Coronary Artery DiseaseNCT05320185

Evaluation on the Effectiveness and Safety of RuiXin-CoronaryAI for Diagnosis of Coronary Artery Stenosis

An observational study of Coronary Artery Disease and Artificial Intelligence, run by Shenzhen Raysight Intelligent Medical Technology Co., Ltd. at 3 sites in China.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

With the emergence of advanced technology to date in the artificial intelligence (AI), computer aided diagnosis has gradually gained its popularity in the field of healthcare. Particularly, in the clinical practice of coronary artery disease diagnosis, the application of AI could be of great implication in alleviating… Read more about this study

Who is it for?
Ages 18 years and older
What is being tested?
RuiXin-CoronaryAI software
Type of treatment
Medical device
How many people
615 participants
Where
3 sitesChina
Study dates
28 Jul 2021to 30 Dec 2022
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Study population: Patients ≥18 years old with suspected or known coronary artery disease who underwent CCTA examination.

Who may be able to join

  • layer thickness of CCTA images should be less than 1mm, image quality should be up to DICOM 3.0 standards;
  • vessels should be clearly developed, contrast medium ought to be well filled, the average of CT value of aortic root cavity should be between 325-600HU in CCTA image;
  • remodeling of vessels should be intact, including coronary artery and branches, without missed or inaccurate slices;
  • CCTA image should be obtained from single- or dual-source computed tomography (CT) scanners with a minimum of 64 detector rows.

Who cannot join

  • CCTA image is of poor quality due to motion artifact, severe calcification, metal coverage, noise, poor contrast medium injection and other variables influencing the diagnosis of stenosis;
  • previous percutaneous coronary intervention (PCI) or coronary artery bypass grafting (CABG);
  • anomalous origin of coronary artery;
  • other non-atherosclerosis-related coronary diseases like coronary artery fistula, aneurysm, coronary artery ectasia, arteritis coronaria, etc.;
  • repeated enrollment;
  • other conditions not suitable for enrollment.

Trial locations and contacts

3 sites in 1 country.

China3 sites, 3 recruiting
  • Beijing Hospital

    Beijing, Beijing Municipality, 100730

    Min Chen

    Recruiting
  • Sun Yat-sen Memorial Hospital

    Guangzhou, Guangdong, 510120

    Jun Shen

    Recruiting
  • Tongji Hospital

    Wuhan, Hubei, 430030

    Liming Xia, 13607176908, lmxia@tjh.tjmu.edu.cn

    Recruiting

Other trials in China

About this study

Official title: Evaluation on the Effectiveness and Safety of AI-based Coronary CT Angiographic Analysis Software (RuiXin-CoronaryAI) for Diagnosis of Coronary Artery Stenosis

With the emergence of advanced technology to date in the artificial intelligence (AI), computer aided diagnosis has gradually gained its popularity in the field of healthcare. Particularly, in the clinical practice of coronary artery disease diagnosis, the application of AI could be of great implication in alleviating the shortage of medical sources. To evaluate the effectiveness and safety of the AI-based coronary CT angiographic analysis software (RuiXin-CoronaryAI) for diagnosis of coronary artery stenosis, a retrospective, multi-center, cross-over designed, blinded, sensitivity superiority and specificity non-inferiority clinical trial will be conducted.

Read the detailed description

Patients ≥18 years old with suspected or known coronary artery disease who underwent CCTA will be included. CCTA images of subjects should be of good quality up to the DICOM 3.0 standards, obtained by CT scan with ≥64-slices. The subjects with unqualified CTA will be excluded. CCTA images will be analyzed in three methods (3 groups). Control group: CCTA images will be visually evaluated by physicians. Experiment group: CCTA images will be evaluated by physicians using RuiXin-CoronaryAI. Reference group: CCTA images will be visually evaluated by cardiologists with at least 10 years experiences, and the conclusions they offer will be used as golden standard. Primary outcomes are diagnostic sensitivity and specificity of RuiXin-CoronaryAI and coronary CTA for diagnosis of ischemia on a per-vessel basis. The effectiveness of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis will be conducted by testing superiority of diagnostic sensitivity and non-inferiority of specificity.

Being tested
RuiXin-CoronaryAI software
Medical subject headings
Coronary Artery Disease, Coronary Stenosis

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • RuiXin-CoronaryAI softwareDevice

    RuiXin-CoronaryAI, based on Computed Tomography Angiography (CTA) and was independently designed by RaysightMed Inc., which has been already authorized by National Medical Products Administration (NMPA).

Study groups in NCT05320185
Study groupRoleWhat they receive
Control groupCCTA images will be visually evaluated by physicians.
Experiment groupCCTA images will be evaluated by physicians using RuiXin-CoronaryAI.
Receives: Device: RuiXin-CoronaryAI software
Reference groupCCTA images will be visually evaluated by cardiologists with at least 10 years experiences, and the conclusions they offer will be used as golden standard.

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Per-vessel diagnostic sensitivity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis

    Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-vessel basis

    Measured: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.

  2. Per-vessel diagnostic specificity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis

    Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-vessel basis

    Measured: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.

Secondary outcomes (2)
  1. Per-patient diagnostic sensitivity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis

    Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-patient basis

    Measured: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.

  2. Per-patient diagnostic specificity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis

    Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-patient basis

    Measured: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted11 Apr 2022
  2. Study start28 Jul 2021
  3. Primary completion30 Jul 2022
  4. Study completion30 Dec 2022 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Retrospective
Enrolment
615 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05320185
Sponsor's protocol ID
RuiXin-CoronaryAI
Status last verified
Mar 2022
First submitted
16 Mar 2022
Last updated
21 Sep 2022
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

An observational study of Coronary Artery Disease and Artificial Intelligence, run by Shenzhen Raysight Intelligent Medical Technology Co., Ltd. at 3 sites in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 21 Sep 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 3 sites in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Liming Xia (13607176908, lmxia@tjh.tjmu.edu.cn), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Per-vessel diagnostic sensitivity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis (time frame: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 21 Sep 2022.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in