Diabetes MellitusMedInPSNCT04223973

MEDTRUM A7+ TouchCare Insulin Patch Pump (MedInPS)

A study testing Medtrum A7+ insulin Pump and Lab A1C in people with Diabetes Mellitus, run by Medtrum France at 1 site in France.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Demonstration study of the interest of the MEDTRUM A7+ TouchCare insulin patch pump versus INSULET Omnipod® patch pump Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Diabetes Mellitus
What is being tested?
Medtrum A7+ insulin Pump and Lab A1C
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
93 participants
Where
1 siteFrance
Study dates
29 Jan 2020to 1 Jun 2021
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patient with type 1 or 2 diabetes, 18 years of age and over
  • Patients already equipped with Omnipod® ® (INSULET) insulin patch pump and Abbott FreeStyleLibre sensor®.
  • A1C ranking from : >= 6,5% to <= 9.5%
  • Any type of rapid insulin except FIASP (which can be substituted if necessary) with 60 IU max per day (unauthorized use of insulin supplements by pen injector)
  • Patient able to receive and understand study information, give written informed consent, and easily participate to the trial
  • Patient affiliated to the French social security system

Who cannot join

  • - Patient already participating in another study
  • Patient under the protection of justice or under guardianship or curatorship
  • Type 2 diabetic patient requiring a daily insulin dose > 60 IU
  • Patients not suitable for suing insulin pump such as: severe psychiatric disorders, rapidly progressing ischemic or proliferative retinopathy before laser treatment, and exposure to high magnetic fields
  • Patients allergic to nickel and adhesive
  • Pregnant women, breastfeeding women
  • Or any other criteria as appreciated by the study investigator.

Trial locations and contacts

1 site in 1 country.

France1 site
  • Centre hospitalier Sud Francilien

    Corbeil-Essonnes, Essones, 91100

Other trials in France

About this study

Official title: Demonstration Study of the Interest of the MEDTRUM A7+ TouchCare Insulin Patch Pump Versus INSULET Omnipod® Patch Pump

Demonstration study of the interest of the MEDTRUM A7+ TouchCare insulin patch pump versus INSULET Omnipod® patch pump

Read the detailed description

Compare the daily use of two single usage patch pumps and demonstrate the interest of use of the A7+TouchCare® (Medtrum) patch pump for type 1 \& 2 diabetes management.

Main objective is an estimation of Patients A1C (Glycated Haemoglobin) based on the average blood glucose (obtained from a Flash glucose monitoring) and to compare it in both groups.

Secondary objectives:

  • comparison of mean A1C (laboratory values) between the two groups at Baseline and study end. Comparison to be established in a non-inferiority
  • Observe Skin \& overall Tolerance, Technical incidents with the device (failure, disconnection, catheter obstruction…)
  • Collection of Glucose metrics (minimum, maximum, average/median, standard deviation, out of range value, percentage of time in range, glucose variability)
Being tested
Medtrum A7+ insulin Pump, Lab A1C
Keywords
diabetes management, Insulin Pump
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Medtrum A7+ insulin PumpDevice

    Use of the Medtrum pump during 3 months in combination with a CGM (Continuous Glucose Measurement)

  • Lab A1CBiological

    Measure of Lab A1C at baseline and end of study

Study groups in NCT04223973
Study groupRoleWhat they receive
Active GroupExperimentalUsing the Medtrum Pump A7+ during 3 months
Receives: Device: Medtrum A7+ insulin Pump; Biological: Lab A1C
Control GroupActive comparatorusing the usual Insulet Patch pump
Receives: Biological: Lab A1C

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Estimation of A1C (%) Based on Average Blood Glucose Measured by Continuous Glucose Sensor on the Last 10 Weeks

    A1C will be estimated by a Continuous Glucose Measurement (FreeStyle Libre®) for all patients, at baseline and end of the study (12 weeks). The main objective is to demonstrate Non-inferiority efficacy (PP then ITT) of Medtrum A7+ insulin patch pump vs Insulet Omnipod® insulin patch pump (2 randomized groups). For the primary outcome, A1C will be estimated and appear in percentage based on assessment done by a Continuous Glucose Monitoring system.

    The A1C obtained in real life for the group using the comparator pump (Omnipod®) is estimated at 7.8%. Assuming that blood glucose level will be the same with Medtrum pump and based on a defined non-inferiority margin ∆ = + 0.4 (following FDA Guidance for Industry Diabetes Mellitus: Developing Drugs and Therapeutic Biologics for Treatment and Prevention).

    Measured: Month 3

Secondary outcomes (5)
  1. Report Patients' Laboratory Hemoglobin A1C Difference Between Baseline and 3 Months

    A1C is expressed in percentage based on Laboratory measurement. A1C is obtained \& measured from a blood sample.

    Measured: Month 3

  2. Compare The Time Spent in Range

    Glycemic Time In Range is the Time spent in the target. It is expressed in % of time.

    Time in Range is automatically calculated from the Sensor data wear.

    Measured: Month 3

  3. Compare Patients Pump Satisfaction

    Question: Overall, how satisfied were you with the pump? Satisfaction to be measured by a Patient Self- questionnaire through a Likert scale ranking from 1 (the less) to 5 (the most).

    Measured: Month 3

  4. Compare Patients PDM Satisfaction

    Question: How to qualify the discretion of use of the PDM (Personal Diabetes Manager) of the pump? the satisfaction scale went from 1 (very unsatisfied) to 5 (very satisfied)

    Measured: Month 3

  5. Number of Participants With at Least One Pump Occlusion

    occlusion reported by patient

    Measured: Month 3

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted13 Jan 2020
  2. Study start29 Jan 2020
  3. Primary completion1 Mar 2021
  4. Study completion1 Jun 2021

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Model description
Randomized controlled trial (RCT), open-label with two arms, multicentric national, prospective, non-inferiority methodology versus reimbursed comparator.
Masking
None (open label)
Enrolment
93 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04223973
Sponsor's protocol ID
2019-A02566-51
Acronym
MedInPS
Status last verified
Mar 2024
First submitted
11 Dec 2019
Last updated
16 Aug 2024
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • Amadou C, Melki V, Allain J, Clavel S, Gouet D, Chaillous L, Catargi B, Schaeplynck-Belicard P, Petit C, Thivolet C, Penfornis A. Performance and patients' satisfaction with the A7+TouchCare insulin patch pump system: A randomized controlled non-inferiority study. PLoS One. 2023 Aug 24;18(8):e0289684. doi: 10.1371/journal.pone.0289684. eCollection 2023.

    PubMed 37616289

Questions about this trial

What is this clinical trial studying?

A study testing Medtrum A7+ insulin Pump and Lab A1C in people with Diabetes Mellitus, run by Medtrum France at 1 site in France.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 16 Aug 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 16 Aug 2024). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Estimation of A1C (%) Based on Average Blood Glucose Measured by Continuous Glucose Sensor on the Last 10 Weeks (time frame: Month 3). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT04223973, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 16 Aug 2024.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in
MEDTRUM A7+ TouchCare Insulin Patch Pump (MedInPS) | AzTrials