Diabetes MellitusNCT06587087
A Proof of Concept (PoC), Randomized, Controlled Study to Validate the Algorithm and Evaluate the Accuracy, Safety and Pain Level of the Smartest-AE in Comparison to Self-measure Blood Glucose Levels
A study testing Smartest-AE: The lancet that will be used will be the Vitrex Soft Lite lll 28G. in people with Diabetes Mellitus and Gestational Diabetes Mellitus (GDM), run by Smartest Technologies at 1 site in Israel.
Trial summary
The study is conducted in Israel so plain language for the study is exist in Hebrew only Read more about this study
- Who is it for?
- Ages 18 to 80 yearsHealthy volunteers can take part
- Conditions the study is for
- Diabetes Mellitus, Gestational Diabetes Mellitus (GDM), Pre Diabetes
- What is being tested?
- Smartest-AE: The lancet that will be used will be the Vitrex Soft Lite lll 28G.
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 300 participants
- Where
- 1 siteIsrael
- Study dates
- 20 Sep 2024to 15 Feb 2025
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 80 years
- All sexes
- Healthy volunteers accepted
- Adult, Older adult
Who may be able to join
- Man or Woman, 18 through 80 years of age
- Diagnosed with either:
- Group 1: Type I/II Diabetes Melitus/ GDM who regularly prick his/her fingers to monitor blood glucose levels or,
- Group 2: High risk individuals who do not typically engage in finger pricking but are at high risk of developing prediabetes or diabetes due to one of the following various risk factors:
- BMI \> 25
- Age 45 or older
Show all 13Show fewer
- Have a parent or sibling with type 2 diabetes
- Have ever had gestational diabetes
- Hypercholesterolemia - according to blood level and\or using medication for lowering cholesterol
- Dyslipidemia- according to blood level and\or using medication for lowering lipids
- Or any other risk factor according to the staff's discretion
- Able to self-test using a glucose level device
- Subject is willing and able to comply with protocol requirements and has provided a signed informed consent form (ICF)
Who cannot join
- 1. Subject suffers from neuropathy, chronic pains, pain diseases, impaired sensation in the fingers\' area 2. Subject diagnosed with dementia 3. Subject diagnosed with major psychiatric disorders 4. Subject use Narcotics or Analgesics, chronically or at the time of enrollment to the study and up to 4 hours prior to first test of the study 5. Any other medical condition that is considered not eligible by the medical team members 6. Subject Is not capable to report pain levels, objectively 7. Participation in an interventional study at the time of enrollment to this study.
Trial locations and contacts
1 site in 1 country.
Israel1 site
The BARUCH PADEH Medical Center, Poriya
Poria – Neve Oved, The Lower Galilee, 15208
Enav Yefet, MD, 972-52-386-2160, enyefet@tzmc.gov.il
Keren Lubenfeld, Stduy coordinator, 972-50-5709562, kelubenfel@tzmc.gov.il
About this study
The study is conducted in Israel so plain language for the study is exist in Hebrew only
Read the detailed description
Protocol Title:
A Proof of Concept (PoC), Randomized, Controlled study to validate the algorithm and evaluate the accuracy, safety and pain level of the Smartest-AE in comparison to self-measure Blood Glucose Levels
Investigational Device:
The Smartest-AE
Device Description:
The Smartest's System is intended to measure blood glucose levels in home environment. The Smartest's System consists of a disposable single-use device and a mobile app.
At this study Smartest-AE ("Smartest-AE") will be tested.
Study Overview:
The subjects will be enrolled from the Diabetes Mellitus (DM) clinics in the hospitals, endocrinological clinic, Internal divisions, obstetrics and gynecology department, diabetes in pregnancy and maternal and fetal medicine clinics.
Study subjects will be consecutively screened and enrolled into the study. The study will include up to 300 subjects, men and women aged 18 to 80.
The trial will encompass two distinct populations:
Group 1- Diabetes Mellitus Patients At least 100 subjects who diagnosed with Type I/Type II DM or GDM (Gestational Diabetes Mellitus), who regularly prick their fingers to monitor blood glucose levels and willing to participate in the study, and
Group 2- High risk individuals. at least 100 individuals who do not typically engage in finger pricking but are at high risk of developing prediabetes or diabetes due to to one of the following various risk factors:
- BMI \> 25
- Age 45 or older
- Have a parent or sibling with type 2 diabetes
- Have ever had gestational diabetes
- Hypercholesterolemia - according to blood level and\or using medication for lowering cholesterol
- Dyslipidemia - according to blood level and\or using medication for lowering lipids Or any other risk factor according to the staff's discretion.
The study will be conducted in 2 stages:
Stage I- Algorithm verification- Blood glucose levels will be taken from the first 30 subjects in order to verify the algorithm. At the end of the enrollment of the first 30 cases an analysis will be done by Smartest's R\&D. If the algorithm yields correlation of R2\>0.9, based on the calibration curves, the study will proceed to the second phase. In case the calibration does not show the expected R2 the Smartest's R\&D will re-assess the algorithm. If correction/ modification to the algorithm will be implemented another 30 subjects will be enrolled to verify the updated algorithm.
If the analysis following algorithm update, yields satisfying results (R2\>0.9), the study will proceed to the second stage and the results of the 30 subjects used for verification will be included in the final analysis. If not, the study will be temporarily paused and the device algorithm will be re-assessed by Smartest's R\&D for accuracy and precision. Following that, another 30 subjects will be enrolled for algorithm verification. If the expected R2 is not obtained the study will be terminated.
Blood glucose levels will be measured for all 30/60/90 subjects using 2 devices:
- The Smartest-AE blood sample analyzer and
- A commercial standard reference approved glucometer including approved needles and strips.
The lancet that will be used for all subjects will be the Vitrex Soft Lite lll 28G.
All Subjects will complete a Numeric Rating Scale (NRS) questionnaire (Appendix A) to determine their level of pain from each device. NRS will be completed following each finger's pricking
Following NRS completion, blood will be taken from the subject's veins for blood glucose level that will be measured by the site's local lab.
NRS will not be completed following blood taken from the vein.
Stage II- Following algorithm verification and Smartest's approval to continue to stage II at least 200 subjects will be recruited.
Blood Glucose levels will be taken from each subject using the 2 devices; the Smartest-AE and the commercial glucometer.
The blood levels' results of the Smartest-AE will be compared to the commercial approved glucometer in order to assess accuracy. Subjects' comfort/pain level will be assessed following each finger's pricking using the Numeric Rating Scale (NRS) questionnaire
All Subjects will complete the NRS questionnaire to determine their level of pain from each device.
Following NRS completion, blood will be taken from the subject's veins for blood glucose level that will be measured by the site's local lab.
NRS will not be completed following blood taken from the vein. No follow-up period is expected.
Fingers III (middle) and IV (ring) of the non-dominant hand, will be pricked. Randomization will be used to allocate subjects that will be first tested with the Smartest-AE device and subjects that will be first tested with the commercial device. Finger III will always be the first to be pricked.
It was decided that subject IDs' even numbers will have the first prick with the Smartest-AE and subject IDs' odd numbers will have the first prick with the commercial device at each of the groups (1 and 2).
In case the device shows 'error 'or the results seems objectively, to be unreasonable (for eg: extremely high or low blood glucose level result that does not reflect subject physical condition at the time of test), another attempt will be done to achieve the blood glucose level. If the error will be obtained in the first test, then NRS will be completed by the subject subsequently to the test and the second attempt will be done only after the second (reference device) test will be performed. Up to 2 attempts are allowed for each system. Subject that has measurement results only from one system will be excluded from the analysis.
In case failings will be occurred in the venous blood test (either during the blood taking (problematic veins, etc.) or error received from the lab (hemolytic blood, etc.)), a second attempt will not be done and the result will be indicated as missing.
Study Population:
Adults volunteers, men and women, at least 25% of each gender at each group, at the ages of 18- 80 years old, diagnosed with Type I/Type II DM or GDM (Gestational diabetes Mellitus), who regularly prick their fingers to monitor blood glucose levels (Group 1, Diabetes Mellitus patients) and High risk individuals who do not typically engage in finger pricking but are at high risk of developing prediabetes or diabetes due to various risk factors (Group 2).
Hospital personnel are allowed to participate in this study as long they are not subjected directly to the principal investigator.
All subjects will confirm their willingness to participate in the study by signing an Informed Consent Form (ICF).
Study Objective:
To validate the algorithm and to evaluate the accuracy, safety and pain level of the Smartest-AE
Inclusion Criteria:
- Man or Woman, 18 through 80 years of age
- Diagnosed with either:
- Group 1: Type I/II Diabetes Melitus/ GDM who regularly prick his/her fingers to monitor blood glucose levels or,
- Group 2: High risk individuals who do not typically engage in finger pricking but are at high risk of developing prediabetes or diabetes due to one of the following various risk factors:
- BMI \> 25
- Age 45 or older
- Have a parent or sibling with type 2 diabetes
- Have ever had gestational diabetes
- Hypercholesterolemia - according to blood level and\or using medication for lowering cholesterol
- Dyslipidemia- according to blood level and\or using medication for lowering lipids
- Or any other risk factor according to the staff's discretion
- Able to self-test using a glucose level device
- Subject is willing and able to comply with protocol requirements and has provided a signed informed consent form (ICF)
Exclusion Criteria:
- Subject suffers from neuropathy, chronic pains, pain diseases, impaired sensation in the fingers\' area
- Subject diagnosed with dementia
- Subject diagnosed with major psychiatric disorders
- Subject use Narcotics or Analgesics, chronically or at the time of enrollment to the study and up to 4 hours prior to first test of the study
- Any other medical condition that is considered not eligible by the medical team members
- Subject Is not capable to report pain levels, objectively
- Participation in an interventional study at the time of enrollment to this study.
Study Duration :
One year depending on rate of subject recruitment. Duration for each participant: A single visit for couple of hours in the clinic or at the department during hospitalization.
Primary Endpoints:
Stage I- Smartest-AE system algorithm verification (R2\>0.9) Stage II- Assess the accuracy of the Smartest-AE glucose measurements' results in comparison to the results received from a commercial approved glucometer.
Secondary Efficacy Endpoints:
- Subjects' pain level of the Smartest-AE vs. the commercial lancet using the NRS questionnaire
- Assess the accuracy of the Smartest-AE glucose measurements' results and the commercial glucometer results in comparison to the blood glucose level results of the venous blood test received from the local site's lab
Safety Endpoint:
Any adverse events reported throughout the study following finger pricking will be recorded, including severity (Adverse events such as: infection, exceptional bleeding, blisters, etc.)
Sample size:
The sample size was calculated based on the expected correlation between the commercial glucometer and the Smartest-AE using the G*Power software (version 3.1.9.6), focusing on the Pearson correlation coefficient. The following parameters were used:
Test Type: Exact, Correlation: Bivariate normal model. Tail(s): Two-tailed test. Effect Size (ρ H1): A correlation coefficient of 0.3, reflecting a realistic expectation of the correlation between the variables.
α Error Probability: The alpha level was set at 0.05, maintaining the standard criterion for statistical significance, where there is a 5% chance of a Type I error.
Power (1-β Error Probability): The power was set at 80% (0.8), indicating that there is an 80% probability of correctly rejecting the null hypothesis when it is false.
Correlation ρ H0: The null hypothesis correlation was set to 0, assuming no correlation between the variables.
The calculation results yielded a total sample size of 84 samples.
Smartest company received a grant from the Israel Innovation Authority (IIA). As part of the grant the IIA set a milestone of conducting a clinical trial in 200 subjects for this study purposes. Therefore, the study population size is higher than the necessary sample size.
Statistical Analysis:
Descriptive Statistics: Data will be described by group using mean, standard deviations, median, minimum, and maximum for continuous variables, and counts and proportions for categorical variables.
Subgroup Analysis Stratification by Experience: Differences in reported pain levels between experienced (Group 1) and naïve (Group 2) subjects will be tested using a regression analysis.
Gender Analysis: Differences in reported pain levels will be tested across gender, ensuring at least 25% representation from each gender in both groups.
Software and Data Management Software: All statistical analyses will be conducted using R software. R packages such as "ggplot2" for data visualization and "dplyr" for data manipulation will be utilized to handle and analyze the study data.
Data Management: Data will be coded and stored securely with access restricted to authorized personnel only.
Tables and Figures Table 1: Demographic and Baseline Characteristics by Group. Table 2: Summary Statistics of Blood Glucose Measurements from Smartest-AE, Commercial Glucometer, and Blood Tests.
Table 3: Subgroup Analysis of Pain Levels. Figure 1: correlation plot of Blood Glucose Measurements from Smartest-AE, Commercial Glucometer, and Blood Tests.
Figure 2: Raincloud plots of Pain Levels Reported on NRS by gender.
Handling of Missing Data Missing Data Policy: Subjects missing measurements from either the Smartest-AE or the commercial glucometer will be excluded from the analysis.
Data Imputation: No imputation will be performed for missing glucose measurements; only complete case analysis will be used.
- Being tested
- Smartest-AE: The lancet that will be used will be the Vitrex Soft Lite lll 28G.
- Keywords
- Diabetes, Gestational Diabetes Mellitus (GDM), Pre-Diabetes
- Medical subject headings
- Diabetes Mellitus, Diabetes, Gestational, Glucose Intolerance
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Smartest-AE: The lancet that will be used will be the Vitrex Soft Lite lll 28G.Device
The lancet that will be used for all subjects will be the Vitrex Soft Lite lll 28G.
| Study group | Role | What they receive |
|---|---|---|
| Investigational Glucose level device- Smartest-AE | Experimental | The Smartest's System is intended to measure blood glucose levels in home environment. The Smartest's System consists of a disposable single-use device and a mobile app. Receives: Device: Smartest-AE: The lancet that will be used will be the Vitrex Soft Lite lll 28G. |
| Commercial Glucometer | Active comparator | Not described Receives: Device: Smartest-AE: The lancet that will be used will be the Vitrex Soft Lite lll 28G. |
What is being measured
The primary outcome is the main question the study is designed to answer.
Algorithm verification
Stage I- Smartest-AE system algorithm verification (R2\>0.9)
Measured: 1 months
Stage II- Accuracy
Stage II- Assess the accuracy of the Smartest-AE glucose measurements' results in comparison to the results received from a commercial approved glucometer.
Measured: 3 months
Secondary outcomes (2)
Pain Level
Subjects' pain level of the Smartest-AE vs. the commercial lancet using the NRS questionnaire
Measured: 4 months
accuracy vs. lab results
Assess the accuracy of the Smartest-AE glucose measurements' results and the commercial glucometer results in comparison to the blood glucose level results of the venous blood test received from the local site's lab
Measured: 4 months
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted19 Sep 2024
- Study start20 Sep 2024 (estimated)
- Primary completion31 Jan 2025 (estimated)
- Study completion15 Feb 2025 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Diabetes MellitusStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Screening
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- None (open label)
- Enrolment
- 300 participants (estimated)
Sponsor and investigators
- Lead sponsor
- Smartest Technologies
- Sponsor type
- Industry (pharma, biotech, devices)
- Collaborators
- Israel Innovation Authority
- Responsible party
- Sponsor
- Investigators
- Enav Yefet, MD, Principal investigator, Baruch Padeh Medical center Poriya
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT06587087
- Sponsor's protocol ID
- SMT000018
- Status last verified
- Sep 2024
- First submitted
- 4 Sep 2024
- Last updated
- 19 Sep 2024
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- Undecided
Study at a glance
From the registry record, last updated by the sponsor on 19 Sep 2024. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- A Proof of Concept (PoC), Randomized, Controlled Study to Validate the Algorithm and Evaluate the Accuracy, Safety and Pain Level of the Smartest-AE in Comparison to Self-measure Blood Glucose Levels
- ClinicalTrials.gov ID
- NCT06587087
- Acronym / sponsor ID
- SMT000018
- Keywords
- Diabetes, Gestational Diabetes Mellitus (GDM), Pre-Diabetes
- Study type
- Interventional
- Recruitment status
- Unknown status
- Enrolment
- 300 participants (estimated)
- Lead sponsor
- Smartest TechnologiesIndustry (pharma, biotech, devices)
- Collaborators
- Israel Innovation Authority
- Study start
- 20 Sep 2024 (estimated)
- Primary completion
- 31 Jan 2025 (estimated)
- Study completion
- 15 Feb 2025 (estimated)
- First posted
- 19 Sep 2024 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The study is conducted in Israel so plain language for the study is exist in Hebrew only
Detailed description
Protocol Title:
A Proof of Concept (PoC), Randomized, Controlled study to validate the algorithm and evaluate the accuracy, safety and pain level of the Smartest-AE in comparison to self-measure Blood Glucose Levels
Investigational Device:
The Smartest-AE
Device Description:
The Smartest's System is intended to measure blood glucose levels in home environment. The Smartest's System consists of a disposable single-use device and a mobile app.
At this study Smartest-AE ("Smartest-AE") will be tested.
Study Overview:
The subjects will be enrolled from the Diabetes Mellitus (DM) clinics in the hospitals, endocrinological clinic, Internal divisions, obstetrics and gynecology department, diabetes in pregnancy and maternal and fetal medicine clinics.
Study subjects will be consecutively screened and enrolled into the study. The study will include up to 300 subjects, men and women aged 18 to 80.
The trial will encompass two distinct populations:
Group 1- Diabetes Mellitus Patients At least 100 subjects who diagnosed with Type I/Type II DM or GDM (Gestational Diabetes Mellitus), who regularly prick their fingers to monitor blood glucose levels and willing to participate in the study, and
Group 2- High risk individuals. at least 100 individuals who do not typically engage in finger pricking but are at high risk of developing prediabetes or diabetes due to to one of the following various risk factors:
- BMI \> 25
- Age 45 or older
- Have a parent or sibling with type 2 diabetes
- Have ever had gestational diabetes
- Hypercholesterolemia - according to blood level and\or using medication for lowering cholesterol
- Dyslipidemia - according to blood level and\or using medication for lowering lipids Or any other risk factor according to the staff's discretion.
The study will be conducted in 2 stages:
Stage I- Algorithm verification- Blood glucose levels will be taken from the first 30 subjects in order to verify the algorithm. At the end of the enrollment of the first 30 cases an analysis will be done by Smartest's R\&D. If the algorithm yields correlation of R2\>0.9, based on the calibration curves, the study will proceed to the second phase. In case the calibration does not show the expected R2 the Smartest's R\&D will re-assess the algorithm. If correction/ modification to the algorithm will be implemented another 30 subjects will be enrolled to verify the updated algorithm.
If the analysis following algorithm update, yields satisfying results (R2\>0.9), the study will proceed to the second stage and the results of the 30 subjects used for verification will be included in the final analysis. If not, the study will be temporarily paused and the device algorithm will be re-assessed by Smartest's R\&D for accuracy and precision. Following that, another 30 subjects will be enrolled for algorithm verification. If the expected R2 is not obtained the study will be terminated.
Blood glucose levels will be measured for all 30/60/90 subjects using 2 devices:
- The Smartest-AE blood sample analyzer and
- A commercial standard reference approved glucometer including approved needles and strips.
The lancet that will be used for all subjects will be the Vitrex Soft Lite lll 28G.
All Subjects will complete a Numeric Rating Scale (NRS) questionnaire (Appendix A) to determine their level of pain from each device. NRS will be completed following each finger's pricking
Following NRS completion, blood will be taken from the subject's veins for blood glucose level that will be measured by the site's local lab.
NRS will not be completed following blood taken from the vein.
Stage II- Following algorithm verification and Smartest's approval to continue to stage II at least 200 subjects will be recruited.
Blood Glucose levels will be taken from each subject using the 2 devices; the Smartest-AE and the commercial glucometer.
The blood levels' results of the Smartest-AE will be compared to the commercial approved glucometer in order to assess accuracy. Subjects' comfort/pain level will be assessed following each finger's pricking using the Numeric Rating Scale (NRS) questionnaire
All Subjects will complete the NRS questionnaire to determine their level of pain from each device.
Following NRS completion, blood will be taken from the subject's veins for blood glucose level that will be measured by the site's local lab.
NRS will not be completed following blood taken from the vein. No follow-up period is expected.
Fingers III (middle) and IV (ring) of the non-dominant hand, will be pricked. Randomization will be used to allocate subjects that will be first tested with the Smartest-AE device and subjects that will be first tested with the commercial device. Finger III will always be the first to be pricked.
It was decided that subject IDs' even numbers will have the first prick with the Smartest-AE and subject IDs' odd numbers will have the first prick with the commercial device at each of the groups (1 and 2).
In case the device shows 'error 'or the results seems objectively, to be unreasonable (for eg: extremely high or low blood glucose level result that does not reflect subject physical condition at the time of test), another attempt will be done to achieve the blood glucose level. If the error will be obtained in the first test, then NRS will be completed by the subject subsequently to the test and the second attempt will be done only after the second (reference device) test will be performed. Up to 2 attempts are allowed for each system. Subject that has measurement results only from one system will be excluded from the analysis.
In case failings will be occurred in the venous blood test (either during the blood taking (problematic veins, etc.) or error received from the lab (hemolytic blood, etc.)), a second attempt will not be done and the result will be indicated as missing.
Study Population:
Adults volunteers, men and women, at least 25% of each gender at each group, at the ages of 18- 80 years old, diagnosed with Type I/Type II DM or GDM (Gestational diabetes Mellitus), who regularly prick their fingers to monitor blood glucose levels (Group 1, Diabetes Mellitus patients) and High risk individuals who do not typically engage in finger pricking but are at high risk of developing prediabetes or diabetes due to various risk factors (Group 2).
Hospital personnel are allowed to participate in this study as long they are not subjected directly to the principal investigator.
All subjects will confirm their willingness to participate in the study by signing an Informed Consent Form (ICF).
Study Objective:
To validate the algorithm and to evaluate the accuracy, safety and pain level of the Smartest-AE
Inclusion Criteria:
- Man or Woman, 18 through 80 years of age
- Diagnosed with either:
- Group 1: Type I/II Diabetes Melitus/ GDM who regularly prick his/her fingers to monitor blood glucose levels or,
- Group 2: High risk individuals who do not typically engage in finger pricking but are at high risk of developing prediabetes or diabetes due to one of the following various risk factors:
- BMI \> 25
- Age 45 or older
- Have a parent or sibling with type 2 diabetes
- Have ever had gestational diabetes
- Hypercholesterolemia - according to blood level and\or using medication for lowering cholesterol
- Dyslipidemia- according to blood level and\or using medication for lowering lipids
- Or any other risk factor according to the staff's discretion
- Able to self-test using a glucose level device
- Subject is willing and able to comply with protocol requirements and has provided a signed informed consent form (ICF)
Exclusion Criteria:
- Subject suffers from neuropathy, chronic pains, pain diseases, impaired sensation in the fingers\' area
- Subject diagnosed with dementia
- Subject diagnosed with major psychiatric disorders
- Subject use Narcotics or Analgesics, chronically or at the time of enrollment to the study and up to 4 hours prior to first test of the study
- Any other medical condition that is considered not eligible by the medical team members
- Subject Is not capable to report pain levels, objectively
- Participation in an interventional study at the time of enrollment to this study.
Study Duration :
One year depending on rate of subject recruitment. Duration for each participant: A single visit for couple of hours in the clinic or at the department during hospitalization.
Primary Endpoints:
Stage I- Smartest-AE system algorithm verification (R2\>0.9) Stage II- Assess the accuracy of the Smartest-AE glucose measurements' results in comparison to the results received from a commercial approved glucometer.
Secondary Efficacy Endpoints:
- Subjects' pain level of the Smartest-AE vs. the commercial lancet using the NRS questionnaire
- Assess the accuracy of the Smartest-AE glucose measurements' results and the commercial glucometer results in comparison to the blood glucose level results of the venous blood test received from the local site's lab
Safety Endpoint:
Any adverse events reported throughout the study following finger pricking will be recorded, including severity (Adverse events such as: infection, exceptional bleeding, blisters, etc.)
Sample size:
The sample size was calculated based on the expected correlation between the commercial glucometer and the Smartest-AE using the G*Power software (version 3.1.9.6), focusing on the Pearson correlation coefficient. The following parameters were used:
Test Type: Exact, Correlation: Bivariate normal model. Tail(s): Two-tailed test. Effect Size (ρ H1): A correlation coefficient of 0.3, reflecting a realistic expectation of the correlation between the variables.
α Error Probability: The alpha level was set at 0.05, maintaining the standard criterion for statistical significance, where there is a 5% chance of a Type I error.
Power (1-β Error Probability): The power was set at 80% (0.8), indicating that there is an 80% probability of correctly rejecting the null hypothesis when it is false.
Correlation ρ H0: The null hypothesis correlation was set to 0, assuming no correlation between the variables.
The calculation results yielded a total sample size of 84 samples.
Smartest company received a grant from the Israel Innovation Authority (IIA). As part of the grant the IIA set a milestone of conducting a clinical trial in 200 subjects for this study purposes. Therefore, the study population size is higher than the necessary sample size.
Statistical Analysis:
Descriptive Statistics: Data will be described by group using mean, standard deviations, median, minimum, and maximum for continuous variables, and counts and proportions for categorical variables.
Subgroup Analysis Stratification by Experience: Differences in reported pain levels between experienced (Group 1) and naïve (Group 2) subjects will be tested using a regression analysis.
Gender Analysis: Differences in reported pain levels will be tested across gender, ensuring at least 25% representation from each gender in both groups.
Software and Data Management Software: All statistical analyses will be conducted using R software. R packages such as "ggplot2" for data visualization and "dplyr" for data manipulation will be utilized to handle and analyze the study data.
Data Management: Data will be coded and stored securely with access restricted to authorized personnel only.
Tables and Figures Table 1: Demographic and Baseline Characteristics by Group. Table 2: Summary Statistics of Blood Glucose Measurements from Smartest-AE, Commercial Glucometer, and Blood Tests.
Table 3: Subgroup Analysis of Pain Levels. Figure 1: correlation plot of Blood Glucose Measurements from Smartest-AE, Commercial Glucometer, and Blood Tests.
Figure 2: Raincloud plots of Pain Levels Reported on NRS by gender.
Handling of Missing Data Missing Data Policy: Subjects missing measurements from either the Smartest-AE or the commercial glucometer will be excluded from the analysis.
Data Imputation: No imputation will be performed for missing glucose measurements; only complete case analysis will be used.
Study design
- Primary purpose
- Screening
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- None (open label)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Investigational Glucose level device- Smartest-AE | Experimental | Device: Smartest-AE: The lancet that will be used will be the Vitrex Soft Lite lll 28G.The Smartest's System is intended to measure blood glucose levels in home environment. The Smartest's System consists of a disposable single-use device and a mobile app. |
| Commercial Glucometer | Active comparator | Device: Smartest-AE: The lancet that will be used will be the Vitrex Soft Lite lll 28G. |
Who may qualify
- Ages
- 18 to 80 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (13 listed)
- Man or Woman, 18 through 80 years of age
- Diagnosed with either:
- Able to self-test using a glucose level device
- Subject is willing and able to comply with protocol requirements and has provided a signed informed consent form (ICF)
Key exclusion criteria (1 listed)
- 1. Subject suffers from neuropathy, chronic pains, pain diseases, impaired sensation in the fingers\' area 2. Subject diagnosed with dementia 3. Subject diagnosed with major psychiatric disorders 4. Subject use Narcotics or Analgesics, chronically or at the time of enrollment to the study and up to 4 hours prior to first test of the study 5. Any other medical condition that is considered not eligible by the medical team members 6. Subject Is not capable to report pain levels, objectively 7. Participation in an interventional study at the time of enrollment to this study.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Smartest-AE: The lancet that will be used will be the Vitrex Soft Lite lll 28G.The lancet that will be used for all subjects will be the Vitrex Soft Lite lll 28G. | Device | Commercial Glucometer, Investigational Glucose level device- Smartest-AE |
Primary outcome
Algorithm verification
Time frame: 1 months
Stage I- Smartest-AE system algorithm verification (R2\>0.9)
Stage II- Accuracy
Time frame: 3 months
Stage II- Assess the accuracy of the Smartest-AE glucose measurements' results in comparison to the results received from a commercial approved glucometer.
2 secondary outcomes
Pain Level
Time frame: 4 months
accuracy vs. lab results
Time frame: 4 months
Results and safety
Results are not available in the source record.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 19 Sep 2024.
- Individual participant dataSharing planned: Undecided
Related trials for Diabetes Mellitus
Recruiting
Prevalence of Diabetes-related Distress Among Patients Living With Type 2 Diabetes in a University Hospital Center and Identification of Its Associated Factors.
Diabetes Mellitus, Psychological Distress
Not yet recruiting
The Effect of Individualized Nursing Interventions on Glycemic Variability and Self-Management in Type 2 Diabetes
Type 2 Diabetes Mellitus (T2DM)
Not yet recruitingObservational study
Evaluation of the Clinical Spectrum of Diabetes and Obesity in Youth and Adults
Obesity, Diabetes
Active, not recruitingObservational study
Non-Invasive MASLD Diagnosis & T2D Remission Biomarkers
MASLD, T2DM (Type 2 Diabetes Mellitus)
Recruiting
Effects of Medisinstart for Type 2 Diabetes
Type 2 Diabetes
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Enav Yefet, MD principal investigator
Baruch Padeh Medical center Poriya
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Smartest-AE: The lancet that will be used will be the Vitrex Soft Lite lll 28G. in people with Diabetes Mellitus and Gestational Diabetes Mellitus (GDM), run by Smartest Technologies at 1 site in Israel.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 13 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: status not confirmed, as of the sponsor's last update on 19 Sep 2024. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Israel. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Results are not available in the source record.
How can a doctor refer a patient or contact the research team?
Contact Ariel Horwitz, CEO (972-54-8884330, ariel@smarttest.io), or a site near the patient. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Algorithm verification (time frame: 1 months). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 19 Sep 2024.