Meibomian Gland DysfunctionNCT04052841

Morphological Analysis of Meibomian Glands

A study testing Thermal pulsation, Intense pulsed light therapy and Manual warm compresses in people with Meibomian Gland Dysfunction, run by Zhongshan Ophthalmic Center, Sun Yat-sen University at 2 sites in China.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

An automated quantitative meibomian gland analyzer based on all kinds of infrared meibomian gland images was develop to obtain more detail in meibomian gland, including width, length, area, signal intensity correlated to the quality of meibum, deformation index and ratio of area of each visible specific gland. Read more about this study

Who is it for?
Ages 18 to 70 yearsHealthy volunteers can take part
Conditions the study is for
Meibomian Gland Dysfunction
What is being tested?
Thermal pulsation, Intense pulsed light therapy and Manual warm compresses
Type of treatment
Surgery or procedure
Could I receive a placebo?
No, everyone receives an active treatment
How many people
180 participants
Where
2 sitesChina
Study dates
12 Oct 2020to 1 Dec 2023
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 70 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Clinical diagnosis of MGD: The diagnosis of MGD was based on an altered quality of expressed secretions and/or decreased or absent expression.
  • Patients without ≥2/3 Meibomian glands atrophy.
  • Fitzpatrick skin type 1-4.
  • Negative history or condition of ocular or systemic illness based on evaluation by a research physician.
  • General Exclusion Criteria:
  • Patients and healthy volunteers with ocular allergies, trauma, contact lens wear, continuous medications usage such as tretinoin, isotretinoin, antidepressant medications, photosensitive drugs, glucocorticoids and immunomodulators, or have used them within one month.
Show all 13Show fewer
  • Patients and healthy volunteers who have a history of ocular surface surgery.
  • Patients and healthy volunteers who have active ocular surface infection or have suffered from ocular surface infection within one month.
  • Patients and healthy volunteers who have endophthalmitis or a medical history of endophthalmitis.
  • Patients and healthy volunteers who have a medical history of viral keratitis infection.
  • Women who are pregnant, planning to become pregnant during the course of the study or breast-feeding (women of child-bearing age will be asked by the physician).
  • Meibography images were blurred or with obvious tarsus folds, incomplete exposure and large hyperreflective area.
  • Patients and healthy volunteers who are not suitable for the trial as determined by investigators.

Who cannot join

  • Patients have abnormalities of ocular surface function or eyelid function, or presence of precancerous lesions, cancer or pigmentation in the eyelid area.
  • Patients who have plans to receive ocular surgeries (e.g., cataract, myopic refractive surgery) within 6 months.
  • Patients who have been treated with lacrimal punctum embolization within one month.
  • Patients with disease that could lead to ADDE, such as Sjogren syndrome and a lacrimal gland abnormality.

Trial locations and contacts

2 sites in 1 country.

Yuqing Deng, MD contact

18120557291

15927646647@163.com

China2 sites, 2 recruiting
  • Zhongshan Ophthalmic Center, Sun Yat-Sen University

    Guangzhou, Guangdong, 510080

    Saiqun Li, M.D., Ph.D., 86-013642710612, 123213197@qq.com

    Jin Yuan, M.D., Ph.D.

    Recruiting
  • Deng Yuqing

    Guangzhou, 510060

    Yuqing Deng, MD, 18120557291, 15927646647@163.com

    Jin Yuan, PHD

    Recruiting

Other trials in China

About this study

Official title: Automated Morphological Analysis of Meibomian Glands

An automated quantitative meibomian gland analyzer based on all kinds of infrared meibomian gland images was develop to obtain more detail in meibomian gland, including width, length, area, signal intensity correlated to the quality of meibum, deformation index and ratio of area of each visible specific gland. The purpose of this study is present as separate sections the following points: (1) to compared the detailed characteristics of meibomian glands in normal subjects, Meibomian gland dysfunction (MGD) patients by the automated quantitative analyzer; (2) to identify the inter-examiner and intra-examiner repeatability of the new technique; (3) to explore the correlation among morphological and functional parameters of meibomian gland and risk factors,clinical symptoms and signs; (4) to explore the sensitivity and specificity of meibomian gland morphological and functional parameters in MGD diagnosis. (5) using morphological and functional parameters as new assessment of MGD severity and efficacy indicators for treatment.

Read the detailed description

Meibomian glands are essential for maintaining ocular surface health and integrity secrete various lipid components to forms a lipid layer to prevent excessive tear evaporation. Functional disorders of the meibomian glands, referred to today as meibomian gland dysfunction (MGD), are increasingly recognized as a high incidence disease commonly characterized by terminal duct obstruction and/or abnormal glandular secretion, often results in ocular surface epithelium damage, chronic blepharitis and dry eye disease that significantly reduces quality of life. A wide variation of the prevalence of MGD were reported from 0.39% to 69.3%, which is likely due to lack of diagnostic methods. To identify which clinical features are likely to be predictive of progressive disease in MGD may indicate the early diagnosis and proper treatment strategies.

Histologic section through the normal meibomian glands and the obstructed human meibomian gland revealed that obstruction of orifice in MGD could lead to dilation of the central duct, damage of the secretory meibocytes and finally result in atrophy of dilated meibomian glands and glands drop-out. It was thus be accepted that detailed changes of meibomian glands morphology are key signs to diagnose and evaluate the severity of MGD. The detailed changes including dilation, distortion, shortening and loss of visualisation of glands which can be directly observed and visual assessment by the developed of non-contact meibomian gland infrared imaging technology. Quantitative evaluations of meibomian glands were obtain by developing imaging processing techniques. The most common use is the image editing software Image J (National Institute of Health; http://imagej.nih.gov/ij) which can identify the gland region on the image manually by the users and may lead to inter-observer variability. Koh et al., first applied original algorithms to automatically analysed gland loss in meibography images with a manually pre-processing. Reiko et al., then develop an objective and automatic system to measure the meibomian gland area. However, the existing methods of meibomian gland analysis have been limited to clinical use where large number of images needs to be analyzed efficiently due to the inter-observer variability or time-consuming process.

Meanwhile, the existing quantitative morphological parameters obtain by those imaging processing techniques, including percentage of MG drop-out and gland atrophy area, were suggested to not only be advanced stages or terminal changes in MGD, but also occurs as an age-related atrophic process. The early findings of MGD induced by the primary pathologic obstruction including degenerative gland dilation, irregularly shapes of gland and change of meibum quality are still difficult to be evaluated automatically and quantitively from the image. Moreover, the meibomian gland drop-out is still an approximate assessment without specific pattern. Whether the atrophy or loss occur in upper or lower eyelids, central, distal or proximal, total loss of gland or partial loss of gland has the greatest effect on the pathology progress of MGD will be important to identify. Thus, a comprehensive analysis technique to automatically detect multi-information of meibomian gland morphology will benefit the future early diagnosis and management of MGD.

Recently, an automated quantitative meibomian gland analyzer based on all kinds of infrared meibomian gland images was develop to obtain more detail in meibomian gland, including width, length, area, signal intensity correlated to the quality of meibum, deformation index and ratio of area of each visible specific gland. The purpose of this study is present as separate sections the following points: (1) to compared the detailed characteristics of meibomian glands in normal subjects and MGD patients by the automated quantitative analyzer; (2) to identify the inter-examiner and intra-examiner repeatability of the new technique; (3) to explore the correlation among morphological and functional parameters of meibomian gland and risk factors,clinical symptoms and signs; (4) to explore the sensitivity and specificity of meibomian gland morphological and functional parameters in MGD diagnosis. (5) using morphological and functional parameters as new assessment of MGD severity and efficacy indicators for treatment.

Being tested
Thermal pulsation, Intense pulsed light therapy, Manual warm compresses
Medical subject headings
Meibomian Gland Dysfunction

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Thermal pulsationProcedure or surgery

    Participants underwent a single LipiFlow® thermal pulsation system (TearScience Inc., Morrisville, NC,USA) treatment on the first visit: after lid hygiene with wet cotton swabs(OCuSOFT, Inc., Texas, USA) and instillation of anesthetic eye drops (Alcaine, proparacaine hydrochloride 0.4 ml/2mg) in both eyes, sterile eye cups were placed on to the conjunctival sac as instructed by the manufacturer, after 12 minutes of upper and lower palpebral conjunctival surfaces heat while simultaneously graded pulsatile pressure applying, eye cups were removed slightly.

  • Intense pulsed light therapyProcedure or surgery

    Lid hygiene with wet cotton swabs(OCuSOFT, Inc., Texas, USA) before treatment. Intense pulsed light (IPL) with a range of wavelength (570 or 620 nm) was performed every 3 weeks,3 times in total (0, 3w, 6w). 10 pulses were transmited from one tragus through nose to the other tragus was a single pass, each treatment needed to do 2 passes. Manual lid massage was done after per IPL treatment.

  • Manual warm compressesProcedure or surgery

    Lids hygiene of both eyes with wet cotton swabs(OCuSOFT, Inc., Texas, USA).Commercial heated eye patch was use for 10 min. Manual lid massage every 2 weeks, 5 times in total(0,2w, 4w, 4w, 8w).

Study groups in NCT04052841
Study groupRoleWhat they receive
MGD-thermal pulsation groupExperimentalA 12 minutes LIPIFLOW treatment was performed.
Receives: Procedure: Thermal pulsation
MGD-IPL groupExperimentalIPL was performed every 3 weeks,3 times in total (0, 3w, 6w).
Receives: Procedure: Intense pulsed light therapy
MGD-manual warm compressesExperimentalManual warm compresses were performed every 2 weeks,5 times in total (0, 2w, 4w, 6w, 8w).
Receives: Procedure: Manual warm compresses
Normal health subject groupNo interventionNormal health subject without intervention.

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mophology of meibomian glands

    Infrared photography of inversed upper meibomian glands were measured by Meibography pattern of Keratograph 5M (Oculus, Wetzlar, Germany). The infrared images of Meibography were analysed using the new developed software for identifying the mophology features of meibomian glands in millimeter.

    Measured: 30 days after commencement of treatment

  2. Functional feature of meibomian glands

    Infrared photography of inversed upper and lower meibomian glands were measured by Meibography pattern of Keratograph 5M (Oculus, Wetzlar, Germany). The infrared images of Meibography were analysed using the new developed software for identifying the mean signal intensity of meibomian glands in millimeter.

    Measured: 30 days after commencement of treatment

Secondary outcomes (48)
  1. Non-invasive tear-film break-up time

    Non-invasived tear-film break-up time is measured by tear film pattern of Keratograph 5M (Oculus, Wetzlar, Germany) with a scale of seconds. Higher values represent a better outcome

    Measured: Baseline

  2. Non-invasive tear-film break-up time

    Non-invasived tear-film break-up time is measured by tear film pattern of Keratograph 5M (Oculus, Wetzlar, Germany) with a scale of seconds. Higher values represent a better outcome

    Measured: 15 days after commencement of treatment

  3. Non-invasive tear-film break-up time

    Non-invasived tear-film break-up time is measured by tear film pattern of Keratograph 5M (Oculus, Wetzlar, Germany) with a scale of seconds. Higher values represent a better outcome

    Measured: 30 days after commencement of treatment

  4. Non-invasive tear-film break-up time

    Non-invasived tear-film break-up time is measured by tear film pattern of Keratograph 5M (Oculus, Wetzlar, Germany) with a scale of seconds. Higher values represent a better outcome

    Measured: 90 days after commencement of treatment

  5. Non-invasive tear-film break-up time

    Non-invasived tear-film break-up time is measured by tear film pattern of Keratograph 5M (Oculus, Wetzlar, Germany) with a scale of seconds. Higher values represent a better outcome

    Measured: 180 days after commencement of treatment

  6. Non-invasive tear meniscus height

    Non-invasived tear meniscus height is measured by tear film pattern of Keratograph 5M (Oculus, Wetzlar, Germany) in millimeter. The value higher than 0.20 mm was provided as a normal condition of tear secretion.

    Measured: Baseline

  7. Non-invasive tear meniscus height

    Non-invasived tear meniscus height is measured by tear film pattern of Keratograph 5M (Oculus, Wetzlar, Germany) in millimeter. The value higher than 0.20 mm was provided as a normal condition of tear secretion.

    Measured: 15 days after commencement of treatment

  8. Non-invasive tear meniscus height

    Non-invasived tear meniscus height is measured by tear film pattern of Keratograph 5M (Oculus, Wetzlar, Germany) in millimeter. The value higher than 0.20 mm was provided as a normal condition of tear secretion.

    Measured: 30 days after commencement of treatment

  9. Non-invasive tear meniscus height

    Non-invasived tear meniscus height is measured by tear film pattern of Keratograph 5M (Oculus, Wetzlar, Germany) in millimeter. The value higher than 0.20 mm was provided as a normal condition of tear secretion.

    Measured: 90 days after commencement of treatment

  10. Non-invasive tear meniscus height

    Non-invasived tear meniscus height is measured by tear film pattern of Keratograph 5M (Oculus, Wetzlar, Germany) in millimeter. The value higher than 0.20 mm was provided as a normal condition of tear secretion.

    Measured: 180 days after commencement of treatment

  11. Tear film lipid layer thicknesses

    Tear film lipid layer thicknesses were averaged in nanometer(0-100nm) during 20 seconds by LipiView II (Tear Science, Morrisville, NC).

    Measured: Baseline

  12. Tear film lipid layer thicknesses

    Tear film lipid layer thicknesses were averaged in nanometer(0-100nm) during 20 seconds by LipiView II (Tear Science, Morrisville, NC).

    Measured: 15 days after commencement of treatment

  13. Tear film lipid layer thicknesses

    Tear film lipid layer thicknesses were averaged in nanometer(0-100nm) during 20 seconds by LipiView II (Tear Science, Morrisville, NC).

    Measured: 30 days after commencement of treatment

  14. Tear film lipid layer thicknesses

    Tear film lipid layer thicknesses were averaged in nanometer(0-100nm) during 20 seconds by LipiView II (Tear Science, Morrisville, NC).

    Measured: 90 days after commencement of treatment

  15. Tear film lipid layer thicknesses

    Tear film lipid layer thicknesses were averaged in nanometer(0-100nm) during 20 seconds by LipiView II (Tear Science, Morrisville, NC).

    Measured: 180 days after commencement of treatment

  16. The pattern of eye blinks

    The pattern of eye blinks including numbers of incompleted and completed blinks during 20 seconds were measured by LipiView II (Tear Science, Morrisville, NC).

    Measured: Baseline

  17. The pattern of eye blinks

    Tear film lipid layer thicknesses and numbers of incomplete blinks during 20 seconds were measured by LipiView II (Tear Science, Morrisville, NC).

    Measured: 15 days after commencement of treatment

  18. The pattern of eye blinks

    Tear film lipid layer thicknesses and numbers of incomplete blinks during 20 seconds were measured by LipiView II (Tear Science, Morrisville, NC).

    Measured: 30 days after commencement of treatment

  19. The pattern of eye blinks

    Tear film lipid layer thicknesses and numbers of incomplete blinks during 20 seconds were measured by LipiView II (Tear Science, Morrisville, NC).

    Measured: 90 days after commencement of treatment

  20. The pattern of eye blinks

    Tear film lipid layer thicknesses and numbers of incomplete blinks during 20 seconds were measured by LipiView II (Tear Science, Morrisville, NC).

    Measured: 180 days after commencement of treatment

  21. Lid margin signs

    Three lid margin abnormalities (irregular lid margin, plugging of meibomian gland orifices, and anterior or posterior replacement of the mucocutaneous junction) were scored from 0 through 4 according to the number of these abnormalities that were present in each eye. Higher scores represent a worse outcome.

    Measured: baseline

  22. Lid margin signs

    Three lid margin abnormalities (irregular lid margin, plugging of meibomian gland orifices, and anterior or posterior replacement of the mucocutaneous junction) were scored from 0 through 4 according to the number of these abnormalities that were present in each eye. Higher scores represent a worse outcome.

    Measured: 15 days after commencement of treatment

  23. Lid margin signs

    Three lid margin abnormalities (irregular lid margin, plugging of meibomian gland orifices, and anterior or posterior replacement of the mucocutaneous junction) were scored from 0 through 4 according to the number of these abnormalities that were present in each eye. Higher scores represent a worse outcome.

    Measured: 30 days after commencement of treatment

  24. Lid margin signs

    Three lid margin abnormalities (irregular lid margin, plugging of meibomian gland orifices, and anterior or posterior replacement of the mucocutaneous junction) were scored from 0 through 4 according to the number of these abnormalities that were present in each eye. Higher scores represent a worse outcome.

    Measured: 90 days after commencement of treatment

  25. Lid margin signs

    Three lid margin abnormalities (irregular lid margin, plugging of meibomian gland orifices, and anterior or posterior replacement of the mucocutaneous junction) were scored from 0 through 4 according to the number of these abnormalities that were present in each eye. Higher scores represent a worse outcome.

    Measured: 180 days after commencement of treatment

  26. Meibum expressibility

    Meibum expressibility were measured by firm digital pressure of Meibomian gland diagnostic Expressibility (Tear Sience, Morrisville, NC):

    For each of these glands, the secretion was graded as follows: 0:no secretion; 1: inspissated/ toothpaste consistency; 2: cloudy liquid secretion and 3: clear liquid secretion19. The scores were then summed in 15 glands to a single meibomian gland yield secretion score (MGYSS) with a range from 0 to 45. Higher values represent a better outcome.

    Measured: baseline

  27. Meibum expressibility

    Meibum expressibility were measured by firm digital pressure of Meibomian gland diagnostic Expressibility (Tear Sience, Morrisville, NC):

    For each of these glands, the secretion was graded as follows: 0:no secretion; 1: inspissated/ toothpaste consistency; 2: cloudy liquid secretion and 3: clear liquid secretion19. The scores were then summed in 15 glands to a single meibomian gland yield secretion score (MGYSS) with a range from 0 to 45. Higher values represent a better outcome.

    Measured: 15 days after commencement of treatment

  28. Meibum expressibility

    Meibum expressibility were measured by firm digital pressure of Meibomian gland diagnostic Expressibility (Tear Sience, Morrisville, NC):

    For each of these glands, the secretion was graded as follows: 0:no secretion; 1: inspissated/ toothpaste consistency; 2: cloudy liquid secretion and 3: clear liquid secretion19. The scores were then summed in 15 glands to a single meibomian gland yield secretion score (MGYSS) with a range from 0 to 45. Higher values represent a better outcome.

    Measured: 30 days after commencement of treatment

  29. Meibum expressibility

    Meibum expressibility were measured by firm digital pressure of Meibomian gland diagnostic Expressibility (Tear Sience, Morrisville, NC):

    For each of these glands, the secretion was graded as follows: 0:no secretion; 1: inspissated/ toothpaste consistency; 2: cloudy liquid secretion and 3: clear liquid secretion19. The scores were then summed in 15 glands to a single meibomian gland yield secretion score (MGYSS) with a range from 0 to 45. Higher values represent a better outcome.

    Measured: 90 days after commencement of treatment

  30. Meibum expressibility

    Meibum expressibility were measured by firm digital pressure of Meibomian gland diagnostic Expressibility (Tear Sience, Morrisville, NC):

    For each of these glands, the secretion was graded as follows: 0:no secretion; 1: inspissated/ toothpaste consistency; 2: cloudy liquid secretion and 3: clear liquid secretion19. The scores were then summed in 15 glands to a single meibomian gland yield secretion score (MGYSS) with a range from 0 to 45. Higher values represent a better outcome.

    Measured: 180 days after commencement of treatment

  31. Corneal Fluorescein Staining

    Fluorescein was administered into the conjunctival sac under a cobalt blue light from the slit lamp. Corneal epithelial cell disruption was measured via corneal staining (National Eye Institute (NEI) scale (0-3 scale for each area of 5 areas, total score 15). Higher values represent a worse outcome.

    Measured: baseline

  32. Corneal Fluorescein Staining

    Fluorescein was administered into the conjunctival sac under a cobalt blue light from the slit lamp. Corneal epithelial cell disruption was measured via corneal staining (National Eye Institute (NEI) scale (0-3 scale for each area of 5 areas, total score 15). Higher values represent a worse outcome.

    Measured: 15 days after commencement of treatment

  33. Corneal Fluorescein Staining

    Fluorescein was administered into the conjunctival sac under a cobalt blue light from the slit lamp. Corneal epithelial cell disruption was measured via corneal staining (National Eye Institute (NEI) scale (0-3 scale for each area of 5 areas, total score 15). Higher values represent a worse outcome.

    Measured: 30 days after commencement of treatment

  34. Corneal Fluorescein Staining

    Fluorescein was administered into the conjunctival sac under a cobalt blue light from the slit lamp. Corneal epithelial cell disruption was measured via corneal staining (National Eye Institute (NEI) scale (0-3 scale for each area of 5 areas, total score 15). Higher values represent a worse outcome.

    Measured: 90 days after commencement of treatment

  35. Corneal Fluorescein Staining

    Fluorescein was administered into the conjunctival sac under a cobalt blue light from the slit lamp. Corneal epithelial cell disruption was measured via corneal staining (National Eye Institute (NEI) scale (0-3 scale for each area of 5 areas, total score 15). Higher values represent a worse outcome.

    Measured: 180 days after commencement of treatment

  36. Schirmer I test

    The tear production was measured with Schirmer strips without anaesthesia 15 minutes after corneal staining.

    Measured: baseline

  37. Schirmer I test

    The tear production was measured with Schirmer strips without anaesthesia 15 minutes after corneal staining.

    Measured: 15 days after commencement of treatment

  38. Schirmer I test

    The tear production was measured with Schirmer strips without anaesthesia 15 minutes after corneal staining.

    Measured: 30 days after commencement of treatment

  39. Schirmer I test

    The tear production was measured with Schirmer strips without anaesthesia 15 minutes after corneal staining.

    Measured: 90 days after commencement of treatment

  40. Schirmer I test

    The tear production was measured with Schirmer strips without anaesthesia 15 minutes after corneal staining.

    Measured: 180 days after commencement of treatment

  41. Mophology of meibomian glands

    Infrared photography of inversed upper meibomian glands were measured by Meibography pattern of Keratograph 5M (Oculus, Wetzlar, Germany). The infrared images of Meibography were analysed using the new developed software for identifying the mophology features of meibomian glands in millimeter.

    Measured: 15 days after commencement of treatment

  42. Mophology of meibomian glands

    Infrared photography of inversed upper meibomian glands were measured by Meibography pattern of Keratograph 5M (Oculus, Wetzlar, Germany). The infrared images of Meibography were analysed using the new developed software for identifying the mophology features of meibomian glands in millimeter.

    Measured: 90 days after commencement of treatment

  43. Mophology of meibomian glands

    Infrared photography of inversed upper meibomian glands were measured by Meibography pattern of Keratograph 5M (Oculus, Wetzlar, Germany). The infrared images of Meibography were analysed using the new developed software for identifying the mophology features of meibomian glands in millimeter.

    Measured: 180 days after commencement of treatment

  44. Mophology of meibomian glands

    Infrared photography of inversed upper meibomian glands were measured by Meibography pattern of Keratograph 5M (Oculus, Wetzlar, Germany). The infrared images of Meibography were analysed using the new developed software for identifying the mophology features of meibomian glands in millimeter.

    Measured: Baseline

  45. Functional feature of meibomian glands

    Infrared photography of inversed upper and lower meibomian glands were measured by Meibography pattern of Keratograph 5M (Oculus, Wetzlar, Germany). The infrared images of Meibography were analysed using the new developed software for identifying the mean signal intensity of meibomian glands in millimeter.

    Measured: Baseline

  46. Functional feature of meibomian glands

    Infrared photography of inversed upper and lower meibomian glands were measured by Meibography pattern of Keratograph 5M (Oculus, Wetzlar, Germany). The infrared images of Meibography were analysed using the new developed software for identifying the mean signal intensity of meibomian glands in millimeter.

    Measured: 15 days after commencement of treatment

  47. Functional feature of meibomian glands

    Infrared photography of inversed upper and lower meibomian glands were measured by Meibography pattern of Keratograph 5M (Oculus, Wetzlar, Germany). The infrared images of Meibography were analysed using the new developed software for identifying the mean signal intensity of meibomian glands in millimeter.

    Measured: 90 days after commencement of treatment

  48. Functional feature of meibomian glands

    Infrared photography of inversed upper and lower meibomian glands were measured by Meibography pattern of Keratograph 5M (Oculus, Wetzlar, Germany). The infrared images of Meibography were analysed using the new developed software for identifying the mean signal intensity of meibomian glands in millimeter.

    Measured: 180 days after commencement of treatment

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted12 Aug 2019
  2. Study start12 Oct 2020
  3. Primary completion1 Jul 2023 (estimated)
  4. Study completion1 Dec 2023 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Model description
The recruitment of subjects must meet the diagnosis criteria of MGD the international workshop on meibomian gland dysfunction.
Masking
None (open label)
Enrolment
180 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04052841
Sponsor's protocol ID
20190722
Status last verified
Apr 2023
First submitted
30 Jul 2019
Last updated
25 Apr 2023
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Yes: Study protocol

Publications

  • Arita R, Suehiro J, Haraguchi T, Shirakawa R, Tokoro H, Amano S. Objective image analysis of the meibomian gland area. Br J Ophthalmol. 2014 Jun;98(6):746-55. doi: 10.1136/bjophthalmol-2012-303014. Epub 2013 Jun 27.

    PubMed 23813417
  • Nelson JD, Shimazaki J, Benitez-del-Castillo JM, Craig JP, McCulley JP, Den S, Foulks GN. The international workshop on meibomian gland dysfunction: report of the definition and classification subcommittee. Invest Ophthalmol Vis Sci. 2011 Mar 30;52(4):1930-7. doi: 10.1167/iovs.10-6997b. Print 2011 Mar. No abstract available.

    PubMed 21450914
  • Schein OD, Munoz B, Tielsch JM, Bandeen-Roche K, West S. Prevalence of dry eye among the elderly. Am J Ophthalmol. 1997 Dec;124(6):723-8. doi: 10.1016/s0002-9394(14)71688-5.

    PubMed 9402817
  • Lekhanont K, Rojanaporn D, Chuck RS, Vongthongsri A. Prevalence of dry eye in Bangkok, Thailand. Cornea. 2006 Dec;25(10):1162-7. doi: 10.1097/01.ico.0000244875.92879.1a.

    PubMed 17172891
  • Uchino M, Dogru M, Yagi Y, Goto E, Tomita M, Kon T, Saiki M, Matsumoto Y, Uchino Y, Yokoi N, Kinoshita S, Tsubota K. The features of dry eye disease in a Japanese elderly population. Optom Vis Sci. 2006 Nov;83(11):797-802. doi: 10.1097/01.opx.0000232814.39651.fa.

    PubMed 17106406
  • Chia EM, Mitchell P, Rochtchina E, Lee AJ, Maroun R, Wang JJ. Prevalence and associations of dry eye syndrome in an older population: the Blue Mountains Eye Study. Clin Exp Ophthalmol. 2003 Jun;31(3):229-32. doi: 10.1046/j.1442-9071.2003.00634.x.

    PubMed 12786773
  • Schaumberg DA, Sullivan DA, Buring JE, Dana MR. Prevalence of dry eye syndrome among US women. Am J Ophthalmol. 2003 Aug;136(2):318-26. doi: 10.1016/s0002-9394(03)00218-6.

    PubMed 12888056
  • Wise RJ, Sobel RK, Allen RC. Meibography: A review of techniques and technologies. Saudi J Ophthalmol. 2012 Oct;26(4):349-56. doi: 10.1016/j.sjopt.2012.08.007.

    PubMed 23961019
  • Tomlinson A, Bron AJ, Korb DR, Amano S, Paugh JR, Pearce EI, Yee R, Yokoi N, Arita R, Dogru M. The international workshop on meibomian gland dysfunction: report of the diagnosis subcommittee. Invest Ophthalmol Vis Sci. 2011 Mar 30;52(4):2006-49. doi: 10.1167/iovs.10-6997f. Print 2011 Mar. No abstract available.

    PubMed 21450918
  • Simcock B. The role of the intrauterine device in contraceptive practice. Aust Fam Physician. 1976 Mar;5(2):124-39.

    PubMed 942338
  • Knop E, Knop N, Brewitt H, Pleyer U, Rieck P, Seitz B, Schirra F. [Meibomian glands : part III. Dysfunction - argument for a discrete disease entity and as an important cause of dry eye]. Ophthalmologe. 2009 Nov;106(11):966-79. doi: 10.1007/s00347-009-2043-9. German.

    PubMed 19941140
  • Arita R, Itoh K, Maeda S, Maeda K, Furuta A, Fukuoka S, Tomidokoro A, Amano S. Proposed diagnostic criteria for obstructive meibomian gland dysfunction. Ophthalmology. 2009 Nov;116(11):2058-63.e1. doi: 10.1016/j.ophtha.2009.04.037. Epub 2009 Sep 10.

    PubMed 19744718
  • Koh YW, Celik T, Lee HK, Petznick A, Tong L. Detection of meibomian glands and classification of meibography images. J Biomed Opt. 2012 Aug;17(8):086008. doi: 10.1117/1.JBO.17.8.086008.

    PubMed 23224195
  • Knop E, Knop N, Millar T, Obata H, Sullivan DA. The international workshop on meibomian gland dysfunction: report of the subcommittee on anatomy, physiology, and pathophysiology of the meibomian gland. Invest Ophthalmol Vis Sci. 2011 Mar 30;52(4):1938-78. doi: 10.1167/iovs.10-6997c. Print 2011 Mar. No abstract available.

    PubMed 21450915

Questions about this trial

What is this clinical trial studying?

A study testing Thermal pulsation, Intense pulsed light therapy and Manual warm compresses in people with Meibomian Gland Dysfunction, run by Zhongshan Ophthalmic Center, Sun Yat-sen University at 2 sites in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 13 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 25 Apr 2023. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Yuqing Deng, MD (18120557291, 15927646647@163.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mophology of meibomian glands (time frame: 30 days after commencement of treatment). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT04052841, which also lists 14 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.