CataractNCT06642779

Comparative Analysis of the Visual Performance and Patient Satisfaction After Cataract Surgery With Implantation of a Medicontur Monofocal (877PAY) or a Medicontur Extended Depth of Focus (877PEY) IOL

An observational study of Cataract and Presbyopia, run by Medicontur Medical Engineering Ltd at 3 sites in Hungary.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this multi-center, prospective, randomized, comparative clinical investigation was to evaluate and compare the visual performance at far, intermediate and near distances, the presence of photopic phenomena and patient satisfaction after cataract surgery in patients implanted binocularly with a… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Cataract, Presbyopia, Ametropia
What is being tested?
VFQ- 25 (Visual Function Questionnaire), Optical Biometry and Slit lamps / Fundoscopy / Biomicroscopy
How many people
64 participants
Where
3 sitesHungary
Study dates
2 Dec 2020to 24 Jan 2024
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Adult, Older adult

Study population: Adults (18 years old and older) with cataract and/or ametropia (hyperopia, myopia), who hace signed the informed consent form and undergone binocular implantation with the investigational device (877PEY) or the comparator device (877PAY).

Who may be able to join

  • cataractous patients who wish to be partly spectacle independent
  • cataractous lens changes as demonstrated by best corrected visual acuity of 20/40 or worse either with or without a glare source present (e.g., Brightness Acuity Tester) or with significant cataract-related visual symptoms*
  • best corrected visual acuity to be better than 20/25 (0.1 logMAR) after cataract removal and IOL implantation*
  • ≤ 1.0 D of preoperative keratometric astigmatism;
  • clear intraocular media other than cataract;
  • given written informed consent by subject;
  • subjects are willing and able to comply with schedule for follow-up visits;
  • adult patient

Who cannot join

  • corneal astigmatism > 1.00 D
  • irregular astigmatism
  • diabetic retinopathy
  • iris neovascularisation
  • serious intraoperative complications
  • congenital eye abnormality
Show all 25Show fewer
  • uncontrolled glaucoma or glaucoma with changes in optical nerve and visual field
  • pseudoexfoliation syndrome
  • amblyopia
  • uveitis
  • long-term anti-inflammatory treatment
  • AMD
  • retinal detachment
  • prior ocular surgery in personal medical history
  • corneal diseases
  • severe retinal diseases (dystrophy, degeneration)
  • severe myopia (if required IOL power is lower than 10 D)
  • inadequate visualization of the fundus on preoperative examination
  • patients deemed by the clinical investigator because of any systemic disease.
  • eye trauma in medical history
  • Intraoperative exclusion criteria:
  • tear in capsulorhexis*
  • zonular dehiscence*
  • posterior capsular rupture vitreous loss and other unexpected surgical complication*
  • According to coordinator investigator's decision, the marked inclusion and exclusion criteria were not applied.

Trial locations and contacts

3 sites in 1 country.

Hungary3 sites
  • Győr-Moson-Sopron Vármegyei Petz Aladár Egyetemi Oktató Kórház

    Győr, Győr-Moson-Sopron Vármegye, 9024

  • Borsod-Abaúj-Zemplén Vármegyei Központi Kórház és Egyetemi Oktatókórház

    Miskolc, BAZ Vármegye, 3526

  • Retinaszerviz Kft.

    Veszprém, Veszprém Vármegye, 8200

Other trials in Hungary

About this study

The purpose of this multi-center, prospective, randomized, comparative clinical investigation was to evaluate and compare the visual performance at far, intermediate and near distances, the presence of photopic phenomena and patient satisfaction after cataract surgery in patients implanted binocularly with a Medicontur monofocal (877PAY) or a Medicontur extended depth of focus (877PEY) IOL models.

The visit schedule included a total of 6 visits, in addition to the IOL implantation surgery, with the listed follow-up timeline:

  • screening and baseline assessments
  • day 0 (IOL implantation surgery)
  • day 1
  • month 1
  • month 3
  • month 6
  • month 12
Being tested
VFQ- 25 (Visual Function Questionnaire), Optical Biometry, Slit lamps / Fundoscopy / Biomicroscopy, UDVA, CDVA
Keywords
Intraocular lens, monofocal IOL, EDOF
Medical subject headings
Cataract, Presbyopia, Refractive Errors

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • VFQ- 25 (Visual Function Questionnaire)Behavioural

    VFQ-25 questionnaire was used to evaluate patient satisfaction at the M3 and M12 visits.

  • Optical BiometryDiagnostic test

    Axial length, anterior chamber depth and corneal keratometric values K1-K2. All these parameters were measured at the screening visit and at the month 3 follow-up visit, using the ARGOS® Biometer or the ANTERION® Cataract App

  • Slit lamps / Fundoscopy / BiomicroscopyDiagnostic test

    The fundus and other ocular structures, such as the cornea, anterior chamber, iris, pupil, crystalline lens / IOL alignment, corpus ciliare and choroid were examined at screening visit, on the day of the surgery, and at the follow up visits D1, M1, M3, M6, and M12, using the slit lamp binocular indirect ophthalmoscopy.

  • UDVADiagnostic test

    Monocular and binocular Uncorrected Distance Visual Acuity (UDVA) was evaluated at all visits (preoperative, D1, M1, M3, M6 and M12), using the ETDRS chart.

  • CDVADiagnostic test

    Monocular and binocular Corrected Distance Visual Acuity (UDVA) was evaluated at all visits (preoperative, D1, M1, M3, M6 and M12), using the ETDRS chart.

  • UIVA / DCIVADiagnostic test

    Uncorrected and Distance Corrected Intermediate Visual acuity (UIVA / DCIVA) was evaluated postoperatively at M1, M3, M6 and M12, using an intermediate chart from 67 cm.

  • UNVA / DCNVADiagnostic test

    Uncorrected and Distance Corrected Near Visual acuity (UNVA / DCNVA) was evaluated postoperatively at M1, M3, M6 and M12, using a near chart from 40 cm.

  • Manifest RefractionDiagnostic test

    Manifest refraction was the monocular subjective refraction (Sph, Cyl, Axis) giving the best distance corrected visual acuity. It was measured at the screening visit and the follow-up visits Day 1, Month 1, Month 3, Month 6 and Month 12.

  • IOPDiagnostic test

    Intraocular pressure (IOP) was measured at all visits (preoperative, D1, M1, M3, M6 and M12), using a non-contact tonometer.

  • VADCDiagnostic test

    Visual Acuity Defocus Curve (VADC) was taken monocularly at M3, monocularly and binocularly at M6 and binocularly at M12, using the Multifocal Lens analyzer - MLA application.

  • CSDCDiagnostic test

    Contrast Sinsitivity Defocus Curve (CSDC) was taken under photopic conditions at M6 and M12, using the Multifocal Lens Analyzer - MLA application

  • CSV-1000Diagnostic test

    Contrast Sensitivity was measured using the CSV-1000 with distance correction in place at spatial frequencies of 3, 6, 12, and 18 cycles per degree (cpd). Measurements were taken under four different light conditions:

    • Photopic (M3 monocular and binocular, M6 binocular and M12 monocular)
    • Mesopic (M6 binocular)
    • Photopic with backlight (M6 binocular)
    • Mesopic with backlight (M6 binocular)
  • Tear film stabilityDiagnostic test

    Tear film stability was assessed at screening visit and at follow-up visits M1, M3, M6, and M12 using the non-invasive Tear break Up Time (TBUT) with fluorescein dye.

  • IOL implantationDevice

    The investigational device (877PEY) and the comparator (877PAY) intraocular lens models were implanted, using sutureless, small-incision phacoemulsification surgery.

Study groups in NCT06642779
Study groupRoleWhat they receive
877PAY (monofocal)26 patients (52 eyes) binocularly implanted with the 877PAY monofocal IOL model
Receives: Behavioral: VFQ- 25 (Visual Function Questionnaire); Diagnostic Test: Optical Biometry; Diagnostic Test: Slit lamps / Fundoscopy / Biomicroscopy; Diagnostic Test: UDVA; Diagnostic Test: CDVA; Diagnostic Test: UIVA / DCIVA; Diagnostic Test: UNVA / DCNVA; Diagnostic Test: Manifest Refraction; Diagnostic Test: IOP; Diagnostic Test: VADC; Diagnostic Test: CSDC; Diagnostic Test: CSV-1000; Diagnostic Test: Tear film stability; Device: IOL implantation
877PEY (extended depth of focus)38 patients (76 eyes) binocularly implanted with the 877PEY extended depth of focus IOL model
Receives: Behavioral: VFQ- 25 (Visual Function Questionnaire); Diagnostic Test: Optical Biometry; Diagnostic Test: Slit lamps / Fundoscopy / Biomicroscopy; Diagnostic Test: UDVA; Diagnostic Test: CDVA; Diagnostic Test: UIVA / DCIVA; Diagnostic Test: UNVA / DCNVA; Diagnostic Test: Manifest Refraction; Diagnostic Test: IOP; Diagnostic Test: VADC; Diagnostic Test: CSDC; Diagnostic Test: CSV-1000; Diagnostic Test: Tear film stability; Device: IOL implantation

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. DCIVA

    distance corrected intermediate visual acuity

    Measured: measured at month 1, month 3, month 6 and month 12

Secondary outcomes (5)
  1. UDVA/CDVA

    uncorrected and corrected distance visual acuity

    Measured: measured at month 1, month 3, month 6 and month 12

  2. VADC

    visual acuity defocus curves

    Measured: measured at month 3, month 6 and month 12

  3. CSDC

    contrast sensitivity defocus curves

    Measured: measured at month 6 and month 12

  4. Patient satisfaction

    presence of visual disturbances (glares, halos), evaluated with the VFQ-25 questionnaire.

    Measured: Evaluated at month 3 and month 12

  5. CSV-1000

    Contrast sensitivity measured with the CSV-1000 standardized auto-calibrating vision testing instrument

    Measured: measured at month 3, month 6 and month 12

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted15 Oct 2024
  2. Study start2 Dec 2020
  3. Primary completion21 Nov 2023
  4. Study completion24 Jan 2024

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Case-control
Time perspective
Prospective
Enrolment
64 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06642779
Sponsor's protocol ID
M_877PAY_877PEY_HU_2020
Status last verified
Oct 2024
First submitted
10 Oct 2024
Last updated
15 Oct 2024
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

An observational study of Cataract and Presbyopia, run by Medicontur Medical Engineering Ltd at 3 sites in Hungary.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 8 inclusion and 25 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 15 Oct 2024. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 3 sites in Hungary. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

DCIVA (time frame: measured at month 1, month 3, month 6 and month 12). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 15 Oct 2024.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in