Cardiogenic ShockNCT07825363
Clinical Evaluation of NIRSense Tissue Oximetry Sensor Monitoring to Early Detect Cardiogenic Shock
An observational study of Cardiogenic Shock and Acute Myocardial Infarction, run by NIRSense Inc. at 1 site in United States.
Trial summary
This is a prospective, single-center, observational pilot study of the investigational NIRSense Tissue Oximetry Sensor System (Envello Core), a noninvasive wearable sensor patch that uses near-infrared spectroscopy (NIRS) and photoplethysmography (PPG), combined with an electrocardiography/impedance cardiography… Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Cardiogenic Shock, Acute Myocardial Infarction, Acute Decompensated Heart Failure
- What is being tested?
- NIRSense Tissue Oximetry Sensor System
- Type of treatment
- Pill or oral medicine, Medical device
- How many people
- 20 participants
- Where
- 1 siteUnited States
- Study dates
- 17 Jun 2025to Aug 2027
- Phase
- Observational
- Are there results?
- Not yet: the study is under way
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Study population: Adults admitted to the Cleveland Clinic Florida cardiac intensive care unit with acute myocardial infarction undergoing coronary angiography, or with acutely decompensated heart failure, who are in SCAI shock stage B, C, or D. Participants are recruited from the existing patient population of the recruitment site.
Who may be able to join
- Acutely decompensated symptomatic heart failure (HF) documented by: (a) symptoms of HF requiring current treatment; (b) Society for Cardiovascular Angiography and Interventions (SCAI) stage B, C, or D; and (c) hospital admission with HF as the primary or secondary diagnosis; OR acute myocardial infarction (MI) patient undergoing coronary angiography
- Admitted to the intensive care unit and currently in SCAI stage B, C, or D
- Age 18 years or older
- Weight greater than 40 kg
Who cannot join
- Prior known severe allergy to medical-grade adhesive or tape (self-reported)
- Currently participating in an investigational drug or device study that would interfere with the conduct or results of this study. Trials requiring extended follow-up for products that were investigational but have since become commercially available are not considered investigational.
- The investigator believes the participant is not an appropriate candidate for the study
Trial locations and contacts
1 site in 1 country. Contact the study team or a site near you to ask about taking part.
United States1 site, 1 recruiting
- Recruiting
Cleveland Clinic Florida, Weston Hospital
Weston, Florida, 33331
David A. Baran, MD, 954-659-5290, barand@ccf.org
Maxwell Sapper, sapperm2@ccf.org
About this study
Official title: Clinical Evaluation of Using NIRSense Sensor Patch to Measure Microvascular Perfusion Responses and Predict Decompensation in Cardiogenic Shock Patients
This is a prospective, single-center, observational pilot study of the investigational NIRSense Tissue Oximetry Sensor System (Envello Core), a noninvasive wearable sensor patch that uses near-infrared spectroscopy (NIRS) and photoplethysmography (PPG), combined with an electrocardiography/impedance cardiography (ECG/ICG) sensor, to continuously monitor tissue oxygen saturation (StO2), peripheral oxygen saturation (SpO2), pulse rate, respiratory rate, stroke volume, and related metrics. Up to 20 adults admitted to the Cleveland Clinic Florida cardiac intensive care unit with acute myocardial infarction or acute decompensated heart failure and in Society for Cardiovascular Angiography and Interventions (SCAI) shock stage B, C, or D will wear the sensors on the sternum, forearm, and other locations for approximately 24 to 72 hours during standard care. Sensor data will be compared with standard-of-care hemodynamic and laboratory measurements and with the patient's clinical course to evaluate the feasibility of the device, the correlation of its measurements with established clinical parameters, and its potential to detect deterioration in SCAI shock stage earlier. No treatment decisions are made based on the study device.
Read the detailed description
Background: Cardiogenic shock (CS) carries a poor prognosis and is the leading cause of death in patients with acute myocardial infarction. When decompensation occurs there is an imperative to institute treatment rapidly. Standard intensive care unit monitoring relies on intermittent hemodynamic and laboratory measures. Continuous, noninvasive measurement of tissue perfusion and oxygenation could allow earlier recognition of deterioration and earlier intervention.
Device: The NIRSense Tissue Oximetry Sensor System is an investigational, non-significant-risk, noninvasive wearable optical physiological monitoring system. It consists of a sensor patch with multiple red and near-infrared light sources and detectors (NIRS and PPG), a pre-applied medical-grade adhesive, an ECG/ICG sensor patch, and a display unit (tablet running the NIRSense application) that receives data over Bluetooth and stores it. The sensor estimates StO2 in microvascular tissue up to about 2.5 cm below the skin and derives PPG-based metrics including pulse rate, respiratory rate, and perfusion index. The device has not been cleared or approved by the U.S. FDA.
Design: Prospective data collection, single-center observational study. Two populations are targeted: (1) patients with acute myocardial infarction, with or without shock (SCAI B to D), admitted for coronary angiography and subsequent interventions, and (2) patients with acute decompensated heart failure admitted for pulmonary artery catheter insertion and subsequent interventions. Participants are enrolled from the existing Cleveland Clinic Florida patient population in a way that avoids disruption of clinical workflow. After informed consent, sensors are placed on the sternum, forearm, and other locations according to manufacturer specifications and remain in place while the patient is in the cardiac intensive care unit, for approximately 24 to 72 hours. Sensors may be repositioned during catheterization or imaging at the investigator's discretion. Participation is limited to the single standard-of-care admission; there is no study follow-up visit or call.
Data collected: Continuous sensor data (StO2, SpO2, pulse rate, respiratory rate, ECG/ICG-derived stroke volume, PPG-derived metrics) and sensor placement images. Data collected during standard-of-care treatment and shared with the sponsor: demographics, Fitzpatrick skin type, medical history, concomitant medications, physical examination, vital signs, a single 12-lead ECG, blood pressure, heart rate, cardiac output (thermodilution) where obtained, clinically ordered lactate values, SCAI stage, and clinical course. Adverse events and device deficiencies are recorded from enrollment through sensor removal. Clinical staff complete a usability survey after using the device on 5 patients.
Analysis: Descriptive statistics for demographics and baseline characteristics; adverse events summarized by frequency and percentage, severity, and device relatedness. Effectiveness analyses use all evaluable data and compare device measurements with standard-of-care hemodynamic parameters and with SCAI stage trajectory, as specified in the Statistical Analysis Plan. Following an interim analysis of the 20-patient pilot, enrollment may be expanded to 200 patients under a protocol amendment; that expansion is not part of this record.
- Being tested
- NIRSense Tissue Oximetry Sensor System
- Keywords
- near-infrared spectroscopy, photoplethysmography, impedance cardiography, tissue oxygen saturation, perfusion index, microcirculation, wearable sensor, noninvasive hemodynamic monitoring, SCAI shock stage
- Medical subject headings
- Shock, Cardiogenic
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
NIRSense Tissue Oximetry Sensor SystemDevice
Also called Envello Core, NIRSense Sensor Oximetry System patch
Investigational noninvasive wearable sensor patches using near-infrared spectroscopy and photoplethysmography to continuously measure tissue oxygen saturation, SpO2, pulse rate, respiratory rate, and perfusion index, plus an ECG/ICG sensor patch for stroke volume, with a tablet display and recording application. Sensors are applied to the sternum, forearm, and other locations with a medical-grade adhesive and worn for approximately 24 to 72 hours during standard cardiac ICU care. Sensor data are collected for research only and are not used for treatment decisions.
| Study group | Role | What they receive |
|---|---|---|
| Acute myocardial infarction | Patients with acute myocardial infarction, with or without shock (SCAI stage B to D), admitted for coronary angiography and subsequent interventions. Enrolled at the time of consent for coronary angiography. Receives: Device: NIRSense Tissue Oximetry Sensor System | |
| Acute decompensated heart failure | Patients with acute decompensated heart failure (SCAI stage B to D) admitted for pulmonary artery catheter insertion and subsequent interventions. Consented before or after catheterization. Receives: Device: NIRSense Tissue Oximetry Sensor System |
What is being measured
The primary outcome is the main question the study is designed to answer.
Correlation between NIRSense sensor measurements and standard-of-care hemodynamic parameters
Correlation coefficients (Pearson and Spearman) between time-matched NIRSense measurements (tissue oxygen saturation, perfusion index, SpO2, pulse rate, respiratory rate, ICG-derived stroke volume) and standard-of-care clinical measurements (arterial blood pressure, heart rate, pulmonary artery and central venous pressures where a pulmonary artery catheter is present, cardiac output by thermodilution, and lactate).
Measured: Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Secondary outcomes (4)
Feasibility: proportion of intended monitoring time with valid sensor data
Percentage of the intended monitoring period during which each sensor provided valid StO2 and PPG data, per sensor site, including data recovered from onboard memory after wireless transmission gaps.
Measured: Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Patient tolerance: device-related adverse events and device deficiencies
Number and percentage of participants with adverse events, adverse device effects (including skin irritation or redness at the adhesive site), and device deficiencies, and number of participants requesting early sensor removal.
Measured: From enrollment through sensor removal, approximately 24 to 72 hours
Association between NIRSense measurement trends and change in SCAI shock stage
Change in NIRSense measurements (StO2, StO2 variability, perfusion index, stroke volume) relative to the time of clinically documented change in SCAI shock stage, including the lead time between the first sustained change in device metrics and the clinically recognized deterioration.
Measured: Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Distribution of NIRSense measurements by SCAI shock stage
Summary statistics of StO2, perfusion index, and ICG-derived stroke volume at each sensor site by SCAI shock stage (B, C, D, E), to inform development of a predictive model of shock status.
Measured: Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted17 Sep 2026
- Study start17 Jun 2025
- Primary completionJan 2027 (estimated)
- Study completionAug 2027 (estimated)
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Cardiogenic ShockStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Observational model
- Cohort
- Time perspective
- Prospective
- Enrolment
- 20 participants (estimated)
Sponsor and investigators
- Lead sponsor
- NIRSense Inc.
- Sponsor type
- Industry (pharma, biotech, devices)
- Collaborators
- Cleveland Clinic Florida
- Responsible party
- Sponsor
- Investigators
- David A. Baran, MD, Principal investigator, Cleveland Clinic FloridaThibault Roumengous, PhD, Study director, NIRSense Inc.
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT07825363
- Sponsor's protocol ID
- NSE2401
- Other IDs
- 25-127
- Status last verified
- Sep 2026
- First submitted
- 13 Sep 2026
- Last updated
- 17 Sep 2026
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
- Individual data sharing
- Undecided
Study at a glance
From the registry record, last updated by the sponsor on 17 Sep 2026. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Clinical Evaluation of Using NIRSense Sensor Patch to Measure Microvascular Perfusion Responses and Predict Decompensation in Cardiogenic Shock Patients
- ClinicalTrials.gov ID
- NCT07825363
- Acronym / sponsor ID
- NSE2401
- Keywords
- near-infrared spectroscopy, photoplethysmography, impedance cardiography, tissue oxygen saturation, perfusion index, microcirculation, wearable sensor, noninvasive hemodynamic monitoring
- Study type
- Observational
- Recruitment status
- Recruiting
- Enrolment
- 20 participants (estimated)
- Lead sponsor
- NIRSense Inc.Industry (pharma, biotech, devices)
- Collaborators
- Cleveland Clinic Florida
- Study start
- 17 Jun 2025 (actual)
- Primary completion
- Jan 2027 (estimated)
- Study completion
- Aug 2027 (estimated)
- First posted
- 17 Sep 2026 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
This is a prospective, single-center, observational pilot study of the investigational NIRSense Tissue Oximetry Sensor System (Envello Core), a noninvasive wearable sensor patch that uses near-infrared spectroscopy (NIRS) and photoplethysmography (PPG), combined with an electrocardiography/impedance cardiography (ECG/ICG) sensor, to continuously monitor tissue oxygen saturation (StO2), peripheral oxygen saturation (SpO2), pulse rate, respiratory rate, stroke volume, and related metrics. Up to 20 adults admitted to the Cleveland Clinic Florida cardiac intensive care unit with acute myocardial infarction or acute decompensated heart failure and in Society for Cardiovascular Angiography and Interventions (SCAI) shock stage B, C, or D will wear the sensors on the sternum, forearm, and other locations for approximately 24 to 72 hours during standard care. Sensor data will be compared with standard-of-care hemodynamic and laboratory measurements and with the patient's clinical course to evaluate the feasibility of the device, the correlation of its measurements with established clinical parameters, and its potential to detect deterioration in SCAI shock stage earlier. No treatment decisions are made based on the study device.
Detailed description
Background: Cardiogenic shock (CS) carries a poor prognosis and is the leading cause of death in patients with acute myocardial infarction. When decompensation occurs there is an imperative to institute treatment rapidly. Standard intensive care unit monitoring relies on intermittent hemodynamic and laboratory measures. Continuous, noninvasive measurement of tissue perfusion and oxygenation could allow earlier recognition of deterioration and earlier intervention.
Device: The NIRSense Tissue Oximetry Sensor System is an investigational, non-significant-risk, noninvasive wearable optical physiological monitoring system. It consists of a sensor patch with multiple red and near-infrared light sources and detectors (NIRS and PPG), a pre-applied medical-grade adhesive, an ECG/ICG sensor patch, and a display unit (tablet running the NIRSense application) that receives data over Bluetooth and stores it. The sensor estimates StO2 in microvascular tissue up to about 2.5 cm below the skin and derives PPG-based metrics including pulse rate, respiratory rate, and perfusion index. The device has not been cleared or approved by the U.S. FDA.
Design: Prospective data collection, single-center observational study. Two populations are targeted: (1) patients with acute myocardial infarction, with or without shock (SCAI B to D), admitted for coronary angiography and subsequent interventions, and (2) patients with acute decompensated heart failure admitted for pulmonary artery catheter insertion and subsequent interventions. Participants are enrolled from the existing Cleveland Clinic Florida patient population in a way that avoids disruption of clinical workflow. After informed consent, sensors are placed on the sternum, forearm, and other locations according to manufacturer specifications and remain in place while the patient is in the cardiac intensive care unit, for approximately 24 to 72 hours. Sensors may be repositioned during catheterization or imaging at the investigator's discretion. Participation is limited to the single standard-of-care admission; there is no study follow-up visit or call.
Data collected: Continuous sensor data (StO2, SpO2, pulse rate, respiratory rate, ECG/ICG-derived stroke volume, PPG-derived metrics) and sensor placement images. Data collected during standard-of-care treatment and shared with the sponsor: demographics, Fitzpatrick skin type, medical history, concomitant medications, physical examination, vital signs, a single 12-lead ECG, blood pressure, heart rate, cardiac output (thermodilution) where obtained, clinically ordered lactate values, SCAI stage, and clinical course. Adverse events and device deficiencies are recorded from enrollment through sensor removal. Clinical staff complete a usability survey after using the device on 5 patients.
Analysis: Descriptive statistics for demographics and baseline characteristics; adverse events summarized by frequency and percentage, severity, and device relatedness. Effectiveness analyses use all evaluable data and compare device measurements with standard-of-care hemodynamic parameters and with SCAI stage trajectory, as specified in the Statistical Analysis Plan. Following an interim analysis of the 20-patient pilot, enrollment may be expanded to 200 patients under a protocol amendment; that expansion is not part of this record.
Study design
- Observational model
- Cohort
- Time perspective
- Prospective
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Acute myocardial infarction | – | Device: NIRSense Tissue Oximetry Sensor SystemPatients with acute myocardial infarction, with or without shock (SCAI stage B to D), admitted for coronary angiography and subsequent interventions. Enrolled at the time of consent for coronary angiography. |
| Acute decompensated heart failure | – | Device: NIRSense Tissue Oximetry Sensor SystemPatients with acute decompensated heart failure (SCAI stage B to D) admitted for pulmonary artery catheter insertion and subsequent interventions. Consented before or after catheterization. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
- Study population
- Adults admitted to the Cleveland Clinic Florida cardiac intensive care unit with acute myocardial infarction undergoing coronary angiography, or with acutely decompensated heart failure, who are in SCAI shock stage B, C, or D. Participants are recruited from the existing patient population of the recruitment site.
Key inclusion criteria (4 listed)
- Acutely decompensated symptomatic heart failure (HF) documented by: (a) symptoms of HF requiring current treatment; (b) Society for Cardiovascular Angiography and Interventions (SCAI) stage B, C, or D; and (c) hospital admission with HF as the primary or secondary diagnosis; OR acute myocardial infarction (MI) patient undergoing coronary angiography
- Admitted to the intensive care unit and currently in SCAI stage B, C, or D
- Age 18 years or older
- Weight greater than 40 kg
Key exclusion criteria (3 listed)
- Prior known severe allergy to medical-grade adhesive or tape (self-reported)
- Currently participating in an investigational drug or device study that would interfere with the conduct or results of this study. Trials requiring extended follow-up for products that were investigational but have since become commercially available are not considered investigational.
- The investigator believes the participant is not an appropriate candidate for the study
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| NIRSense Tissue Oximetry Sensor SystemAlso: Envello Core, NIRSense Sensor Oximetry System patchInvestigational noninvasive wearable sensor patches using near-infrared spectroscopy and photoplethysmography to continuously measure tissue oxygen saturation, SpO2, pulse rate, respiratory rate, and perfusion index, plus an ECG/ICG sensor patch for stroke volume, with a tablet display and recording application. Sensors are applied to the sternum, forearm, and other locations with a medical-grade adhesive and worn for approximately 24 to 72 hours during standard cardiac ICU care. Sensor data are collected for research only and are not used for treatment decisions. | Device | Acute decompensated heart failure, Acute myocardial infarction |
Primary outcome
Correlation between NIRSense sensor measurements and standard-of-care hemodynamic parameters
Time frame: Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Correlation coefficients (Pearson and Spearman) between time-matched NIRSense measurements (tissue oxygen saturation, perfusion index, SpO2, pulse rate, respiratory rate, ICG-derived stroke volume) and standard-of-care clinical measurements (arterial blood pressure, heart rate, pulmonary artery and central venous pressures where a pulmonary artery catheter is present, cardiac output by thermodilution, and lactate).
4 secondary outcomes
Feasibility: proportion of intended monitoring time with valid sensor data
Time frame: Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Patient tolerance: device-related adverse events and device deficiencies
Time frame: From enrollment through sensor removal, approximately 24 to 72 hours
Association between NIRSense measurement trends and change in SCAI shock stage
Time frame: Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Distribution of NIRSense measurements by SCAI shock stage
Time frame: Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application
Results and safety
This study is still under way. Results are not available in the source record yet.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 17 Sep 2026.
- Companion reference-oximetry validation study of the same sensor systemLink provided by the sponsor.
- Individual participant dataSharing planned: Undecided
Related trials for Cardiogenic Shock
RecruitingObservational study
Evaluation of Acid-base Disorders and Ventilation Settings in Cardiogenic Shock and Post-cardiac Arrest Patients: Comparison With VentilO Algorithm
Cardiac Arrest (CA), Cardiogenic Shock
Not yet recruitingObservational study
Hemodynamic Evaluation of Left Atrial Pressure in Relationship to Pulmonary Capillary Wedge Pressure in Cardio Thoracic Patients
Cardiogenic Shock, Pulmonary Edema Cardiac Cause
Not yet recruitingPhase 4
Capital Offload-Shock
Hypervolemia, Cardiogenic Shock
Recruiting
CRP Apheresis in Infarct-Related Cardiogenic Shock
Cardiogenic Shock, Acute Myocardial Infarction (AMI)
RecruitingObservational study
Cardiogenic Shock Working Group Registry
Cardiogenic Shock
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part.
David A. Baran, MD principal investigator
Cleveland Clinic Florida
Thibault Roumengous, PhD study director
NIRSense Inc.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
An observational study of Cardiogenic Shock and Acute Myocardial Infarction, run by NIRSense Inc. at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 4 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: looking for participants now, as of the sponsor's last update on 17 Sep 2026. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
This study is still under way. Results are not available in the source record yet.
How can a doctor refer a patient or contact the research team?
Contact Thibault Roumengous, PhD (804-625-1135, thibault@nirsense.com), or a site near the patient. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Correlation between NIRSense sensor measurements and standard-of-care hemodynamic parameters (time frame: Continuous monitoring during the cardiac ICU stay, approximately 24 to 72 hours from sensor application). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 Sep 2026.