Upper Limb Amputation Below Elbow (Injury)CYBORGNCT07325708

Evaluation of the Phantom X Implantable EMG Sensor System for Myoelectric Prosthesis Control

A study of Upper Limb Amputation Below Elbow (Injury) and Wrist Disarticulation, run by Phantom Neuro Inc. at 2 sites in Australia.

Recruiting

Looking for participants now. Contact the study team or a site below to ask about taking part.

Sites and contacts
Official record
Search all trials

Trial summary

The CYBORG Study is a first-in-human clinical investigation evaluating the safety and performance of the Phantom X System, an implantable electromyography (EMG) sensor array designed to provide control of upper-limb prostheses. Read more about this study

Who is it for?
Ages 18 years and older
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
10 participants
Where
2 sitesAustralia
Study dates
17 Apr 2026to Jun 2027
Phase
Interventional
Are there results?
Not yet: the study is under way

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patient has a unilateral upper-limb difference (congenital or acquired) at the transradial or wrist level.
  • The patient currently uses a myoelectric prosthesis or is deemed a suitable candidate for fitting with a myoelectric prosthesis.
  • Patient is a suitable candidate for the implantation of the Phantom X sensor array as assessed by the implanting physician.
  • Patient is at least 18 years old and can provide written informed consent.
  • Patient is willing to comply with study protocol and make required study visits.
  • Patient received amputation ≥ 12 months prior to consent.

Who cannot join

  • Patient has a significant cognitive deficit resulting in inability to follow study directions.
  • Patient has a neuro-muscular deficit and is unable to initiate substantial muscle contraction in the residual limb as determined by the implanting physician or investigator.
  • Patient has another medical condition, which, in the opinion of the Investigator, may cause him/her to be non-compliant with the protocol, confound the data interpretation, or is associated with a life expectancy insufficient to allow for the completion of study procedures and follow-up.
  • The patient has a planned medical procedure (e.g., surgery) during the study period that could interfere with compliance with study visits or assessments.
  • Patient has an allergy to anesthesia or other medication used for sedation in the study.
  • Patient has an active implantable device (e.g. cardiac pacemaker, neuromodulation device) that can potentially interfere with the performance of Phantom X system.
Show all 9Show fewer
  • Patient may require an MRI during the study duration.
  • Implantation of the device poses a health risk to the patient as determined by the implanting physician.
  • Women who are breastfeeding, currently pregnant or planning to become pregnant during the duration of the study or have a positive urine pregnancy test at screening.

Trial locations and contacts

2 sites in 1 country. Contact the study team or a site near you to ask about taking part.

Vinod Sharma, PhD contact

+1 512-763-0152

vinod@phantomneuro.com

Australia2 sites, 2 recruiting
  • Cabrini Health Ltd

    Malvern, Victoria, 3189

    Michael Lo, MBBS, FRACS, BSc (Hons), +61 039-509-7802, mlo@bigpond.net.au

    Recruiting
  • ProMotion Prosthetics Pty Ltd

    Moorabbin, Victoria, 3189

    Michael Lo, MSc, Dip.App.Sci, +61 039-532-5098, davidlg@pmprosthetics.com.au

    Recruiting

Other trials in Australia

About this study

Official title: Clinical Evaluation of System for Bionic Arm Gesture Control (CYBORG)

The CYBORG Study is a first-in-human clinical investigation evaluating the safety and performance of the Phantom X System, an implantable electromyography (EMG) sensor array designed to provide control of upper-limb prostheses. The study will assess surgical feasibility, device function, signal quality, and the system's ability to support intuitive prosthetic control during functional tasks. Participants will undergo implantation of the Phantom X sensors and complete a series of in-clinic visits to assess functional performance of the Phantom X system. Patient-reported outcomes will also be collected at various timepoints.

Read the detailed description

This early-feasibility study will evaluate the safety and functional performance of the Phantom X Implantable EMG Sensor System in individuals with unilateral transradial amputation and wrist disarticulation. The system consists of an implanted EMG sensor array, an implanted telemetry module, and an external wearable interface that provides real-time muscle signals to a myoelectric prosthesis.

The study follows a prospective, single-arm, open label design. Participants will undergo surgical implantation followed by device activation, calibration, and integration with a compatible prosthetic hand. Study procedures include periodic assessments of EMG signal quality, standardized functional assessments, and patient reported outcomes including quality of life measures and user experience. Safety will be monitored through adverse-event reporting and system performance checks.

Being tested
Phantom X multielectrode EMG sensor array implant for upper limb prosthesis control
Keywords
Upper limb amputation, Prosthesis, Myoelectric
Medical subject headings
Wounds and Injuries

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Phantom X multielectrode EMG sensor array implant for upper limb prosthesis controlDevice

    Participants receive the Phantom X implantable EMG sensor array and telemetry module to enable myoelectric control of an upper-limb prosthesis.

Study groups in NCT07325708
Study groupRoleWhat they receive
Phantom XExperimentalNot described
Receives: Device: Phantom X multielectrode EMG sensor array implant for upper limb prosthesis control

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Safety of Phantom X system during the implantation procedure, peri-procedurally, and 9 weeks thereafter

    Safety is characterized by collecting device and procedure related adverse events during device implantation and 9-week study follow-up.

    Measured: 9 weeks

  2. Functionality of Phantom X system using standardized Assessment of Capacity for Myoelectric Control (ACMC) instrument

    Average and standard deviation of ACMC scores will be computed for all participants

    Measured: 9 weeks

  3. Functionality of Phantom X system using standardized Targeted Box and Blocks Test (tBBT) instrument

    Average and standard deviation of tBBT scores will be computed for all participants

    Measured: 9 weeks

Secondary outcomes (13)
  1. Safety of Phantom X system during the implantation procedure, peri-procedurally, and 21 weeks thereafter

    Safety is characterized by collecting device and procedure related adverse events during device implantation and 21-week study follow-up.

    Measured: 21 weeks

  2. Functionality of Phantom X system using standardized Assessment of Capacity for Myoelectric Control (ACMC) instrument

    Average and standard deviation of ACMC scores will be computed for all participants

    Measured: 21 weeks

  3. Functionality of Phantom X system using standardized Targeted Box and Blocks Test (tBBT) instrument

    Average and standard deviation of tBBT scores will be computed for all participants

    Measured: 21 weeks

  4. Phantom X Signal-to-Noise Ratio (SNR)

    Characterize signal-to-noise ratio (SNR) of electromyography (EMG) signals obtained using Phantom X system. Average and standard deviation of SNR for all participants will be computed

    Measured: 9 weeks and 21 weeks

  5. Phantom X algorithm gesture accuracy

    Classification accuracy of the Phantom X algorithm will be computed based on the number of correctly and incorrectly classified gestures and reported as a percentage

    Measured: 9 weeks and 21 weeks

  6. Phantom X at-home wear duration

    Average and standard deviation of at-home prosthesis wear duration for all participants will be computed

    Measured: 9 weeks and 21 weeks

  7. At-home limb usage pattern using Inertial Measurement Unit (IMU) data

    Active and inactive durations in hours per day for each limb will be computed for each participant

    Measured: 9 weeks and 21 weeks

  8. Prosthesis control resolution

    Resolution of Phantom X control system for controlling a myoelectric hand prosthesis will be characterized using Myoelectric Control Resolution Assessment (MCRA) instrument

    Measured: 9 weeks and 21 weeks

  9. Number of system recalibration events

    Number of Phantom X recalibrations per participant will be collected and used to compute average and standard deviation of recalibration events for all participants

    Measured: 9 weeks and 21 weeks

  10. User Satisfaction

    User Satisfaction with Phantom X system is characterized using following three standardized instruments: modified version of Orthotics Prosthetics User Survey (mOPUS, range: 0-100, Higher score is better), 5-level Quality of Life questionnaire (EQ-5D-5L, Higher score is better), Canadian Occupational Performance Measure (COPM, range:1-10, Higher score is better), and a study User Satisfaction Survey.

    Measured: 9 weeks and 21 weeks

  11. Robustness of Phantom X system

    Robustness of Phantom X system is characterized by recording problematic activities of daily living (REC-ADL) that participants have problem performing with their baseline myoelectric control system.

    Measured: 9 weeks and 21 weeks

  12. Phantom X ease of use

    Ease of use and intuitiveness of Phantom X system is characterized using NASA-Task Load Index (NASA-TLX) instrument.

    Measured: 9 weeks and 21 weeks

  13. Phantom Limb Pain

    Change in phantom limb pain is characterized post implant and compared with baseline using Questionnaire for Post-Amputation Pain (Q-PAP).

    Measured: 9 weeks and 21 weeks

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted8 Jan 2026
  2. Study start17 Apr 2026
  3. Primary completionDec 2026 (estimated)
  4. Study completionJun 2027 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Device feasibility
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
10 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT07325708
Sponsor's protocol ID
PN03
Acronym
CYBORG
Status last verified
Jun 2026
First submitted
7 Dec 2025
Last updated
24 Jun 2026
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
Undecided: We may not share IPD because of concerns about protecting participants' privacy and confidentiality. Even though our participant data would be de-identified, there is a risk (especially with sensitive, rare, or detailed clinical data as would be the case for amputees in our study) that the individuals could be re-identified. Maintaining confidentiality and minimizing potential harm to participants may outweigh the benefits of data sharing.

Questions about this trial

What is this clinical trial studying?

A study of Upper Limb Amputation Below Elbow (Injury) and Wrist Disarticulation, run by Phantom Neuro Inc. at 2 sites in Australia.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 9 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: looking for participants now, as of the sponsor's last update on 24 Jun 2026. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in Australia. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

This study is still under way. Results are not available in the source record yet.

How can a doctor refer a patient or contact the research team?

Contact Vinod Sharma, PhD (+1 512-763-0152, vinod@phantomneuro.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Safety of Phantom X system during the implantation procedure, peri-procedurally, and 9 weeks thereafter (time frame: 9 weeks). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 24 Jun 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in