Long Term Results With the Vibrant Soundbridge in Patients With Mixed Hearing Loss: a 60-month Longitudinal Study

A study testing Vibrant Soundbridge active middle ear implant in people with Mixed Hearing Loss, run by MED-EL Elektromedizinische Geräte GmbH at 1 site in France.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Vibrant Soundbridge (VSB, MED-EL, Innsbruck) Active middle ear implants (AMEI) can benefit patients who are unable to use conventional hearing aids (HA) due to medical reasons, discomfort or unsuccessful rehabilitation. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Mixed Hearing Loss
What is being tested?
Vibrant Soundbridge active middle ear implant
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
31 participants
Where
1 siteFrance
Study dates
1 Sep 2008to 1 Sep 2018
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Adult patient (>= 18 years old) speaking French
  • with mixed hearing loss on at least one side
  • with a maximum bone conduction (BC) threshold of 45 dB hearing level (HL) at 500 Hz, 50 dB HL at 1000 Hz, 55 dB HL at 1500 Hz, and 60 dB HL at 2000 Hz.
  • with a 20 dB minimum air-bone gap (defined as the difference between bone- and air conduction thresholds).
  • with stable BC thresholds over the previous 24 months, and no sufficient benefit with the HA.

Who cannot join

  • fluctuating hearing loss,
  • cutaneous disease
  • subjects who need an MRI.

Trial locations and contacts

1 site in 1 country.

France1 site
  • CHU Rennes

    Rennes, 35000

Other trials in France

About this study

Official title: Long Term Results With the Vibrant Soundbridge in Patients With Mixed Hearing Loss: a Prospective Multicenter Longitudinal Study Over 60-months

Vibrant Soundbridge (VSB, MED-EL, Innsbruck) Active middle ear implants (AMEI) can benefit patients who are unable to use conventional hearing aids (HA) due to medical reasons, discomfort or unsuccessful rehabilitation. Long-term prospective longitudinal studies are lacking on the VSB notably for conductive and mixed hearing loss. The main aim of the present study was to prospectively assess aided hearing benefits in a 60-month, long-term study including conductive and mixed hearing loss adults implanted with the VSB. The secondary objective was to compare the hearing results according to the Floating Mass Transducer (FMT) site (Round/oval window (RW/OW) vs incus/stapes) and the type of pathology responsible for the hearing loss (Inflammatory vs non-inflammatory disease, ID vs NID).

Read the detailed description

Introduction:

Active middle ear implants (AMEI) can benefit patients who are unable to use conventional hearing aids (HA) due to medical reasons, discomfort or unsuccessful rehabilitation.

AMEIs provide more hearing gain and consistency compared to middle ear surgery plus conventional hearing aids. Coletti et al. (2013) reported on long-term outcomes of patients implanted at the round window (RW) with the VSB. Brkic et al. (2019) presented a large number of subjects (n = 23) with a 20-year follow-up. The limitations of these studies were their retrospective design.

The best location for the surgeon to place the VSB's floating mass transducer (FMT) depends on various factors. A meta-analysis showed the hearing results at different FMT sites to be similar. None of these studies compared a FMT placed to a FMT clipped.

The impact of the etiology of deafness on a patient's outcome with an AMEI has been poorly studied. Only one study has shown that the difference between inflammatory disease (ID) and non-inflammatory disease (NID) groups was not significant.

The main objective:

Prospectively assess aided hearing benefits in a 60-month, long-term study including conductive and mixed hearing loss adults implanted with the VSB.

Measures are tonal audiometry, pure-tone average (PTA), vocal audiometry in quiet, 50% speech recognition threshold (SRT) in quiet, word recognition in quiet (WRS) at 65 dB sound pressure level (SPL), word recognition in noise (WRSN ; speech at 65 dB SPL and 0°, white noise at 55 dB SPL and 180°. signal-to-noise ratio = 10 dB).

The secondary objectives:

Comparison of the hearing results according to the FMT site (RW/OW to incus/stapes)

Comparison of the hearing results according the type of pathology responsible for the hearing loss (ID to NID).

Plan of the study:

It is a multicentric, prospective, longitudinal study with 6 tertiary references centres with a follow up over a period of 60 months.

Measures will be done on the patient preoperatively and 1, 3, 6, 12 and 60 months post-implantation.

Preoperative measures will be done also with hearing aid (HA) and bone-anchored hearing aid on a headband (HBAHA).

Being tested
Vibrant Soundbridge active middle ear implant
Keywords
Active middle ear implant, Vibrant Soundbridge
Medical subject headings
Hearing Loss, Mixed Conductive-Sensorineural

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Vibrant Soundbridge active middle ear implantDevice

    active middle ear implant

Study groups in NCT05793580
Study groupRoleWhat they receive
VSB armOtherlongitudinal follow-up of VSB users preoperatively and at 1, 3, 6, 12 and 60 months post-implantation.
Receives: Device: Vibrant Soundbridge active middle ear implant

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Tonal audiometry

    Pure-tone audiograms

    Measured: preoperatively

  2. Tonal audiometry

    Pure-tone audiograms

    Measured: at 1 month post-implantation

  3. Tonal audiometry

    Pure-tone audiograms

    Measured: at 3 months post-implantation

  4. Tonal audiometry

    Pure-tone audiograms

    Measured: at 6 months post-implantation

  5. Tonal audiometry

    Pure-tone audiograms

    Measured: at 12 months post-implantation

  6. Tonal audiometry

    Pure-tone audiograms

    Measured: at 60 months post-implantation

Secondary outcomes (18)
  1. Word recognition in quiet

    Speech recognition test of disyllabic words (Fournier lists).

    Measured: preoperatively

  2. Word recognition in quiet

    Speech recognition test of disyllabic words (Fournier lists).

    Measured: at 1 month post-implantation

  3. Word recognition in quiet

    Speech recognition test of disyllabic words (Fournier lists).

    Measured: at 3 months post-implantation

  4. Word recognition in quiet

    Speech recognition test of disyllabic words (Fournier lists).

    Measured: at 6 months post-implantation

  5. Word recognition in quiet

    Speech recognition test of disyllabic words (Fournier lists).

    Measured: at 12 months post-implantation

  6. Word recognition in quiet

    Speech recognition test of disyllabic words (Fournier lists).

    Measured: at 60 months post-implantation

  7. Word recognition in noise

    Speech recognition test of disyllabic words (Fournier lists) in noise. Noise = white noise at 55 dB SPL presented at 180°. Speech at 65 dB SPL presented at 0°.

    Measured: preoperatively

  8. Word recognition in noise

    Speech recognition test of disyllabic words (Fournier lists) in noise. Noise = white noise at 55 dB SPL presented at 180°. Speech at 65 dB SPL presented at 0°.

    Measured: at 1 month post-implantation

  9. Word recognition in noise

    Speech recognition test of disyllabic words (Fournier lists) in noise. Noise = white noise at 55 dB SPL presented at 180°. Speech at 65 dB SPL presented at 0°.

    Measured: at 3 months post-implantation

  10. Word recognition in noise

    Speech recognition test of disyllabic words (Fournier lists) in noise. Noise = white noise at 55 dB SPL presented at 180°. Speech at 65 dB SPL presented at 0°.

    Measured: at 6 months post-implantation

  11. Word recognition in noise

    Speech recognition test of disyllabic words (Fournier lists) in noise. Noise = white noise at 55 dB SPL presented at 180°. Speech at 65 dB SPL presented at 0°.

    Measured: at 12 months post-implantation

  12. Word recognition in noise

    Speech recognition test of disyllabic words (Fournier lists) in noise. Noise = white noise at 55 dB SPL presented at 180°. Speech at 65 dB SPL presented at 0°.

    Measured: at 60 months post-implantation

  13. Speech reception threshold

    Speech reception threshold to have 50% of intelligibility on disyllabic words.

    Measured: preoperatively

  14. Speech reception threshold

    Speech reception threshold to have 50% of intelligibility on disyllabic words.

    Measured: at 1 month post-implantation

  15. Speech reception threshold

    Speech reception threshold to have 50% of intelligibility on disyllabic words.

    Measured: at 3 months post-implantation

  16. Speech reception threshold

    Speech reception threshold to have 50% of intelligibility on disyllabic words.

    Measured: at 6 months post-implantation

  17. Speech reception threshold

    Speech reception threshold to have 50% of intelligibility on disyllabic words.

    Measured: at 12 months post-implantation

  18. Speech reception threshold

    Speech reception threshold to have 50% of intelligibility on disyllabic words.

    Measured: at 60 months post-implantation

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted31 Mar 2023
  2. Study start1 Sep 2008
  3. Primary completion1 Sep 2018
  4. Study completion1 Sep 2018

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Model description
Prospective longitudinal study
Masking
None (open label)
Enrolment
31 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT05793580
Sponsor's protocol ID
VSB_longterm_Rennes_study
Status last verified
Mar 2023
First submitted
20 Mar 2023
Last updated
3 Apr 2023
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Vibrant Soundbridge active middle ear implant in people with Mixed Hearing Loss, run by MED-EL Elektromedizinische Geräte GmbH at 1 site in France.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 3 Apr 2023. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Tonal audiometry (time frame: preoperatively). The study lists 6 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 3 Apr 2023.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in