Comparison of Subjective Refraction Measurement With SiVIEW Software and by an Expert

A study testing SiVIEW software and Expert in people with Eye Diseases and Ametropia, run by SiVIEW at 1 site in France.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Main objective Show that the equivalent sphere refraction measurements obtained by the SiVIEW software have no significant clinical difference with those obtained by an optometrist expert. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Eye Diseases, Ametropia
What is being tested?
SiVIEW software and Expert
Could I receive a placebo?
No, everyone receives an active treatment
How many people
111 participants
Where
1 siteFrance
Study dates
1 Nov 2022to 15 Jun 2023
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • subject with visual AV on each eye >= 8/10 and no amblyopia
  • subject with normal environmental transparency
  • subject with a healthy fundus that does not reveal major abnormalities that may affect vision
  • subject able to read the letters of the Latin alphabet
  • subject without heavy drug treatment that may affect vision as synthetic antimalarials, corticosteroids
  • non-diabetic subject
Show all 9Show fewer
  • non nystagmic subject
  • non-strabic subject
  • subject who has not had eye surgery less than 1 year old

Who cannot join

  • subject with a topography showing an anomaly (keratoconus type or other)

Trial locations and contacts

1 site in 1 country.

France1 site
  • Ophthalmological Foundation A. de Rothschild

    Paris

Other trials in France

About this study

Official title: Study of the Subjective Refraction Measured With SiVIEW Software Compared to That Obtained Manually by an Expert. Monocentric Cross-sectional Study

Main objective

Show that the equivalent sphere refraction measurements obtained by the SiVIEW software have no significant clinical difference with those obtained by an optometrist expert.

Read more

Secondary objectives

Show that the sphere refraction measurements obtained by the SiVIEW software have no significant clinical difference with those obtained by an optometrist expert.

Show that the cylinder refraction measurements obtained by the SiVIEW software have no significant clinical difference with those obtained by an optometrist expert.

Show that the visual acuity measurements obtained by the SiVIEW software do not have a significant clinical difference with those obtained by an optometrist expert.

Demonstrate that the report delivered for each eye exam by SiVIEW software is consistent and relevant to a vision health practitioner.

Show that the differences obtained between two eye examination experts are consistent with those found in the literature.

Read the detailed description

Introduction:

Ametropia is an eye refractive disorder typically leading to myopia, hyperopia, astigmatism... 4.5 billion people are currently affected worldwide, of whom only 2 billion are corrected. Of these people, about 30% (or +/- 600,000) do not benefit from an adequate correction. From an economic point of view, the cost on productivity of the absence or poor vision correction is estimated at more than 270 billion euros / year worldwide.

Any correction of an ametropic disorder necessarily requires a precise and relevant eye examination in order to identify the disorder and determine the best possible visual correction. The procedure for this examination is identical in all countries. It must follow specific steps[Benjamin 2006; Grosvenor 2007].

At present the refraction measurement is done manually by an expert. The optical correction may vary from one examination to another, for multiple physiological and/or examination-related reasons. Several factors can cause fluctuations in subjective refraction. These include examiner and subject characteristics [Grein et al. 2014].

In addition, there is also variability due to the measurement itself, depending on each examiner. The repeatability (on the same subject) and, in addition, the reproducibility (inter-examiners) of the subjective refraction measurement related to the examiner's interpretation, examination process and experience, among other things, are therefore important data to be defined in order to know the tolerance of the measurement determined between two examinations.

In studies involving a large number of subjects, the repeatability/reproducibility limit is greater than or equal to 0.71D [Zadnik 1994, Bullimore 1998, MacKenzie 2008], and more particularly during inter-examiner measurements.

The objective of the SiVIEW solution is therefore to make the measurement of subjective refraction both repeatable and reproducible regardless of the practitioner and his expertise, by completely automating the search for the best visual prescription using an expert system that aims to make this examination simple, intuitive and reliable.

It is therefore essential to compare the SiVIEW measurement with that of an optometrist expert.

The interest of the investigator's study will be to demonstrate that the results between a refraction expert and the measurement obtained by SiVIEW are clinically similar.

Main objective:

Show that the equivalent sphere refraction measurements obtained by the SiVIEW software have no significant clinical difference with those obtained by an optometrist expert.

Secondary objectives

Show that the sphere refraction measurements obtained by the SiVIEW software have no significant clinical difference with those obtained by an optometrist expert.

Show that the cylinder refraction measurements obtained by the SiVIEW software have no significant clinical difference with those obtained by an optometrist expert.

Show that the visual acuity measurements obtained by the SiVIEW software do not have a significant clinical difference with those obtained by an optometrist expert.

Demonstrate that the report delivered for each eye exam by SiVIEW software is consistent and relevant to a vision health practitioner.

Show that the differences obtained between two eye examination experts are consistent with those found in the literature.

Plan of the study:

It is a comparative prospective open monocentric cross-sectionnal study.

Being tested
SiVIEW software, Expert
Keywords
automatic sphere measurements, eye examination, ametropia
Medical subject headings
Eye Diseases, Refractive Errors

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • SiVIEW softwareOther

    Eye examination by a technician with the SiVIEW software

  • ExpertOther

    Eye examination by an expert including optimal correction in distance and near vision

Study groups in NCT04185701
Study groupRoleWhat they receive
Eye examination by expert and SiVIEW softwareExperimentalEye examination by an expert and by a technician with the SiVIEW system.
Receives: Other: SiVIEW software; Other: Expert

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Equivalent sphere measurement

    The subject will have an eye exam with SiVIEW and an examination by an optometrist expert. The equivalent sphere value will be evaluated in both cases.

    Measured: At time of inclusion

Secondary outcomes (5)
  1. Sphere measurement

    The subject will have an eye exam performed with SiVIEW and by an optometrist expert. The sphere value will be evaluated in both cases. The examination will be started in both cases on the objective measurement value obtained during the routine care examination.

    Measured: At time of inclusion

  2. Cylinder measurement

    The subject will have an eye exam performed with SiVIEW and by an optometrist expert. The cylinder value will be evaluated in both cases. The examination will be started in both cases on the objective measurement value obtained during the routine care examination.

    Measured: At time of inclusion

  3. Visual acuity

    The subject will have an eye exam performed with SiVIEW and by an optometrist expert. The value of visual acuity will be assessed in both cases.

    Measured: At time of inclusion

  4. Consistency and relevance of the SiVIEW report

    After each examination, the report edited by the SiVIEW software will be scored consistently on a Likert scale from 1 to 5 by an expert optometrist:

    1: very inconsistent / 2: inconsistent / 3: no opinion / 4: consistent / 5: very consistent

    After each examination, the report edited by SiVIEW software will be scored in relevance on a 5 item Likert scale by an expert optometrist:

    1: not relevant / not very relevant / no opinion / relevant / very relevant.

    Measured: At time of inclusion

  5. Measurement between two optometrist experts

    Each subject will have an eye examination conducted with an optometrist from the current care centre and by an optometrist expert mandated by SiVIEW (who is used to working at the Rothschild Foundation in clinical studies, among others). The equivalent sphere value will be evaluated in both cases.

    Measured: At time of inclusion

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted4 Dec 2019
  2. Study start1 Nov 2022
  3. Primary completion15 Jun 2023
  4. Study completion15 Jun 2023

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Sequential
Model description
* Eye examination by an expert including optimal correction in distance and near vision * Eye examination by a technician with the SiVIEW system * Analysis of the SiVIEW report by the expert who conducted the second review
Masking
None (open label)
Enrolment
111 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04185701
Sponsor's protocol ID
SiVIEW_refraction_Paris_study
Status last verified
Mar 2023
First submitted
30 Nov 2019
Last updated
24 Nov 2023
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing SiVIEW software and Expert in people with Eye Diseases and Ametropia, run by SiVIEW at 1 site in France.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 9 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 24 Nov 2023. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Equivalent sphere measurement (time frame: At time of inclusion). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 24 Nov 2023.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in