End Stage Renal Failure on DialysisNCT04292210

Human Factors Validation Testing for the Peripal System; a Manual Connection Assist Device for Peritoneal Dialysis Devices

A study testing Perisafe in people with End Stage Renal Failure on Dialysis and End Stage Renal Disease on Dialysis, run by Peripal AG at 1 site in United Kingdom.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

It is estimated that there are currently over 3 million patients receiving dialysis treatment worldwide. With effective pre-dialysis counselling, a majority of patients choose the home-based therapy peritoneal dialysis (PD) but only approximately 11% of prevalent dialysis patients use this modality. Read more about this study

Who is it for?
Any ageHealthy volunteers can take part
What is being tested?
Perisafe
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
24 participants
Where
1 siteUnited Kingdom
Study dates
19 Feb 2018to 30 Apr 2018
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Any age
  • All sexes
  • Healthy volunteers accepted
  • Child, Adult, Older adult

Fifteen patients and nine carers volunteered in this study, ranging from 23 to 86 years in age and from 0.3 to 24 years in experience in the PD therapy.

Trial locations and contacts

1 site in 1 country.

United Kingdom1 site
  • NIHR Trauma Management MedTech Co-operative

    Birmingham

Other trials in United Kingdom

About this study

Official title: Evaluating Patient Safety And Ease Of Use Of A Novel Connection-Assist Device For Peritoneal Dialysis

It is estimated that there are currently over 3 million patients receiving dialysis treatment worldwide. With effective pre-dialysis counselling, a majority of patients choose the home-based therapy peritoneal dialysis (PD) but only approximately 11% of prevalent dialysis patients use this modality. Connection-assist devices can overcome the challenges posed by decreased manual dexterity and/or visual acuity, and can allow more patients to be treated with home-based therapies. As part of the CE marking authorization, a connection device has been evaluated for safety and ease of use in a usability study.

Read the detailed description

The growth in the renal replacement therapy population is primaryly in the elderly age group, a population with a high frailty and comorbidity burden. Facilitating PD in this group of patients requires improved PD technology with greater accessibility tailored to patient needs. Following evaluation of specific barriers to the use of PD including feedback from established PD patients, a novel connection-assist device has been developed.

As part of the CE marking authorization, the connection device has been evaluated for safety and ease of use in a study at University Hospitals Birmingham NHS Foundation Trust involving 24 patients and carers.

The evaluated device is designed to assist peritoneal dialysis patients performing their daily dialysis treatment either for APD or for CAPD. The device facilitates the connection and disconnection of the dialysis line to the PD catheter transfer set. The connections are performed inside the device in a protected area, and the patient does not need to touch the transfer set or the dialysis line while they are unprotected. In addition, the breaking of the dialysis line frangible and the clamping of the dialysis line are replaced by pushing a button.

The operating procedure of the device consists of multiple tasks including preparing and loading the transfer set and dialysis line into the device, performing the handling steps related to the PD therapy with the device and finally unloading the used material from the device. Each task comprises several handling steps. During the study, the operating procedure with the device was simulated and the subjects were connected to a dummy catheter, which was attached to an apron.

This study recruited established PD patients and carers involved in the delivery of PD. Carers included healthcare professionals and lay carers.

The study design was divided into three one hour long parts. Firstly, a training session introduced the operation of the device. Secondly, a break of one hour allowed relaxation and recovery. Thirdly, a test was carried out to determine whether the subject could use the device independently with the aid of a quick start guide and the device manual.

The training consisted of two cycles in which the moderator guided the subject and three cycles in which the subject guided the moderator through the complete handling of the device. After the break, the subjects were asked to carry out a complete handling cycle themselves. Subjects could use the quick start guide and the device manual as help. After the unsupported handling cycle, subjects were asked to complete the NASA Task Load Index (N-TLX) questionnaire. In the meantime, the moderator left the room to obtain information on observed use errors. In a final semi-structured interview, the subjects were asked about the reasons for the observed use errors.

For each subject, one moderator and one observer were involved in the study. The moderator was the interface between the subject and the device both in training and in testing. This person was employed by an external agency to avoid any influence on the subject. The observer saw the handling with the help of a live feed from a separate room. He evaluated the handling steps in the categories of "safe use" and "use error" according to IEC 62366-1. For each use error, the observer described the observed situation from his point of view. In addition to evaluating the effectiveness of the task performance, the subjects were interviewed in a semi structured way. Here the observations of technique errors were discussed with the subjects in order to identify the specific causes. In this interview, the subjects were also asked for their feedback on the general ease of use of the device.

In addition, the subjects were asked about the task load in six different dimensions in the standardized N-TLX. Task load is a hypothetical construct that represents the cost incurred by a human operator to achieve a particular level of performance. The dimensions of the N-TLX are mental, physical and temporal demands as well as performance, effort and frustration in the task. Each dimension is evaluated on a 20-point scale. This scale is then transferred to the task load index, which is a scale between 0 and 100 points. To classify the results of the N-TLX, Eitrheim and Fernandes (2016) state "Workload levels below 50 were perceived as acceptable."

Being tested
Perisafe
Keywords
patient safety, usability study, connection-assist device, ease of use, peritoneal dialysis, NASA Task Load Index, renal replacement therapy
Medical subject headings
Renal Insufficiency

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • PerisafeDevice

    A medical device simplifying the connection of tubing as well as breaking of the frangible which is done in a sterile environment with the help of levers and buttons.

Study groups in NCT04292210
Study groupRoleWhat they receive
CAPD handlingExperimentalA connecting device simplifying the steps during a cycle of Peritoneal dialysis. Instead of directly doing a manual connection and manually breaking a frangible, the device assist in connecting and breaking the frangible. This study was done demonstrating a continuous ambulatory peritoneal dialysis (CAPD)
Receives: Device: Perisafe

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Safety and ease of use of a connection device during a continuous ambulatory peritoneal dialysis

    Firstly, a training session introduced the operation of the device. The normal steps of a manual connection and breaking of the frangible is replaced by the levers and the buttons in the device. The connections are thereby done in a 'safe' environment which reduces infection risk. A whole cycle of continuous ambulatory peritoneal dialysis is simulated during each training. Secondly, a break of one hour allowed relaxation and recovery. Thirdly, a test was carried out to determine whether the subject could use the device independently with the aid of a quick start guide and the device manual.

    The training consisted of two cycles in which the moderator guided the subject and three cycles in which the subject guided the moderator through the complete handling of the device.

    Each of the 24 subjects performed 52 handling steps resulting in a total of 1248 evaluated handling steps.

    Measured: 3 hours

  2. Nasa Task Load Index (N-TLX)

    The operating procedure of the device consists of multiple tasks comprising several handling steps. the subjects were asked about the task load in six different dimensions in the standardized N-TLX. Task load is a hypothetical construct that represents the cost incurred by a human operator to achieve a particular level of performance. The dimensions of the N-TLX are mental, physical and temporal demands as well as performance, effort and frustration in the task. Each dimension is evaluated on a 20-point scale. This scale is then transferred to the task load index, which is a scale between 0 and 100 points. Workload levels below 50 were perceived as acceptable.

    Measured: 3 hours

  3. Observation of technique

    Here the observations of technique errors were discussed with the subjects in order to identify the specific causes. In this interview, the subjects were also asked for their feedback on the general ease of use of the device.

    Measured: 3 hours.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted3 Mar 2020
  2. Study start19 Feb 2018
  3. Primary completion12 Mar 2018
  4. Study completion30 Apr 2018

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Model description
The growth in the RRT population is primary in the elderly age group, a population with a high frailty and comorbidity burden. Facilitating PD in this group of patients requires improved PD technology with greater accessibility tailored to patient needs. It is proposed that by making PD treatment simpler, more standardized and better protected from infections, patient access to PD could be increased significantly. Following evaluation of specific barriers to the use of PD including feedback from established PD patients, a novel connection-assist device has been developed. As part of the CE marking authorization, the connection device has been evaluated for safety and ease of use in a study at University Hospitals Birmingham NHS Foundation Trust involving 24 patients and carers.
Masking
None (open label)
Enrolment
24 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT04292210
Sponsor's protocol ID
IRAS ID :235715
Status last verified
Feb 2020
First submitted
19 Feb 2020
Last updated
3 Mar 2020
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Questions about this trial

What is this clinical trial studying?

A study testing Perisafe in people with End Stage Renal Failure on Dialysis and End Stage Renal Disease on Dialysis, run by Peripal AG at 1 site in United Kingdom.

The full registered description is in the About this study section and the official record.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 3 Mar 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Safety and ease of use of a connection device during a continuous ambulatory peritoneal dialysis (time frame: 3 hours). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 3 Mar 2020.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.