Comparative Effects Of Dialysate Flow Rate And Membrane Packing On The Performance Of Dialyzers Used For Hemodialysis

A study testing Polyflux HD-C4 Big, HD-C4 Small and Optiflux 160NR in people with Chronic Kidney Disease and Chronic Renal Disease, run by Vantive Health LLC at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purposes of this study are to determine if the performance of a dialyzer depends on how tightly the hollow fiber membranes are packed in the housing of the dialyzer (the membrane packing density) and if that dependence is a function of the dialysate flow rate. Read more about this study

Who is it for?
Ages 19 years and older
Conditions the study is for
Chronic Kidney Disease, Chronic Renal Disease
What is being tested?
Polyflux HD-C4 Big, HD-C4 Small and Optiflux 160NR
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
12 participants
Where
1 siteUnited States
Study dates
Feb 2008to Apr 2008
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 19 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Stable hemodialysis patients dialyzing through a native fistula or Gore-Tex graft. The access must be capable of delivering a stable blood flow of 400 ml/min.
  • Age older than 18 years.
  • Fluid removal requirement less than 3 liters per treatment.

Who cannot join

  • Noncompliance with dialysis regimen.
  • Hematocrit less than 28%.
  • Active infection

Trial locations and contacts

1 site in 1 country.

United States1 site
  • University of Louisville, Kidney Disease Program

    Louisville, Kentucky, 40202-1718

Other trials in United States

About this study

Official title: Effect Of Dialysate Flow Rate On Mass Transfer Coefficient - Area (KoA) In Dialyzers With Different Membrane Packing Densities

The purposes of this study are to determine if the performance of a dialyzer depends on how tightly the hollow fiber membranes are packed in the housing of the dialyzer (the membrane packing density) and if that dependence is a function of the dialysate flow rate. The study will examine how efficiently three different sized molecules pass through a dialyzer membrane at different dialysate flow rates in dialyzers with different membrane packing densities. Transfer of urea, phosphorus and beta-2-microglobulin from blood to dialysate will be measured during clinical hemodialysis using four different dialyzers, each used at three different dialysate flow rates. The data derived from these measurements may provide insight into the importance of membrane packing density as a design parameter for hemodialyzers and if changing the membrane packing density might provide equivalent performance at a lower dialysate flow rate.

Being tested
Polyflux HD-C4 Big, HD-C4 Small, Optiflux 160NR, Optiflux 200NR
Keywords
Hemodialysis, Dialysis, Renal
Medical subject headings
Renal Insufficiency, Chronic

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Polyflux HD-C4 BigDevice

    Also called Gambro Polyflux HD-C4 Big dialyzer, Gambro Revaclear MAX

    Three consecutive treatments with the HD-C4 Big. During the third treatment, dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.

  • HD-C4 SmallDevice

    Also called Gambro Polyflux HD-C4 Small dialyzer, Gambro Revaclear dialyzer

    Three consecutive treatments with the HD-C4 Small. During the third treatment dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.

  • Optiflux 160NRDevice

    Also called Fresenius Optiflux 160NR

    Three consecutive treatments with the HD-C4 Small. During the third treatment dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.

  • Optiflux 200NRDevice

    Also called Fresenius Optiflux 200NR

    Three consecutive treatments with the HD-C4 Small. During the third treatment dialyzer clearances of urea, phosphorus and β2-microglobulin will be determined at a blood flow rate of 400 mL/min and dialysate flow rates of 350 mL/min, 500 mL/min and 800 mL/min.

Study groups in NCT00636077
Study groupRoleWhat they receive
HD-C4 BigOther3 consecutive treatments with the HD-C4 Big dialyzer.
Receives: Device: Polyflux HD-C4 Big
HD-C4 SmallOther3 consecutive treatments with the HD-C4 Small dialyzer.
Receives: Device: HD-C4 Small
F160NROther3 consecutive treatments with the F160NR dialyzer.
Receives: Device: Optiflux 160NR
F200NROther3 consecutive treatments with the F200NR dialyzer.
Receives: Device: Optiflux 200NR

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Effect of Increased Dialysate Flow Rate on KoA for Urea Between 4 Dialyzers With Different Membrane Packing Densities.

    KoA is a constant that describes the efficiency of a dialyzer in removing urea. KoA is determined by surface area of the membrane, the thickness of the membrane and pore size.

    Measured: During the third treatment with each dialyzer (one time during each trial period week)

  2. Effect of Increased Dialysate Flow Rate on Whole Blood Urea Clearance Between 4 Dialyzers With Different Membrane Packing Densities.

    Measured: During the third treatment with each dialyzer (one time during each trial period week)

  3. Effect of Increased Dialysate Flow Rate on KoA for Phosphorus Between 4 Dialyzers With Different Membrane Packing Densities.

    Measured: During the third treatment with each dialyzer (one time during each trial period week)

  4. Effect of Increased Dialysate Flow Rate on Whole Blood Phosphorus Clearance Between 4 Dialyzers With Different Membrane Packing Densities.

    Measured: During the third treatment with each dialyzer (one time during each trial period week)

  5. Effect of Increased Dialysate Flow Rate on KoA for b2-microglobulin Between 4 Dialyzers With Different Membrane Packing Densities.

    Measured: During the third treatment with each dialyzer (one time during each trial period week)

  6. Effect of Increased Dialysate Flow Rate on Whole Blood B2-microglobulin Clearance Between 4 Dialyzers With Different Membrane Packing Densities.

    Measured: During the third treatment with each dialyzer (one time during each trial period week)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted14 Mar 2008 (estimated)
  2. Study startFeb 2008
  3. Primary completionApr 2008
  4. Study completionApr 2008

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Crossover
Masking
None (open label)
Enrolment
12 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT00636077
Sponsor's protocol ID
Gambro 1461
Status last verified
Jul 2025
First submitted
20 Feb 2008
Last updated
14 Jul 2025
Data monitoring committee
No

Publications

  • Leypoldt JK, Cheung AK, Agodoa LY, Daugirdas JT, Greene T, Keshaviah PR. Hemodialyzer mass transfer-area coefficients for urea increase at high dialysate flow rates. The Hemodialysis (HEMO) Study. Kidney Int. 1997 Jun;51(6):2013-7. doi: 10.1038/ki.1997.274.

    PubMed 9186896
  • Ouseph R, Ward RA. Increasing dialysate flow rate increases dialyzer urea mass transfer-area coefficients during clinical use. Am J Kidney Dis. 2001 Feb;37(2):316-20. doi: 10.1053/ajkd.2001.21296.

    PubMed 11157372
  • Michaels AS. Operating parameters and performance criteria for hemodialyzers and other membrane-separation devices. Trans Am Soc Artif Intern Organs. 1966;12:387-92. No abstract available.

    PubMed 5960730
  • Ward RA, Ouseph R: Comparative evaluation of small and large molecule removal during hemodialysis with Polyflux-S and Fresenius polysulfone membranes. Report to Gambro Renal Products, February 6, 2003.

  • Ward RA, Ouseph R: A comparison of urea, phosphorus and b2-microglobulin removal during hemodialysis with dialyzers containing Polyflux H and Fresenius Optiflux membranes. Report to Gambro Renal Products, August 29, 2005.

  • Leypoldt JK, Cheung AK, Chirananthavat T, Gilson JF, Kamerath CD, Deeter RB. Hollow fiber shape alters solute clearances in high flux hemodialyzers. ASAIO J. 2003 Jan-Feb;49(1):81-7. doi: 10.1097/00002480-200301000-00013.

    PubMed 12558312
  • Ward RA, Ouseph R: Modification of membrane characteristics allows a reduction in dialyzer membrane area without loss of performance. J Am Soc Nephrol 18:453A, 2007.

  • Chen PS, Toribara TY, Warner H. Microdetermination of phosphorus. Anal Chem 1956; 28: 1756-1758

  • Bhimani JP, Ouseph R, Ward RA. Effect of increasing dialysate flow rate on diffusive mass transfer of urea, phosphate and beta2-microglobulin during clinical haemodialysis. Nephrol Dial Transplant. 2010 Dec;25(12):3990-5. doi: 10.1093/ndt/gfq326. Epub 2010 Jun 13.

    PubMed 20543211

Questions about this trial

What is this clinical trial studying?

A study testing Polyflux HD-C4 Big, HD-C4 Small and Optiflux 160NR in people with Chronic Kidney Disease and Chronic Renal Disease, run by Vantive Health LLC at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 14 Jul 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 1 Jul 2011). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Effect of Increased Dialysate Flow Rate on KoA for Urea Between 4 Dialyzers With Different Membrane Packing Densities. (time frame: During the third treatment with each dialyzer (one time during each trial period week)). The study lists 6 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT00636077, which also lists 9 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 14 Jul 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in