Evaluation of Calcified Coronary Lesion Preparation With the Naviscore Scoring Balloon

A study testing Scoring Balloon PTCA Catheter and Non-Compliant (NC) Angioplasty Balloon in people with Coronary Arterial Disease (CAD) and Calcific Coronary Arteriosclerosis, run by University of Mons at 14 sites in 4 countries.

Not yet recruiting

Opening soon. The study is approved but enrolment has not started yet. Sites may already take enquiries.

Sites and contacts
Official record
Search all trials

Trial summary

The purpose of this clinical study is to evaluate the effectiveness and safety of a specialized medical device, the Naviscore scoring balloon, in preparing calcified coronary artery narrowings before the implantation of a drug-eluting stent. Read more about this study

Who is it for?
Ages 18 years and older
What is being tested?
Scoring Balloon PTCA Catheter and Non-Compliant (NC) Angioplasty Balloon
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
200 participants
Where
14 sitesin 4 countries
Study dates
Jul 2026to Jun 2028
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patient written consent is given, ≥18 years.
  • A de novo lesion to be treated in a vessel between 2.5 and 4.0 mm.
  • Moderate to heavily calcified lesions detected by coronary angiography on two orthogonal views, and confirmed by IVUS if the lesion can be crossed with the catheter.

Who cannot join

  • Patient <18 years old.
  • Pregnant female.
  • Contraindication to dual antiplatelet therapy.
  • Thrombocytopenia (under 100 000).
  • Major surgical intervention planned within one year.
  • Significant left main lesion.
Show all 10Show fewer
  • Chronic total occlusion.
  • Lesion in a graft.
  • In-stent restenosis lesion.
  • Lesion responsible of a ST elevated Myocardial Infarction (STEMI)

Trial locations and contacts

14 sites in 4 countries. Contact the study team or a site near you to ask about taking part.

Stéphane Carlier, MD, PhD contact

+3265373381

stephane.carlier@umons.ac.be

14 sites in 4 countries

Belgium6 sites
  • Imelda

    Bonheiden, 2820

  • Hôpital Erasme - Cliniques universitaires de Bruxelles

    Brussels, 1070

    Gabriela Flores Vivian, MD, +3225554290, g.floresvivian@hubruxelles.be

    Gabriela Flores Vivian, MD

  • Hôpital Vivalia

    Libramont, 6800

    Chadi Ghafari, MD, PhD, +32483128996, chadi_ghafari@yahoo.com

    Chadi Ghafari, MD, PhD

  • CHR de la Citadelle

    Liège, 4000

    Georges Saad, MD, +3243217155, georges.saad@citadelle.be

    Sophie Jacquet, +3243217820, rc-cardio@citadelle.be

  • Humani CHU Charleroi - Chimay

    Lodelinsart, 6042

    Cellule de recherche clinique, +32071920545, cellule.recherche.clinique@humani.be

    Mohamed Ben Yedder, MD

  • CHU - HELORA site Kennedy

    Mons, Hainaut, 7000

    Stéphane Carlier, MD, PhD, +3265373381, stephane.carlier@umons.ac.be

    Stéphane Carlier, MD, Phd

Other trials in Belgium

Spain6 sites
  • Hospital de Sant Pau

    Barcelona, 08025

    Estefanía Fernández-Peregrina, MD, +34628839525, efernandezpe@santpau.cat

    Elena Jiménez Xarrié, +34628839525

  • Hospital Universitari de Bellvitge

    Barcelona, 08901-08908

    Andrea Ruberti, +34 932 60 75 00, andreaalbertoruberti@bellvitgehospital.cat

    Andrea Ruberti, MD

  • Hospital San Pedro de Alcantara

    Cáceres, 10003

    Juan Carlos Rama, +34615611413, ramamerchan@hotmail.com

    Juan Carlos Rama, MD

  • Hospital Puerta Del Mar

    Cadiz, 11009

    Alejandro Gutiérrez-Barrios, MD, +34 620688877, aleklos@hotmail.com

    Alejandro Gutiérrez-Barrios, MD

  • Hospital La Fe

    Valencia, 46026

    Jorge Sanz Sanchez, +34 660360598, sjorge4@gmx.com

    Ines Roche Fortea, ines_roche@islafe.es

  • Hospital Clínico Universitario de Valladolid

    Valladolid, 47003

    Ignacio J Amat Santos, MD, +34983420000 ext. 86571, ijamat@gmail.com

    Carlos Baladrón Zorita, +34983 42 00 00 ext. 86571, cbaladron@icicor.es

Other trials in Spain

Luxembourg1 site
  • INCCI Haerz Zenter

    Luxembourg, 1210

    Bruno Pereira, MD, + 352 691 86 31 86, dr.pereira@pt.lu

    Bruno Pereira, MD

Other trials in Luxembourg

Portugal1 site
  • Centro Hospitalar de Vila Nova de Gaia/Espinho

    Vila Nova de Gaia, 4430-999

    Alberto Rodrigues, +351919589753, rodrigues.jal@gmail.com

    Alberto Rodrigues, MD

Other trials in Portugal

About this study

Official title: MILOU: Minimize Intra Luminal Obstructive Underexpansion: A Prospective, Multi-Center Randomized Trial for the Evaluation of Calcified Coronary Lesion Preparation With the Naviscore Scoring Balloon

The purpose of this clinical study is to evaluate the effectiveness and safety of a specialized medical device, the Naviscore scoring balloon, in preparing calcified coronary artery narrowings before the implantation of a drug-eluting stent. During percutaneous coronary interventions, the presence of calcified plaques in the heart arteries represents a major challenge because it can prevent stents from expanding fully. When a stent remains under-expanded, it significantly increases the long-term risk of arterial re-narrowing or blood clot formation. To optimize stent expansion, appropriate preparation of the diseased vessel section before stent insertion is a critical phase.

This study is a prospective, multi-center randomized trial designed to test the hypothesis that treating calcified coronary lesions with the Naviscore scoring balloon will achieve a better stent expansion and a larger final minimal stent area compared to standard lesion preparation using regular non-compliant balloons. Eligible participants will be randomized in a one-to-one ratio to one of these two lesion preparation strategies. For all included patients, standard drug-eluting stents will be deployed. The study will use intravascular ultrasound imaging to evaluate the final minimum area of the stent directly inside the treated artery at the site of the highest initial calcium burden. Participant health and clinical outcomes will be monitored for up to twelve months following the procedure.

Read the detailed description

Percutaneous transluminal coronary angioplasty (PTCA) is the main technique in percutaneous coronary intervention (PCI). It aims to restore coronary blood flow by mechanically enlarging a stenotic arterial segment through controlled balloon inflation within the vessel lumen. The acute luminal gain with balloon angioplasty is the result of multiple mechanisms, ranging from percolation and longitudinal redistribution of atheroma, to plaque fracture, and overstretching of the vessel wall, often accompanied by intimal tearing and medial dissection. While contemporary drug-eluting technologies have significantly improved long-term clinical outcomes, fibrocalcified atheromatous intracoronary plaques still present a significant challenge for stent crossing and optimal deployment.

Research has demonstrated that suboptimal stent expansion and malapposition are significant contributing factors to restenosis and stent thrombosis rates, regardless of the type of stent used. Intravascular imaging registries have identified that twenty to thirty percent of deployed stents remain under-expanded or malapposed in daily clinical practice. Severe coronary calcification can make it difficult to deliver devices and can limit the expansion of stents, which can result in suboptimal improvement in blood flow and an increased risk of adverse events, including target lesion failure. Furthermore, heavily calcified lesions have been found to be associated with higher rates of vessel dissection, perforation and impaired anti-proliferative drug delivery, resulting from the mechanical barrier of the calcium burden.

In order to achieve successful stent deployment, optimal preparation of fibro-calcific lesions is essential and involves a number of plaque modification strategies. Focused force angioplasty is a procedure that targets specific locations of the plaque, focusing inflation forces on the area to induce localized stress and facilitate cracking. It is crucial to ensure that the pre-dilatation balloon is correctly sized to maximise safety by minimising the risk of advanced coronary perforation or severe medial damage. It is imperative that sizing compliance chart data is followed rigorously, as compliant balloons can display an unpredictable non-linear expansion behaviour at higher pressures, whereas non-compliant balloons maintain a stable volume, concentrating their dilating force directly at the calcified site up to their rated burst pressure. Semi-compliant pre-dilatation balloons often expand asymmetrically in resistant lesions, increasing the risk of edge dissections. The sizing and performance of scoring devices and non-compliant balloons remains limited in contemporary clinical literature. This randomised trial is designed to evaluate the post-market clinical performance, safety and comparative effectiveness of the Naviscore scoring balloon catheter manufactured by iVascular.

Being tested
Scoring Balloon PTCA Catheter, Non-Compliant (NC) Angioplasty Balloon
Keywords
Percutaneous Coronary Intervention, Lesion Preparation, Scoring Balloon, Intravascular Ultrasound, Stent Expansion, Coronary Artery Calcification

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Scoring Balloon PTCA CatheterDevice

    Also called Naviscore Scoring Balloon

    Percutaneous coronary preparation of calcified coronary stenosis using the rapid exchange Naviscore scoring balloon catheter before drug-eluting stent deployment. The device incorporates an over-the-balloon metallic nitinol element designed to focally concentrate dilation forces, minimizing balloon slippage and scoring the atheromatous plaque at lower pressures to facilitate optimal stent expansion. Sizing is based on a 0.8:1 to 1:1 ratio relative to the reference lumen. Operators must execute a specific sequence of inflating the scoring balloon to nominal pressure, fully deflating it, mobilizing it within the lesion, and reinflating it, repeated three to four times across the target segment to ensure multi-focal plaque disruption before stenting.

  • Non-Compliant (NC) Angioplasty BalloonDevice

    Also called Conventional PTCA Balloon

    Conventional non-compliant (NC) PTCA balloon catheter used for standard lesion pre-dilatation according to routine clinical practice, inflated below the rated burst pressure without parallel guidewires in place.

Study groups in NCT07650578
Study groupRoleWhat they receive
Standard NC groupActive comparatorLesion preparation of calcified coronary stenosis performed using standard non-compliant (NC) balloons. Balloons are sized according to a 0.8:1 to 1:1 ratio relative to the reference lumen and inflated below the rated burst pressure. To maintain a strict comparison and eliminate any scoring effect, no parallel guidewire is allowed in the target vessel during inflation.
Receives: Device: Non-Compliant (NC) Angioplasty Balloon
Naviscore groupExperimentalLesion preparation of calcified coronary stenosis performed using the Naviscore scoring balloon catheter before drug-eluting stent implantation. The device is sized according to a 0.8:1 to 1:1 ratio relative to the reference lumen. To maximize the plaque-modification effect of the external nitinol wires, a mandatory mechanical mobilization sequence must be performed: the balloon is inflated to nominal pressure, fully deflated, mobilized within the lesion, and reinflated. This specific sequence must be repeated 3 to 4 times across the target segment.
Receives: Device: Scoring Balloon PTCA Catheter

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Final Minimal Stent Area (MSA)

    The minimal stent area (MSA) will be measured in square millimeters by intravascular ultrasound (IVUS) after final procedural optimization. The primary comparison will evaluate the MSA specifically at the exact site of the initial heaviest calcium burden to demonstrate the superiority of the Naviscore scoring balloon over standard non-compliant (NC) balloons.

    Measured: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure.

Secondary outcomes (14)
  1. Minimal Stent Area (MSA) Across the Entire Stented Segment

    The minimal stent area (MSA) in square millimeters measured across the entire length of the deployed segment using intravascular ultrasound (IVUS) pullbacks to evaluate overall stent expansion expansion

    Measured: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure

  2. Stent Eccentricity at the Site of the Minimal Stent Area

    Stent eccentricity index calculated as the ratio of the minimum stent diameter over the maximum stent diameter, measured via intravascular ultrasound (IVUS) at the exact cross-section of the minimal stent area segment

    Measured: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure

  3. Stent Eccentricity at the Initial Heaviest Calcium Burden Site

    Stent eccentricity index calculated as the ratio of the minimum stent diameter over the maximum stent diameter, measured via intravascular ultrasound (IVUS) specifically at the cross-section matching the initial heaviest baseline calcium plaque burden

    Measured: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure

  4. Device Success Rate

    Percentage of cases achieving successful delivery of the assigned pre-dilatation balloon, complete inflation with adequate lesion preparation, and the absolute absence of vessel rupture or advanced coronary perforation categorized by the Ellis classification

    Measured: During the index percutaneous coronary intervention (PCI) procedure

  5. Major Adverse Cardiac Events (MACE) Rate

    Percentage of participants experiencing Major Adverse Cardiac Events (MACE), defined as a composite clinical endpoint of myocardial infarction, stroke or transient ischemic attack, or cardiac death

    Measured: At 1 and 12 months post-procedure

  6. Target Lesion Failure (TLF)

    Percentage of participants experiencing Target Lesion Failure (TLF), defined as a composite of cardiac death, target-vessel myocardial infarction defined according to ARC-2 criteria, or clinically indicated Target Lesion Revascularization

    Measured: At 1 and 12 months post-procedure

  7. Target Vessel Failure (TVF)

    Defined as a composite of cardiac death, target-vessel myocardial infarction, or clinically indicated Target Vessel Revascularization (TVR).

    Measured: At 1 and 12 months post-procedure

  8. Clinically Indicated Target Lesion Revascularization (TLR)

    Percentage of participants requiring a repeat percutaneous coronary intervention (PCI) or coronary artery bypass graft surgery (CABG) driven by a recurrent narrowing inside the originally treated target lesion segment (including the stent and its five-millimeter proximal and distal borders), which is accompanied by clinical symptoms of ischemia or objective functional evidence of myocardial ischemia.

    Measured: At 1 and 12 months post-procedure

  9. Clinically Indicated Target Vessel Revascularization (TVR)

    Percentage of participants requiring any repeat percutaneous coronary intervention (PCI) or coronary artery bypass graft surgery (CABG) in any segment of the originally treated target vessel, driven by angiographic restenosis or disease progression, and accompanied by clinical symptoms of ischemia or objective functional evidence of myocardial ischemia.

    Measured: At 1 and 12 months post-procedure

  10. Procedural Success Rate

    Percentage of cases achieving Device Success (successful delivery of the assigned balloon and complete inflation with adequate lesion preparation) combined with a final residual stenosis of less than thirty percent as measured by quantitative coronary angiography (QCA) and a final Thrombolysis in Myocardial Infarction (TIMI) grade 3 flow, without the occurrence of in-hospital Major Adverse Cardiac Events (MACE).

    Measured: During the index percutaneous coronary intervention (PCI) procedure and until hospital discharge (up to 24 hours post-procedure).

  11. Final Residual Stent Underexpansion

    Evaluation of the final residual underexpansion of the drug-eluting stent after deployment and high-pressure post-dilatation, measured in millimeters using motion-corrected X-ray stent visualization software (StentBoost or equivalent stent enhancement method) in two orthogonal views without contrast.

    Measured: At the end of the index percutaneous coronary intervention (PCI) procedure (immediate post-stenting optimization).

  12. Final Stent Diameter Stenosis

    The final percentage of stent diameter stenosis evaluated by quantitative coronary angiography (QCA) at the index procedure after final stent deployment and post-dilatation optimization.

    Measured: At the end of the index percutaneous coronary intervention (PCI) procedure.

  13. Rate of Procedural Complications

    The cumulative rate of procedure-related complications during the index procedure, defined as a composite of cardiac death, acute myocardial infarction, emergency coronary artery bypass graft surgery (CABG), coronary artery perforation (Ellis classification), and cardiac tamponade directly related to the use of the Naviscore scoring balloon or the regular non-compliant (NC) balloon.

    Measured: During the index percutaneous coronary intervention (PCI) procedure (immediate intra-procedural period).

  14. Procedural Cost-Effectiveness

    Evaluation of the procedural cost-effectiveness of using the Naviscore scoring balloon compared to standard non-compliant (NC) balloons for calcified lesion preparation, calculated by factoring device costs, overall material utilisation during the intervention, procedure duration, and the need for bailout additional plaque-modification tools or complementary high-pressure devices.

    Measured: During the index percutaneous coronary intervention (PCI) procedure.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted16 Jun 2026
  2. Study startJul 2026 (estimated)
  3. Primary completionJun 2027 (estimated)
  4. Study completionJun 2028 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
200 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07650578
Sponsor's protocol ID
009
Other IDs
Unrestricted grant
Acronym
MILOU
Status last verified
Jun 2026
First submitted
10 Jun 2026
Last updated
16 Jun 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Yes: The individual participant data (IPD) that underlie the results reported in this publication, including the final minimal stent area (MSA) measurements, procedural success data, and relevant clinical adverse event records, will be shared. Supporting documents, including the study protocol, statistical analysis plan, and informed consent forms, will be provided to support the interpretation of the IPD. All data shared will be de-identified to maintain participant confidentiality in accordance with GDPR requirements.

Publications

  • Garcia-Garcia HM, McFadden EP, Farb A, Mehran R, Stone GW, Spertus J, Onuma Y, Morel MA, van Es GA, Zuckerman B, Fearon WF, Taggart D, Kappetein AP, Krucoff MW, Vranckx P, Windecker S, Cutlip D, Serruys PW; Academic Research Consortium. Standardized End Point Definitions for Coronary Intervention Trials: The Academic Research Consortium-2 Consensus Document. Eur Heart J. 2018 Jun 14;39(23):2192-2207. doi: 10.1093/eurheartj/ehy223.

    PubMed 29897428
  • Jujo K, Saito K, Ishida I, Kim A, Suzuki Y, Furuki Y, Ouchi T, Ishii Y, Sekiguchi H, Yamaguchi J, Ogawa H, Hagiwara N. Intimal disruption affects drug-eluting cobalt-chromium stent expansion: A randomized trial comparing scoring and conventional balloon predilation. Int J Cardiol. 2016 Oct 15;221:23-31. doi: 10.1016/j.ijcard.2016.07.002. Epub 2016 Jul 4.

    PubMed 27400292
  • Riley RF, Patel MP, Abbott JD, Bangalore S, Brilakis ES, Croce KJ, Doshi D, Kaul P, Kearney KE, Kerrigan JL, McEntegart M, Maehara A, Rymer JA, Sutton NR, Shah B. SCAI Expert Consensus Statement on the Management of Calcified Coronary Lesions. J Soc Cardiovasc Angiogr Interv. 2024 Jan 31;3(2):101259. doi: 10.1016/j.jscai.2023.101259. eCollection 2024 Feb.

    PubMed 39132214
  • Barbato E, Gallinoro E, Abdel-Wahab M, Andreini D, Carrie D, Di Mario C, Dudek D, Escaned J, Fajadet J, Guagliumi G, Hill J, McEntegart M, Mashayekhi K, Mezilis N, Onuma Y, Reczuch K, Shlofmitz R, Stefanini G, Tarantini G, Toth GG, Vaquerizo B, Wijns W, Ribichini FL. Management strategies for heavily calcified coronary stenoses: an EAPCI clinical consensus statement in collaboration with the EURO4C-PCR group. Eur Heart J. 2023 Nov 1;44(41):4340-4356. doi: 10.1093/eurheartj/ehad342.

    PubMed 37208199
  • Zhang M, Matsumura M, Usui E, Noguchi M, Fujimura T, Fall KN, Zhang Z, Nazif TM, Parikh SA, Rabbani LE, Kirtane AJ, Collins MB, Leon MB, Moses JW, Karmpaliotis D, Ali ZA, Mintz GS, Maehara A. Intravascular Ultrasound-Derived Calcium Score to Predict Stent Expansion in Severely Calcified Lesions. Circ Cardiovasc Interv. 2021 Oct;14(10):e010296. doi: 10.1161/CIRCINTERVENTIONS.120.010296. Epub 2021 Oct 19.

    PubMed 34665658
  • Mangieri A, Nerla R, Castriota F, Reimers B, Regazzoli D, Leone PP, Gasparini GL, Khokhar AA, Laricchia A, Giannini F, Casale F, Bezzeccheri A, Briguori C, Colombo A. Cutting balloon to optimize predilation for stent implantation: The COPS randomized trial. Catheter Cardiovasc Interv. 2023 Mar;101(4):798-805. doi: 10.1002/ccd.30603. Epub 2023 Feb 25.

    PubMed 36841945
  • Penaranda AS, Peregrina EF, Kockar MJ, Del Blanco BG, Romani S, Moreiras JM, Bermudez EP, Rodrigues A, Ojeda S, Lopez NG, Regueiro A, Frutos AS. New scoring balloon to treat moderate-to-severe calcified coronary lesions. The first-in-man Naviscore study. REC Interv Cardiol. 2025 Feb 3;7(2):91-98. doi: 10.24875/RECICE.M24000484. eCollection 2025 Apr-Jun.

    PubMed 40438639

Questions about this trial

What is this clinical trial studying?

A study testing Scoring Balloon PTCA Catheter and Non-Compliant (NC) Angioplasty Balloon in people with Coronary Arterial Disease (CAD) and Calcific Coronary Arteriosclerosis, run by University of Mons at 14 sites in 4 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 10 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: opening soon, as of the sponsor's last update on 16 Jun 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 14 sites in Belgium, Spain, Luxembourg, Portugal. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Stéphane Carlier, MD, PhD (+3265373381, stephane.carlier@umons.ac.be), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Final Minimal Stent Area (MSA) (time frame: Immediately post-stenting optimization during the index percutaneous coronary intervention (PCI) procedure.). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT07650578, which also lists 7 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 16 Jun 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in