Corneal TopographyNCT01861925
Study of Clinical Performance of Corneal Topography Measurement Using an Optical Biometer
An observational study of Corneal Topography, run by Haag-Streit AG at 1 site in Switzerland.
Trial summary
The purpose of this study is to quantify in-vivo repeatability of a new corneal topography measurement mode of an optical biometer, and the clinical performance of measurement in comparison to reference devices. Read more about this study
- Who is it for?
- Ages 18 years and olderHealthy volunteers can take part
- Conditions the study is for
- Corneal Topography
- How many people
- 50 participants
- Where
- 1 siteSwitzerland
- Study dates
- Jun 2013to Jul 2013
- Phase
- Observational
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- Healthy volunteers accepted
- Adult, Older adult
Study population: Volunteers recruited at University Hospital
Who may be able to join
- 18 years or older
Who cannot join
- No corneal transplants
- No corneal scar(s)
- No acute infections of the eye
- No weak fixation
- No suspected lack of compliance
- No vulnerable population
Trial locations and contacts
1 site in 1 country.
About this study
Official title: Lenstar LS 900 Topography
The purpose of this study is to quantify in-vivo repeatability of a new corneal topography measurement mode of an optical biometer, and the clinical performance of measurement in comparison to reference devices.
- Conditions
- Corneal Topography
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
| Study group | Role | What they receive |
|---|---|---|
| Normal eye | Astigmatism smaller than 1.5 diopters | |
| Large regular astigmatism | Astigmatism of > 1.5 diopters, regular astigmatism. | |
| Large irregular astigmatism | Astigmatism of > 1.5 diopters, irregular astigmatism. |
What is being measured
The primary outcome is the main question the study is designed to answer.
In-vivo Repeatability of Corneal Topography Measurements With Lenstar LS 900 Topography: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences Between Two Consecutive Measurements
Corneal topography is a measurement of the shape of the anterior cornea. The shape of a cornea can be fully quantified by providing a map of local power. "Diopter" is the unit of refractive power of a lens. In case of the cornea, the power K [diopter] is related to the radius (curvature) R [mm] of the best fitting sphere by the relation K=337.5/R.
Here, corneal topography measurements are implemented by the Placido method, i.e. by analyzing the reflection image of a ring-shaped illumination.
According to International Standards Organization (ISO) 19980-2012, repeatability of corneal topography is assessed on the central cornea: area with diameter d<=3mm, and middle cornea: 3mm<d<=6mm.
"power difference (2 rep. meas.)": sample mean and 95% C.I. of differences of spatial mean of corneal power between two consecutive measurements.
"power difference 2 std.dev.": sample mean of twice the deviation of local power differences (two consecutive measurements).
Measured: 1 day of examination
Secondary outcomes (1)
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences for One Measurement Per Device.
For definition of corneal topography, corneal power in "diopter", areas of evaluation and methods, see outcome 1.
This outcome measure aims at testing the equivalence of corneal topography measurement between Lenstar LS 900 Topography (Haag Streit) and Atlas 9000 (Zeiss) by analyzing differences in measurement results for the same eye between both devices.
"power difference (2 devices)" is the sample mean and 95% C.I. of differences of spatial mean of corneal power between two devices. This value quantifies systematic differences between devices (e.g. calibration), ignoring local variations of the power measurement.
"power difference (2 std.dev.)" is the sample mean of twice the standard deviation of local power differences between measurements with both devices. This value quantifies the spatially resolved agreement of corneal shape measurement of the two devices.
Measured: 1 day of examination
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted24 May 2013 (estimated)
- Study startJun 2013
- Primary completionJul 2013
- Study completionJul 2013
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Corneal TopographyStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Observational model
- Cohort
- Time perspective
- Prospective
- Enrolment
- 50 participants (actual)
Sponsor and investigators
- Lead sponsor
- Haag-Streit AG
- Sponsor type
- Industry (pharma, biotech, devices)
- Collaborators
- Insel Gruppe AG, University Hospital Bern
- Responsible party
- Sponsor
- Investigators
- Beatrice Frueh, Prof. Fr., Principal investigator, Insel Gruppe AG, University Hospital Bern
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT01861925
- Sponsor's protocol ID
- HS-LS900-TOPO-01
- Status last verified
- May 2014
- First submitted
- 21 May 2013
- Last updated
- 5 Jun 2014
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
Study at a glance
From the registry record, last updated by the sponsor on 5 Jun 2014. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Lenstar LS 900 Topography
- ClinicalTrials.gov ID
- NCT01861925
- Acronym / sponsor ID
- HS-LS900-TOPO-01
- Conditions
- Corneal Topography
- Study type
- Observational
- Recruitment status
- Completed
- Enrolment
- 50 participants (actual)
- Lead sponsor
- Haag-Streit AGIndustry (pharma, biotech, devices)
- Collaborators
- Insel Gruppe AG, University Hospital Bern
- Study start
- Jun 2013
- Primary completion
- Jul 2013 (actual)
- Study completion
- Jul 2013 (actual)
- First posted
- 24 May 2013 (estimated)
- Results
- Posted 5 Jun 2014
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The purpose of this study is to quantify in-vivo repeatability of a new corneal topography measurement mode of an optical biometer, and the clinical performance of measurement in comparison to reference devices.
Study design
- Observational model
- Cohort
- Time perspective
- Prospective
- Number of arms
- 3
| Arm | Type | Receives |
|---|---|---|
| Normal eye | – | –Astigmatism smaller than 1.5 diopters |
| Large regular astigmatism | – | –Astigmatism of > 1.5 diopters, regular astigmatism. |
| Large irregular astigmatism | – | –Astigmatism of > 1.5 diopters, irregular astigmatism. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
- Study population
- Volunteers recruited at University Hospital
Key inclusion criteria (1 listed)
- 18 years or older
Key exclusion criteria (6 listed)
- No corneal transplants
- No corneal scar(s)
- No acute infections of the eye
- No weak fixation
- No suspected lack of compliance
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Primary outcome
In-vivo Repeatability of Corneal Topography Measurements With Lenstar LS 900 Topography: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences Between Two Consecutive Measurements
Time frame: 1 day of examination
Corneal topography is a measurement of the shape of the anterior cornea. The shape of a cornea can be fully quantified by providing a map of local power. "Diopter" is the unit of refractive power of a lens. In case of the cornea, the power K [diopter] is related to the radius (curvature) R [mm] of the best fitting sphere by the relation K=337.5/R. Here, corneal topography measurements are implemented by the Placido method, i.e. by analyzing the reflection image of a ring-shaped illumination. According to International Standards Organization (ISO) 19980-2012, repeatability of corneal topography is assessed on the central cornea: area with diameter d<=3mm, and middle cornea: 3mm<d<=6mm. "power difference (2 rep. meas.)": sample mean and 95% C.I. of differences of spatial mean of corneal power between two consecutive measurements. "power difference 2 std.dev.": sample mean of twice the deviation of local power differences (two consecutive measurements).
1 secondary outcome
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences for One Measurement Per Device.
Time frame: 1 day of examination
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 5 Jun 2014.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
Related trials for Corneal Topography
Not yet recruiting
Performance Evaluation of the OCUTOP Hand-held Corneal Topographer
Corneal Topography, Dry Eye Syndromes
Recruiting
PRoTECT: Corneal Wavefront Guided PRK + Epi-off CXL vs. PTK+Epi-off CXL in Keratoconus
Keratoconus
Recruiting
CONTOURA vs WFO Ablation PRK & LASIK
Refractive Surgery
Active, not recruitingObservational study
Predicting Manifest Astigmatism in Keratoconus Patients.
Keratoconus, Keratoconus, Artificial Intelligence, Support Vector Machine
RecruitingObservational study
The Effect of Upper Blepharoplasty on Vision Quality in the Early Postoperative Period
Excess Skin of Both Eyelids, Contrast Sensitivity
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Beatrice Frueh, Prof. Fr. principal investigator
Insel Gruppe AG, University Hospital Bern
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
An observational study of Corneal Topography, run by Haag-Streit AG at 1 site in Switzerland.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 1 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 5 Jun 2014. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Switzerland. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 5 Jun 2014). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
In-vivo Repeatability of Corneal Topography Measurements With Lenstar LS 900 Topography: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences Between Two Consecutive Measurements (time frame: 1 day of examination). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 5 Jun 2014.