Atrial Synchrony Evaluation of Bachmann Bundle Pacing in Sick Sinus Syndrome

A study testing Bachmann Bundle Pacing and Right Atrial Appendage Pacing in people with Sick Sinus Syndrome, run by Shiyan City Renmin Hospital at 1 site in China.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This is a prospective, randomized study to compare the effects of Bachmann bundle pacing (BBP) versus conventional right atrial appendage (RAA) pacing on atrial electromechanical synchrony in patients with sick sinus syndrome (SSS). Read more about this study

Who is it for?
Ages 30 to 90 years
Conditions the study is for
Sick Sinus Syndrome
What is being tested?
Bachmann Bundle Pacing and Right Atrial Appendage Pacing
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
72 participants
Where
1 siteChina
Study dates
1 Jan 2023to 31 Jan 2025
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 30 to 90 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Meets diagnostic criteria for sick sinus syndrome (e.g., persistent sinus bradycardia <50 bpm, sinus arrest, sinoatrial block, or tachy-brady syndrome).
  • Age between 30 and 90 years.
  • Has not taken any other antiarrhythmic drugs within one month prior to enrollment.
  • Has not participated in any other related clinical trials.
  • Voluntarily signs the informed consent form.

Who cannot join

  • Patients with severe primary diseases causing unstable vital signs (e.g., severe circulatory, digestive, or neurological system diseases).
  • Patients with a history of recurrent syncope, Adams-Stokes syndrome, severe bradycardia (lowest sinus rhythm <30 bpm), or long RR interval >3s.
  • Pregnant or lactating women.
  • Patients with mental illness.

Trial locations and contacts

1 site in 1 country.

China1 site
  • Beijing Anzhen Hospital, Capital Medical University

    Shiyan, Hubei, 442000

Other trials in China

About this study

Official title: Clinical Study on the Application of Two-dimensional Speckle Tracking Imaging to Evaluate the Atrial Synchrony of Bachmann Bundle Pacing in Patients With Sick Sinus Syndrome

This is a prospective, randomized study to compare the effects of Bachmann bundle pacing (BBP) versus conventional right atrial appendage (RAA) pacing on atrial electromechanical synchrony in patients with sick sinus syndrome (SSS). The study utilizes two-dimensional speckle tracking imaging (2D-STI) to assess improvements in cardiac function and synchrony, aiming to determine a more optimal pacing site for SSS patients.

Read the detailed description

Sick sinus syndrome (SSS) is a common indication for permanent pacemaker implantation. Conventional right atrial appendage (RAA) pacing, however, is non-physiological and may lead to atrial conduction delays, atrial remodeling, and an increased risk of atrial fibrillation. Bachmann bundle pacing (BBP) is emerging as a more physiological pacing strategy that may correct atrial conduction delays and improve electromechanical synchronization. This study prospectively enrolled and randomized 72 patients with SSS to receive either BBP or RAA pacing. The primary objective is to evaluate the differences in atrial synchrony, cardiac function, and pacing parameters between the two groups at baseline, 1-month, and 3-month follow-ups, using electrocardiography and two-dimensional speckle tracking imaging (2D-STI). The study aims to provide evidence for BBP as a superior treatment option for improving long-term outcomes in SSS patients.

Being tested
Bachmann Bundle Pacing, Right Atrial Appendage Pacing
Medical subject headings
Sick Sinus Syndrome

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Bachmann Bundle PacingDevice

    Implantation of a permanent pacemaker system. The atrial lead (Medtronic 3830 active electrode) is surgically positioned and fixed in the Bachmann bundle region, located at the angle between the superior vena cava and the atrium in the mid-posterior position of the high atrial septum. Successful placement is confirmed by achieving a P-wave width shortening of at least 30ms compared to the preoperative P-wave. The ventricular lead is positioned in the mid-to-lower portion of the ventricular septum. Pacing parameters are measured and confirmed to be good.

  • Right Atrial Appendage PacingDevice

    Implantation of a permanent pacemaker system using the conventional technique. The atrial active electrode is surgically positioned and fixed in the right atrial appendage. The ventricular lead is placed in the mid-to-lower portion of the ventricular septum. Pacing parameters (sensing > 2.0mV; impedance between 300 to 1000Ω; threshold < 1.5mV) are confirmed to be good.

Study groups in NCT07055035
Study groupRoleWhat they receive
Bachmann Bundle Pacing (BBP) GroupExperimentalPatients were randomized to receive a permanent pacemaker with the atrial lead placed in the Bachmann bundle region. They were followed up at 1 month and 3 months post-procedure.
Receives: Device: Bachmann Bundle Pacing
Right Atrial Appendage (RAA) Pacing GroupActive comparatorPatients were randomized to receive a permanent pacemaker with the atrial lead placed in the conventional right atrial appendage. They were followed up at 1 month and 3 months post-procedure.
Receives: Device: Right Atrial Appendage Pacing

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Change in Inter-atrial Electromechanical Delay (Inter AEMD)

    Measured by two-dimensional speckle tracking imaging (2D-STI) as the time difference between the onset of the P-wave on the electrocardiogram (ECG) and the peak longitudinal strain of the lateral walls of the right and left atria. A smaller value indicates better inter-atrial synchrony.

    Measured: Baseline, 1 Month Post-procedure, 3 Months Post-procedure

  2. Change in Intra-atrial Mechanical Delay (IAMD)

    Measured by 2D-STI as the standard deviation of the time-to-peak longitudinal strain of six segments of the left atrium. A smaller value indicates better intra-atrial synchrony.

    Measured: Baseline, 1 Month Post-procedure, 3 Months Post-procedure

  3. Change in Intra-right Atrial Electromechanical Delay (Intra-right AEMD)

    Measured by 2D-STI as the time difference between the earliest and latest peak longitudinal strain within the right atrium segments.

    Measured: Baseline, 1 Month Post-procedure, 3 Months Post-procedure

Secondary outcomes (13)
  1. Change in P-wave Duration (PWD) and Morphology

    Measured from the 12-lead surface ECG, including P-wave duration (P-AT, P-Am) and P-wave axis (P-aI, P-aL).

    Measured: Baseline, 1 Month Post-procedure, 3 Months Post-procedure

  2. Change in QRS Duration and Axis

    Measured from the 12-lead surface ECG.

    Measured: Baseline, 1 Month Post-procedure, 3 Months Post-procedure

  3. Change in Left Atrial (LA) and Right Atrial (RA) Dimensions

    Measured by standard 2D echocardiography.

    Measured: Baseline, 1 Month Post-procedure, 3 Months Post-procedure

  4. Change in Left Ventricular End-Systolic Diameter (LVESD) and End-Diastolic Diameter (LVEDD)

    Measured by standard 2D echocardiography.

    Measured: Baseline, 1 Month Post-procedure, 3 Months Post-procedure

  5. Change in Left Atrial Strain Parameters

    Includes left atrial reservoir strain (LASr), conduit strain (LAScd), and contraction strain (LASct), measured by 2D-STI.

    Measured: Baseline, 1 Month Post-procedure, 3 Months Post-procedure

  6. Recurrence of Atrial Arrhythmia

    Number of patients experiencing newly developed or recurrent atrial fibrillation or atrial tachycardia.

    Measured: Up to 1 year

  7. Change in Atrial Pacing Threshold

    Measured via pacemaker programmer interrogation. Unit: millivolts (mV).

    Measured: 1 Month Post-procedure, 3 Months Post-procedure

  8. Change in Atrial Sensing (P-wave amplitude)

    Measured via pacemaker programmer interrogation. Unit: millivolts (mV).

    Measured: 1 Month Post-procedure, 3 Months Post-procedure

  9. Change in Atrial Pacing Impedance

    Measured via pacemaker programmer interrogation. Unit: Ohms (Ω).

    Measured: 1 Month Post-procedure, 3 Months Post-procedure

  10. Change in Ventricular Pacing Threshold

    Measured via pacemaker programmer interrogation. Unit: millivolts (mV).

    Measured: 1 Month Post-procedure, 3 Months Post-procedure

  11. Change in Ventricular Sensing (R-wave amplitude)

    Measured via pacemaker programmer interrogation. Unit: millivolts (mV).

    Measured: 1 Month Post-procedure, 3 Months Post-procedure

  12. Change in Ventricular Pacing Impedance

    Measured via pacemaker programmer interrogation. Unit: Ohms (Ω).

    Measured: 1 Month Post-procedure, 3 Months Post-procedure

  13. Percentage of Ventricular Pacing

    Measured via pacemaker programmer interrogation. Unit: Percentage (%).

    Measured: 1 Month Post-procedure, 3 Months Post-procedure

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted8 Jul 2025
  2. Study start1 Jan 2023
  3. Primary completion30 Apr 2024
  4. Study completion31 Jan 2025

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
Single blind
Who is masked
Outcomes assessor
Enrolment
72 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07055035
Sponsor's protocol ID
2025126x
Status last verified
Jun 2025
First submitted
26 Jun 2025
Last updated
8 Jul 2025
Results posted
No
FDA-regulated drug
No
FDA-regulated device
No

Questions about this trial

What is this clinical trial studying?

A study testing Bachmann Bundle Pacing and Right Atrial Appendage Pacing in people with Sick Sinus Syndrome, run by Shiyan City Renmin Hospital at 1 site in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 8 Jul 2025. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Change in Inter-atrial Electromechanical Delay (Inter AEMD) (time frame: Baseline, 1 Month Post-procedure, 3 Months Post-procedure). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.