AzTrials
AzTrials

Dialysis; ComplicationsNCT04285775

A Novel Device for Surveillance of Vascular Access Sites for Bleeding

An observational study of Dialysis; Complications and Wound Complication, run by Changi General Hospital at 1 site in Singapore.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Background Bleeding is often encountered after vascular access procedures for dialysis and fatal episodes of haemorrhage has been reported. A technological solution was sought to monitor for such rare but potentially catastrophic incidents. A novel device, BWATCH, was developed to detect fresh blood from wounds. Read more about this study

Who is it for?
Ages 21 years and older
What is being tested?
Application of device which monitors for bleeding - Blood Warning Technology with Continuous Hemoglobin sensor (BWATCH)
Type of treatment
Medical device
How many people
250 participants
Where
1 siteSingapore
Study dates
1 Mar 2019to 30 Jun 2020
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 21 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: Patients admitted to an acute care hospital with renal failure requiring dialysis. These patients were planned for dialysis catheter insertion or removal, based on standard clinical care and indications as identified by the primary nephrologist-in-charge.

Who may be able to join

  • Hospital inpatients planned for dialysis catheter insertion or removal, based on standard clinical care and indications as identified by the primary nephrologist-in-charge.

Who cannot join

  • Patients who declined participation, or
  • Patients with no legal representatives who were unable to provide consent

Trial locations and contacts

1 site in 1 country.

Singapore1 site
  • Changi General Hospital

    Singapore, 529889

Other trials in Singapore

About this study

Background Bleeding is often encountered after vascular access procedures for dialysis and fatal episodes of haemorrhage has been reported. A technological solution was sought to monitor for such rare but potentially catastrophic incidents. A novel device, BWATCH, was developed to detect fresh blood from wounds.

Aims The aim of this clinical trial was to evaluate the sensitivity and specificity of the device in detecting bleeding in a clinical environment.

Methods This a prospective, observational study on inpatients who will have a dialysis catheter inserted or removed. The device will be placed over the dressing for at least 6 hours. An alarm will be triggered if the device detects moisture and wavelength of light reflected specific for haemoglobin.

Read the detailed description

Background

Post-procedural wound hemorrhage is a potential complication following any invasive procedure. In the setting of hemodialysis, with multiple repeated vascular access procedures performed on uremic patients prone to bleeding, post-procedural bleeding would not be infrequent.

Serious adverse events following central venous catheter removal, while rare, were not isolated incidents. The British National Reporting and Learning System, reported 6 incidents of late bleeding following femoral line removal over 3 years. Of these, 3 resulted in deaths and 2 suffered more than one litre of blood loss. The UK Renal Association, British Renal Society and Intensive Care Society subsequently released a recommendation for the safe removal of a temporary femoral dialysis line in 2019. Protocols implemented in some units required an intense period of monitoring every 5-15 minutes in the first 2 hours.

However, frequent monitoring protocols places additional demands on already scarce nursing manpower. As catastrophic bleeding is rare, it is difficult to expect healthcare staff to maintain a consistent high level of vigilance for such a low probability event. The possibility of missed catastrophic bleeding is always present, particularly if the patient is incapacitated and unable to call for assistance.

A technological solution to safely monitor patients for bleeding following catheter removal is needed. Ideally, this technology should also be applicable for other wounds at risk of severe bleeding, for example, after arterial access for coronary angiogram or vascular interventions. There were no commercially available devices which fully met the user requirements, which included ease of use, wireless, compact form factor and specific detection of bleeding with no direct contact with the wound or blood. A compact device was developed

Through a research collaboration between a hospital and a technology and design university, supported by a national innovation grant, a device was developed to monitor wound for bleeding. By the end of 2018, a working prototype - Blood Warning Technology with Continuous Hemoglobin sensor (BWATCH) - was available for first-in-human clinical trials. The technological concept has been presented in electrical engineering conferences. This is a stand-alone disc-shaped device (approximately 5 cm in diameter and <1 cm thick) which can be placed over dressings and continuous monitoring for bleeding will take place. The presence of fresh blood triggers a loud alarm, alerting medical staff to attend to the patient.

Objectives

The primary aim of this clinical trial was to evaluate the sensitivity of the device in detecting bleeding in a clinical environment. The secondary aim is to determine the specificity of the device for actual bleeding (false positive rates).

Materials and Methods

This study has been reviewed and approved by the institutional review board (Singhealth cIRB 2014-2036).

This a prospective, observational study performed in an acute care hospital. Inpatients planned for dialysis catheter insertion or removal based on standard clinical care and indications will be identified fro recruitment.

Consent will be obtained from the patient or their legal representative if cognitively impaired. As part of standard care, the procedurist will secure haemostasis following the procedure and apply a bandage. The investigators will then place the device over the dressing and secure it in place. The device will then be left in-situ for at least 6 hours. Any alarms triggered and bleeding episodes will be documented in a case report form.

For safety, throughout the observation period, routine monitoring and inspection for bleeding by the nurse-in-charge would continue as per hospital clinical protocol.

Baseline demographic data of the patients will be collected. Laboratory test results collected will include the platelet count, international normalized ratio (INR) and partial thromboplastin time (PTT). The use of any anti-coagulation or anti-platelet medication will be recorded.

From the bleeding incidences and device alarm activation rates, the sensitivity and specificity of the device for detection of true bleeding will be calculated.

Being tested
Application of device which monitors for bleeding - Blood Warning Technology with Continuous Hemoglobin sensor (BWATCH)
Keywords
Dialysis, Catheter, Access, Bleeding, Haemorrhage, Mortality, Monitoring, Detection
Medical subject headings
Hemorrhage

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Application of device which monitors for bleeding - Blood Warning Technology with Continuous Hemoglobin sensor (BWATCH)Device

    A stand-alone disc-shaped device (approximately 5 cm in diameter and <1 cm thick) which can be placed over dressings. Once switched on and put in place, continuous monitoring for bleeding takes place. Detection occurs in 2 phases. First, the system monitors for the appearance of fluid in the dressing by detecting a change in capacitance. Next, a light sensor differentiates blood from other fluids by detecting that a unique spectrum of light (525nm) is absorbed, a property unique for hemoglobin. The presence of fresh blood triggers a loud alarm, alerting medical staff to attend to the patient. The device can be used over transparent plastic dressings and does not require direct contact with blood.

Study groups in NCT04285775
Study groupRoleWhat they receive
Insertion of Dialysis CatheterHospital inpatients planned for dialysis catheter insertion, based on standard clinical care and indications as identified by the primary nephrologist-in-charge.
Receives: Device: Application of device which monitors for bleeding - Blood Warning Technology with Continuous Hemoglobin sensor (BWATCH)
Removal of Dialysis CatheterHospital inpatients planned for dialysis catheter removal, based on standard clinical care and indications as identified by the primary nephrologist-in-charge.
Receives: Device: Application of device which monitors for bleeding - Blood Warning Technology with Continuous Hemoglobin sensor (BWATCH)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Number of bleeding events detected by device

    Number of bleeding episodes and detected by the device

    Measured: 0 to 6 hours after procedure

Secondary outcomes (2)
  1. Number of False Alarms

    Number of times the device alarm was triggered with no bleeding is observed.

    Measured: 0 till 6 hours after procedure or till alarm triggered (whichever is earlier)

  2. Number of Missed Bleeding Events

    Number of episodes of bleeding but no device alarm was triggered.

    Measured: 0 to 6 hours after procedure

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted26 Feb 2020
  2. Study start1 Mar 2019
  3. Primary completion30 Jun 2020
  4. Study completion30 Jun 2020

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
250 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04285775
Sponsor's protocol ID
NHIC-I2D-1608124
Status last verified
Jan 2021
First submitted
24 Feb 2020
Last updated
28 Jan 2021
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

An observational study of Dialysis; Complications and Wound Complication, run by Changi General Hospital at 1 site in Singapore.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 28 Jan 2021. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Singapore. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Number of bleeding events detected by device (time frame: 0 to 6 hours after procedure). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 28 Jan 2021; AzTrials last refreshed its copy on 7 Nov 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in