AzTrials
AzTrials

Pulse OximetryNCT07680530

SpO2 Verification Study - Philips Nasal Alar With Nellcor OxiMax SpO2 Board (453564624031)

An observational study of Pulse Oximetry, run by Philips Clinical & Medical Affairs Global at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The primary purpose of the clinical investigation is to verify the SpO2 accuracy of the Philips AlarX5 SpO2 sensor with Nellcor OxiMax SpO2 Board. Read more about this study

Who is it for?
Ages 18 to 50 yearsHealthy volunteers can take part
Conditions the study is for
Pulse Oximetry
What is being tested?
Evaluation of Oxygen Sensors
How many people
43 participants
Where
1 siteUnited States
Study dates
28 May 2026to 29 Jun 2026
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 50 years
  • All sexes
  • Healthy volunteers accepted
  • Adult

Study population: The study population will include a minimum of 36 to 72 healthy adult volunteers.

Who may be able to join

  • Participant must have the ability to understand and provide written informed consent
  • Participant is an adult between 18-50 years of age
  • Participant must be willing and able to comply with study procedures and duration
  • Participant is a non-smoker or who has not smoked within 2 days prior to the study

Who cannot join

  • Participant is considered as being morbidly obese (defined as BMI >39.5)
  • Compromised circulation (i.e., Raynaud's Syndrome), injury, or physical malformation of fingers, hands, nose, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. Tattoos (e.g. permanent, Henna) or artificial dyes (e.g. spray tan, artificial tanning lotion) in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow participants to participate if the condition is noted and would not affect the sites utilized.)
  • Females who are pregnant
  • Females who are trying to get pregnant with confirmation of positive urine pregnancy test
  • Participants who have smoked in the last 2 days or participants who have refrained, for at least 48 hours, with COHb levels >3% as assessed with a Masimo Radical 7 (Rainbow)
  • Participants with known respiratory conditions such as: (self-reported)
Show all 58Show fewer
  • uncontrolled / severe asthma
  • flu
  • pneumonia / bronchitis
  • shortness of breath / respiratory distress
  • unresolved respiratory or lung surgery
  • emphysema, COPD, lung disease
  • recent COVID with hospitalization
  • Participants with self-reported heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review)
  • high blood pressure: systolic >140 mmHg or diastolic >90 mmHg on 3 consecutive readings (reviewed during health screen)
  • have had cardiovascular surgery, except successful minor surgery without clinical symptoms (i.e., PFO, PDA)
  • chest pain (angina)
  • heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen)
  • previous heart attack
  • blocked artery
  • unexplained shortness of breath
  • congestive heart failure (CHF)
  • history of stroke
  • transient ischemic attack
  • carotid artery disease
  • myocardial ischemia
  • myocardial infarction
  • cardiomyopathy
  • implantable active medical device such as pacemaker or automatic defibrillator
  • Participants with self-reported health conditions
  • diabetes
  • uncontrolled thyroid disease
  • kidney disease / chronic renal impairment,
  • history of seizures (except childhood febrile seizures)
  • epilepsy
  • history of unexplained syncope
  • recent history of frequent migraine headaches
  • recent symptomatic head injury (within the last 2 months) with continued symptoms
  • cancer requiring chemotherapy, radiation, or current treatment - Participants with known clotting disorders (self reported)
  • history of bleeding disorders or personal history of prolonged bleeding from injury
  • history of blood clots
  • hemophilia
  • current use of blood thinner: prescription or daily use of aspirin
  • Sickle Cell Trait or Disease - Participants with self-reported dermatological conditions at sensor application sites
  • severe dermatitis
  • hyperkeratosis
  • nail fungus - Participants with severe contact allergies to standard adhesives, latex, silicone or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self reported)
  • Participants with severe allergies to iodine (only applicable if iodine is used)
  • Participants with severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocaine)
  • Participants with allergies to ultrasound gel
  • Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test (Ratio <0.4)
  • Unwillingness or inability to remove nail polish, nail jewelry, or artificial nails from test digits. Bruised nail bed of test digits.
  • Unwillingness or inability to remove makeup, creams, lotions, sunscreen, or powders from application sites on the nose or ear.
  • Unwillingness or inability to cut/trim fingernail(s) of test digits if determined length will interfere with correct application of the sensor
  • Piercing at application site of the ear or nose sensors that may interfere with correct placement of sensor
  • Participants received colored intravascular dye within the past 48 hours (e.g. Indocyanine green, methylene blue, dyes used in cardiac output monitoring)
  • Surgical hardware in pathway of Device Under Test
  • Other known health conditions should be considered upon disclosure

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Element Materials Technology

    Louisville, Colorado, 80027

Other trials in United States

About this study

The primary purpose of the clinical investigation is to verify the SpO2 accuracy of the Philips AlarX5 SpO2 sensor with Nellcor OxiMax SpO2 Board.

Read the detailed description

To verify the SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) with the Nellcor OxiMax pulse oximetry technology during non-motion conditions over the range of 70-100% SaO2, when compared with the reference SaO2 obtained from CO-Oximetry analysis of arterial blood samples.

Conditions
Pulse Oximetry
Being tested
Evaluation of Oxygen Sensors

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Evaluation of Oxygen SensorsOther

    The experimental Philips AlarX5 SpO2 sensor will be tested with Nellcor OxiMax pulse oximetry technology. The AlarX5 SpO2 sensor will be connected via an adapter cable to an IntelliVue X3 Patient Monitor integrated with a Nellcor OxiMax SpO2 Board. Within the study, biomedical/health related outcomes as it relates to the healthy volunteers are not evaluated. The data collected will be used to evaluate how accurate the pulse oximeter (DUT) is to the reference.

    Devices under test are listed below:

    • Nellcor OxiMax SpO2 Board (453564624031) integrated within Philips IntelliVue X3 Patient Monitor (867030)
    • Philips AlarX5, Experimental SpO2 sensor based on the Nasal Alar design (nasal ala and earlobe placement)
    • Philips M1943NL SpO2 Adapter Cable (3m)
Study groups in NCT07680530
Study groupRoleWhat they receive
Healthy VolunteersParticipants are healthy, non-smoking (or has refrained from smoking for 2 days) competent adults, ages 18 to 50 years.
Receives: Other: Evaluation of Oxygen Sensors

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation.

    Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation.

    Measured: Up to 6 weeks

Secondary outcomes (1)
  1. Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (earlobe placement) during non-motion conditions.

    Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (earlobe placement) during non-motion conditions.

    Measured: Up to 6 weeks

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted2 Jul 2026
  2. Study start28 May 2026
  3. Primary completion29 Jun 2026
  4. Study completion29 Jun 2026

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Other
Time perspective
Prospective
Enrolment
43 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07680530
Sponsor's protocol ID
SEF-2025-301123
Status last verified
Jul 2026
First submitted
16 Jun 2026
Last updated
16 Jul 2026
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
Undecided: This is currently under consideration.

Questions about this trial

What is this clinical trial studying?

An observational study of Pulse Oximetry, run by Philips Clinical & Medical Affairs Global at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 58 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 16 Jul 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Measurement and comparison of SpO2 accuracy of the AlarX5 SpO2 sensor (nasal ala application) compared to arterial oxygen saturation. (time frame: Up to 6 weeks). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 16 Jul 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in