Closed-loop in Adults With T2D Requiring Dialysis
A study testing CamAPS HX Closed-loop in people with Type 2 Diabetes and End Stage Renal Disease, run by University of Cambridge at 2 sites in 2 countries.
Trial summary
The main objective of this study is to determine the efficacy, safety and utility of fully automated closed-loop glucose control in the home setting over a 20 day period in adults with type 2 diabetes (T2D) requiring maintenance dialysis. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Type 2 Diabetes, End Stage Renal Disease
- What is being tested?
- CamAPS HX Closed-loop
- Type of treatment
- Injection or infusion, Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 27 participants
- Where
- 2 sitesin 2 countries
- Study dates
- 1 Aug 2019to 1 Mar 2021
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- The subject is age 18 years or over
- Diagnosis of type 2 diabetes using standard diagnostic practice
- The subject requires maintenance dialysis
- The subject requires current treatment with subcutaneous insulin
- Screening HbA1c ≤ 11% (97mmol/mol) on analysis from local laboratory
- Subject is willing to perform regular finger-prick blood glucose monitoring
- The subject is literate in English (UK) or German (Switzerland)
- The subject is willing to wear study devices 24/7 during intervention arm and follow study specific instructions
Who cannot join
- Physical or psychological condition likely to interfere with the normal conduct of the study and interpretation of the study results as judged by the investigator
- Known or suspected allergy to insulin
- Lack of reliable telephone facility for contact
- Pregnancy, planned pregnancy, or breast feeding
- Severe visual impairment
- Severe hearing impairment
Show all 11Show fewer
- Medically documented allergy towards the adhesive (glue) of plasters or unable to tolerate tape adhesive in the area of sensor placement
- Serious skin diseases located at places of the body, which potentially are possible to be used for localisation of the glucose sensor
- Illicit drugs abuse
- Prescription drugs abuse
- Alcohol abuse
Trial locations and contacts
2 sites in 2 countries.
2 sites in 2 countries
United Kingdom1 site
Cambridge University Hospitals NHS Foundation Trust
Cambridge, CB2 0QQ
About this study
Official title: An Open-label, Two-centre, Randomised, 2-period Cross-over Study to Assess the Efficacy, Safety and Utility of Fully Closed-loop Insulin Delivery in Comparison With Standard Care, in Adults With Type 2 Diabetes Requiring Maintenance Dialysis
The main objective of this study is to determine the efficacy, safety and utility of fully automated closed-loop glucose control in the home setting over a 20 day period in adults with type 2 diabetes (T2D) requiring maintenance dialysis. This study builds on previous and on-going studies of closed-loop systems that have been performed in Cambridge in adults with type 1 diabetes in the home setting, and in adults with type 2 diabetes in the inpatient setting.
This is an open-label, two-centre, randomised, cross-over study, involving two home study periods during which glucose levels will be controlled either by a fully automated closed-loop system or by participants' usual insulin therapy in random order. Each treatment arm is 20 days long with a 2-4 week washout period between treatments. A total of up to 40 adults with T2D requiring maintenance dialysis will be recruited through outpatient clinics or the dialysis unit, to allow for 32 completed participants available for assessment.
Read more
Participants will receive appropriate training by the research team on the safe use of the study devices (insulin pump and continuous glucose monitoring (CGM) and closed-loop insulin delivery system). Participants in the control arm will continue with standard therapy and will wear a blinded CGM system.
The primary outcome is time spent with glucose levels in the target range between 5.6 and 10.0 mmol/L as recorded by CGM. Secondary outcomes are the time spent with glucose levels above and below target, as recorded by CGM, and other CGM-based metrics in addition to insulin requirements. Safety evaluation comprises the tabulation of severe hypoglycaemic episodes.
Read the detailed description
Purpose of clinical trial To determine the efficacy, safety and utility of fully automated closed-loop insulin delivery in the home setting in adults with type 2 diabetes requiring maintenance dialysis.
Study objectives The study objective is to compare fully automated closed-loop insulin delivery with usual care in adults with type 2 diabetes requiring maintenance dialysis.
- EFFICACY: The objective is to assess the ability of fully-automated closed-loop insulin delivery in maintaining CGM glucose levels within the target range from 5.6 to 10.0 mmol/l as compared to usual care in adults with type 2 diabetes requiring maintenance dialysis.
- SAFETY: The objective is to evaluate the safety of fully automated closed-loop insulin delivery in terms of episodes and severity of hypoglycaemia, and nature and severity of other adverse events.
- UTILITY: The objective is to determine the acceptability and duration of use of the closed-loop system.
Study design An open-label, two-centre, randomised, two-period crossover study comparing fully automated closed-loop insulin delivery with usual care in adults with type 2 diabetes requiring dialysis. Two intervention periods in the home setting will last 20 days each with a 2-4 week washout period. The order of the two interventions will be random.
Study endpoints The primary endpoint is the time spent in the target glucose range from 5.6 to 10.0 mmol/l based on CGM glucose levels during the 20 day home stay.
Other key endpoints:
- Time spent with sensor glucose above target (10.0 mmol/l)
- Time spent with sensor glucose <3.9 mmol/l
- Average of sensor glucose levels
Secondary endpoints include:
- Time spent with sensor glucose below target (5.6 mmol/l)
- Time spent with sensor glucose <3.0 mmol/l)
- Time spent with sensor glucose levels in significant hyperglycaemia (glucose levels > 20 mmol/l)
- Standard deviation and coefficient of variation of sensor glucose levels
- AUC of glucose below 3.5 mmol/l (63 mg/dl)
- Total daily insulin requirements
- Average inter-dialytic weight gain
Safety evaluation Assessment of frequency and severity of hypoglycaemic episodes and nature and severity of other adverse events.
Utility evaluation Assessment of the acceptability and duration of use of the closed-loop system.
Participating clinical centres
UK 1. Addenbrooke's Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge,
Switzerland
- Bern University Hospital, Bern
Sample size 32 adults completing the study. Recruitment will target up to 40 adults to allow for drop-outs.
Maximum duration of study for a participant 12 weeks
Recruitment Participants will be recruited through the adult diabetes outpatient clinics or dialysis units at participating centres.
Consent Written informed consent will be obtained from participants according to Research Ethics Committee (REC) requirements.
Screening and baseline assessment Eligible participants will undergo a baseline evaluation including medical (diabetes) history and current therapy.
Randomisation Eligible participants will be randomised in a 1:1 ratio using randomisation software to the use of fully automated closed-loop insulin delivery or to usual care for 20 days, with a 2-4 week washout period between the two interventions.
- Closed loop arm Following randomisation, participants in the closed-loop group will receive training to cover key aspects of insulin pump use and CGM. Competency on the use of study devices will be evaluated Once competent in the use of the study pump and CGM, participants will receive training required for safe and effective use of the closed-loop system. During a 2-4 hour session participants will operate the system under the supervision of the clinical team. Competency on the use of closed-loop system will be evaluated. Thereafter, participants are expected to use closed-loop for 20 days without supervision or remote monitoring. All participants will be provided with 24 hour telephone helpline and will also be given written instructions about when to contact clinical team.
- Standard therapy (control arm) Participants in the control group will continue with standard insulin therapy with blinded CGM for 20 days.
Study contacts Follow up contacts will be conducted within 24 hours of starting each treatment arm and then at weekly intervals thereafter.
End of study assessments Validated questionnaires evaluating the impact of the technology on diabetes management and quality of life will be completed. Participants will resume usual care.
Procedures for safety monitoring during trial Standard operating procedures for monitoring and reporting of all adverse events (AE) will be in place, including serious adverse events (SAE), serious adverse device effects (SADE).
A data safety and monitoring board (DSMB) will be informed of all serious adverse events and any unanticipated serious adverse device effects that occur during the study and will review compiled adverse event data at periodic intervals.
Criteria for withdrawal of patients on safety grounds
A participant may terminate participation in the study at any time without necessarily giving a reason and without any personal disadvantage. An investigator can stop the participation of a subject after consideration of the benefit/risk ratio. Possible reasons are:
- Serious adverse events
- Significant protocol violation or non-compliance
- Failure to satisfy competency assessment
- Decision by the investigator, or the Sponsor, that termination is in the participant's best medical interest
- Pregnancy, planned pregnancy, or breast feeding
- Allergic reaction to insulin
Efforts will be made to retain participants in follow up for the final primary outcome assessment even if the intervention is discontinued, unless the investigator believes that it will be harmful for the participant to continue in the trial.
- Conditions
- Type 2 Diabetes, End Stage Renal Disease
- Being tested
- CamAPS HX Closed-loop, Multiple Daily Insulin Injections (Control)
- Medical subject headings
- Diabetes Mellitus, Type 2, Kidney Failure, Chronic
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
CamAPS HX Closed-loopDevice
The CamAPS HX closed-loop system comprises
- Dana insulin pump (Diabecare, Sooil, Seoul, South Korea)
- Dexcom G6 real-time CGM sensor (Dexcom, Northridge, CA, USA)
- An Android smartphone hosting CamAPS HX Application with the Cambridge model predictive control algorithm and communicating wirelessly with the insulin pump
- Glooko/Diasend cloud upload system to monitor CGM/insulin data.
Multiple Daily Insulin Injections (Control)Other
Multiple Daily Insulin Injections with masked Dexcom G6 CGM sensor (Dexcom, Northridge, CA, USA)
| Study group | Role | What they receive |
|---|---|---|
| Closed loop insulin delivery | Experimental | Unsupervised home use of day and night fully automated closed loop insulin delivery system (CamAPS HX) for 20 days
The CamAPS HX closed-loop system comprises
* Dana insulin pump (Diabecare, Sooil, Seoul, South Korea)
* Dexcom G6 real-time CGM sensor (Dexcom, Northridge, CA, USA)
* An Android smartphone hosting CamAPS HX Application with the Cambridge model predictive control algorithm and communicating wirelessly with the insulin pump
* Glooko/Diasend cloud upload system to monitor CGM/insulin data. Receives: Device: CamAPS HX Closed-loop |
| Standard therapy | Active comparator | Participants in the control arm will continue to follow their current diabetes management plan for the 20 day study period.Participants will be wear a masked continuous glucose monitoring (CGM) system during the 20 day study period. Receives: Other: Multiple Daily Insulin Injections (Control) |
What is being measured
The primary outcome is the main question the study is designed to answer.
Time in the target glucose range (5.6 to 10.0 mmol/l)
Percentage of time spent with sensor glucose readings in the target range from 5.6 to 10.0
Measured: 20 day intervention period
Secondary outcomes (10)
Time spent above target glucose (10.0 mmol/l)
Both armsPercentage of time spent with sensor glucose readings above target glucose (10.0 mmol/l)
Measured: 20 day intervention period
Mean glucose
Mean sensor glucose level
Measured: 20 day intervention period
Time spent in hypoglycaemia (<3.9 mmol/l)
Percentage of time spent with sensor glucose readings <3.9 mmol/l
Measured: 20 day intervention period
Time spent below target glucose (5.6 mmol/l)
Percentage of time spent with sensor glucose readings below target glucose (5.6 mmol/l)
Measured: 20 day intervention period
Standard deviation of glucose levels
Standard deviation of glucose levels
Measured: 20 day intervention period
Coefficient of variation of glucose levels
Coefficient of variation of glucose levels
Measured: 20 day intervention period
Time spent below <3.0 mmol/l
Percentage of time spent with sensor glucose readings <3.0 mmol/l
Measured: 20 day intervention period
Time in significant hyperglycaemia (> 20 mmol/l)
Percentage of time with glucose levels in significant hyperglycaemia (> 20 mmol/l)
Measured: 20 day intervention period
Total, basal and bolus insulin dose
Average daily total, basal and bolus insulin requirements
Measured: 20 day intervention period
AUC of glucose below 3.5 mmol/l
Area under the curve of sensor glucose readings below 3.5mmol/l
Measured: 20 day intervention period
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted19 Jul 2019
- Study start1 Aug 2019
- Primary completion31 Dec 2020
- Study completion1 Mar 2021
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (1)Papers the sponsor has linked to this study.
- Other trials for Type 2 DiabetesStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- None (open label)
- Enrolment
- 27 participants (actual)
Sponsor and investigators
- Lead sponsor
- University of Cambridge
- Sponsor type
- University, hospital or other
- Collaborators
- Insel Gruppe AG, University Hospital Bern
- Responsible party
- Principal investigator, Dr Roman Hovorka, University of Cambridge
- Investigators
- Roman Hovorka, PhD, Study director, University of Cambridge
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT04025775
- Sponsor's protocol ID
- AP-Renal
- Acronym
- AP-Renal
- Status last verified
- May 2022
- First submitted
- 16 Jul 2019
- Last updated
- 17 May 2022
- Results posted
- No
- Data monitoring committee
- Yes
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- Yes: Fully anonymised data may be shared with third parties (EU or non-EU based) for the purposes of advancing management and treatment of diabetes.
Publications
Boughton CK, Tripyla A, Hartnell S, Daly A, Herzig D, Wilinska ME, Czerlau C, Fry A, Bally L, Hovorka R. Fully automated closed-loop glucose control compared with standard insulin therapy in adults with type 2 diabetes requiring dialysis: an open-label, randomized crossover trial. Nat Med. 2021 Aug;27(8):1471-1476. doi: 10.1038/s41591-021-01453-z. Epub 2021 Aug 4.
PubMed 34349267
Study at a glance
From the registry record, last updated by the sponsor on 17 May 2022. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- An Open-label, Two-centre, Randomised, 2-period Cross-over Study to Assess the Efficacy, Safety and Utility of Fully Closed-loop Insulin Delivery in Comparison With Standard Care, in Adults With Type 2 Diabetes Requiring Maintenance Dialysis
- ClinicalTrials.gov ID
- NCT04025775
- Acronym / sponsor ID
- AP-Renal / AP-Renal
- Conditions
- Type 2 Diabetes, End Stage Renal Disease
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 27 participants (actual)
- Lead sponsor
- University of CambridgeUniversity, hospital or other
- Collaborators
- Insel Gruppe AG, University Hospital Bern
- Study start
- 1 Aug 2019 (actual)
- Primary completion
- 31 Dec 2020 (actual)
- Study completion
- 1 Mar 2021 (actual)
- First posted
- 19 Jul 2019 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The main objective of this study is to determine the efficacy, safety and utility of fully automated closed-loop glucose control in the home setting over a 20 day period in adults with type 2 diabetes (T2D) requiring maintenance dialysis. This study builds on previous and on-going studies of closed-loop systems that have been performed in Cambridge in adults with type 1 diabetes in the home setting, and in adults with type 2 diabetes in the inpatient setting.
This is an open-label, two-centre, randomised, cross-over study, involving two home study periods during which glucose levels will be controlled either by a fully automated closed-loop system or by participants' usual insulin therapy in random order. Each treatment arm is 20 days long with a 2-4 week washout period between treatments. A total of up to 40 adults with T2D requiring maintenance dialysis will be recruited through outpatient clinics or the dialysis unit, to allow for 32 completed participants available for assessment.
Participants will receive appropriate training by the research team on the safe use of the study devices (insulin pump and continuous glucose monitoring (CGM) and closed-loop insulin delivery system). Participants in the control arm will continue with standard therapy and will wear a blinded CGM system.
The primary outcome is time spent with glucose levels in the target range between 5.6 and 10.0 mmol/L as recorded by CGM. Secondary outcomes are the time spent with glucose levels above and below target, as recorded by CGM, and other CGM-based metrics in addition to insulin requirements. Safety evaluation comprises the tabulation of severe hypoglycaemic episodes.
Detailed description
Purpose of clinical trial To determine the efficacy, safety and utility of fully automated closed-loop insulin delivery in the home setting in adults with type 2 diabetes requiring maintenance dialysis.
Study objectives The study objective is to compare fully automated closed-loop insulin delivery with usual care in adults with type 2 diabetes requiring maintenance dialysis.
- EFFICACY: The objective is to assess the ability of fully-automated closed-loop insulin delivery in maintaining CGM glucose levels within the target range from 5.6 to 10.0 mmol/l as compared to usual care in adults with type 2 diabetes requiring maintenance dialysis.
- SAFETY: The objective is to evaluate the safety of fully automated closed-loop insulin delivery in terms of episodes and severity of hypoglycaemia, and nature and severity of other adverse events.
- UTILITY: The objective is to determine the acceptability and duration of use of the closed-loop system.
Study design An open-label, two-centre, randomised, two-period crossover study comparing fully automated closed-loop insulin delivery with usual care in adults with type 2 diabetes requiring dialysis. Two intervention periods in the home setting will last 20 days each with a 2-4 week washout period. The order of the two interventions will be random.
Study endpoints The primary endpoint is the time spent in the target glucose range from 5.6 to 10.0 mmol/l based on CGM glucose levels during the 20 day home stay.
Other key endpoints:
- Time spent with sensor glucose above target (10.0 mmol/l)
- Time spent with sensor glucose <3.9 mmol/l
- Average of sensor glucose levels
Secondary endpoints include:
- Time spent with sensor glucose below target (5.6 mmol/l)
- Time spent with sensor glucose <3.0 mmol/l)
- Time spent with sensor glucose levels in significant hyperglycaemia (glucose levels > 20 mmol/l)
- Standard deviation and coefficient of variation of sensor glucose levels
- AUC of glucose below 3.5 mmol/l (63 mg/dl)
- Total daily insulin requirements
- Average inter-dialytic weight gain
Safety evaluation Assessment of frequency and severity of hypoglycaemic episodes and nature and severity of other adverse events.
Utility evaluation Assessment of the acceptability and duration of use of the closed-loop system.
Participating clinical centres
UK 1. Addenbrooke's Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge,
Switzerland
- Bern University Hospital, Bern
Sample size 32 adults completing the study. Recruitment will target up to 40 adults to allow for drop-outs.
Maximum duration of study for a participant 12 weeks
Recruitment Participants will be recruited through the adult diabetes outpatient clinics or dialysis units at participating centres.
Consent Written informed consent will be obtained from participants according to Research Ethics Committee (REC) requirements.
Screening and baseline assessment Eligible participants will undergo a baseline evaluation including medical (diabetes) history and current therapy.
Randomisation Eligible participants will be randomised in a 1:1 ratio using randomisation software to the use of fully automated closed-loop insulin delivery or to usual care for 20 days, with a 2-4 week washout period between the two interventions.
- Closed loop arm Following randomisation, participants in the closed-loop group will receive training to cover key aspects of insulin pump use and CGM. Competency on the use of study devices will be evaluated Once competent in the use of the study pump and CGM, participants will receive training required for safe and effective use of the closed-loop system. During a 2-4 hour session participants will operate the system under the supervision of the clinical team. Competency on the use of closed-loop system will be evaluated. Thereafter, participants are expected to use closed-loop for 20 days without supervision or remote monitoring. All participants will be provided with 24 hour telephone helpline and will also be given written instructions about when to contact clinical team.
- Standard therapy (control arm) Participants in the control group will continue with standard insulin therapy with blinded CGM for 20 days.
Study contacts Follow up contacts will be conducted within 24 hours of starting each treatment arm and then at weekly intervals thereafter.
End of study assessments Validated questionnaires evaluating the impact of the technology on diabetes management and quality of life will be completed. Participants will resume usual care.
Procedures for safety monitoring during trial Standard operating procedures for monitoring and reporting of all adverse events (AE) will be in place, including serious adverse events (SAE), serious adverse device effects (SADE).
A data safety and monitoring board (DSMB) will be informed of all serious adverse events and any unanticipated serious adverse device effects that occur during the study and will review compiled adverse event data at periodic intervals.
Criteria for withdrawal of patients on safety grounds
A participant may terminate participation in the study at any time without necessarily giving a reason and without any personal disadvantage. An investigator can stop the participation of a subject after consideration of the benefit/risk ratio. Possible reasons are:
- Serious adverse events
- Significant protocol violation or non-compliance
- Failure to satisfy competency assessment
- Decision by the investigator, or the Sponsor, that termination is in the participant's best medical interest
- Pregnancy, planned pregnancy, or breast feeding
- Allergic reaction to insulin
Efforts will be made to retain participants in follow up for the final primary outcome assessment even if the intervention is discontinued, unless the investigator believes that it will be harmful for the participant to continue in the trial.
Study design
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- None (open label)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Closed loop insulin delivery | Experimental | Device: CamAPS HX Closed-loopUnsupervised home use of day and night fully automated closed loop insulin delivery system (CamAPS HX) for 20 days The CamAPS HX closed-loop system comprises * Dana insulin pump (Diabecare, Sooil, Seoul, South Korea) * Dexcom G6 real-time CGM sensor (Dexcom, Northridge, CA, USA) * An Android smartphone hosting CamAPS HX Application with the Cambridge model predictive control algorithm and communicating wirelessly with the insulin pump * Glooko/Diasend cloud upload system to monitor CGM/insulin data. |
| Standard therapy | Active comparator | Other: Multiple Daily Insulin Injections (Control)Participants in the control arm will continue to follow their current diabetes management plan for the 20 day study period.Participants will be wear a masked continuous glucose monitoring (CGM) system during the 20 day study period. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (8 listed)
- The subject is age 18 years or over
- Diagnosis of type 2 diabetes using standard diagnostic practice
- The subject requires maintenance dialysis
- The subject requires current treatment with subcutaneous insulin
- Screening HbA1c ≤ 11% (97mmol/mol) on analysis from local laboratory
Key exclusion criteria (11 listed)
- Physical or psychological condition likely to interfere with the normal conduct of the study and interpretation of the study results as judged by the investigator
- Known or suspected allergy to insulin
- Lack of reliable telephone facility for contact
- Pregnancy, planned pregnancy, or breast feeding
- Severe visual impairment
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| CamAPS HX Closed-loopThe CamAPS HX closed-loop system comprises * Dana insulin pump (Diabecare, Sooil, Seoul, South Korea) * Dexcom G6 real-time CGM sensor (Dexcom, Northridge, CA, USA) * An Android smartphone hosting CamAPS HX Application with the Cambridge model predictive control algorithm and communicating wirelessly with the insulin pump * Glooko/Diasend cloud upload system to monitor CGM/insulin data. | Device | Closed loop insulin delivery |
| Multiple Daily Insulin Injections (Control)Multiple Daily Insulin Injections with masked Dexcom G6 CGM sensor (Dexcom, Northridge, CA, USA) | Other | Standard therapy |
Primary outcome
Time in the target glucose range (5.6 to 10.0 mmol/l)
Time frame: 20 day intervention period
Percentage of time spent with sensor glucose readings in the target range from 5.6 to 10.0
10 secondary outcomes
Time spent above target glucose (10.0 mmol/l)
Time frame: 20 day intervention period
Mean glucose
Time frame: 20 day intervention period
Time spent in hypoglycaemia (<3.9 mmol/l)
Time frame: 20 day intervention period
Time spent below target glucose (5.6 mmol/l)
Time frame: 20 day intervention period
Standard deviation of glucose levels
Time frame: 20 day intervention period
Coefficient of variation of glucose levels
Time frame: 20 day intervention period
Time spent below <3.0 mmol/l
Time frame: 20 day intervention period
Time in significant hyperglycaemia (> 20 mmol/l)
Time frame: 20 day intervention period
Total, basal and bolus insulin dose
Time frame: 20 day intervention period
AUC of glucose below 3.5 mmol/l
Time frame: 20 day intervention period
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 17 May 2022.
- Derived from this study: PubMed 34349267Boughton CK, Tripyla A, Hartnell S, Daly A, Herzig D, Wilinska ME, Czerlau C, Fry A, Bally L, Hovorka R. Fully automated closed-loop glucose control compared with standard insulin therapy in adults with type 2 diabetes requiring dialysis: an open-label, randomized crossover trial. Nat Med. 2021 Aug;27(8):1471-1476. doi: 10.1038/s41591-021-01453-z. Epub 2021 Aug 4.
- Individual participant dataSharing planned: Yes. Fully anonymised data may be shared with third parties (EU or non-EU based) for the purposes of advancing management and treatment of diabetes.
Related trials for Type 2 Diabetes
Recruiting
Prevalence of Diabetes-related Distress Among Patients Living With Type 2 Diabetes in a University Hospital Center and Identification of Its Associated Factors.
Diabetes Mellitus, Psychological Distress
Not yet recruiting
The Effect of Individualized Nursing Interventions on Glycemic Variability and Self-Management in Type 2 Diabetes
Type 2 Diabetes Mellitus (T2DM)
Active, not recruitingObservational study
Non-Invasive MASLD Diagnosis & T2D Remission Biomarkers
MASLD, T2DM (Type 2 Diabetes Mellitus)
Recruiting
Effects of Medisinstart for Type 2 Diabetes
Type 2 Diabetes
Not yet recruiting
Supplementation of Inulin + L. Acidophilus in Elders With T2DM
Diabetes Mellitus, Microbiome
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Roman Hovorka, PhD study director
University of Cambridge
2 sites in 2 countries: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing CamAPS HX Closed-loop in people with Type 2 Diabetes and End Stage Renal Disease, run by University of Cambridge at 2 sites in 2 countries.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 8 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 17 May 2022. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 2 sites in Switzerland, United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Time in the target glucose range (5.6 to 10.0 mmol/l) (time frame: 20 day intervention period). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT04025775, which also lists 1 linked publication.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 May 2022.