Age-Related Macular DegenerationNCT05686421

Comfortable and Stabilizing Chin & Forehead Rest Attachment for Slit Lamp Configurations

A study testing OCT Imaging Using Device N and OCT Imaging Using Device C in people with Age-Related Macular Degeneration and Diabetic Retinopathy, run by NYU Langone Health at 1 site in United States.

Terminated

Stopped early. The study ended before it was planned to and will not resume.

Official record
Search all trials

Trial summary

In this study, participants will be imaged using two Optical Coherence Tomography (OCT) devices: device N, a standard conventional OCT device with an invention (comfortable chin and forehead rest that can be adjusted to fit each individual's size) attached to the device; and device C, the standard conventional OCT… Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
What is being tested?
OCT Imaging Using Device N and OCT Imaging Using Device C
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
42 participants
Where
1 siteUnited States
Study dates
30 Mar 2023to 28 Apr 2023
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Healthy volunteers; OR,
  • AMD, diabetic retinopathy, central serous chorioretinopathy, and/or glaucoma or glaucoma suspects or any other suspected eye disease
  • Group-Specific Inclusion Criteria:
  • -- There are no group-specific inclusion criteria for patients with AMD, diabetic retinopathy, and central serous chorioretinopathy.
  • Healthy Volunteers
  • A normal clinical ophthalmic examination.
Show all 14Show fewer
  • Reliable VF, reproducible glaucoma hemifield tests labeled within normal limits on at least one test.
  • Good quality (adequate signal strength ≥ 6/10, without segmentation algorithm failure and motion artifacts > 1 vessel diameter) RNFL and Ganglion Cell Inner Plexiform Layer (GCIPL) OCT, within normal limits.
  • Primary Open Angle Glaucoma (POAG)
  • Clinical characteristics of glaucoma: optic nerve head (ONH) abnormalities: global rim thinning, rim notch, or disc hemorrhage; retinal nerve fiber layer (RNFL) defect.
  • Typical glaucomatous field loss in reliable VF, reproducible glaucoma hemifield tests labeled outside normal limits on at least two consecutive tests.
  • Good quality (adequate signal strength ≥ 6/10, without segmentation algorithm failure and motion artifacts > 1 vessel diameter) RNFL and GCIPL OCT, labeled outside normal limits with typical glaucomatous RNFL and GCIL thinning.
  • Normal Tension Glaucoma (NTG)
  • Identical to POAG criteria with IOP recorded at ≤ 21 mmHg at any time point.

Who cannot join

  • An individual who meets any of the following criteria will be excluded from participation in this study:
  • Any medical treatment (e.g., chronic corticosteroid, hydroxychloroquine, chloroquine, thioridazine, canthaxanthine) or conditions that affect VF (e.g., stroke) and retinal thickness other than glaucoma.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • NYU Langone Health

    New York, New York, 10017

Other trials in United States

About this study

Official title: A Single-Center Proof Of Concept Study of a Novel Comfortable and Stabilizing Chin & Forehead Rest Attachment for Slit Lamp Configuration Devices

In this study, participants will be imaged using two Optical Coherence Tomography (OCT) devices: device N, a standard conventional OCT device with an invention (comfortable chin and forehead rest that can be adjusted to fit each individual's size) attached to the device; and device C, the standard conventional OCT device with no invention attached. The investigators will assess whether the chin and forehead rest attachment (invention) provides a more comfortable experience for patients.

Being tested
OCT Imaging Using Device N, OCT Imaging Using Device C
Medical subject headings
Macular Degeneration, Diabetic Retinopathy, Central Serous Chorioretinopathy, Glaucoma, Ocular Hypertension, Eye Diseases

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • OCT Imaging Using Device NDevice

    Standard conventional OCT imaging device equipped with extendable and comfortable chin and forehead rest.

  • OCT Imaging Using Device CDevice

    Standard conventional OCT imaging device.

Study groups in NCT05686421
Study groupRoleWhat they receive
Device C (OD-OS), then Device N (OD-OS)ExperimentalParticipants will be randomized to receive 1 macular scan on each eye and 1 optic nerve head (ONH) scan on each eye from two devices: device C, a standard conventional device with no attachment; and device N, the standard conventional device with the invention (comfortable chin and forehead rest that can be adjusted to fit each individual's size) attached to the device. Participants will first be imaged using device C starting with the right eye (OD), then left eye (OS). They will then be imaged using device N, starting with OD, then OS.
Receives: Device: OCT Imaging Using Device N; Device: OCT Imaging Using Device C
Device C (OS-OD), then Device N (OS-OD)ExperimentalParticipants will be randomized to receive 1 macular scan on each eye and 1 optic nerve head (ONH) scan on each eye from two devices: device C, a standard conventional device with no attachment; and device N, the standard conventional device with the invention (comfortable chin and forehead rest that can be adjusted to fit each individual's size) attached to the device. Participants will first be imaged using device C starting with the left eye (OS), then right eye (OD). They will then be imaged using device N, starting with OS, then OD.
Receives: Device: OCT Imaging Using Device N; Device: OCT Imaging Using Device C
Device N (OD-OS), then Device C (OD-OS)ExperimentalParticipants will be randomized to receive 1 macular scan on each eye and 1 optic nerve head (ONH) scan on each eye from two devices: device N, a standard conventional device with the invention (comfortable chin and forehead rest that can be adjusted to fit each individual's size) attached to the device; and device C, the standard conventional device with no attachment. Participants will first be imaged using device N starting with the right eye (OD), then left eye (OS). They will then be imaged using device C, starting with OD, then OS.
Receives: Device: OCT Imaging Using Device N; Device: OCT Imaging Using Device C
Device N (OS-OD), then Device C (OS-OD)ExperimentalParticipants will be randomized to receive 1 macular scan on each eye and 1 optic nerve head (ONH) scan on each eye from two devices: device N, a standard conventional device with the invention (comfortable chin and forehead rest that can be adjusted to fit each individual's size) attached to the device; and device C, the standard conventional device with no attachment. Participants will first be imaged using device N starting with the left eye (OS), then right eye (OD). They will then be imaged using device C, starting with OS, then OD.
Receives: Device: OCT Imaging Using Device N; Device: OCT Imaging Using Device C

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Percentage of Participants Who Indicate Device N Provided a More Comfortable Experience, Compared With Device C

    Participants will be asked which device provided a more comfortable experience: Device C or Device N.

    Measured: Post-Imaging Session (Day 1)

  2. Comfort Level Rating of Device N on 0-5 Scale

    Participants will be asked to rate how comfortable Device N was, on a scale of 0-5. Higher scores indicate higher levels of comfort.

    Measured: Post-Imaging Session (Day 1)

  3. Comfort Level Rating of Device C on 0-5 Scale

    Participants will be asked to rate how comfortable Device C was, on a scale of 0-5. Higher scores indicate higher levels of comfort.

    Measured: Post-Imaging Session (Day 1)

Secondary outcomes (2)
  1. Time Duration to Adjust for Number of Motion Artifacts Present With Device N Imaging

    The OCT imaging device includes an "eye tracking" feature that selects a specific location within the eye and constantly adjusts, i.e. tracks, for fine fixation adjustments that occur when a patient's eye is moving very slightly during imaging. The time to adjust for the number of motion artifacts present will be defined as the duration it takes from the moment the imager clicks "Acquire" on the device to when the progress bar is "Full."

    Measured: Imaging Session (Day 1)

  2. Time Duration to Adjust for Number of Motion Artifacts Present With Device C Imaging

    The OCT imaging device includes an "eye tracking" feature that selects a specific location within the eye and constantly adjusts, i.e. tracks, for fine fixation adjustments that occur when a patient's eye is moving very slightly during imaging. The time to adjust for the number of motion artifacts present will be defined as the duration it takes from the moment the imager clicks "Acquire" on the device to when the progress bar is "Full."

    Measured: Imaging Session (Day 1)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted17 Jan 2023
  2. Study start30 Mar 2023
  3. Primary completion28 Apr 2023
  4. Study completion28 Apr 2023

Why it stopped: Study was stopped prior to collection of data from participants due to investigator transfer.

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Supportive care
Allocation
Randomised
Intervention model
Crossover
Masking
Double blind
Who is masked
Participant, Investigator
Enrolment
42 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05686421
Sponsor's protocol ID
22-01420
Status last verified
Oct 2024
First submitted
6 Jan 2023
Last updated
5 Nov 2024
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Yes: The de-identified participant data from the final research dataset used in the published manuscript will be shared upon reasonable request beginning 9 months and ending 36 months following article publication or as required by a condition of awards and agreements supporting the research provided the investigator who proposes to use the data executes a data use agreement with NYU Langone Health. Requests may be directed to: \[Gadi.Wollstein@nyulangone.org\]. The protocol and statistical analysis plan will be made available on Clinicaltrials.gov only as required by federal regulation or as a condition of awards and agreements supporting the research.

Questions about this trial

What is this clinical trial studying?

A study testing OCT Imaging Using Device N and OCT Imaging Using Device C in people with Age-Related Macular Degeneration and Diabetic Retinopathy, run by NYU Langone Health at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 14 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: stopped early, as of the sponsor's last update on 5 Nov 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 5 Nov 2024). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Percentage of Participants Who Indicate Device N Provided a More Comfortable Experience, Compared With Device C (time frame: Post-Imaging Session (Day 1)). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 5 Nov 2024.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in