Blood PressureNCT06609486

Clinical Evaluation of the Performance of Bionet Sphygmomanometer Modules Based on Protocol Presented by the International Organization for Standardization

A study testing Blood pressure measurement group in people with Blood Pressure, run by Hoon-Ki Park at 1 site in South Korea.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

According to INTERNATIONAL STANDARD (hereafter ISO) blood pressure (BP) monitor clinical evaluation protocol (ISO 81060-2:2018, Non-invasive sphygmomanometers- Part2: Clinical investigation of intermittent automated measurement type) for BN1 (BNiBP), an electronic sphygmomanometer module newly developed by Bionet Co.… Read more about this study

Who is it for?
Ages 3 years and olderHealthy volunteers can take part
Conditions the study is for
Blood Pressure
What is being tested?
Blood pressure measurement group
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
86 participants
Where
1 siteSouth Korea
Study dates
1 Jun 2020to 23 Jul 2021
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 3 years and older
  • All sexes
  • Healthy volunteers accepted
  • Child, Adult, Older adult

Who may be able to join

  • Recruitment is made through the recruitment notice for patients and their guardians who visit Hanyang University Hospital. Subjects who meet the criteria presented in ISO 81060-2:2018 Protocol 5.1 Subject requirements are recruited.

Who cannot join

  • Emergency surgery patient
  • Persons who have mental disorders such as mental retardation, autism, etc. and who have difficulty measuring blood pressure or are expected to have difficulties
  • A person who disagreed with the study

Trial locations and contacts

1 site in 1 country.

South Korea1 site
  • Hanyang University Seoul Hospital

    Seoul, 04763

Other trials in South Korea

About this study

Official title: Clinical Evaluation of Bionet Sphygmomanometer Module BN1 Based on ISO 81060-2:2018 Protocol

According to INTERNATIONAL STANDARD (hereafter ISO) blood pressure (BP) monitor clinical evaluation protocol (ISO 81060-2:2018, Non-invasive sphygmomanometers- Part2: Clinical investigation of intermittent automated measurement type) for BN1 (BNiBP), an electronic sphygmomanometer module newly developed by Bionet Co., Ltd., the clinical group and evaluation criteria presented in, the effectiveness of blood pressure measurement performance was evaluated.

The auscultation method used as a reference (gold standard) of a non-invasive blood pressure (NIBP) measurement method was used. The first blood pressure measurement was performed once each (reference BP not used for evaluation, R0; test BP not used for evaluation, T0) with auscultation and the test medical device (BM5 equipped with BN1 (BNiBP)). The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each blood pressure measurement.

Three data pairs (mean value of R1, R2 \& T1 / R2, mean value of R3 \& T2 / R3, mean value of R4 \& T3) were obtained from one study subject. This study aims to verify that the mean value of the differences and standard deviations of data pairs meet the criteria of ISO 81060-2:2018.

Conditions
Blood Pressure
Being tested
Blood pressure measurement group
Keywords
Hypotension, Hypertension, Prehypertension, Sphygmomanometers
Medical subject headings
Hypotension, Hypertension, Prehypertension

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Blood pressure measurement groupDevice

    In this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device.

Study groups in NCT06609486
Study groupRoleWhat they receive
Blood pressure measurement groupExperimentalIn this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device.
Receives: Device: Blood pressure measurement group

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Test Device's Blood Pressure of Resting Condition

    The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each BP measurement. Three data pairs (mean value of R1, R2 \& T1 / R2, mean value of R3 \& T2 / R3, mean value of R4 \& T3) were obtained from one study subject.

    Criterion 1 is that for valid data pairs, the mean value of the difference between the test medical device and the auscultation method BP (test device BP - auscultation BP) should be 5 mmHg or less and the standard deviation (SD) should be 8 mmHg or less.

    Criterion 2 is that, for study subjects, the ISO 81060-2:2018 for presenting different SD thresholds depending on the mean value of the difference should be satisfied. When mean value of the difference is ±1.6, ±0.0 mmHg, the SD should be less than 6.76 and 6.95 mmHg, respectively.

    Measured: Right after the experiment

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted24 Sep 2024
  2. Study start1 Jun 2020
  3. Primary completion21 Jul 2021
  4. Study completion23 Jul 2021

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Model description
In this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device.
Masking
None (open label)
Enrolment
86 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06609486
Sponsor's protocol ID
HKPark
Status last verified
Feb 2025
First submitted
20 Sep 2024
Last updated
3 Mar 2025
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • Alpert BS. Validation of the Welch Allyn Spot Vital Signs blood pressure device according to the ANSI/AAMI SP10: 2002. Accuracy and cost-efficiency successfully combined. Blood Press Monit. 2007 Oct;12(5):345-7. doi: 10.1097/MBP.0b013e3282c9abf7.

    PubMed 17890975
  • Coleman A, Freeman P, Steel S, Shennan A. Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Press Monit. 2006 Feb;11(1):27-32. doi: 10.1097/01.mbp.0000189788.05736.5f.

    PubMed 16410738
  • Guidelines for Accuracy Evaluation of Electronic Sphygmomanometers - KFDA Food and Drug Administration

  • O'Brien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R, Imai Y, Wang J, Mengden T, Shennan A; Working Group on Blood Pressure Monitoring of the European Society of Hypertension. European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit. 2010 Feb;15(1):23-38. doi: 10.1097/MBP.0b013e3283360e98. No abstract available.

    PubMed 20110786
  • Stergiou GS, Alpert B, Mieke S, Asmar R, Atkins N, Eckert S, Frick G, Friedman B, Grassl T, Ichikawa T, Ioannidis JP, Lacy P, McManus R, Murray A, Myers M, Palatini P, Parati G, Quinn D, Sarkis J, Shennan A, Usuda T, Wang J, Wu CO, O'Brien E. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. J Hypertens. 2018 Mar;36(3):472-478. doi: 10.1097/HJH.0000000000001634.

    PubMed 29384983
  • International Organization for Standardization. Noninvasive sphygmomanometers: part 2: clinical investigation of intermittent automated measurement type. (2018): 36.

Questions about this trial

What is this clinical trial studying?

A study testing Blood pressure measurement group in people with Blood Pressure, run by Hoon-Ki Park at 1 site in South Korea.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 3 Mar 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in South Korea. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 3 Mar 2025). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Test Device's Blood Pressure of Resting Condition (time frame: Right after the experiment). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT06609486, which also lists 6 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 3 Mar 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in