Trial summary

The aim of this study is to determine the accuracy of devices called pulse oximeters, which measure blood oxygen by shining light through fingers, ears or other skin, without requiring blood sampling. Study will be used with patients at rest Read more about this study

Who is it for?
Ages 18 to 50 yearsHealthy volunteers can take part
Conditions the study is for
Hypoxia
What is being tested?
Pulse oximeter
Type of treatment
Medical device
How many people
12 participants
Where
1 siteUnited States
Study dates
23 Mar 2022to 29 Oct 2026
Phase
Observational
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 50 years
  • All sexes
  • Healthy volunteers accepted
  • Adult

Study population: Healthy adult volunteers. At least 10 subjects will be recruited to complete the study. Of these 2 will be dark pigmented. If the number of samples is less than the required 200 or if less than 2 of dark pigment subjects successfully complete the study, subjects will be added to include 2 of dark pigmented subjects and at least 200 data points.

Who may be able to join

  • Both male and female subjects who can give written informed consent
  • Healthy subjects capable of undergoing controlled hypoxemia to the levels outlined in the desaturation profile
  • Meeting the demographic requirements

Who cannot join

  • Pregnant women
  • Significant arrhythmia
  • Blood pressure above 150 systolic or 90 diastolic
  • Carboxyhemoglobin levels over 3%
  • Subjects whom the investigator consider ineligible for the study

Trial locations and contacts

1 site in 1 country.

United States1 site
  • UCSF Hypoxia Research Laboratory

    San Francisco, California, 94133

Other trials in United States

About this study

Official title: Accuracy of Pulse Oximeters With Profound Hypoxia NIHO 14

The aim of this study is to determine the accuracy of devices called pulse oximeters, which measure blood oxygen by shining light through fingers, ears or other skin, without requiring blood sampling. Study will be used with patients at rest

Read the detailed description

This study is intended to evaluate performance of a new design pulse oximeter (test device) manufactured by Nihon Kohden Corporation sufficiently to support performance claims for an FDA 510K submission or ISO technical file. Specifically, SpO2 and pulse rate accuracy will be assessed for a Nihon Kohden OLV-4202 pulse oximeter with adult patients under a controlled setting of varying levels of inhaled oxygen concentration levels for patients at rest.

Conditions
Hypoxia
Being tested
Pulse oximeter
Medical subject headings
Hypoxia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Pulse oximeterDevice

    OLV-4202 pulse oximeter (SW version: 01-12)

Study groups in NCT05297500
Study groupRoleWhat they receive
Adult healthy subjectsAdult healthy subjects capable of undergoing controlled hypoxemia to the levels outlined in the desaturation profile in an at rest state
Receives: Device: Pulse oximeter

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Root-mean-square (Rms) of device SpO2 measurement with room air stabilization is within +-2% [%SpO2] of reference co-oximeter SaO2 measurement at varying desaturation levels (70% to 100%)

    Each subject will have an arterial catheter inserted in the left or right wrist and the pulse oximeter study device placed on different fingers of the left or right hand, or the left or right earlobe or the forehead. Measurement position is randomized across subjects. For each subject, a series of desaturation runs are performed.

    Run 1 stabilized room air period -> stabilized plateaus at saturation level targets:

    room air, 92%, 87%, 82%, 77%, 72%

    Run 2 stabilized 100% O2 period -> stabilized plateaus at saturation level targets:

    100%, 93%, 88%, 83%, 78%, 73%

    SpO2 is observed breath by breath arterial saturation, which is in turn computed from end tidal PO2 and PCO2. FiO2 is reduced suddenly at first then adjusted to achieve a stable plateau value at the desired target. Each desaturation plateaus should not exceed 10 minutes. A blood draw is taken at stable FiO2 and +30 seconds. The gas mixture is then changed for the next value and repeated for each plateau value.

    Measured: 30 minutes

  2. Root-mean-square (Rms) of device Pulse Rate measurement with room air stabilization is within +-3 [1/min] of reference ECG Pulse Rate measurement at heart rates (30 to 300 [1/min])

    Each subject will have an arterial catheter inserted in the left or right wrist and the pulse oximeter study device placed on different fingers of the left or right hand, or the left or right earlobe or the forehead. Measurement position is randomized across subjects. For each subject, a series of desaturation runs are performed.

    Run 1 stabilized room air period -> stabilized plateaus at saturation level targets:

    room air, 92%, 87%, 82%, 77%, 72%

    Run 2 stabilized 100% O2 period -> stabilized plateaus at saturation level targets:

    100%, 93%, 88%, 83%, 78%, 73%

    Pulse rate is recorded from the test device and from the reference ECG monitor. PR data measurements are gathered during the stable plateau value at the desired saturation target. Each desaturation plateaus should not exceed 10 minutes. The gas mixture is then changed for the next value and repeated for each plateau value.

    Measured: 30 minutes

Secondary outcomes (1)
  1. Root-mean-square (Rms) of device SpO2 measurement with room air stabilization is within +-2% [%SpO2] of reference co-oximeter SaO2 measurement at desaturation levels (80% to 100%) and within +-3% [%SpO2] at desaturation levels (70% to 80%)

    Each subject will have an arterial catheter inserted in the left or right wrist and the pulse oximeter study device placed on different fingers of the left or right hand, or the left or right earlobe or the forehead. Measurement position is randomized across subjects. For each subject, a series of desaturation runs are performed.

    Run 1 stabilized room air period -> stabilized plateaus at saturation level targets:

    room air, 92%, 87%, 82%, 77%, 72%

    Run 2 stabilized 100% O2 period -> stabilized plateaus at saturation level targets:

    100%, 93%, 88%, 83%, 78%, 73%

    SpO2 is observed breath by breath arterial saturation, which is in turn computed from end tidal PO2 and PCO2. FiO2 is reduced suddenly at first then adjusted to achieve a stable plateau value at the desired target. Each desaturation plateaus should not exceed 10 minutes. A blood draw is taken at stable FiO2 and +30 seconds. The gas mixture is then changed for the next value and repeated for each plateau value.

    Measured: 30 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted28 Mar 2022
  2. Study start23 Mar 2022
  3. Primary completion28 Mar 2022
  4. Study completion29 Oct 2026 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Case only
Time perspective
Prospective
Enrolment
12 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05297500
Sponsor's protocol ID
71Ag_Vital-0031
Status last verified
Nov 2024
First submitted
17 Mar 2022
Last updated
24 Nov 2025
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No: Data to be used by Sponsor only

Questions about this trial

What is this clinical trial studying?

An observational study of Hypoxia, run by Nihon Kohden at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: under way, not looking for participants, as of the sponsor's last update on 24 Nov 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Root-mean-square (Rms) of device SpO2 measurement with room air stabilization is within +-2% [%SpO2] of reference co-oximeter SaO2 measurement at varying desaturation levels (70% to 100%) (time frame: 30 minutes). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 24 Nov 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in