Non-invasive Continuous Blood Pressure Monitoring

A study testing Boppli Infant Blood Pressure Monitor in people with Blood Pressure, run by PyrAmes Inc. at 1 site in Canada.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The Boppli device was designed to provide continuous non-invasive blood pressure (cNIBP) monitoring for infants, including neonates, in intensive care settings by detecting systolic, diastolic, and mean arterial blood pressure values comparable to that of an invasive arterial line (IAL) within published accuracy… Read more about this study

Who is it for?
Ages Up to 12 monthsHealthy volunteers can take part
Conditions the study is for
Blood Pressure
What is being tested?
Boppli Infant Blood Pressure Monitor
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
8 participants
Where
1 siteCanada
Study dates
22 Jun 2022to 22 Sep 2022
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Up to 12 months
  • All sexes
  • Healthy volunteers accepted
  • Child

Who may be able to join

  • Groups 1 - 4:
  • Pediatric patients less than 1 year of age and less than 5000 g in weight (primary target)
  • Requires intra-arterial blood pressure monitoring
  • Group 5:
  • Pediatric patients
  • Requires intra-arterial blood pressure monitoring
  • NOTE: Under no circumstance should an invasive arterial line be implanted in a patient due to this study, if not medically necessary.

Who cannot join

  • Skin conditions that limit the placement of the device
  • Patients connected to or treated by other devices that may alter the pulse waveform
  • Patients with a cardiac or vascular malformation that results in an abnormal waveform inappropriate to be considered a reference standard.

Trial locations and contacts

1 site in 1 country.

Canada1 site
  • Foothills Medical Center Neonatal Intensive Care Unit, NICU 51

    Calgary, Alberta, T2N 4N1

Other trials in Canada

About this study

Official title: A Prospective Study to Evaluate the Safety and Effectiveness of the Boppli™ in Measuring Mean Arterial Pressure (MAP), Systolic Blood Pressure (SBP), and Diastolic Blood Pressure (DBP) as Compared to an Invasive Arterial Reference

The Boppli device was designed to provide continuous non-invasive blood pressure (cNIBP) monitoring for infants, including neonates, in intensive care settings by detecting systolic, diastolic, and mean arterial blood pressure values comparable to that of an invasive arterial line (IAL) within published accuracy criteria. The device consists of an array of capacitive sensors contained in a soft, flexible, biocompatible band. It applies proprietary algorithms to capacitance detections to generate accurate systolic, diastolic, and mean arterial blood pressure and waveform data, then communicates the data to an external device via Bluetooth Low Energy for storage and display. It may be applied to the patient's arm or foot on either side of the body, which is advantageous for infants whose bodies provide limited surface area for medical devices or equipment. It does not require high skin pressure when correctly placed. The device is intended for single use of up to 72 hours and is discarded following use.

Read the detailed description

The purpose of this study is to validate PyrAmes's non-invasive blood pressure measurements against invasive blood pressure measurements taken with a reference device that is routinely used in a hospital setting, is generally regarded as accurate, and is routinely used in the diagnosis and treatment of patients. In this case, the reference device is the invasive arterial line, which is routinely used in the care of infants in intensive care settings to collect accurate, continuous invasive blood pressure measurements.

The specific aim of this study is to record data obtained by use of the Boppli device, along with invasive measurements of blood pressure, for use in validating the Boppli device measurements and algorithm. The changes in capacitance will serve as an input for the Boppli algorithm and systolic, diastolic, and mean arterial pressure will be output. The measurements generated from the non-invasive Boppli Device will be validated against the invasive measurements from the reference device. This is a prospective, multi-center clinical study to validate the effectiveness of the Boppli Blood Pressure Monitor as compared to an invasive arterial reference. The study will evaluate the accuracy and precision of the Boppli system for continuous blood pressure monitoring by comparing SBP (systolic blood pressure), DBP (diastolic blood pressure), and MAP (mean arterial pressure) against the reference methodology.

Invasive arterial lines (IAL) were selected as the comparator for this study. IALs are the current gold standard for continuous blood pressure monitoring for infants in intensive care settings. Sphygmomanometers are widely used for non-invasive blood pressure monitoring among other populations but have documented issues with accuracy in patients with very low blood pressures, which includes many neonates and pediatric patients. Due to their potentially low accuracy in detecting blood pressure in infants, the manual sphygmomanometer is not regarded as an appropriate comparator for the Boppli device.

As stated in reference standards ISO 81060-2 and ANSI SP10, intra-arterial methods should not be used for subjects solely for the purpose of validating instrument performance. However, the SP10 standard states that in validation studies in children less than 3 years of age, the intra-arterial method should be used as the reference standard. In addition, it is noted that ISO 81060-2 explicitly states that children less than 3 years of age should not be included in clinical investigations using auscultatory reference readings. Therefore, this study will be conducted on clinical subjects (neonates/infants) in whom an intra-arterial line has already been placed for reasons other than validating instrument performance.

Primary objective: To assess the accuracy and precision of the Boppli system in measuring SBP, DBP, and MAP as compared to an invasive arterial line for continuous blood pressure monitoring.

Secondary Objective: To assess the safety of the Boppli system in a clinical setting.

Methods: Data to assess device efficacy, subject safety, and subject demographics will be collected from the Investigational sites. Data will be collected using an electronic data collection system (EDC) designed to comply with 21 CFR Part 11 and will be recorded into the electronic Case Report Form (eCRF). Primary source documentation will come directly from the subject's medical record

Conditions
Blood Pressure
Being tested
Boppli Infant Blood Pressure Monitor
Keywords
Neonates, Blood Pressure monitoring, Medical Device, Non-Invasive Blood Pressure Monitoring, Invasive Arterial Lines

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Boppli Infant Blood Pressure MonitorDevice

    The intervention is a medical device for the non-invasive measurement of blood pressure in infants

Study groups in NCT05224011
Study groupRoleWhat they receive
ExperimentalExperimentalThe Boppli device will be applied to the patients' arm or foot on either side of the body and is intended for single use of up to 72 hours and is discarded following use. The device will only be used in patients requiring continuous blood pressure monitoring, through the use of an invasive arterial line (IAL), to compare the data obtained by the Boppli to that obtained by the IAL.
Receives: Device: Boppli Infant Blood Pressure Monitor

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Assess the accuracy and precision of the Boppli System in continuous blood pressure monitoring

    The primary objective of this pivotal study is to assess the accuracy and precision of the Boppli system in measuring systolic blood pressure (SBP), diastolic blood pressure (DBP) and mean arterial pressure (MAP), as compared to an invasive arterial line for continuous blood pressure monitoring in infants.

    The following steps will be performed specifically to collect data for this study:

    • Boppli will be used alongside IALs onsite to collect blood pressure measurements, as well as pulse waveforms.
    • Continuous data from the IAL and Boppli will be paired for comparison
    • The total duration of data collection is expected to be approximately 6 hours with some patients remaining in the study up to 72 hours. However, subjects will not be required to remain in the study should clinical care indicate that the IAL should be removed.
    • Observed rapid changes in blood pressure will be collected and the Boppli's accuracy in detecting these changes will be compared against IAL

    Measured: 6 - 72 Hours

Secondary outcomes (1)
  1. Assess the safety of the Boppli System in a clinical setting

    The secondary objective is to assess the safety of the Boppli System in a clinical setting. This will be done by collecting and analyzing all adverse events observed during the study. Information on the usability of the device and its impact on workflow will also be collected.

    The following steps will be performed specifically to collect data for this study:

    • Boppli will be used alongside IALs onsite to collect blood pressure measurements, as well as pulse waveforms.
    • Continuous data from the IAL and Boppli will be paired for comparison
    • The total duration of data collection is expected to be approximately 6 hours with some patients remaining in the study up to 72 hours. However, subjects will not be required to remain in the study should clinical care indicate that the IAL should be removed.
    • Observed rapid changes in blood pressure will be collected and the Boppli's accuracy in detecting these changes will be compared against IAL

    Measured: 6 - 72 Hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted4 Feb 2022
  2. Study start22 Jun 2022
  3. Primary completion22 Sep 2022
  4. Study completion22 Sep 2022

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Model description
The study is a single-arm study in which patient measurements will be obtained using both the investigational device-Boppli-as well as traditional blood pressure measurement collected via IAL.
Masking
None (open label)
Enrolment
8 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05224011
Sponsor's protocol ID
PyrAmes-2021-01
Status last verified
Jun 2022
First submitted
4 Jan 2022
Last updated
22 Mar 2023
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Boppli Infant Blood Pressure Monitor in people with Blood Pressure, run by PyrAmes Inc. at 1 site in Canada.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 22 Mar 2023. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Canada. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Assess the accuracy and precision of the Boppli System in continuous blood pressure monitoring (time frame: 6 - 72 Hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 22 Mar 2023.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in