Visual AcuityNCT05259163

LFR-260 vs Traditional Phoropter in Visual Acuity Testing

A study testing LFR-260 portable phoropter and Traditional phoropter (SOC) in people with Visual Acuity and Astigmatism, run by Evolution Optiks Limited at 2 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to establish if LFR-260 (investigational device) is not inferior in effectiveness to a traditional phoropter (control device) when applied in visual acuity test in subjects undergoing a full routine eye examination. Read more about this study

Who is it for?
Ages 12 to 65 yearsHealthy volunteers can take part
Conditions the study is for
Visual Acuity, Astigmatism, Anisometropia
What is being tested?
LFR-260 portable phoropter and Traditional phoropter (SOC)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
112 participants
Where
2 sitesUnited States
Study dates
7 Jun 2022to 6 Jan 2023
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 12 to 65 years
  • All sexes
  • Healthy volunteers accepted
  • Child, Adult, Older adult

Who may be able to join

  • The participant or legal guardian is willing and able to understand, sign and date the Ethics committee approved study specific Informed Consent Form.
  • The participant is a male or female between the ages of 12 and 65 (inclusive).
  • The participant has no more than mild to moderate far-sightedness, near-sightedness, astigmatism or anisometropia.
  • The participant is free of ocular and systemic abnormalities that might affect visual functions.

Who cannot join

  • The participant has diabetes mellitus (Type 1 or 2).
  • The participant has an autoimmune condition.
  • The participant is pregnant (self-reported).
  • The participant has an active corneal or conjunctival infection.
  • The participant has an active corneal, conjunctival, or intraocular inflammation (i.e. - uveitis).
  • The participant has diabetic retinopathy.
Show all 16Show fewer
  • The participant has glaucoma or ocular hypertension.
  • The participant has macular degeneration.
  • The participant has had a previous ocular surgery.
  • The participant has ocular and systemic diseases or abnormalities that might affect visual functions.
  • The participant has a history of amblyopia, strabismus, or any other binocular vision abnormality.
  • The participant has a history of AMD (age macular degeneration).
  • The participant is a prisoner, a transient or has been treated for alcohol and/or drug abuse in an inpatient substance abuse program within 6 months prior to proposed study enrolment.
  • The participant will not be able to complete questionnaires.
  • The participant is currently in an investigational study for a similar purpose.
  • The participant, in the judgment of the Investigator, may be inappropriate for the intended study procedures.

Trial locations and contacts

2 sites in 1 country.

United States2 sites
  • Gold Coast Optometric Vision

    Oyster Bay, New York, 11771

  • MCPHS

    Worcester, Massachusetts, 01608

Other trials in United States

About this study

Official title: An Evaluation of the LFR-260 Against a Traditional Phoropter in Visual Acuity Testing

The purpose of this study is to establish if LFR-260 (investigational device) is not inferior in effectiveness to a traditional phoropter (control device) when applied in visual acuity test in subjects undergoing a full routine eye examination. LFR-260 proposes to provide information of the visual capabilities of the patient. The test will be provided and supervised by a qualified eye care provider. The informed consent, screening, randomization (into the order of visual test device used), (visual) device testing (to include precision testing, remote and offsite testing) will all occur at a single visit.

Being tested
LFR-260 portable phoropter, Traditional phoropter (SOC)
Keywords
Ocular
Medical subject headings
Astigmatism, Anisometropia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • LFR-260 portable phoropterDevice

    Portable unit to evaluate visual refractive state of the patient

  • Traditional phoropter (SOC)Device

    Standard unit for evaluating visual refractive state of the patient

Study groups in NCT05259163
Study groupRoleWhat they receive
Group 1: LFR-260 first, then traditional phoropterExperimentalThe LFR-260 was administered first in this group, followed by the traditional phoropter. The LFR-260 is a portable digital refractor which allows for determination of refractive error as well as for fully remote refractions during eye exams. The traditional phoropter is intended to be used for distance vision testing in any environment.
Receives: Device: LFR-260 portable phoropter; Device: Traditional phoropter (SOC)
Group II: Traditional Phoropter first, then LFR-260Active comparatorThe traditional phoropter was administered first in this group, followed by the traditional phoropter. The traditional phoropter is intended to be used for distance vision testing in any environment. The LFR-260 is a portable digital refractor which allows for determination of refractive error as well as for fully remote refractions during eye exams.
Receives: Device: LFR-260 portable phoropter; Device: Traditional phoropter (SOC)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Effectiveness of LFR260 on Visual Acuity Test

    Agreement of LFR-260 (investigational device) to traditional phoropter (comparator device) when applied in visual acuity test (eye chart test) in subjects undergoing a full routine eye examination. Agreement will be determined by demonstrating that the LFR-260 is within 0.5D (+/-) in measuring sphere and cylinder results and within 10 degrees for axis results among subjects with low astigmatism and within 5 degrees among subjects with moderate to high astigmatism from the traditional phoropter.

    Measured: Day 0

  2. Precision of LFR260 in Repeated Testing

    Precision testing will also be conducted to confirm repeatability and reproducibility of the LFR-260.

    Measured: Day 0

Secondary outcomes (2)
  1. Patient Satisfaction Survey

    4-question survey (1=worst; 5=best) given to patients to determine satisfaction with LFR-260 device Quad view.

    Q1: Was the LFR-260 device Quad view comfortable to use for your eye exam? (1=very uncomfortable - 5=very comfortable) Q2: Did the LFR-260 Quad view feel the same as the traditional phoropter when taking your eye exam? (1=LFR-260 definitely more uncomfortable - 5=LFR-260 definitely more comfortable) Q3: Did you prefer to read the eye chart using the LFR-260 device Quad view or traditional phoropter? (1=definitely prefer phoropter - 5=definitely prefer LFR-260) Q4: If taking an eye exam in the future, would you prefer your doctor use the LFR-260 Quad view or traditional phoropter? (1=definitely prefer phoropter - 5=definitely prefer LFR-260)

    Measured: Day 0

  2. Provider Satisfaction Survey

    4-question survey (1=worst; 5=best) given to eye care provider to determine satisfaction with LFR-260 related to usability, convenience.

    Q1: LFR-260 device performed as expected according to the Instructions for Use (IFU)? (1=definitely did not - 5=definitely) Q2: LFR-260 device was easy to use? (1=very difficult to use - 5=very easy to use) Q3: LFR-260 device was intuitive to operate? (1=definitely counterintuitive - 5=very intuitive) Q4: LFR-260 was more or less convenient to use than the phoropter? (1=definitely less convenient- 5=definitely more convenient)

    Measured: Day 0

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted28 Feb 2022
  2. Study start7 Jun 2022
  3. Primary completion6 Jan 2023
  4. Study completion6 Jan 2023

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Screening
Allocation
Randomised
Intervention model
Crossover
Model description
The investigational product was compared to the traditional phoropter. Subjects were randomized to the order of testing.
Masking
None (open label)
Enrolment
112 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05259163
Sponsor's protocol ID
LFR-260-2021
Status last verified
Aug 2024
First submitted
9 Feb 2022
Last updated
25 Sep 2024
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
Undecided

Questions about this trial

What is this clinical trial studying?

A study testing LFR-260 portable phoropter and Traditional phoropter (SOC) in people with Visual Acuity and Astigmatism, run by Evolution Optiks Limited at 2 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 16 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 25 Sep 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 25 Sep 2024). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Effectiveness of LFR260 on Visual Acuity Test (time frame: Day 0). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 25 Sep 2024.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in