ECMO TreatmentNCT06593756

Hemodynamic Monitoring and Fluid Responsiveness in Venoarterial Extracorporeal Membrane Oxygenation (VA ECMO) - "HemodynamECMOnitoring-VA Study"

A study testing Transthoracic Echocardiography, Uncalibrated Pulse Contour Analysis and Transpulmonary Thermodilution/Calibrated Pulse Contour Analysis in people with ECMO Treatment and ARDS, run by Medical University of Vienna at 1 site in Austria.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

In extracorporeal membrane oxygenation (ECMO), blood is drawn out of the body via tubes, oxygenated in an artificial lung; and then pumped back into the blood vessels. This allows the supply of oxygen-rich blood to the organs (brain, heart, lungs, kidneys, liver, intestines, etc.) to be maintained. Read more about this study

Who is it for?
Ages 18 to 75 years
Conditions the study is for
ECMO Treatment, ARDS, Pneumonia
What is being tested?
Transthoracic Echocardiography, Uncalibrated Pulse Contour Analysis and Transpulmonary Thermodilution/Calibrated Pulse Contour Analysis
Could I receive a placebo?
No, everyone receives an active treatment
How many people
30 participants
Where
1 siteAustria
Study dates
3 May 2024to Jan 2027
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 75 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patient receiving VA ECMO support
  • Age 18 - 75 years

Who cannot join

  • Pregnancy
  • Conditions not allowing for passive leg raising maneuvers, e.g. "open abdomen", known or suspected elevation of intracranial pressure, recent leg or spinal trauma or orthopedic conditions not permitting leg raising
  • Known ischemic or hemorrhagic stroke within 3 months prior to study enrollment.

Trial locations and contacts

1 site in 1 country. Contact the study team or a site near you to ask about taking part.

Austria1 site, 1 recruiting
  • Medical University of Vienna

    Vienna, 1090

    Thomas Staudinger, MD, +4314040044920, thomas.staudinger@meduniwien.ac.at

    Bernhard Nagler, MD, +4314040044920, bernhard.nagler@meduniwien.ac.at

    Recruiting

Other trials in Austria

About this study

Official title: Validation of Techniques for Hemodynamic Monitoring and Prediction of Fluid Responsiveness in Patients Undergoing Venoarterial Extracorporeal Membrane Oxygenation - A Prospective Diagnostic Accuracy Study ("HemodynamECMOnitoring-VA Study")

In extracorporeal membrane oxygenation (ECMO), blood is drawn out of the body via tubes, oxygenated in an artificial lung; and then pumped back into the blood vessels. This allows the supply of oxygen-rich blood to the organs (brain, heart, lungs, kidneys, liver, intestines, etc.) to be maintained. Continuous monitoring of cardiac function and circulatory status (blood pressure, blood flow to organs) is very important in intensive care medicine in order to control the administration of circulation-supporting medication and infusions. Various devices are routinely used for this task. However, in the specific situation of ECMO treatment, the measurements of these devices could be affected due to the artificial circulation; outside the body. The purpose of this study is therefore to test the accuracy of different methods of circulation monitoring during ECMO treatment.

Read the detailed description

Hemodynamic monitoring and tests for fluid responsiveness are cornerstones of intensive care medicine. Generally, hemodynamic measurements can be obtained, for instance, with the following methods: pulmonary artery catheter, transthoracic echocardiography (TTE), esophageal doppler, transpulmonary thermodilution, pulse contour analysis and bioreactance, amongst others. Maneuvers for assessing volume responsiveness include passive leg raising (PLR), respiratory pulse pressure variation (PPV), stroke volume variation (SVV), inferior vena cava ultrasound (IVC), and end-inspiratory or end-expiratory occlusion tests. While these commonly used methods of hemodynamic assessment have been validated in various clinical scenarios, data are lacking in the setting of venoarterial extracorporeal membrane oxygenation (VA ECMO). VA ECMO is commonly used for circulatory support in patients with severe hemodynamic failure or cardiac arrest. Blood is most commonly drained from a femoral vein, pumped through an oxygenator, where it is oxygenated and decarboxylated, and thereafter reinfused into the patient via an arterial, most commonly femoral, return cannula. Theoretically, the artificial circulation with its blood drainage and return flows may interfere with common hemodynamic monitoring techniques and lead to erroneous measurements. The aim of this study therefore is to validate select techniques of hemodynamic monitoring and assessment of fluid responsiveness in patients on VA ECMO. In the context of this study, the performance of different hemodynamic monitoring tools and techniques for predicting fluid responsiveness will be compared.

Being tested
Transthoracic Echocardiography, Uncalibrated Pulse Contour Analysis, Transpulmonary Thermodilution/Calibrated Pulse Contour Analysis, Esophageal Doppler, Bioreactance
Keywords
VV ECMO, Extracorporeal Membrane Oxygenation, ARDS, Intensive Care Medicine, Critical Care Medicine, Hemodynamic Monitoring, Fluid Responsiveness, Passive Leg Raising, Pulse Contour Analysis, Transpulmonary Thermodilution, Bioreactance, Pulse Pressure Variation, End-expiratory Occlusion Test, End-inspiratory Occlusion Test, Cardiac Output, LVOT VTI, Stroke Volume, Vena Cava Ultrasound
Medical subject headings
Pneumonia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Transthoracic EchocardiographyDevice

    Transthoracic echocardiography (TTE) is used for intermittent non-invasive stroke volume (SV) measurements. It is calculated by multiplication of left ventricular out flow tract (LVOT) and LVOT velocity time integral (VTI), obtained in a parasternal long axis view and apical five chamber view, respectively.

  • Uncalibrated Pulse Contour AnalysisDevice

    Pulse Contour Analysis allows an automated and continuous measurement of stroke volume (SV). Its underlying principle is that the integral of the systolic arterial pressure curve directly correlates with stroke volume.

  • Transpulmonary Thermodilution/Calibrated Pulse Contour AnalysisDevice

    Transpulmonary thermodilution (TPTD) involves the administration of a cold saline bolus into a central venous catheter. A special thermistor catheter placed in the femoral or brachial artery detects the successive changes in blood temperature. The resulting heat dissipation curve is analyzed to estimate stroke volume, cardiac output and other hemodynamic variables such as intrathoracic thermal volume (ITTV), pulmonary thermal volume (PTV), global end-diastolic volume (GEDV), intrathoracic blood volume (ITBV) and extravascular lung water (EVLW). Intermittent TPTD-derived cardiac output measurements (typically performed 1-3x/d) are used to calibrate pulse contour analysis.

  • Esophageal DopplerDevice

    In esophageal Doppler, a thin ultrasound probe, coated with aqueous ultrasound gel, is orally or nasally inserted into the esophagus and orientated towards the aorta. By emission and detection of continuous wave Doppler signals, real time spectral waveforms of red blood cell velocity in the aorta are obtained, from which cardiac indices can be derived.

  • BioreactanceDevice

    Bioreactance is a noninvasive hemodynamic monitoring technique, in which four double electrode sensors are placed on the skin of the chest. A high frequency sine wave is transmitted across the thorax. Pulsatile flow in the aorta causes phase shifts and amplitude changes of this signal, which are measured across the different electrodes and used to compute cardiac output.

  • Passive Leg RaisingDiagnostic test

    Passive Leg Raising (PLR) is a maneuver that mimics a fluid challenge by shifting about 300 ml of venous blood from the lower body to the heart. Thereby, it can help to predict fluid responsiveness without actual fluid infusion. To start with, the patient is placed in a semi-recumbent position. Then, the bed is adjusted so that the patient's torso is moved to a horizontal position and the lower limbs are raised to an angle of 45°. Hemodynamic effects occur and can be measured within one minute.

  • Vena Cava UltrasoundDiagnostic test

    Inferior Vena Cava (IVC) Ultrasound has become a popular technique for assessing volume status. IVC diameter is measured in a subcostal long-axis IVC view 1-2 cm from the junction with the right atrium. The magnitude of distensibility during mechanical ventilation cycles or collapsibility during spontaneous breathing has been proposed to correlate with fluid responsiveness

  • End-expiratory /-inspiratory occlusion testDiagnostic test

    In preload-dependent patients, mechanical ventilation induces periodic changes in cardiac output. Standardized maneuvers of end-expiratory or end-inspiratory interruption over 15 seconds may increase or decrease stroke volume, respectively, which is a valid predictor of fluid responsiveness

  • Fluid bolusDrug

    To verify fluid responsiveness, 500 ml of balanced crystalloids will be infused over a time of 15-20 min (25-33.33 ml/min) after completion of passive leg raising and restoration of baseline patient positioning

Study groups in NCT06593756
Study groupRoleWhat they receive
Entire Study PopulationExperimentalThe entire study population will undergo serial hemodynamic assessments throughout the course of ECMO therapy. Hemodynamic variables are obtained using transthoracic echocardiography, uncalibrated pulse contour analysis, and optionally - depending on device availability - transpulmonary thermodilution, bioreactance and esophageal doppler. Maneuvers for assessing volume responsiveness include passive leg raising (PLR), respiratory pulse pressure variation (PPV), stroke volume variation (SVV), inferior vena cava ultrasound (IVC), and end-inspiratory or end-expiratory occlusion tests.
Receives: Device: Transthoracic Echocardiography; Device: Uncalibrated Pulse Contour Analysis; Device: Transpulmonary Thermodilution/Calibrated Pulse Contour Analysis; Device: Esophageal Doppler; Device: Bioreactance; Diagnostic Test: Passive Leg Raising; Diagnostic Test: Vena Cava Ultrasound; Diagnostic Test: End-expiratory /-inspiratory occlusion test; Drug: Fluid bolus

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Agreement of receiver operating characteristic (ROC) curves for predicting fluid responsiveness using the passive leg-raising test between different cardiac output measurement techniques (echocardiography, pulse contour analysis, thermodilution).

    Cardiac Output (L/min) will be measured using transthoracic echocardiography, uncalibrated pulse contour analysis, and thermodilution before, during, and after a passive leg-raising test, as well as after administration of a fluid bolus of 500 ml balanced crystalloids over 15-20 min. A cardiac output increase of > 15% will be the cut-off for defining fluid responsiveness. Receiver operating characteristic (ROC) curves will be generated for each cardiac output measurement technique and compared using the Hanley-McNeil method. The agreement between the ROC curves (Hanley-McNeil test statistic) will serve as the primary outcome.

    Measured: Repeated measurements throughout ECMO therapy (duration ranging from a few days to up to 24 weeks) and within up to 7 days after ECMO removal. Separate analysis for controlled and assisted mechanical ventilation.

Secondary outcomes (5)
  1. Diagnostic performance (receiver operating characteristic (ROC) area under the curve) of an inspiratory and expiratory occlusion test in conjunction with pulse contour analysis for the prediction of fluid responsiveness during ECMO.

    Cardiac Output (L/min) will be measured using calibrated and uncalibrated pulse contour analysis before, during, and after an end-inspiratory and end-expiratory occlusion test (15 s), as well as after administration of a fluid bolus of 500 ml balanced crystalloids over 15-20 min. A cardiac output increase of > 15% after fluid infusion will be the cut-off for defining fluid responsiveness. Receiver operating characteristic (ROC) curves will be generated to assess the performance of the end-inspiratory and end-expiratory occlusion tests and the best threshold for predicting fluid responsiveness during ECMO.

    Measured: Repeated measurements throughout ECMO therapy (duration ranging from a few days to several weeks) and within a few days after ECMO removal. Separate analysis for controlled and assisted mechanical ventilation.

  2. Changes of cardiac output (L/min) over the course of ECMO therapy

    Cardiac output (L/min) will be measured at different time points (at least at the beginning of ECMO therapy and after ECMO removal) throughout ECMO therapy using transthoracic echocardiography, uncalibrated pulse contour analysis, and thermodilution.

    Measured: Repeated measurements throughout ECMO therapy (duration ranging from a few days to up to 24 weeks) and within up to 7 days after ECMO removal.

  3. Changes of tricuspid annular plane systolic excursion (TAPSE, mm) over the course of ECMO therapy

    Tricuspid annular plane systolic excursion (TAPSE, mm) will be measured at different time points (at least at the beginning of ECMO therapy and after ECMO removal) throughout ECMO therapy using transthoracic echocardiography.

    Measured: Repeated measurements throughout ECMO therapy (duration ranging from a few days to up to 24 weeks) and within up to 7 days after ECMO removal.

  4. Changes of tissue doppler imaging tricuspid annular velocity (cm/s) over the course of ECMO therapy

    Tissue doppler imaging tricuspid annular velocity (cm/s) will be measured at different time points (at least at the beginning of ECMO therapy and after ECMO removal) throughout ECMO therapy using transthoracic echocardiography.

    Measured: Repeated measurements throughout ECMO therapy (duration ranging from a few days to up to 24 weeks) and within up to 7 days after ECMO removal.

  5. Changes in cardiac output (L/min, measured by transthoracic echocardiography, uncalibrated pulse contour analysis, thermodilution) at different ECMO blood flow rates

    During the first days of VA ECMO therapy, cardiac output (L/min) will be measured repeatedly at different ECMO blood flow rates (baseline blood flow, blood flow reduced by 1 l/min, return to baseline blood flow, blood flow increased by 1 l/min) using transthoracic echocardiography, uncalibrated pulse contour analysis, and thermodilution.

    Measured: During the first (up to 7) days of VA ECMO therapy.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted19 Sep 2024
  2. Study start3 May 2024
  3. Primary completionJan 2027 (estimated)
  4. Study completionJan 2027 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
30 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT06593756
Sponsor's protocol ID
1613/2023
Status last verified
Sep 2025
First submitted
3 May 2024
Last updated
29 Sep 2025
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Yes: All individual participant data that underlie results in a publication may be provided to qualified researchers with academic interest in hemodynamic monitoring. Data or samples shared will be coded, with no PHI included.

Questions about this trial

What is this clinical trial studying?

A study testing Transthoracic Echocardiography, Uncalibrated Pulse Contour Analysis and Transpulmonary Thermodilution/Calibrated Pulse Contour Analysis in people with ECMO Treatment and ARDS, run by Medical University of Vienna at 1 site in Austria.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 29 Sep 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Austria. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Bernhard Nagler, MD (+4314040044920, bernhard.nagler@meduniwien.ac.at), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Agreement of receiver operating characteristic (ROC) curves for predicting fluid responsiveness using the passive leg-raising test between different cardiac output measurement techniques (echocardiography, pulse contour analysis, thermodilution). (time frame: Repeated measurements throughout ECMO therapy (duration ranging from a few days to up to 24 weeks) and within up to 7 days after ECMO removal. Separate analysis for controlled and assisted mechanical ventilation.). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 29 Sep 2025.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in