Clinical Evaluation of the Performance of Bionet Sphygmomanometer Modules Based on Protocol Presented by the International Organization for Standardization
A study testing Blood pressure measurement group in people with Blood Pressure, run by Hoon-Ki Park at 1 site in South Korea.
Trial summary
According to INTERNATIONAL STANDARD (hereafter ISO) blood pressure (BP) monitor clinical evaluation protocol (ISO 81060-2:2018, Non-invasive sphygmomanometers- Part2: Clinical investigation of intermittent automated measurement type) for BN1 (BNiBP), an electronic sphygmomanometer module newly developed by Bionet Co.… Read more about this study
- Who is it for?
- Ages 3 years and olderHealthy volunteers can take part
- Conditions the study is for
- Blood Pressure
- What is being tested?
- Blood pressure measurement group
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 86 participants
- Where
- 1 siteSouth Korea
- Study dates
- 1 Jun 2020to 23 Jul 2021
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 3 years and older
- All sexes
- Healthy volunteers accepted
- Child, Adult, Older adult
Who may be able to join
- Recruitment is made through the recruitment notice for patients and their guardians who visit Hanyang University Hospital. Subjects who meet the criteria presented in ISO 81060-2:2018 Protocol 5.1 Subject requirements are recruited.
Who cannot join
- Emergency surgery patient
- Persons who have mental disorders such as mental retardation, autism, etc. and who have difficulty measuring blood pressure or are expected to have difficulties
- A person who disagreed with the study
Trial locations and contacts
1 site in 1 country.
About this study
Official title: Clinical Evaluation of Bionet Sphygmomanometer Module BN1 Based on ISO 81060-2:2018 Protocol
According to INTERNATIONAL STANDARD (hereafter ISO) blood pressure (BP) monitor clinical evaluation protocol (ISO 81060-2:2018, Non-invasive sphygmomanometers- Part2: Clinical investigation of intermittent automated measurement type) for BN1 (BNiBP), an electronic sphygmomanometer module newly developed by Bionet Co., Ltd., the clinical group and evaluation criteria presented in, the effectiveness of blood pressure measurement performance was evaluated.
The auscultation method used as a reference (gold standard) of a non-invasive blood pressure (NIBP) measurement method was used. The first blood pressure measurement was performed once each (reference BP not used for evaluation, R0; test BP not used for evaluation, T0) with auscultation and the test medical device (BM5 equipped with BN1 (BNiBP)). The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each blood pressure measurement.
Three data pairs (mean value of R1, R2 \& T1 / R2, mean value of R3 \& T2 / R3, mean value of R4 \& T3) were obtained from one study subject. This study aims to verify that the mean value of the differences and standard deviations of data pairs meet the criteria of ISO 81060-2:2018.
- Conditions
- Blood Pressure
- Being tested
- Blood pressure measurement group
- Keywords
- Hypotension, Hypertension, Prehypertension, Sphygmomanometers
- Medical subject headings
- Hypotension, Hypertension, Prehypertension
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Blood pressure measurement groupDevice
In this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device.
| Study group | Role | What they receive |
|---|---|---|
| Blood pressure measurement group | Experimental | In this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device. Receives: Device: Blood pressure measurement group |
What is being measured
The primary outcome is the main question the study is designed to answer.
Test Device's Blood Pressure of Resting Condition
The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each BP measurement. Three data pairs (mean value of R1, R2 \& T1 / R2, mean value of R3 \& T2 / R3, mean value of R4 \& T3) were obtained from one study subject.
Criterion 1 is that for valid data pairs, the mean value of the difference between the test medical device and the auscultation method BP (test device BP - auscultation BP) should be 5 mmHg or less and the standard deviation (SD) should be 8 mmHg or less.
Criterion 2 is that, for study subjects, the ISO 81060-2:2018 for presenting different SD thresholds depending on the mean value of the difference should be satisfied. When mean value of the difference is ±1.6, ±0.0 mmHg, the SD should be less than 6.76 and 6.95 mmHg, respectively.
Measured: Right after the experiment
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted24 Sep 2024
- Study start1 Jun 2020
- Primary completion21 Jul 2021
- Study completion23 Jul 2021
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Publications (6)Papers the sponsor has linked to this study.
- Other trials for Blood PressureStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Model description
- In this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device.
- Masking
- None (open label)
- Enrolment
- 86 participants (actual)
Sponsor and investigators
- Lead sponsor
- Hoon-Ki Park
- Sponsor type
- University, hospital or other
- Collaborators
- Bionet Co., Ltd
- Responsible party
- Sponsor-investigator, Hoon-Ki Park, Hanyang University Seoul Hospital
- Investigators
- Hoon-Ki Park, M.D., Ph.D., Principal investigator, Department of Family Medicine, Hanyang University Seoul Hospital
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT06609486
- Sponsor's protocol ID
- HKPark
- Status last verified
- Feb 2025
- First submitted
- 20 Sep 2024
- Last updated
- 3 Mar 2025
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- No
Publications
Alpert BS. Validation of the Welch Allyn Spot Vital Signs blood pressure device according to the ANSI/AAMI SP10: 2002. Accuracy and cost-efficiency successfully combined. Blood Press Monit. 2007 Oct;12(5):345-7. doi: 10.1097/MBP.0b013e3282c9abf7.
PubMed 17890975Coleman A, Freeman P, Steel S, Shennan A. Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Press Monit. 2006 Feb;11(1):27-32. doi: 10.1097/01.mbp.0000189788.05736.5f.
PubMed 16410738Guidelines for Accuracy Evaluation of Electronic Sphygmomanometers - KFDA Food and Drug Administration
O'Brien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R, Imai Y, Wang J, Mengden T, Shennan A; Working Group on Blood Pressure Monitoring of the European Society of Hypertension. European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit. 2010 Feb;15(1):23-38. doi: 10.1097/MBP.0b013e3283360e98. No abstract available.
PubMed 20110786Stergiou GS, Alpert B, Mieke S, Asmar R, Atkins N, Eckert S, Frick G, Friedman B, Grassl T, Ichikawa T, Ioannidis JP, Lacy P, McManus R, Murray A, Myers M, Palatini P, Parati G, Quinn D, Sarkis J, Shennan A, Usuda T, Wang J, Wu CO, O'Brien E. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. J Hypertens. 2018 Mar;36(3):472-478. doi: 10.1097/HJH.0000000000001634.
PubMed 29384983International Organization for Standardization. Noninvasive sphygmomanometers: part 2: clinical investigation of intermittent automated measurement type. (2018): 36.
Study at a glance
From the registry record, last updated by the sponsor on 3 Mar 2025. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Clinical Evaluation of Bionet Sphygmomanometer Module BN1 Based on ISO 81060-2:2018 Protocol
- ClinicalTrials.gov ID
- NCT06609486
- Acronym / sponsor ID
- HKPark
- Conditions
- Blood Pressure
- Keywords
- Hypotension, Hypertension, Prehypertension, Sphygmomanometers
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 86 participants (actual)
- Lead sponsor
- Hoon-Ki ParkUniversity, hospital or other
- Collaborators
- Bionet Co., Ltd
- Study start
- 1 Jun 2020 (actual)
- Primary completion
- 21 Jul 2021 (actual)
- Study completion
- 23 Jul 2021 (actual)
- First posted
- 24 Sep 2024 (actual)
- Results
- Posted 3 Mar 2025
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
According to INTERNATIONAL STANDARD (hereafter ISO) blood pressure (BP) monitor clinical evaluation protocol (ISO 81060-2:2018, Non-invasive sphygmomanometers- Part2: Clinical investigation of intermittent automated measurement type) for BN1 (BNiBP), an electronic sphygmomanometer module newly developed by Bionet Co., Ltd., the clinical group and evaluation criteria presented in, the effectiveness of blood pressure measurement performance was evaluated.
The auscultation method used as a reference (gold standard) of a non-invasive blood pressure (NIBP) measurement method was used. The first blood pressure measurement was performed once each (reference BP not used for evaluation, R0; test BP not used for evaluation, T0) with auscultation and the test medical device (BM5 equipped with BN1 (BNiBP)). The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each blood pressure measurement.
Three data pairs (mean value of R1, R2 \& T1 / R2, mean value of R3 \& T2 / R3, mean value of R4 \& T3) were obtained from one study subject. This study aims to verify that the mean value of the differences and standard deviations of data pairs meet the criteria of ISO 81060-2:2018.
Study design
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
- Model description
- In this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device.
| Arm | Type | Receives |
|---|---|---|
| Blood pressure measurement group | Experimental | Device: Blood pressure measurement groupIn this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device. |
Who may qualify
- Ages
- 3 years and older
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Child, Adult, Older adult
Key inclusion criteria (1 listed)
- Recruitment is made through the recruitment notice for patients and their guardians who visit Hanyang University Hospital. Subjects who meet the criteria presented in ISO 81060-2:2018 Protocol 5.1 Subject requirements are recruited.
Key exclusion criteria (3 listed)
- Emergency surgery patient
- Persons who have mental disorders such as mental retardation, autism, etc. and who have difficulty measuring blood pressure or are expected to have difficulties
- A person who disagreed with the study
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Blood pressure measurement groupIn this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device. | Device | Blood pressure measurement group |
Primary outcome
Test Device's Blood Pressure of Resting Condition
Time frame: Right after the experiment
The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each BP measurement. Three data pairs (mean value of R1, R2 \& T1 / R2, mean value of R3 \& T2 / R3, mean value of R4 \& T3) were obtained from one study subject. Criterion 1 is that for valid data pairs, the mean value of the difference between the test medical device and the auscultation method BP (test device BP - auscultation BP) should be 5 mmHg or less and the standard deviation (SD) should be 8 mmHg or less. Criterion 2 is that, for study subjects, the ISO 81060-2:2018 for presenting different SD thresholds depending on the mean value of the difference should be satisfied. When mean value of the difference is ±1.6, ±0.0 mmHg, the SD should be less than 6.76 and 6.95 mmHg, respectively.
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 3 Mar 2025.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Background: PubMed 17890975Alpert BS. Validation of the Welch Allyn Spot Vital Signs blood pressure device according to the ANSI/AAMI SP10: 2002. Accuracy and cost-efficiency successfully combined. Blood Press Monit. 2007 Oct;12(5):345-7. doi: 10.1097/MBP.0b013e3282c9abf7.
- Background: PubMed 16410738Coleman A, Freeman P, Steel S, Shennan A. Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol. Blood Press Monit. 2006 Feb;11(1):27-32. doi: 10.1097/01.mbp.0000189788.05736.5f.
- BackgroundGuidelines for Accuracy Evaluation of Electronic Sphygmomanometers - KFDA Food and Drug Administration
- Background: PubMed 20110786O'Brien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R, Imai Y, Wang J, Mengden T, Shennan A; Working Group on Blood Pressure Monitoring of the European Society of Hypertension. European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit. 2010 Feb;15(1):23-38. doi: 10.1097/MBP.0b013e3283360e98. No abstract available.
- Background: PubMed 29384983Stergiou GS, Alpert B, Mieke S, Asmar R, Atkins N, Eckert S, Frick G, Friedman B, Grassl T, Ichikawa T, Ioannidis JP, Lacy P, McManus R, Murray A, Myers M, Palatini P, Parati G, Quinn D, Sarkis J, Shennan A, Usuda T, Wang J, Wu CO, O'Brien E. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. J Hypertens. 2018 Mar;36(3):472-478. doi: 10.1097/HJH.0000000000001634.
- BackgroundInternational Organization for Standardization. Noninvasive sphygmomanometers: part 2: clinical investigation of intermittent automated measurement type. (2018): 36.
- Individual participant dataSharing planned: No
Related trials for Blood Pressure
Active, not recruiting
Digital Reinforcement of Motivational Interviewing Micro-Skills Among Primary Care Healthcare Professionals
Hypertension (Without Type 2 Diabetes Mellitus)
RecruitingPhase 3
A Phase III Renal Outcomes and Cardiovascular Mortality Study to Investigate the Efficacy and Safety of Baxdrostat in Combination With Dapagliflozin in Participants With Chronic Kidney Disease and High Blood Pressure
Chronic Kidney Disease and Hypertension
Recruiting
Guiding Optimal Blood Pressure Management in Peritoneal Dialysis Pilot Trial
End Stage Kidney Disease (ESKD), Hypertension (HTN)
RecruitingPhase 4
Hematological Safety of Dipyrone Versus Paracetamol After Dental Extractions
Hematologic Test, Agranulocytosis
Not yet recruiting
Acute Effects of Mat Pilates on Blood Pressure Response to Mental Stress in Higher Education Teachers
Behavior
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Hoon-Ki Park, M.D., Ph.D. principal investigator
Department of Family Medicine, Hanyang University Seoul Hospital
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Blood pressure measurement group in people with Blood Pressure, run by Hoon-Ki Park at 1 site in South Korea.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 1 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 3 Mar 2025. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in South Korea. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 3 Mar 2025). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Test Device's Blood Pressure of Resting Condition (time frame: Right after the experiment). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT06609486, which also lists 6 linked publications.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 3 Mar 2025.