Perioperative CareNCT07260890

Paired Comparison of SVV and PVI Accuracy

An observational study of Perioperative Care and Fluid Therapy, run by Wonkwang University Hospital at 1 site in South Korea.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This study will compare two operating-room monitors-stroke volume variation (SVV, from the arterial line) and the Pleth Variability Index (PVI, from the pulse oximeter)-to see which one more accurately predicts whether giving a small fluid bolus will improve the heart's pumping during laparoscopic major abdominal… Read more about this study

Who is it for?
Ages 19 years and olderHealthy volunteers can take part
What is being tested?
Stroke Volume Variation (SVV) monitoring and Pleth Variability Index (PVI) monitoring
Type of treatment
Medical device
How many people
300 participants
Where
1 siteSouth Korea
Study dates
1 Jul 2026to 11 Aug 2026
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 19 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Study population: Adults undergoing elective laparoscopic major abdominal surgery at a single academic hospital.

Who may be able to join

  • Age ≥18 years
  • ASA physical status I-III
  • Elective laparoscopic major abdominal surgery under general anesthesia
  • Arterial line in place for clinical care
  • Able to provide informed consent
  • Arterial and plethysmographic waveforms adequate for measurement

Who cannot join

  • linically significant arrhythmia (e.g., atrial fibrillation with irregular R-R)
  • Severe valvular heart disease or intracardiac shunt
  • Pregnancy
  • Severe right-heart failure or condition where a fluid bolus is unsafe
  • Ongoing hemodynamic instability requiring rapid vasoactive titration at assessment
  • Persistently unreliable arterial/pleth signals despite optimization
  • Any situation the anesthesiologist judges that the fluid challenge is unsafe

Trial locations and contacts

1 site in 1 country.

South Korea1 site
  • Wonkwang University School of Medicine Hospital

    Iksan, Jeollabuk-do, 54538

Other trials in South Korea

About this study

Official title: Diagnostic Accuracy of Stroke Volume Variation (SVV) vs Pleth Variability Index (PVI) for Predicting Fluid Responsiveness in Laparoscopic Major Abdominal Surgery-A Prospective, Paired, Cohort Study

This study will compare two operating-room monitors-stroke volume variation (SVV, from the arterial line) and the Pleth Variability Index (PVI, from the pulse oximeter)-to see which one more accurately predicts whether giving a small fluid bolus will improve the heart's pumping during laparoscopic major abdominal surgery. Adults having elective surgery under general anesthesia will receive two small, timed 250-mL crystalloid infusions as part of routine care (one before and one after creation of the pneumoperitoneum). The research team will record SVV and PVI values just before and three minutes after each infusion while keeping these readings hidden from the clinicians so that usual care is not changed. No experimental drugs or devices are used. The main goal is to learn which index better identifies "fluid responsiveness," so future care can be safer and more consistent. Potential risks are minimal and relate to the small fluid boluses (temporary changes in blood pressure or heart rate); the test stops if the anesthesiologist has any safety concerns. There is no direct benefit to participants, but results may help guide fluid therapy for similar patients in the future. The study is being conducted at a single academic hospital in the Republic of Korea and plans to enroll about 300 adults.

Read the detailed description

Background and rationale. Goal-directed fluid therapy during laparoscopic major abdominal surgery remains variable because dynamic preload indices perform inconsistently under pneumoperitoneum and positional changes. Stroke Volume Variation (SVV, derived from the arterial pressure waveform) and the Pleth Variability Index (PVI, derived from the pulse oximeter) are both widely available and non-drug, non-investigational monitors. A head-to-head, paired accuracy study using standardized fluid challenges can clarify which index better identifies fluid responsiveness in this setting.

Objectives. Primary: Compare the diagnostic accuracy (area under the ROC curve, AUROC) of SVV vs PVI for predicting fluid responsiveness to a small crystalloid bolus during laparoscopic major abdominal surgery.

Key secondary objectives: (1) determine optimal decision thresholds and report sensitivity, specificity, predictive values, and likelihood ratios; (2) compare calibration and classification performance (Youden index, net reclassification); (3) assess effect modification by pneumoperitoneum status, body mass index, ventilation parameters, and vasoactive use; (4) describe peri-bolus hemodynamic changes.

Design and setting. Prospective, single-center, paired diagnostic-accuracy cohort. Adults undergoing elective laparoscopic major abdominal surgery under general anesthesia are enrolled. Each participant undergoes up to two standardized 250-mL crystalloid fluid challenges delivered over ~3 minutes at two time points when feasible: (A) before insufflation and (B) after establishment of pneumoperitoneum at a hemodynamic steady state. Because both indices are recorded around the same fluid challenges, each participant serves as his/her own control.

Index tests. SVV (from the arterial pressure monitor) and PVI (from the pulse-oximetry monitor) are captured immediately before ("pre-bolus") and at ~3 minutes after each bolus. Index values are recorded by trained research staff and kept masked from anesthesia clinicians to avoid influencing usual care.

Reference standard and definition of fluid responsiveness. The reference standard is change in stroke volume index (SVI) or cardiac index derived from arterial pulse contour analysis. Fluid responsiveness is defined a priori as a ≥10% increase in SVI after the 250-mL bolus. Sensitivity analyses will use a 15% threshold and will repeat analyses with cardiac index. If signal quality is inadequate or a clinical safety concern arises, the test is aborted and data flagged.

Anesthesia and peri-bolus standardization. Ventilation is volume-controlled with tidal volume ~6-8 mL/kg predicted body weight and PEEP according to institutional practice; settings and pneumoperitoneum pressure/position are recorded. Vasoactive drugs and additional fluids are allowed as clinically indicated but are time-stamped; fluid challenges for research are not given when active titration of vasoactives makes a stable baseline impossible.

Participants. Inclusion: adults (≥18 years), ASA I-III, elective laparoscopic major abdominal surgery, arterial line in place, informed consent. Exclusion: clinically significant arrhythmia (e.g., atrial fibrillation with irregular R-R), severe valvular disease, intracardiac shunts, pregnancy, severe right-heart failure, inability to obtain reliable arterial or plethysmographic waveforms, or any condition in which a fluid challenge is judged unsafe by the anesthesiologist.

Outcomes. Primary outcome: AUROC of SVV vs PVI for predicting fluid responsiveness. Secondary outcomes: sensitivity/specificity at prespecified thresholds (e.g., SVV 12-13%, PVI 13-15%), decision-curve analysis, subgroup AUROC by pneumoperitoneum status, and immediate hemodynamic responses (MAP, HR).

Sample size and power. The study plans to enroll ~300 adults, yielding up to ~600 evaluable fluid challenges. Assuming a fluid-responsive rate of ~40%, this sample provides >80% power to detect a paired AUROC difference of ~0.08-0.10 between SVV and PVI at α=0.05 using DeLong's test with clustering by subject.

Statistical analysis. ROC curves for SVV and PVI are constructed using clustered bootstrap resampling at the patient level (≥2000 replicates). AUROCs are compared with paired DeLong tests. Threshold performance metrics include exact 95% CIs with patient-clustered bootstrap. Mixed-effects logistic regression (random intercept for patient) will estimate adjusted odds of fluid responsiveness per unit change of each index; prespecified covariates include pneumoperitoneum status, BMI, tidal volume per kg, PEEP, and vasoactive use. Missing index or reference values are minimized; if >5% are missing, multiple imputation will be performed in sensitivity analyses. All analyses follow a prespecified statistical analysis plan.

Safety and risk/benefit. Risks relate to small fluid boluses (transient blood pressure or heart-rate changes, fluid accumulation in vulnerable patients). The bolus is withheld or stopped for any safety concern. No experimental drugs or investigational devices are used; all monitors are standard of care. Participants are unlikely to benefit directly, but the results may improve future intraoperative fluid management.

Confidentiality and data handling. Data are captured on a secure electronic case-report form with coded identifiers. Only the research team has access. Results will be presented in aggregate without personal identifiers and disseminated in peer-reviewed venues.

Regulatory status. This is not an FDA-regulated drug or device investigation and is conducted under institutional review board approval at a single academic hospital in the Republic of Korea.

Being tested
Stroke Volume Variation (SVV) monitoring, Pleth Variability Index (PVI) monitoring
Keywords
stroke volume variation; SVV, Pleth Variability Index; PVI, fluid responsiveness, goal-directed fluid therapy; GDFT, pneumoperitoneum, diagnostic accuracy; ROC; AUROC
Medical subject headings
Pneumoperitoneum

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Stroke Volume Variation (SVV) monitoringDevice

    Also called SVV, Arterial pressure waveform analysis, FloTrac, EV1000

    Non-investigational physiologic monitoring. SVV derived from the invasive arterial pressure waveform using a commercially available platform. Used as an index test; values recorded immediately before and ~3 minutes after a standardized 250-mL crystalloid bolus at up to two time points (pre- and post-pneumoperitoneum). Readings are masked to clinicians; operation per usual care.

  • Pleth Variability Index (PVI) monitoringDevice

    Also called PVI, Pleth variability Index, Pulse oximetry-derived variability, Masimo

    Non-investigational pulse-oximetry-derived index. PVI obtained from a commercial monitor. Used as a paired index test; values recorded immediately before and ~3 minutes after each 250-mL crystalloid bolus (up to two time points). Readings masked; monitor used within standard care.

Study groups in NCT07260890
Study groupRoleWhat they receive
Laparoscopic Surgery CohortAdults undergoing elective laparoscopic major abdominal surgery under general anesthesia at a single academic hospital. Each participant may receive up to two standardized 250-mL crystalloid fluid challenges (pre- and post-pneumoperitoneum). SVV and PVI are recorded around each bolus; no assignment or experimental treatment.
Receives: Device: Stroke Volume Variation (SVV) monitoring; Device: Pleth Variability Index (PVI) monitoring

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Area under the ROC curve (AUROC) of stroke volume variation (SVV) and Pleth Variability Index (PVI) for predicting fluid responsiveness to a 250-mL crystalloid bolus

    Index tests: pre-bolus SVV (arterial waveform-derived) and PVI (pulse-oximetry-derived). Reference standard: fluid responsiveness defined as a >=10% increase in stroke volume index measured about 3 minutes after the 250-mL crystalloid bolus using arterial pulse contour analysis. ROC curves will be built for SVV and PVI to discriminate responders vs non-responders; AUROCs will be compared with a paired DeLong test with patient-level clustering. Unit: AUROC (0-1; higher indicates better discrimination).

    Measured: Intraoperative: from the pre-bolus baseline to ~3 minutes after each standardized 250-mL crystalloid bolus (up to two episodes per participant: before and after pneumoperitoneum).

Secondary outcomes (10)
  1. Sensitivity of prespecified SVV thresholds for predicting fluid responsiveness

    Proportion of fluid challenges in which SVV > 13% correctly identifies fluid responsiveness (≥10% increase in stroke volume index after a 250 mL bolus).Unit of Measure: Percent of fluid challenges (%)

    Measured: Intraoperative: pre-bolus baseline to about 3 minutes after each 250 mL bolus (up to two boluses).

  2. Specificity of prespecified SVV thresholds for predicting fluid responsiveness

    Proportion of fluid challenges in which SVV ≤ 13% correctly identifies non-responsiveness.

    Unit of Measure: Percent of fluid challenges (%)

    Measured: Intraoperative: pre bolus baseline to about 3 minutes after each 250 mL bolus (up to two episodes per participant).

  3. Sensitivity of prespecified PVI threshold for predicting fluid responsiveness

    Description: Proportion of fluid challenges in which PVI > 15% correctly identifies fluid responsiveness (≥10% increase in stroke volume index).

    Unit of Measure: Percent of fluid challenges (%)

    Measured: Intraoperative: pre bolus baseline to about 3 minutes after each 250 mL bolus (up to two episodes per participant).

  4. Specificity of prespecified PVI threshold for predicting fluid responsiveness

    Proportion of fluid challenges in which PVI ≤ 15% correctly identifies non-responsiveness.

    Unit of Measure: Percent of fluid challenges (%)

    Measured: Intraoperative: pre bolus baseline to about 3 minutes after each 250 mL bolus (up to two episodes per participant).

  5. Optimal cutoffs for SVV using Youden index

    Data-driven SVV threshold (%) that maximizes the Youden index (sensitivity + specificity - 1) for predicting fluid responsiveness.

    Unit of Measure: Percent (%)

    Measured: Intraoperative: pre bolus to about 3 minutes post bolus.

  6. Optimal cutoffs for PVI using Youden index

    Data-driven PVI threshold (%) that maximizes the Youden index for predicting fluid responsiveness.

    Unit of Measure: Percent (%)

    Measured: Intraoperative: pre bolus to about 3 minutes post bolus.

  7. Percent change in stroke volume index after a fluid bolus

    Percent change in stroke volume index from pre-bolus baseline to about 3 minutes after each 250 mL bolus; subgroup analyses will compare periods before vs. after pneumoperitoneum. Unit of Measure: Percent change (%)

    Measured: Intraoperative: pre bolus baseline to about 3 minutes after each 250 mL bolus.

  8. Change in mean arterial pressure after fluid challenge

    Absolute change in mean arterial pressure (mmHg) from pre-bolus baseline to about 3 minutes after each 250 mL bolus.

    Unit of Measure: Millimeters of mercury (mmHg)

    Measured: Intraoperative: pre bolus to about 3 minutes post bolus.

  9. Change in heart rate after fluid challenge

    AAbsolute change in heart rate (beats per minute) from pre-bolus baseline to about 3 minutes after each 250 mL bolus.

    Measured: Intraoperative: pre bolus to about 3 minutes post bolus.

  10. Incidence of any predefined hemodynamic adverse event within 10 minutes of a study bolus

    Number of participants experiencing at least one predefined hemodynamic adverse event (hypotension requiring vasopressor, tachycardia requiring treatment, new arrhythmia, etc.) within 10 minutes after each study bolus, divided by the total number of participants. Unit of Measure: Percent of participants (%)

    Measured: From bolus start to 10 minutes after each bolus during surgery.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted3 Dec 2025
  2. Study start1 Jul 2026
  3. Primary completion15 Jul 2026
  4. Study completion11 Aug 2026

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
300 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07260890
Sponsor's protocol ID
Wonkwang UH 22
Status last verified
Aug 2026
First submitted
13 Nov 2025
Last updated
14 Aug 2026
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

An observational study of Perioperative Care and Fluid Therapy, run by Wonkwang University Hospital at 1 site in South Korea.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 14 Aug 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in South Korea. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Area under the ROC curve (AUROC) of stroke volume variation (SVV) and Pleth Variability Index (PVI) for predicting fluid responsiveness to a 250-mL crystalloid bolus (time frame: Intraoperative: from the pre-bolus baseline to ~3 minutes after each standardized 250-mL crystalloid bolus (up to two episodes per participant: before and after pneumoperitoneum).). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 14 Aug 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in