Hearing Outcomes With the Osia 3 Sound Processor Compared to the Osia 2 Sound Processor in Adult Osia Implant Recipients

A study testing Cochlear™ Osia® System (Osia® 2) and Cochlear™ Osia® System (Osia® 3) in people with Hearing Loss, Mixed and Hearing Loss, Conductive, run by Cochlear at 3 sites in 2 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The goal of this interventional study is to evaluate the clinical performance and sound quality of the Osia 3 Sound Processor and to demonstrate its superiority compared with the Osia 2 Sound Processor in adults with mixed hearing loss, conductive hearing loss, or single-sided deafness. Read more about this study

Who is it for?
Ages 18 years and older
What is being tested?
Cochlear™ Osia® System (Osia® 2) and Cochlear™ Osia® System (Osia® 3)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
29 participants
Where
3 sitesin 2 countries
Study dates
9 Sep 2025to 24 Nov 2025
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Implanted with a Cochlear Osia Implant (OSI100, OSI200, OSI300)
  • Conductive or mixed hearing loss in the implanted ear. Bone conduction thresholds up to 65 dB HL (up to 40 dB HL at 0.5 kHz, up to 55 dB HL at 1kHz and up to 65 dB HL at 2 and 4 kHz) or Single-Sided Deafness in the implanted ear. The pure tone average air conduction hearing thresholds of the hearing ear should be better than or equal to 20 dB HL (measured at 0.5, 1, 2 and, 3 kHz).
  • Aged 18 years or older, at time of consent.
  • Minimum experience of 1 month with the Osia 2 Sound Processor.
  • Fluent speaker in the language used to assess speech perception performance.
  • Willing and able to provide written informed consent.

Who cannot join

  • Ongoing infection at or around the sound processor area.
  • Bilaterally implanted with a Cochlear Osia Implant.
  • Unable or unwilling to comply with the requirements of the clinical investigation as determined by the Investigator.
  • Use of ototoxic drugs that could be harmful to the hearing, as judged by the investigator.
  • Investigator site personnel directly affiliated with this study and/or their immediate families; immediate family is defined as a spouse, parent, child, or sibling.
  • Cochlear employees or employees of Contract Research Organisations or contractors engaged by Cochlear for the purposes of this investigation.
  • Current participation, or participation in another interventional clinical study/trial in the past 30 days, involving an investigational drug or device (unless the other investigation was/is a Cochlear sponsored investigation and determined by the investigator or Sponsor to not impact this investigation).

Trial locations and contacts

3 sites in 2 countries.

3 sites in 2 countries

Australia2 sites
  • HEARnet Clinical Studies

    Melbourne, Victoria, 3010

  • Cochlear Macquarie, Macquarie University

    Sydney, New South Wales, 2109

Other trials in Australia

United States1 site
  • Cochlear Americas, Denver Research and Technology Lab

    Denver, Colorado, 80124

Other trials in United States

About this study

Official title: A Pivotal, Interventional, Prospective, Multi-centre Clinical Investigation of Hearing Outcomes With the Osia 3 Sound Processor Compared to the Osia 2 Sound Processor in Adult Osia Implant Recipients With Mixed Hearing Loss, Conductive Hearing Loss or Single-Sided Deafness

The goal of this interventional study is to evaluate the clinical performance and sound quality of the Osia 3 Sound Processor and to demonstrate its superiority compared with the Osia 2 Sound Processor in adults with mixed hearing loss, conductive hearing loss, or single-sided deafness.

The main questions this study aims to answer are:

  • Do the findings confirm the clinical performance of the Osia 3 Sound Processor?
  • Is the Osia 3 Sound Processor superior compared to the Osia 2 Sound Processor?

Participants will:

  • Undergo speech performance testing in both quiet and noisy environments
  • Provide ratings for a self-reported questionnaire
Being tested
Cochlear™ Osia® System (Osia® 2), Cochlear™ Osia® System (Osia® 3)
Keywords
Bone Conduction
Medical subject headings
Hearing Loss, Mixed Conductive-Sensorineural, Hearing Loss, Conductive

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Cochlear™ Osia® System (Osia® 2)Device

    Consists of Osia 2 Sound Processor, Osia 2 Sound Processor Magnets, and Osia Fitting Software 2.

  • Cochlear™ Osia® System (Osia® 3)Device

    Consists of Osia 3 Sound Processor, Osia 3 Sound Processor Magnets, Osia 3 Charger, and Osia Fitting Software 3.

Study groups in NCT07156461
Study groupRoleWhat they receive
Cochlear™ Osia® 2 Sound ProcessorActive comparatorParticipants randomized to group 1 will first receive a study Osia 2 Sound Processor to use at home for 1-2 weeks, before returning to the clinic to complete various assessments. These participants will then switch to the Osia 3 Sound Processor and complete the same protocol.
Receives: Device: Cochlear™ Osia® System (Osia® 2); Device: Cochlear™ Osia® System (Osia® 3)
Cochlear™ Osia® 3 Sound ProcessorExperimentalParticipants randomized to group 2 will first receive a Osia 3 Sound Processor to use at home for 1-2 weeks, before returning to the clinic to complete various assessments. These participants will then switch to the Osia 2 Sound Processor and complete the same protocol.
Receives: Device: Cochlear™ Osia® System (Osia® 2); Device: Cochlear™ Osia® System (Osia® 3)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mean difference in speech reception threshold in noise with the Osia 3 Sound Processor, 1-2 weeks after fitting compared to the unaided condition at baseline

    Speech perception in noise will be assessed using the Australian Speech Test in Noise (AuSTIN) at the Australian sites and the Bamford-Kowal-Bench Speech-In-Noise (BKB-SIN) test at the US site. The AuSTIN uses BKB-SIN like target sentences presented in noise and comprises of 80 lists of 20 sentences each. The BKB-SIN test contains 18 list pairs of sentences, each containing 3-4 key words, presented at 65 dB SPL with adaptive noise levels. Both tests aim to determine the signal-to-noise ratio (SNR) at which 50% of the key words are correctly repeated. For the unaided and aided measurements, speech performance in noise will be measured with speech and noise presented from a front speaker, positioned at 1 m from the subject.

    Measured: At Baseline (Day 0), Visit 1 (7-14 days after Baseline), and Visit 2 (7-14 days after Visit 1)

Secondary outcomes (9)
  1. Mean difference in the word recognition score in quiet with the Osia 3 Sound Processor measured at 65 dB SPL, 1-2 weeks after fitting compared to the unaided condition at baseline

    Speech perception in quiet will be measured using the consonant-nucleus-consonant (CNC) monosyllabic words presented at 65 dB SPL from the front speaker. The CNC word test consists of 30 lists each with 50 words per list. The goal of the speech perception assessment in quiet is to obtain the percentage of correctly repeated words in a list, i.e., the word recognition score.

    Measured: At Baseline (Day 0), Visit 1 (7-14 days after Baseline), and Visit 2 (7-14 days after Visit 1)

  2. Mean difference in the four-frequency sound field threshold average (PTA4, mean of sound field thresholds at 0.5, 1, 2, and 4 kHz) with the Osia 3 Sound Processor, 1-2 weeks after fitting compared to the unaided condition at baseline

    Thresholds in sound field will be measured using warble tones presented from a front speaker positioned at 1 m from the subject following the modified Hughson-Westlake procedure. Frequencies to be tested are 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, and 8000 Hz.

    Measured: At Baseline (Day 0), Visit 1 (7-14 days after Baseline), and Visit 2 (7-14 days after Visit 1)

  3. Mean difference in speech perception in noise with the Osia 3 Sound Processor compared with the Osia 2 Sound Processor measured at a fixed signal-to-noise ratio (SNR) of 0 dB SNR, 1-2 weeks after fitting

    Speech perception in noise with a fixed SNR will be measured using the AzBio sentence test, which consists of 15 lists of 20 sentences each. AzBio sentences are spoken by different talkers in a conversational style with limited contextual cues that the listener can use to predict or 'fill in' unintelligible words. The sentences will be presented at a fixed level of 65 dB SPL at a fixed SNR of 0 dB SNR using the accompanying noise stimulus. Each word in the sentence counts towards the overall score (percent correct). Speech will be presented from the front, while noise will be presented from the rear.

    Measured: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)

  4. Mean difference in speech perception in noise with the Osia 3 Sound Processor compared with the Osia 2 Sound Processor measured at a fixed signal-to-noise ratio (SNR) of +5 dB SNR, 1-2 weeks after fitting

    Speech perception in noise with a fixed SNR will be measured using the AzBio sentence test, which consists of 15 lists of 20 sentences each. AzBio sentences are spoken by different talkers in a conversational style with limited contextual cues that the listener can use to predict or 'fill in' unintelligible words. The sentences will be presented at a fixed level of 65 dB SPL at a fixed SNR of +5 dB SNR using the accompanying noise stimulus. Each word in the sentence counts towards the overall score (percent correct). Speech will be presented from the front, while noise will be presented from the rear.

    Measured: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)

  5. Mean difference in the Speech Spatial and Qualities of Hearing Scale (SSQ-12) global score with the Osia 3 Sound Processor compared with the Osia 2 Sound Processor, 1-2 weeks after fitting

    The SSQ-12 involves a self-reported hearing function and the impact of hearing treatment in daily life across a variety of listening situations. It includes 3 subcategories: Speech Hearing, Spatial Hearing, and Qualities of Hearing, with each question rated on a Likert scale from 0-10 (minimum - maximum hearing abilities).

    Measured: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)

  6. Mean difference in the word recognition score in quiet with the Osia 3 Sound Processor compared with the Osia 2 Sound Processor measured at 50 dB SPL, 1-2 weeks after fitting

    Speech perception in quiet will be measured using the consonant-nucleus-consonant (CNC) monosyllabic words presented at 50 dB SPL from the front speaker. The CNC word test consists of 30 lists each with 50 words per list. The goal of the speech perception assessment in quiet is to obtain the percentage of correctly repeated words in a list, i.e., the word recognition score.

    Measured: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)

  7. Mean difference in the word recognition score in quiet with the Osia 3 Sound Processor compared with the Osia 2 Sound Processor measured at 65 dB SPL, 1-2 weeks after fitting

    Speech perception in quiet will be measured using the consonant-nucleus-consonant (CNC) monosyllabic words presented at 65 dB SPL from the front speaker. The CNC word test consists of 30 lists each with 50 words per list. The goal of the speech perception assessment in quiet is to obtain the percentage of correctly repeated words in a list, i.e., the word recognition score.

    Measured: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)

  8. Mean difference in speech reception threshold in noise with the Osia 3 Sound Processor compared with the Osia 2 Sound Processor, 1-2 weeks after fitting

    Speech perception in noise will be assessed using the Australian Speech Test in Noise (AuSTIN) at the Australian sites and the Bamford-Kowal-Bench-Speech in Noise (BKB-SIN) test at the US site. The AuSTIN uses BKB-SIN like target sentences presented in noise and comprises of 80 lists of 20 sentences each. The BKB-SIN test contains 18 list pairs of sentences, each containing 3-4 key words, presented at 65 dB SPL with adaptive noise levels. Both tests aim to determine the signal-to-noise ratio (SNR) at which 50% of the key words are correctly repeated. For the unaided and aided measurements, speech performance in noise will be measured with speech and noise presented from a front speaker, positioned at 1 m from the subject.

    Measured: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)

  9. Mean difference in the four-frequency sound field threshold average (PTA4, mean of sound field thresholds at 0.5, 1, 2, and 4 kHz) with the Osia 3 Sound Processor compared to the Osia 2 Sound Processor, 1-2 weeks after fitting

    Thresholds in sound field will be measured using warble tones presented from a front speaker positioned at 1 m from the subject following the modified Hughson-Westlake procedure. Frequencies to be tested are 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000 and 8000 Hz.

    Measured: At Visit 1 (7-14 days after Baseline) and Visit 2 (7-14 days after Visit 1)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted5 Sep 2025
  2. Study start9 Sep 2025
  3. Primary completion24 Nov 2025
  4. Study completion24 Nov 2025

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Crossover
Model description
The investigation uses a repeated-measures, cross-over design where study recipients will be randomized into one of two groups (1:1 ratio)
Masking
None (open label)
Enrolment
29 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07156461
Sponsor's protocol ID
CLTD5861
Acronym
SONUS
Status last verified
Sep 2025
First submitted
26 Aug 2025
Last updated
8 Jan 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Cochlear™ Osia® System (Osia® 2) and Cochlear™ Osia® System (Osia® 3) in people with Hearing Loss, Mixed and Hearing Loss, Conductive, run by Cochlear at 3 sites in 2 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 8 Jan 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 3 sites in Australia, United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mean difference in speech reception threshold in noise with the Osia 3 Sound Processor, 1-2 weeks after fitting compared to the unaided condition at baseline (time frame: At Baseline (Day 0), Visit 1 (7-14 days after Baseline), and Visit 2 (7-14 days after Visit 1)). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 8 Jan 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in